Donning/Doffing Instruction, Return Demonstration & Product Care

Key Takeaways

  • Patients must always insert the breast prosthesis into the internal bra pocket using flat finger pads before putting the bra on the body, and must never twist a tight fastened bra 180° around the waist across fragile chest wall scars.

  • Before completing any visit, the CMF must observe a live patient return demonstration (independent doffing, pocket re-insertion, donning, and fastening) to verify manual dexterity, shoulder ROM, and anatomical orientation skills.

  • Silicone prostheses must be hand-washed daily with mild non-moisturizing soap and lukewarm water, patted dry with a lint-free towel, and stored horizontally in their molded protective cradle; bras and compression sleeves must be washed without bleach or fabric softeners and air-dried flat.

  • Silicone forms typically carry a 2-year manufacturer warranty for defects in materials and workmanship; pin, nail, or claw punctures cannot be repaired and are not covered.

Last updated: October 2026

Instructing the Patient on Proper Donning and Doffing (DCO 3.B.1.c)

Even a biomechanically ideal mastectomy bra and prosthesis will fail clinically if the patient applies the device incorrectly at home, twists the prosthesis inside the pocket, or damages fragile chest wall skin while dressing. Under BOC Detailed Content Outline task 3.B.1.c, the fitter must provide clear, step-by-step verbal and physical instruction tailored to the patient's musculoskeletal mobility:

  • Inserting the Breast Prosthesis into the Bra Pocket: The patient must always insert the breast prosthesis into the internal bra pocket before putting the bra on the body. Attempting to push a weighted silicone form into the pocket while wearing the bra strains the glenohumeral joint, stretches the pocket opening, and frequently results in dropping and puncturing the prosthesis. Using both hands with flat finger pads—never pinching or clawing the silicone with fingernails—the patient slides the tapered edge gently through the pocket slit, confirms that the concave back faces the chest wall and any lateral extension points toward the axilla, and smooths the pocket fabric flat across the rear surface.
  • Donning the Mastectomy Bra & the "Scoop and Swoop" Technique:
    • Patients with Intact Shoulder Range of Motion (ROM): The patient places both arms through the shoulder straps, leans forward slightly (30°–45°) so the natural breast and pocketed prosthesis settle fully into the lower cup cradles, fastens the back hooks on the loosest column, stands upright, and performs a gentle "scoop and swoop" on the natural side—sweeping lateral breast tissue forward from the underarm into the cup while smoothing the prosthetic cup border flush against the sternum.
    • Patients with Restricted Shoulder ROM, Post-Op Precautions, or Neuropathy: Patients who cannot reach behind their back due to axillary dissection fibrosis, adhesive capsulitis, arthritis, or chemotherapy-induced peripheral neuropathy should be fitted with a front-closure mastectomy bra (hooks, Velcro, magnetic clasp, or zipper with an underlying starter hook) or taught the "step-in" method (fastening the hooks in front before donning, stepping into the bra, and sliding it gently up the legs and torso). Patients must be warned never to fasten a tight mastectomy bra backward at the waist and twist it 180° around the ribcage. Rotating a snug under-band across the chest creates severe abrasive shear over radiated or surgical scars, flips the prosthesis inside the pocket, and prematurely stretches out the band elastic.
  • Donning and Doffing Lymphedema Compression Sleeves:
    • Apply the sleeve first thing in the morning when limb edema is minimal, over clean, completely dry skin (moisturizers must be applied at bedtime so they absorb overnight; applying compression directly over wet lotion degrades elastane fibers and causes the top band to roll).
    • Wear textured rubber donning gloves (or use a low-friction fabric/wire donning butler) to grip and distribute the knit evenly along the limb without snagging fibers on fingernails or rings.
    • Never pull a compression sleeve up by its top band like a sock. Instead, fold the upper half of the sleeve inside-out down to the wrist crease, slide the folded cuff over the hand and wrist, and unroll the upper half smoothly up the forearm and humerus, using rubber-gloved palms to stroke out any wrinkles that could create a tourniquet effect. To doff, fold the top edge downward and peel the sleeve off inside-out toward the wrist.

