23.2 Intraoral Projection, Tube-Shift Localization, and Panoramic Imaging
Key Takeaways
For paralleling periapicals, position the receptor parallel to the tooth and direct the central ray perpendicular to both. If placement is limited, consider an alternative projection.
Use the direction of tube movement and a fixed anatomical reference for parallax. Same-direction relative movement indicates lingual or palatal position.
Check overlap, foreshortening, elongation and cone-cutting before diagnosing from an intraoral image. Backward receptor artifacts depend on the receptor.
Panoramic distortion, air spaces and ghost images can mimic disease or hide anatomy. Follow the unit's positioning instructions.
Use CBCT for a justified three-dimensional question and an appropriate field of view when lower-dose images cannot supply the needed information.
Intraoral Projection Geometry
For a paralleling periapical, place the receptor parallel to the tooth's long axis and direct the central ray perpendicular to both. An alignment holder helps reproducibility. Moving the receptor farther from the tooth can make parallel placement easier, but it increases magnification. A longer source-to-object distance reduces the magnification and geometric blur caused by that separation. The geometry does not depend on one fixed position-indicating device length. Paralleling usually reduces length distortion and is useful when comparing crestal bone levels or working-length views. It can reduce superimposition of the zygomatic process in maxillary molar views. Superimposition can still occur.
If anatomy or patient tolerance prevents parallel placement, a bisecting-angle projection is one option. Place the receptor near the tooth and aim the central ray perpendicular to the imaginary bisector of the tooth–receptor angle. Length and shape are more sensitive to positioning. A bitewing should show the relevant crowns, approximal contacts and crestal bone. Receptor coverage, occlusal orientation and horizontal alignment through the contacts matter. An occlusal projection places the receptor between the occlusal surfaces and directs the beam through the region of interest. It can give a wider intraoral view for selected localization or trauma questions. IAEA dental report, sections 2.1 and 5.1 · Original comparison of projection geometry · Periapical lesion simulation
For example, in an idealized projection with a source-to-receptor distance of 400 mm and a source-to-object distance of 380 mm, the magnification factor is 400 ÷ 380 ≈ 1.05, about 5%. With the same 20 mm object–receptor separation but a source-to-receptor distance of 200 mm, it is 200 ÷ 180 ≈ 1.11, about 11%. Compare geometry before interpreting small differences in length. IAEA diagnostic radiology physics
| Error | Clue and correction |
|---|---|
| Vertical angulation in bisecting-angle imaging | A beam that is too steep can foreshorten teeth. An insufficient angle can elongate them. Recheck the receptor–tooth angle and aim perpendicular to its bisector. |
| Horizontal angulation | Approximal surfaces overlap when the beam misses the contact openings. Align through the contacts of interest. |
| Cone cut | A sharply bounded unexposed region indicates that part of the receptor lay outside the beam. Center the beam on the receptor and use a compatible holder. |
| Backward receptor | Reversed film can show a lighter image and the backing foil's pattern. Reversed phosphor plates may display reversal or a device-specific marker/artifact. Check orientation for the actual receptor. |
Assess whether an error obscures the region needed. Repeat the image only when the missing information is clinically important. Clinical film and phosphor plate error study · Film foil artifact study · Phosphor plate artifact study
Tube Shift and the SLOB Rule
On two justified projections, change the position of the source relative to the patient and compare the object's apparent movement with a fixed adjacent landmark. Use the direction the tube moved, not the direction the central ray points after angulation.
- Same lingual (or palatal). When the tube shifts distally and the object shifts distally relative to the landmark, the object lies lingual or palatal to it.
- Opposite buccal (or labial). When the tube shifts distally and the object shifts mesially relative to the landmark, the object lies buccal or labial to it.
