11.3 CO-oximetry / Hemoximetry Procedures
Key Takeaways
CO-oximetry (hemoximetry) uses multiple wavelengths to measure FO2Hb, FCOHb, FMetHb, FHHb, and total hemoglobin—unlike pulse oximetry SpO2 or PO2-based calculated SaO2.
Order or perform co-oximetry for smoke inhalation/CO exposure, unexplained low SpO2 with relatively preserved PO2, methemoglobin concerns, and therapy monitoring when dyshemoglobins matter.
Pre-analytical requirements mirror ABG anaerobic technique with adequate volume and mixing; hemolysis, clots, and air contamination threaten optical validity.
DCO maps co-oximetry to II.A.7 (select), II.B.7 (perform), and II.C.7 (validity); exam forms may include only a small maximum of co-ox analysis items (often about 2).
SpO2 cannot distinguish oxyhemoglobin from carboxyhemoglobin; a “normal” pulse ox does not rule out significant COHb.
Hemoximetry versus “the sat on the ABG”
CO-oximetry (also called hemoximetry) is a multiwavelength spectrophotometric measurement of hemoglobin species in a blood sample. On the NBRC PFT Detailed Content Outline it appears as a distinct procedure cluster: select co-oximetry (II.A.7), perform it (II.B.7), and evaluate validity (II.C.7). Form construction often limits pure co-ox analysis items (commonly discussed as a maximum of about 2 on a form), but RPFT candidates still need operational fluency because dyshemoglobin errors are classic clinical traps.
Do not confuse three different “saturations”:
| Term | What it is | Limitation |
|---|---|---|
| SpO2 | Pulse oximeter functional estimate using limited wavelengths | Misreads in CO exposure; motion/perfusion artifacts |
| Calculated SaO2 (many ABG printers) | Estimated from measured PO2 (±pH) | Assumes normal dyshemoglobin load; ignores COHb/MetHb |
| FO2Hb (oxyhemoglobin fraction) from CO-ox | Measured fraction of total Hb that is O2Hb | Requires proper hemoximetry sample and device |
Fractional versus functional saturation language appears on some reports: know that FO2Hb is a fraction of total hemoglobin species, while functional saturation excludes dyshemoglobins from the denominator—read the report legend rather than assuming labels are interchangeable.
Multiwavelength measurement: what is reported
A laboratory CO-oximeter passes light of multiple wavelengths through a hemolyzed or optically prepared blood sample and solves for concentrations of major hemoglobin species. Exam-level parameters:
| Parameter | Meaning |
|---|---|
| FO2Hb | Oxyhemoglobin fraction |
| FCOHb | Carboxyhemoglobin fraction |
| FMetHb | Methemoglobin fraction |
| FHHb | Reduced (deoxy) hemoglobin fraction |
| tHb / total Hb | Total hemoglobin concentration |
| Sometimes FHbF, sulfhemoglobin flags | Context-specific; follow device capability |
Because each species has a characteristic absorption spectrum, multiwavelength systems can separate O2Hb from COHb—something two-wavelength pulse oximeters cannot do. That is why smoke-inhalation assessment requires blood CO-oximetry, not SpO2 reassurance alone.
Many blood gas platforms include a co-ox module in the same cartridge or sample path; others are standalone. Either way, QC and calibration for the optical module must be in control (Domain I knowledge from Chapter 3) before patient results are released.
When CO-oximetry is required (selection — II.A.7)
Select hemoximetry when the clinical question involves hemoglobin species, not merely dissolved oxygen tension:
- Smoke inhalation / known or suspected carbon monoxide exposure — headache, cherry-red skin is unreliable; measure FCOHb.
- Unexplained low SpO2 with ABG PO2 that is not correspondingly low (or clinical suspicion of dyshemoglobin) — consider MetHb or COHb.
- Therapy monitoring — e.g., following COHb decline after oxygen therapy; monitoring methemoglobinemia treatment when ordered.
- Discrepancy workups — SpO2 versus calculated SaO2 versus clinical cyanosis that does not fit PO2.
- Orders specifying co-ox panel — do not substitute “ABG only” when the order or indication needs fractions.
Selection trap: Running only PO2/calculated sat in a fire victim and reporting “oxygenation adequate” because PO2 is normal. PO2 can be normal while FO2Hb is reduced by COHb occupying heme sites.
Performing the procedure (II.B.7)
Pre-analytical sample for hemoximetry
- Prefer arterial samples when simultaneous blood gases are needed; venous co-ox may be acceptable for COHb/MetHb screening in some protocols because those dyshemoglobins are not primarily diffusion-limited the way PO2 is—but follow lab policy and the clinical order.
