5.2 Test Instruments: Selection, Calibration & Accuracy

Key Takeaways

  • Each test requires a specific instrument: low-resistance ohmmeter (test current ≥0.2 A, resolution ≤0.01 Ω) for continuity, insulation resistance tester (selectable 250/500/1000 V DC), loop impedance tester, RCD tester, PFC tester, phase sequence indicator, and an approved two-pole voltage indicator for proving dead
  • GS38 requires fused probes, finger guards, tip length ≤4 mm, insulated leads, and CAT III/IV rating — a multimeter is not an approved voltage indicator
  • Annual calibration to a national standard is required AND the inspector must verify each instrument against a known reference before use — calibration alone is not sufficient
  • Test-lead resistance must be nulled or subtracted for low-resistance continuity measurements to avoid inflated R1+R2 readings
  • Instrument serial numbers and calibration dates must be recorded on the schedule of test results for traceability
Last updated: August 2026

Quick Answer: The 2391-52 requires a family of test instruments, each matched to a specific test. A low-resistance ohmmeter (test current ≥0.2 A, resolution ≤0.01 Ω) does continuity; an insulation resistance tester (selectable 250/500/1000 V DC) does insulation; a loop impedance tester does Ze and Zs; an RCD tester does RCD trip times; a PFC tester (often combined with the loop tester) does prospective fault current; and an approved two-pole voltage indicator proves dead. All leads and probes must comply with HSE GS38. Instruments need annual calibration and in-use verification before each job — calibration alone is not sufficient.

Instrument Categories and Their Tests

Each test in the sequence demands an instrument with the right range, resolution, and test current:

InstrumentTestRange / Resolution
Low-resistance ohmmeterContinuity of protective conductors (R1+R2), ring final continuity (r1, rn, r2), polarity0.01 Ω resolution, test current ≥0.2 A
Insulation resistance testerInsulation resistance; verification of SELV/PELV and electrical separation250 / 500 / 1000 V DC selectable, MΩ range
Earth fault loop impedance testerZe, Zs0.01 Ω resolution, 230/400 V range
PFC / PSCC testerProspective fault currentkA range; often combined with loop tester
RCD testerRCD operating time at IΔn (plus optional 0.5× and 5× diagnostics)ms resolution
Earth electrode resistance testerTT electrode resistance (dead method)Fall-of-potential, Ω range
Phase sequence indicatorCheck of phase sequence on polyphase circuits (Reg 643.9)Rotating-disc or electronic L1-L2-L3 indication
Approved voltage indicator (two-pole)Proving deadCAT III/IV, no reliance on internal battery alone

Instrument Notes

  • Low-resistance ohmmeter — must deliver a test current of at least 0.2 A to ensure the measured value is meaningful at the low resistances involved. Resolution of 0.01 Ω is required so that small changes in R1+R2 are detectable. Instruments with poorer resolution are not suitable for continuity testing to BS 7671.
  • Insulation resistance tester — selectable DC test voltage is essential because different circuit voltages require different test pressures. Applying 1000 V to a 24 V SELV circuit would damage it. The tester must indicate in megohms and flag circuits below the minimum insulation resistance.
  • RCD tester — connects between line and earth on the load side of the device and injects a controlled residual current. The verification BS 7671 requires is an alternating-current test at IΔn: a general (non-delay) type must operate within 300 ms, and a Type S must operate between 130 ms and 500 ms. Test at both the 0° and 180° start points and record the longer time. Most instruments also offer 0.5× and 5× settings and a ramp test that finds the actual tripping current; these are useful diagnostics but are no longer BS 7671 requirements.
  • Earth electrode resistance tester — for the dead fall-of-potential method. Uses current and potential spikes driven into the ground; the electrode under test is disconnected from the main earthing terminal. Live loop-impedance methods (Ze on a TT system) can also determine electrode resistance, since Ze on a TT system is effectively Ra plus the return path through the general mass of earth.
  • PFC tester — often combined with the loop impedance tester in a single instrument. Measures prospective short-circuit current (line-to-neutral) and prospective earth fault current (line-to-earth); the higher of the two is recorded as PFC.
  • Approved voltage indicator — a two-pole device that does not rely on a single internal battery. If the battery fails, the indicator must still show a reading or fail safe. A proving unit confirms the indicator is working before and after the proving-dead test (the live-dead-live sequence). A multimeter is not acceptable for proving dead because a wrong range selection can display a false zero.