Note

Mandatory Observation of Patient Return Demonstration (DCO 3.B.1.d): Under BOC DCO 3.B.1.d, verbal explanation and written brochures are never sufficient to complete a fitting. Before the patient leaves the fitting room, the CMF must have the patient independently remove (doff) the garment, take the prosthesis out of the pocket, re-insert the prosthesis in the correct anatomical orientation, re-apply (don) the bra or compression sleeve, and fasten all closures while the fitter observes. This live return demonstration verifies that the patient possesses the visual acuity, fine motor pinch strength, shoulder ROM, and cognitive sequencing required for safe home wear. If the patient struggles to orient an asymmetrical prosthesis or fasten back hooks during the return demonstration, the fitter must immediately adapt the plan—such as switching to a symmetrical form, marking the pocket orientation, selecting an adaptive front closure, or training an accompanying caregiver.

Daily Hygiene, Laundering & Storage Protocols (DCO 3.B.1.c)

Proper daily hygiene protects compromised post-mastectomy and post-radiation skin from bacterial or fungal maceration while preserving the structural integrity of silicone gels and medical elastics:

Product CategoryDaily Cleaning / Washing ProtocolDrying MethodStorage & Handling StandardStrictly Prohibited Actions
Standard Silicone Prosthesis (L8030 / L8035)Hand-wash daily with lukewarm water and mild, non-moisturizing liquid soap or manufacturer cleanserPat dry gently with a soft, lint-free towelStore horizontally inside the original molded storage cradle/case when not wornNever wring, machine-wash, tumble-dry, soak in bleach/alcohol, or expose to direct heat
Attachable Contact Prosthesis (L8031)Clean adhesive back daily with warm water, mild cleanser, and manufacturer's soft brush to remove skin oilsAir-dry adhesive surface (never blot adhesive with a lint towel)Place protective film liner over adhesive and rest in molded cradleNever use alcohol, acetone, or moisturizing soaps on the adhesive surface; never peel off chest dry without water/release spray
Pocketed Mastectomy Bra (L8000)Hand-wash or machine-wash gentle cycle inside a mesh lingerie bag with hooks fastened; mild detergentAir-dry flat or hang away from heatStore flat or folded without crushing molded cupsNever use bleach, fabric softeners, or dryer sheets (destroys spandex recovery); never put in a heated dryer
Lymphedema Compression Sleeve (Medicare codes such as A6578)Hand-wash or delicate wash daily/every other day in cool/lukewarm water with mild soap to restore knit tensionRoll in towel to blot excess water, then air-dry flatStore flat away from direct sunlight and radiatorsNever wring, twist, iron, bleach, use fabric softener, or dry in a heated clothes dryer

Preventing Polyurethane Film Punctures & Reviewing Manufacturer Warranties (DCO 3.B.1.e)

Definitive silicone breast prostheses (L8030, L8031, L8035) consist of medical-grade silicone gel encased within an ultra-thin, semi-permeable polyurethane outer membrane engineered to mimic human skin softness and thermal conductivity. The CMF must emphasize to every patient that if the outer polyurethane film is punctured, cut, or torn, the prosthesis cannot be repaired. Once breached, uncrosslinked sticky silicone gel leaks continuously from the opening, causing skin irritation, staining clothing, and permanently ruining the device.

To prevent accidental membrane failure, patients must strictly avoid:

  1. Sharp Fasteners & Jewelry: Safety pins, pin-back employee identification badges, lapel brooches, corsage pins, and sharp necklaces worn over the surgical chest wall.
  2. Fingernails & Rings: Clawing or pinching the prosthesis with long fingernails or pronged diamond rings when inserting it into a snug bra pocket.
  3. Household Pets: Allowing cats or dogs with sharp claws to jump onto the patient's chest or lap, or leaving an uncovered prosthesis on a bed or dresser where a pet can bite or scratch it.
  4. Topical Chemicals & Solvents: Applying perfumes, colognes, alcohol-based sprays, insect repellents, self-tanning lotions, suntan oils, or topical steroid/medicated creams directly onto the skin immediately before donning (all topical products must absorb and dry completely before the prosthesis touches the chest wall).
  5. Pool & Ocean Precautions: Standard silicone forms may be worn in chlorinated pools, hot tubs, or salt water if rinsed thoroughly with fresh lukewarm water immediately afterward, though dedicated swim prostheses are best for frequent swimmers. Reassure patients who travel that commercial airplane cabin pressure changes and high altitudes do not cause silicone breast prostheses to rupture (lightweight aerated forms may develop temporary microscopic air bubbles at altitude that disappear after returning to sea level).