- Use the same anaerobic, heparinized, adequately filled, promptly analyzed standards as ABG when gases are also reported.
- Mix thoroughly immediately before analysis; co-ox measures Hb species in whole blood—settling causes error.
- Ensure volume meets co-ox minimum (often higher than a minimal pH/gas-only aspiration).
- Avoid prolonged tourniquet and traumatic draws that hemolyze when possible.
Instrument steps (exam-level)
- Confirm co-ox module ready (wavelength checks/QC per manufacturer).
- Identify sample type in software if prompted.
- Introduce well-mixed specimen without bubbles/clots.
- Review fraction sum approximately totaling ~100% (device-dependent presentation) and flag messages.
- Report FO2Hb, FCOHb, FMetHb, FHHb, tHb as ordered; attach clinical context (e.g., smoke exposure).
Smokers may have mildly elevated baseline COHb (often a few percent); toxic exposures are much higher—know that interpretation cutoffs are clinical and lab-defined, while the technologist’s job is accurate measurement and validity.
Validity threats (II.C.7)
| Problem | Effect on co-ox / related gases | Action |
|---|---|---|
| Air contamination | Biases PO2/PCO2 if gases run; optical fractions may be less air-sensitive than PO2 but sample still compromised for combined panels | Expel air; redraw if gases needed |
| Clots | Incomplete aspiration, path occlusion, erroneous tHb/fractions | Reject; redraw; flush path per SOP |
| Hemolysis | Optical interference; potassium elevation if chemistry linked | Evaluate flags; redraw if results inconsistent |
| Incomplete mixing | Non-representative tHb and fractions | Remix and repeat |
| Lipemia / extreme turbidity / dyes (device-specific) | Spectral interference | Follow manufacturer limitations; alternative method |
| Out-of-control optical QC | All fractions suspect | Repair/calibrate; do not report patient co-ox |
Validity vignette pattern: Normal SpO2, patient from house fire, ABG PO2 normal, no CO-ox performed → incomplete testing. Conversely, CO-ox with clot flag and wild tHb → do not release; fix pre-analytics.
Integrating sampling, ABG, and co-ox on the exam
Think of one clinical story across Chapter 11:
- 11.1 — Obtain an anaerobic arterial sample safely (Allen test, technique, heparin, complications).
- 11.2 — Run pH, PCO2, PO2 on an in-control 37 °C analyzer; treat HCO3/BE/calculated sat as calculated; verify criticals.
- 11.3 — Add multiwavelength fractions when dyshemoglobins or total Hb species matter.
Domain II verbs stay constant: select → perform → validate. Instrumentation QC remains Domain I; clinical pattern interpretation expands in Domain III.
For timed mixed practice across PFT procedures, use /practice/rpft.
Which set of values requires multiwavelength CO-oximetry (hemoximetry) rather than a standard ABG electrode panel alone?
pH, PCO2, and PO2
Calculated HCO3− and base excess only
FO2Hb, FCOHb, FMetHb, and total hemoglobin fractions/concentration
Pulse oximeter SpO2 averaged over one minute
A patient rescued from a structural fire has SpO2 98% on a pulse oximeter and ABG PO2 92 mm Hg on oxygen. Which action best matches RPFT selection of procedures?
Conclude carbon monoxide poisoning is excluded because SpO2 is normal
Perform or ensure CO-oximetry for FCOHb (and related fractions) because SpO2 and PO2 can miss COHb
Report only calculated SaO2 from the ABG as proof of normal oxyhemoglobin fraction
Cancel blood testing and rely on cherry-red skin examination alone
Before running a co-oximetry sample, which pre-analytical step is most important for valid hemoglobin fraction results?
Vigorously shake until foamy to fully oxygenate the specimen
Leave the sample unmixed so red cells settle and plasma is analyzed preferentially
Add extra liquid heparin until the syringe is completely full regardless of design volume
Gently remix the anaerobic, adequately filled sample and exclude clots or gross hemolysis per policy
On the NBRC PFT content outline, evaluating whether a co-oximetry result is acceptable to report is best classified as which Domain II task family?
II.A.7 Select only—validity is not tested
II.C.7 Evaluate validity of co-oximetry / hemoximetry results
I.A instrumentation setup exclusively, with no procedure component
III.C clinical interpretation of COPD severity only
Sections you finish are checked off in the contents.