5.1 Analytical Measurements: Apparatus, Units, Concentration, Sampling, and Sample Preparation
Key Takeaways
Class A volumetric glassware has tight tolerances: about +/-0.03 mL for a 25 mL pipet, +/-0.05 mL for a 50 mL burette, and +/-0.08 mL for a 100 mL volumetric flask.
Concentrated reagent molarity equals density x 1000 x mass fraction / molar mass, so 37.2% HCl (1.19 g/mL) is about 12.1 M and 96% H2SO4 (1.84 g/mL) about 18.0 M.
When a sample is diluted to volume and only an aliquot is titrated, multiply the aliquot result by the total volume divided by the aliquot volume.
Total variance is the sum of the sampling and method variances; once the method standard deviation is below about one-third of the sampling standard deviation, improving the method barely helps.
In the Philippines, ISO/IEC 17025 laboratory accreditation is granted by the DTI Philippine Accreditation Bureau, while the DOST-ITDI National Metrology Laboratory maintains national measurement standards.
5.1 Analytical Measurements: Apparatus, Units, Concentration, Sampling, and Sample Preparation
Quick Answer: Analytical area A (10 items) covers basic concepts: instruments and apparatus, laboratory safety and disposal (Section 1.3), units, and stoichiometric, concentration, dilution, and aliquot calculations. Area B adds sampling and sample preparation, significant figures, and standards organizations. These calculations sit inside almost every quantitative question, so practise them until they are automatic.
Weighing and Volumetric Apparatus
The Analytical Balance
- A standard analytical balance reads to 0.1 mg; semimicro balances read to 0.01 mg.
- Weighing by difference (weigh the container, transfer the sample, reweigh) avoids exposing hygroscopic solids and cancels constant balance errors.
- Dry primary standards in an oven (typically 105-110 C) and cool them in a desiccator before weighing. Never weigh warm objects: convection currents make them read light.
- A buoyancy correction matters for low-density samples: , with g/mL and g/mL for steel weights.
Volumetric Glassware
| Item | Calibration | Class A Tolerance (Typical Size) |
|---|---|---|
| Volumetric (transfer) pipet | To deliver (TD) | mL (25 mL); mL (10 mL) |
| Burette | To deliver (TD) | mL (50 mL) |
| Volumetric flask | To contain (TC) | mL (100 mL); mL (250 mL) |
| Graduated cylinder, beaker | Approximate only | Not for quantitative work |
Read the bottom of the meniscus at eye level to avoid parallax. Rinse pipets and burettes with the solution they will hold, never with water alone. Calibrate glassware by weighing the water it delivers: at 25 C, 1.0000 g of water weighed in air with steel weights corresponds to 1.0040 mL.
Units and Concentration
The SI has seven base units: metre, kilogram, second, ampere, kelvin, mole, and candela. Since 2019 all seven are defined by fixed constants.
| Expression | Definition | Comment |
|---|---|---|
| Molarity, M | mol solute / L solution | Changes slightly with temperature |
| Molality, m | mol solute / kg solvent | Temperature-independent; used for colligative properties |
| Formality, F | formula weights / L | Used for salts without regard to dissociation |
| Normality, N | equivalents / L | Depends on the reaction; IUPAC discourages it |
| Mass percent, % w/w | g solute / 100 g sample | Concentrated reagents |
| ppm, ppb | , mass fractions | In dilute water, 1 ppm is about 1 mg/L and 1 ppb about 1 µg/L |
| p-function | pH, pCa, pAg |
Concentrated reagent molarity:
For 37.2% HCl with density 1.19 g/mL, M. Concentrated (96%, 1.84 g/mL) is 18.0 M.
Unit conversion: 2.5 ppm Fe in water is 2.5 mg/L, which equals M.
Dilutions and Aliquots
For serial dilutions, multiply the dilution factors: three successive 1-to-10 dilutions give 1-to-1000.
Worked aliquot problem: A 0.4000 g iron ore sample is dissolved, reduced to , and diluted to 250.0 mL. A 25.00 mL aliquot needs 12.40 mL of 0.004000 M ().
