Free RPFT Exam Flashcards
Memorize 50 essential terms and definitions for the NBRC Registered Pulmonary Function Technologist (RPFT). See the term, recall the definition, then flip to check yourself.
What maximum permissible error must a spirometer meet under the 2019 ATS/ERS standard, and what tolerance do you actually apply at the bench?
The device must be within +/-2.5% when tested with a 3-L syringe using the ISO 26782 profiles. The syringe itself is allowed +/-0.5%, so the combined daily verification tolerance you apply is +/-3%. On a 3-L syringe that means 2.91 to 3.09 L. The older +/-3% device figure came from the 2005 standard.
Filter by Topic
Jump to Card
About These RPFT Flashcards
These 50 flashcards are designed to help you memorize key terms and definitions for the NBRC Registered Pulmonary Function Technologist (RPFT). Each card shows a term on the front and its definition on the back—the classic flashcard format for vocabulary memorization. Use these alongside our practice questions to build both recall and comprehension.
Topics Covered
Complete Flashcard Reference
Review every term in this set. Open any term to reveal its definition.
What maximum permissible error must a spirometer meet under the 2019 ATS/ERS standard, and what tolerance do you actually apply at the bench?
The device must be within +/-2.5% when tested with a 3-L syringe using the ISO 26782 profiles. The syringe itself is allowed +/-0.5%, so the combined daily verification tolerance you apply is +/-3%. On a 3-L syringe that means 2.91 to 3.09 L. The older +/-3% device figure came from the 2005 standard.
What accuracy must a 3-L calibration syringe hold, and why should you avoid gripping its barrel?
Accuracy must be +/-0.015 L or +/-0.5% of full scale, verified by the manufacturer on delivery and at the recommended intervals. The syringe must sit at room temperature. Holding the barrel to steady it warms the trapped gas, expands it, and adds measurement error, so support the syringe rather than wrapping your hand around it.
Why do gas analyzer circuits use absorbers such as Drierite or Nafion tubing?
They remove water vapor, and CO2 absorbers remove carbon dioxide, before sample gas reaches the analyzer. Infrared and thermal conductivity analyzers read water vapor and CO2 as interfering gases, so a saturated or CO2-rich sample gives a falsely shifted reading. Exhausted absorber (Drierite turns from blue to pink) is a routine cause of analyzer drift.
A patient with severe kyphoscoliosis cannot stand for a stadiometer height. What do you measure instead?
Use arm span or ulnar length and convert to an estimated standing height. Reference equations are driven by height, so using a compressed standing height would inflate the percent predicted and hide a restrictive defect. Record on the report that height was estimated and by which method, because it changes every predicted value.
When must the spirometry system perform a zero-flow setting?
Before calibration, before every calibration verification, and before patient tests, with the spirometer blocked so no gas moves. If flow varies during the procedure or the zero level has shifted noticeably, repeat it. A drifting zero adds a false baseline flow that accumulates into volume error across the whole maneuver.
Daily verification passes at low flow but fails at high flow. What does that pattern point to?
A linearity or flow-dependence problem rather than a simple gain error. Suspect a partially occluded or damaged flow sensor, debris or condensate on a screen, or a filter that was fitted for verification but not for patient testing. This is exactly why the daily check cycles the syringe at least three times across a range of flows instead of once.
During a nitrogen washout the expired N2 concentration suddenly climbs instead of falling. What do you check?
A leak. Expired N2 should fall smoothly breath by breath. A sudden large rise, or any change in the inspired tracer concentration, means room air is entering the circuit, usually at the mouthpiece seal, the nose clip, or a valve, and the 2023 ERS/ATS standard says to stop the test. The leak dilutes the washout and falsely raises the calculated FRC. Reseal and wait at least twice the washout time before repeating.
In closed-circuit helium dilution, the helium concentration drops abruptly and never plateaus. What has happened, and which way is FRC wrong?
A system leak is admitting room air, which dilutes the helium further. Because FRC is calculated from how far the helium was diluted, the extra dilution makes FRC read falsely high. Contrast this with the classic dilution error in obstruction, where poorly ventilated units are never reached and FRC reads falsely low.
