15.2 Mechanical Circulatory Support (LVAD), Pacemakers & Home Ventilators

Key Takeaways

  • A patient with a continuous-flow LVAD can be adequately perfused without a palpable pulse. Assess perfusion and use Doppler if routine cuff measurement fails.

  • In an unresponsive patient with a durable LVAD and impaired perfusion, start chest compressions promptly. Have a second rescuer check the device concurrently when available.

  • Avoid placing external defibrillation pads directly over an implanted generator, and do not delay an indicated shock. A magnet response is device- and programming-specific.

  • In tracheostomy distress, oxygenate and check for obstruction or displacement before applying positive pressure through an uncertain tube.

  • A total laryngectomy separates the upper airway from the lungs. Oxygenation and ventilation must occur through the neck stoma.

Last updated: November 2026

Durable LVADs, rhythm devices and assisted ventilation at home

Identify the device and its usual function, ask what changed and obtain the patient’s emergency plan when available. A caregiver may know normal settings or how to locate spare equipment. Assign device checks to a second clinician when possible so that assessment and life support continue.

Assess perfusion with a durable LVAD

A durable left ventricular assist device (LVAD) pumps blood from the left ventricle into the arterial circulation. A patient with continuous flow may have a weak or absent pulse even when adequately perfused. Low pulsatility can also make automated blood pressure and pulse oximetry readings unreliable. Assess color, temperature, capillary refill, responsiveness and breathing. If an automated cuff does not read, a manual cuff with Doppler can provide a useful opening pressure. Document the method and result and interpret them with the clinical findings. Doppler opening pressure often approximates MAP in low pulsatility but is not automatically equivalent to invasive MAP. Ask the VAD team about patient-specific targets. AHA special circumstances guidance · HeartMate 3 Doppler pressure study

For an unresponsive patient with a durable LVAD, support ventilation as needed and assess perfusion promptly. The 2025 AHA algorithm lists reassuring findings including normal skin color and temperature, normal capillary refill, MAP above 50 mmHg when measurable, or waveform end-tidal CO₂ above 20 mmHg when available. A pump hum indicates device activity and alone does not establish adequate perfusion. When perfusion is adequate, continue reassessment and treat other causes of unresponsiveness. When perfusion is impaired, start chest compressions promptly with BLS or ALS as indicated. Do not wait for a silent pump, an unavailable measurement or a fixed troubleshooting interval. A second rescuer can check device-related causes concurrently. AHA adult and pediatric durable LVAD algorithm

Inspect the controller alarms, power cables and driveline connections under the model-specific plan. Do not disconnect both external power connections for troubleshooting. Connect reliable power if it has been lost. A controller exchange should follow manufacturer instructions and specialist direction within the service scope. Contact the VAD center promptly and bring the spare equipment. Also consider bleeding, hypovolemia, right-heart failure, infection, dysrhythmia and other causes of poor perfusion. Abbott HeartMate 3 patient guide

Manage pacemaker and ICD emergencies

A pacemaker provides pacing impulses, while an implantable cardioverter-defibrillator (ICD) can treat certain ventricular arrhythmias. Assess the patient, monitor the rhythm, and compare suspected electrical capture with mechanical circulation. Pacing artifacts without the expected subsequent depolarization suggest failure to capture. Treat instability under the applicable bradycardia or arrest pathway and seek device advice.

Treat VF or pulseless VT promptly. Avoid placing pads over an implanted generator. The 2025 Resuscitation Council UK guidance advises placing the pad more than 8 cm from the generator or using an alternative position. Follow local procedures and avoid delaying a necessary external shock. For suspected inappropriate ICD shocks, a magnet can suspend shock therapy on some devices. Effects vary with programming and device. Use only under an authorized monitored pathway with external defibrillation ready. RCUK adult ALS guidance · EHRA electrical storm consensus

Restore ventilation after a tracheostomy or home ventilator problem

Use the patient-specific emergency plan and identify whether the upper airway connects to the lungs. Look for displacement, obstruction, pneumothorax and equipment failure. Assess the patient and breathing first. Check the alarm and circuit, oxygen and power. Keep alternative ventilation available. If the ventilator fails and the airway is confirmed patent, use the appropriate backup ventilation with a suitable interface and reassess chest rise and capnography if available. ResMed Astral clinical guide

When the tube may be obstructed or displaced, avoid forcing breaths through it. Give oxygen at the stoma and also at the face if upper-airway continuity is possible or uncertain. Remove an external cap or speaking valve and a removable inner cannula as appropriate, then check whether a suction catheter passes. If it cannot pass, deflate a cuff if present and reassess. If the patient does not improve, trained responders remove the failed tube under the emergency plan and oxygenate or ventilate by a confirmed route. Avoid blind reinsertion into a fresh or uncertain tract. Get airway expertise and use the authorized rescue pathway. A total laryngectomy means the face is disconnected from the lungs; provide oxygen and ventilation at the neck stoma. NTSP tracheostomy emergency algorithm · NTSP laryngectomy algorithm · DHP Paramedic scopes

Loading diagram...
Initial decision for an unresponsive patient with a durable LVAD
Test Your Knowledge

An adult with a continuous-flow LVAD is awake and speaking but feels lightheaded after vomiting. Skin is warm with normal refill. No radial pulse is palpable and the automated cuff does not read. What should the crew do?

A

Start CPR because a radial pulse is absent

B

Check Doppler opening pressure if available, reassess perfusion and investigate causes of symptoms

C

Disconnect both power connections to reset the controller

D

Wait for the automated cuff to provide a value before assessing the patient

Test Your Knowledge

A patient with an implanted ICD remains in ventricular fibrillation arrest. The generator is below the left clavicle. Under 2025 Resuscitation Council UK guidance, how should the crew place the pads?

A

Place one pad directly over the generator

B

Withhold external defibrillation because a device is implanted

C

Use a pad position more than 8 cm from the generator or an appropriate alternative position without delaying the shock

D

Apply a magnet over the device before preparing an external shock

Test Your Knowledge

A ventilator-dependent adult with a tracheostomy has severe distress and a high-pressure alarm. The upper airway is known to be patent. The crew removes the attachments and inner cannula but cannot pass a suction catheter. Deflating the cuff does not improve breathing. Under the emergency plan, what should the trained crew do next?

A

Remove the failed tube under the plan and restore oxygenation or ventilation by a confirmed route

B

Force manual ventilation through the tube to clear the resistance

C

Reconnect the ventilator and increase pressure through the same tube

D

Seal the upper airway and wait for the alarm to stop