For example, if the crown of unerupted 13 moves mesially relative to 12 after a distal tube shift, the crown lies labial to the reference. This establishes relative depth. It does not determine root resorption or the surgical approach. A shift can also separate projected roots or canals in endodontic views. The second mesiobuccal canal in a maxillary molar is called MB2. Clark's original localization method · Clinical parallax localization study · Endodontic imaging guidance
Panoramic Image Layer and Artifacts
In panoramic imaging the source and receptor move around the patient. The image layer is the curved region reproduced with acceptable sharpness. Structures outside it can be blurred or distorted. The anterior part is particularly sensitive to positioning. An object can create a real image, a duplicated real image when intersected twice in the appropriate part of the projection, or a ghost image. A typical ghost is projected toward the opposite side, higher, more blurred and enlarged. The pattern depends on the object's position and the acquisition program. Review external objects and normal anatomy before labeling an opacity as pathology. Clinical positioning study · Original ghost-image case · Study of panoramic program artifacts
| Finding | Possible positioning error or preparation issue |
|---|---|
| Exaggerated smile curve | Chin too low. Correct the vertical orientation using the unit's reference guides. |
| Flat or reversed curve and hard palate over upper apices | Chin too high. Reposition according to the unit's vertical reference. |
| Narrow, blurred anterior teeth | Patient too far forward. Check incisor placement on the bite block or other positioning aid. |
| Wide, blurred anterior teeth | Patient too far back. Check the anterior positioning reference. |
| Dark band over maxillary apices | The tongue was not against the palate, allowing a palatoglossal air space. Ask the patient to hold the tongue to the palate during acquisition. |
| Dense midline band over the anterior jaws | The patient may be slumped and the cervical spine projected over the region. Check upright positioning. |
| Unequal left and right widths | Head rotation can make one side appear wider. Check the midsagittal position before assigning true asymmetry. |
| Unexpected dense opacity in the path of the beam | Consider removable metal, external objects or shielding that entered the field. Follow the equipment instructions and applicable protection protocol. |
Use the unit's positioning aids and instructions. Evaluate whether the image answers the clinical question before deciding to repeat it. Dentsply Sirona Orthophos S instructions · Clinical positioning error study
When Three-Dimensional Imaging Adds Value
CBCT provides multiplanar reconstructions. Movement, scatter and metal can cause artifacts and measurements have limits. Choose the smallest useful field and the resolution that answers the question. Use two-dimensional views where they are adequate. Examples of justified questions include the position of an unerupted tooth when conventional views are inadequate for treatment, cross-sectional bone anatomy for implant planning and selected endodontic problems when clinical assessment and intraoral radiographs are inconclusive. A suspected vertical root fracture may warrant limited-field CBCT in that setting. A negative scan does not exclude a fracture. For an impacted mandibular third molar, panoramic signs of nerve proximity prompt case assessment. Reserve CBCT for a specific question that conventional views cannot answer and that can affect care. Routine caries screening or an asymptomatic baseline does not justify CBCT. 2026 ADA–AAOMR selection guidance · AAE–AAOMR endodontic position statement · EADMFR third-molar imaging position
For tissue beyond the dentoalveolar structures, choose the modality with the specialist. Medical multidetector CT can show wider anatomy and has better soft-tissue contrast than dental CBCT. MRI uses no ionizing radiation and can demonstrate the temporomandibular joint disc when internal derangement is the question. Consider the indication, contraindications and existing images. IAEA dental report, section 2.5 · ADA–AAOMR imaging selection, temporomandibular joint recommendations
You need a reproducible periapical image of a maxillary molar and adequate space is available for a holder. Which setup uses the paralleling technique?
Receptor at an angle to the tooth and central ray perpendicular to the angle bisector
Receptor parallel to the tooth's long axis and central ray perpendicular to both
Receptor parallel to the occlusal plane and central ray directed obliquely for an occlusal view
Receptor parallel to the tooth's long axis and central ray aimed along that axis
On a second periapical view taken with a distal tube shift, the crown of unerupted 13 appears to move mesially relative to the root of 12. Which position does this support?
Palatal to the root of 12
In the same buccolingual plane as the root of 12
A superior position whose depth cannot be assessed despite the observed relative shift
Labial to the root of 12
A panoramic image shows a reversed occlusal curve and a broad hard-palate opacity that obscures the upper tooth apices. What most likely explains this pattern?
The chin was positioned too high
The chin was positioned too low
The tongue was held away from the palate
The anterior teeth were positioned too far forward
Sections you finish are checked off in the contents.