4.2 Equipment & Troubleshooting

Key Takeaways

  • Oxygen analyzers are calibrated daily at two points: room air (21%) and 100% O2; galvanic fuel cell, polarographic (Clark electrode), and paramagnetic types appear on the exam
  • Thorpe tube flowmeters: pressure-compensated reads accurately despite back pressure; non-compensated reads falsely HIGH with back pressure
  • Bourdon gauge flowmeters are gravity-independent (work in any position) for transport but read falsely high with back pressure and miss downstream obstructions
  • Cylinder duration = (gauge PSI x cylinder factor) / flow (LPM); subtract a 200 PSI safe residual for the strict formula
  • Memorize cylinder factors: D = 0.16, E = 0.28, G = 2.41, H/K = 3.14; a full cylinder reads about 2200 PSI
  • Ventilator and equipment alarms are worked patient-first: assess the patient, manually ventilate if in distress, then circuit, then machine
  • Suction: Yankauer (rigid, oropharyngeal), suction catheter sized at no more than half the ETT internal diameter, and closed in-line systems that preserve PEEP and FiO2
  • CPAP/BiPAP troubleshooting centers on mask leak (most common), skin breakdown, aerophagia, claustrophobia, eye irritation, and tubing rainout
Last updated: June 2026

Oxygen Analyzers, Flowmeters, and Regulators

Equipment quality control and troubleshooting form a dedicated scored section of the TMC exam, and the items are intensely practical: you calibrate a device, read a flowmeter, calculate how long a tank lasts, or decide what to do when an alarm sounds. Memorizing the device physics turns these into easy points.

Oxygen Analyzers

An oxygen analyzer measures the fraction of inspired oxygen (FiO2) delivered to the patient.

TypeMechanismResponseMaintenance
Galvanic fuel cellElectrochemical reaction self-generates voltage proportional to O210-30 sReplace cell every 6-12 months
Polarographic (Clark electrode)External battery drives current proportional to O2FasterBattery + electrode upkeep
ParamagneticExploits oxygen's paramagnetismVery fastBuilt into ventilators/analyzers

Two-point calibration is performed daily or with each new patient: the low point is room air (21%) and the high point is 100% O2. If the analyzer still reads inaccurately after calibration, replace the sensor — a fuel cell that cannot be calibrated is depleted. Galvanic cells are valued for needing no external power; polarographic electrodes respond faster but need batteries.

Flowmeters and Regulators

Thorpe tube flowmeters use a tapered glass tube with a floating ball or bobbin that rises with flow; read at the center of the ball (or the top of a bobbin). The critical exam distinction is compensation:

  • Pressure-compensated: the needle valve sits downstream of the float tube, so the tube always sees 50 psi and reads accurately even with back pressure from a nebulizer or other resistance.
  • Non-compensated (uncompensated): the valve sits upstream, so downstream back pressure makes the float read falsely HIGH — the patient actually receives less than the dial shows.

A quick bedside test: with the flowmeter set to zero and plugged into a 50-psi outlet, a pressure-compensated unit shows a momentary surge of the ball before settling at zero; a non-compensated unit does not.

Bourdon gauge flowmeters use a fixed orifice with a pressure gauge calibrated in LPM. They are gravity-independent and work in any position, making them ideal for transport and ambulances. Their weaknesses are tested: they read falsely high with back pressure and cannot detect a downstream obstruction (a kinked line still shows flow on the gauge).

Cylinders, Suction, and Systematic Troubleshooting

Gas Cylinder Calculations

The duration formula is a guaranteed exam item:

Minutes Remaining = (Gauge PSI x Cylinder Factor) / Flow (LPM)

SizeFactorCapacity (full at ~2200 PSI)
D0.16356 L
E0.28622 L
G2.415,300 L
H/K3.146,900 L

Worked example: an E-cylinder reads 1,600 PSI and the patient is on 4 LPM nasal cannula. Minutes = (1,600 x 0.28) / 4 = 448 / 4 = 112 minutes (about 1 h 52 min). The strict version subtracts a 200 PSI safe residual first: (1,600 - 200) x 0.28 / 4 = 98 minutes. Always change cylinders before they reach ~200 PSI so the regulator never reads empty and you keep a safety margin during transport.

Suction Equipment

EquipmentUseKey Feature
Wall (piped) suctionICU, patient roomsContinuous or intermittent; most powerful
Portable suctionTransport, fieldBattery-powered; weaker
Yankauer tipOropharyngeal secretionsRigid, large-bore
Suction catheterTracheal suction via ETTFlexible; size <= half the ETT internal diameter
Closed in-line systemVentilated patientsMaintains PEEP/FiO2; lowers infection risk

To limit hypoxemia and mucosal trauma, preoxygenate, keep each pass under 10-15 seconds, and apply suction only on withdrawal. Adult wall vacuum is typically -100 to -150 mmHg.

CPAP/BiPAP Troubleshooting

ProblemCauseSolution
Mask leak (most common)Poor fit, mouth breathingRefit/resize, chin strap, full-face mask
Skin breakdownExcess pressure, poor fitLoosen straps, add cushion/liner, rotate interface
AerophagiaPressure too high; air swallowingLower pressure, reposition, reassess settings
ClaustrophobiaInterface discomfortNasal pillows, desensitization, reassurance
Eye irritationTop-of-mask leak into eyesReseat at nasal bridge, tighten upper straps
RainoutTubing condensationAdjust humidifier temp, use heated tubing

The Patient-First Troubleshooting Sequence

For any alarm or malfunction, follow this order — the most heavily tested concept in the section:

  1. Assess the patient FIRST — distress, color, SpO2, chest rise.
  2. Manually ventilate with a bag-valve device if a ventilated patient is in distress; disconnecting from a suspect machine is acceptable and safe.
  3. Check the circuit — disconnections, kinks, water, leaks.
  4. Check the machine — alarms, settings, displays.
  5. Correct the identified problem.
  6. Document the event, findings, and interventions.

Never silence an alarm before identifying its cause; a high-pressure alarm may signal a mucus plug or kink, and a low-pressure or apnea alarm may signal a disconnection or accidental extubation.

Oxygen Cylinder Capacities (Liters at Full 2200 PSI)
Test Your Knowledge

An E-cylinder reads 900 PSI. A patient requires oxygen at 3 LPM for transport. Using the standard formula (no residual subtraction), how long will the cylinder last?

A
B
C
D
Test Your Knowledge

A Bourdon gauge flowmeter is preferred over a Thorpe tube flowmeter for patient transport because:

A
B
C
D
Test Your Knowledge

A pressure-compensated Thorpe tube flowmeter differs from a non-compensated Thorpe tube in that it:

A
B
C
D
Test Your Knowledge

When troubleshooting a ventilator alarm, what should the respiratory therapist do FIRST?

A
B
C
D
Test Your KnowledgeFill in the Blank

The cylinder factor for an E-cylinder is _____, and for an H-cylinder it is _____.

Type your answer below

Test Your Knowledge

An oxygen analyzer is calibrated using which two reference points?

A
B
C
D
Test Your Knowledge

A CPAP patient complains of air leaking into their eyes during sleep. The BEST solution is to:

A
B
C
D