GS38 Compliance for Leads and Probes

HSE Guidance Note GS38 governs the design of test leads and probes. The key requirements:

  • Fused probes (typically 0.5 A or 1 A fuse) to protect against short-circuit if the probe slips onto an adjacent conductor.
  • Finger guards to prevent accidental contact with live parts.
  • Tip length ≤4 mm to reduce short-circuit risk between adjacent conductors or terminals.
  • Insulated leads and probes, rated for the working voltage.
  • CAT-rated to the appropriate measurement category — CAT III or CAT IV for installation-level work.
  • Damaged leads must be replaced before use — a visual check is part of the pre-test routine.

GS38 is not advisory for the 2391-52 — the examiner treats it as a hard requirement. Leads that do not meet GS38 must not be used on an inspection.

Calibration vs In-Use Checking

This distinction is heavily examined. Calibration is the annual process of verifying the instrument against a national standard (typically UKAS-accredited) and issuing a calibration certificate. It confirms the instrument was accurate on the day it was calibrated.

In-use checking is what the inspector does before and during use, every job:

  • Zero the leads for continuity measurements — clip the test leads together and press zero, so lead resistance is subtracted from every subsequent reading.
  • Check the ohmmeter against a known resistor — confirm the reading is within tolerance.
  • Check the insulation tester on a known value — confirm it reads correctly.
  • Check the voltage indicator on a known live source, then a proving unit, then the known live source again (the live-dead-live method).

Calibration alone is not sufficient — an instrument can be in calibration but drift between annual checks, or have a damaged lead. Regular in-use verification catches these. The 2391-52 examiner regularly tests this distinction by asking whether an annual calibration certificate alone is enough to trust a reading — it is not.

Test-Lead Resistance Nulling

For continuity measurements, the resistance of the test leads themselves (typically 0.05–0.2 Ω) can be significant relative to the measured value of R1+R2. The leads must be nulled or subtracted — either by the instrument's auto-zero function or by measuring the leads separately and subtracting. Failing to null leads inflates R1+R2 values and can produce false fail readings, leading to unnecessary remedial work and an incorrect schedule of test results.

Battery Condition

Instrument accuracy depends on adequate battery voltage. Low batteries can cause false readings — particularly on insulation testers, where the output voltage may drop below the nominal test voltage and give a falsely high resistance reading. Check battery condition before use as part of the pre-test routine. Carry spare batteries to site.

Recording Instrument Details

The schedule of test results requires the inspector to record:

  • Instrument serial numbers — each instrument used on the job.
  • Calibration dates — from the current calibration certificate.
  • Instrument type or model — where the schedule form provides for it.

This provides traceability — the results can be defended against challenge by reference to the calibration certificate. An instrument without a valid calibration certificate must not be used on an inspection.

Summary

The 2391-52 examiner expects you to match each instrument to its test, know its range and resolution, understand GS38 lead requirements, and distinguish calibration from in-use checking. Recording serial numbers and calibration dates on the schedule of test results is not optional — it is part of producing a valid certificate.

Test Your Knowledge

What is the minimum test current and resolution for a low-resistance ohmmeter used for continuity testing to BS 7671?

A
B
C
D
Test Your Knowledge

Which statement correctly distinguishes calibration from in-use checking?

A
B
C
D
Test Your Knowledge

Which GS38 requirement applies to test probe tips used on electrical installations?

A
B
C
D