Manufacturer Warranty Coverage vs. Patient Misuse

External silicone breast prostheses typically carry a 24-month (2-year) limited manufacturer warranty from the date of dispensing—matching Medicare's 2-year reasonable useful lifetime. This warranty covers manufacturing defects in materials and workmanship, such as spontaneous perimeter seam delamination, internal gel separation, or defective adhesive backing under normal use. Conversely, punctures from pins or pet claws, scissor cuts, ballpoint ink stains, dye transfer, and chemical degradation from perfumes or solvents are excluded as accidental damage or misuse. The fitter should record the manufacturer lot/serial number on the patient's warranty card and instruct the patient to return to the accredited facility if any defect arises.

Warning

Pin-Back Name Badges: A Leading Cause of Preventable Warranty Voiding: Nurses, teachers, retail staff, and conference attendees frequently pin an identification badge or brooch onto their blouse on the mastectomy side because they feel no pain when the pin pierces the prosthesis beneath the fabric. Even a microscopic pinhole in the polyurethane skin allows viscous silicone gel to ooze out under body weight, and the damage is not covered by the manufacturer warranty. Always instruct patients to wear magnetic badge holders, belt clips, or neck lanyards.

Test Your Knowledge

During the delivery of a definitive weighted silicone breast prosthesis (L8030) and a pocketed mastectomy bra (L8000), which action listed in the BOC Detailed Content Outline verifies that the patient can safely manage the devices at home?

A

Have the patient sign a written attestation stating she watched an online manufacturer instructional video prior to her appointment

B

Provide the patient with a laminated care brochure and schedule a telephone quiz 30 days after delivery

C

Watch her remove the bra and form, re-insert the form correctly, and put the bra back on and fasten it herself

D

Apply the bra and prosthesis for the patient and instruct her not to remove the garment until her two-week follow-up visit

Test Your Knowledge

A unilateral mastectomy patient with mild postoperative right shoulder stiffness asks the Certified Mastectomy Fitter for the easiest way to put on her back-hook pocketed mastectomy bra at home. Which instruction should the fitter provide?

A

Put the empty bra on, fasten it, then push the form up into the pocket from below

B

Leave the back hooks permanently fastened on the tightest column and pull the bra overhead like a tight crewneck sweater while holding the prosthesis in one hand

C

Put the form in the pocket first; if reaching back hurts, step into the pre-fastened bra or use a front-closure bra

D

Fasten the back hooks in front at the natural waistline, twist the snug band 180 degrees around the ribcage, and pull the straps up over the shoulders

Test Your Knowledge

Which daily care and laundering instruction should a Certified Mastectomy Fitter give a patient who is receiving both a standard silicone breast prosthesis (L8030) and a post-mastectomy lymphedema compression sleeve?

A

Hand-wash the form daily in mild soap, pat dry, and cradle it; wash the sleeve gently without bleach or softener and dry flat

B

Machine-wash the silicone form and the sleeve together on a heavy-duty cycle, then hang the form from a clothespin overnight to dry

C

Wipe the silicone prosthesis daily with 70% isopropyl rubbing alcohol to sterilize the polyurethane film, and tumble-dry the compression sleeve on high heat for 20 minutes to shrink the elastic fibers back to size

D

Soak the silicone prosthesis overnight in liquid fabric softener to keep the gel supple, and apply heavy body lotion to the arm immediately before pulling the compression sleeve up by its top silicone border

Test Your Knowledge

A patient returns to the mastectomy facility 8 months after receiving a standard silicone breast prosthesis (L8030) because clear, sticky silicone gel is oozing from a tiny hole on the anterior surface where she pinned her workplace security badge through her blouse. How does a typical 2-year manufacturer warranty apply to this situation?

A

The manufacturer will repair the puncture free of charge by vulcanizing a new polyurethane patch over the anterior surface

B

The puncture cannot be repaired and is not covered, because the warranty covers manufacturing defects, not accidental damage

C

The warranty automatically replaces any prosthesis punctured by pins, fingernails, or pet claws at no cost to the patient

D

The fitter should seal the pinhole in the office using cyanoacrylate superglue so the warranty remains active for the remaining 16 months