- mol
- mol in the aliquot
- mol in the sample
- %Fe
Significant Figures
- Addition and subtraction: keep the fewest decimal places: .
- Multiplication and division: keep the fewest significant figures: .
- Logarithms: the digits after the decimal point in the log equal the significant figures in the number. M (two figures) gives pH 4.37; going back, pH 4.37 gives M.
- Exact numbers (stoichiometric coefficients, defined conversions) do not limit significant figures. Round only at the end, carrying at least one extra digit through intermediate steps.
Sampling
An analysis can be no better than its sample. The sampling chain runs from the population (lot) to a gross sample, then a laboratory sample, and finally the test portion actually analysed.
- Random sampling gives every portion an equal chance of selection. Systematic (grid or time-interval) sampling is practical but can miss periodic variation. Composite samples average a lot; discrete samples reveal its variability.
- Heterogeneous solids are reduced by coning and quartering or with a riffle splitter, and ground to a uniform particle size.
- Variance adds: . With 3% sampling and 1% method standard deviations, the overall value is . Halving the method error to 0.5% only lowers it to 3.0%, so collecting better samples is the priority.
- The number of samples needed for a target error is approximately .
- Preserve samples appropriately: acidify water to pH below 2 for metals, refrigerate for nutrients and BOD, and avoid contamination from containers.
Sample Preparation and Dissolution
| Approach | Used For | Notes |
|---|---|---|
| Drying and grinding | Solids | Report on a dry basis or state moisture separately |
| Acid dissolution (, , aqua regia) | Metals, alloys, carbonates | Aqua regia (3 HCl : 1 ) dissolves Au and Pt |
| Silicates | Extremely hazardous; needs calcium gluconate first aid | |
| Closed-vessel microwave digestion | Biological and environmental samples | Fast, retains volatile elements, low contamination |
| Wet ashing (/, ) | Organic matrices | only with present and a perchloric-rated hood |
| Dry ashing in a muffle furnace | Organic matrices | Simple, but volatile elements (Hg, As, Se) may be lost |
| Fusion with fluxes (, , ) | Refractory oxides, silicates | Molten flux at high temperature |
| Kjeldahl digestion (, catalyst) | Organic nitrogen | Converts N to before distillation |
Standards Organizations and Accreditation
- International: ISO (standards, including ISO/IEC 17025 for testing and calibration laboratories); IUPAC (nomenclature and terminology); BIPM (the SI); NIST (Standard Reference Materials); AOAC INTERNATIONAL (official food and agricultural methods); ASTM International; APHA, AWWA, and WEF (Standard Methods for the Examination of Water and Wastewater); the USP (pharmaceutical monographs); US EPA methods; and Eurachem/CITAC guides on method validation and measurement uncertainty.
- Philippines: the DTI Philippine Accreditation Bureau (PAB) accredits laboratories to ISO/IEC 17025. The DTI Bureau of Philippine Standards issues Philippine National Standards (PNS). The National Metrology Laboratory at DOST-ITDI maintains national measurement standards.
- Certified reference materials (CRMs) with traceable certified values are used to establish accuracy and metrological traceability; Section 1.3 covers the quality system.
Concentrated hydrochloric acid is 37.2% HCl by mass and has a density of 1.19 g/mL. What is its approximate molarity? (HCl = 36.46 g/mol)
12.1 M
37.2 M
16.3 M
10.2 M
A 0.4000 g iron ore sample is dissolved, reduced to Fe2+, and diluted to 250.0 mL. A 25.00 mL aliquot needs 12.40 mL of 0.004000 M KMnO4. What is the percentage of iron in the ore? (Fe = 55.845 g/mol; 1 MnO4- reacts with 5 Fe2+)
3.46%
6.92%
0.69%
34.6%
In a soil survey, the sampling standard deviation is 3.0% and the analytical method's standard deviation is 1.0%. Which change most improves the precision of the overall result?
Run every test portion in triplicate on the same instrument
Replace the method with one whose standard deviation is 0.5%
Report more decimal places in the final result
Improve the sampling with more or larger increments