A blood gas analyzer passes its one-point calibration but fails the two-point check. What is the difference between them?
One-point calibration sets the offset using a single known gas or buffer. Two-point calibration also sets the slope, using a second value at the other end of the range. Passing one point but failing two means the electrode responds but its gain has drifted, which typically signals an aging membrane or depleted electrolyte rather than a simple zero shift.
Plethysmograph box pressure drifts steadily during quiet breathing before the shutter closes. What is the usual cause?
Thermal disequilibrium. Body heat and warm expired gas raise cabin temperature until the box is equilibrated with the door closed, and the resulting pressure drift corrupts the volume signal. Allow the stabilization period with the patient seated and the door shut, and confirm a stable end-expiratory level over several breaths before closing the shutter.
A smoke-inhalation patient looks cyanotic but the pulse oximeter reads SpO2 98%. Why, and what measures the truth?
A two-wavelength pulse oximeter cannot distinguish carboxyhemoglobin from oxyhemoglobin, so COHb is counted as saturated and SpO2 reads falsely high. Only multi-wavelength CO-oximetry (hemoximetry) on a blood sample reports the true oxyhemoglobin fraction along with COHb and methemoglobin.
You suspect the CO analyzer in a DLCO system is nonlinear. How do you demonstrate it?
Run a serial dilution check: deliver known fractions of the test gas, plot analyzer output against the expected concentration, and look for departure from a straight line. The 2017 ERS/ATS DLCO standard calls for this linearity check monthly, separately from the daily zero and span. Nonlinearity biases DLCO in a concentration-dependent way, so an analyzer can pass a single-point check and still misreport patients.
State the full daily spirometer verification procedure required by the 2019 standard.
Use a 3-L syringe cycled at least three times to give flows spanning roughly 0.5 to 12 L/s, which corresponds to 3-L injection times of 0.5 to 6 seconds. The measured volume at each flow must be within +/-3%, for both inspiration and expiration on flow-based devices. Verification is required daily even on factory-calibrated devices that cannot be field-adjusted.
What does a biological control add that a syringe verification cannot?
A syringe tests only volume accuracy on a rigid, dry, room-temperature signal. A biological control is a healthy repeat subject with known stable values, so it also exercises BTPS correction, software, reference equation setup, sensor hygiene, and operator technique end to end. Track the subject's results on a chart and investigate when a value falls outside the established range.
What is an isothermal lung analog, and which instrument does it verify?
It verifies the body plethysmograph, not the DLCO system. A rigid container of known volume (ideally two sizes) packed with copper wool keeps the gas isothermal; pumped at 0.5 to 1 Hz to mimic panting, the box must recover the known volume, corrected to ambient saturated conditions rather than BTPS. Check it monthly, after software updates, and when a problem is suspected (2023 ERS/ATS). DLCO systems are checked with a 3-L syringe (VA 3 +/- 0.3 L, DLCO under 0.5) or a DLCO simulator instead.
Beyond the daily volume check, what routine maintenance does the calibration syringe itself need?
Inspect it daily for a displaced piston stop and for smooth travel with no sticking or catching, and perform a leak test monthly. A syringe that no longer delivers a true 3 L, or that leaks past the seal, will quietly certify a spirometer as accurate when it is not, so the reference standard has to be verified as well as the device.
Your lab uses in-line barrier filters for every patient. What does that require during calibration verification?
The same filter must be in place during recalibration and verification. Filters add resistance and a small dead space, so verifying without one certifies a configuration you never actually use on patients. Filters are the main engineering control against cross-contamination in the PFT lab, alongside hand hygiene, single-patient mouthpieces, and surface disinfection between subjects.
When do you add supine spirometry to a standard upright study, and what size of drop matters?
When diaphragm weakness is suspected: neuromuscular disease, phrenic nerve injury, or unexplained orthopnea. Lying flat lets the abdominal contents push against a weak diaphragm. Healthy adults lose under about 10% of VC supine. A fall greater than 25% is the usual threshold for diaphragm dysfunction (about 79% sensitivity and 90% specificity); unilateral weakness typically drops VC a further 10 to 30% and bilateral weakness 30 to 50%.
A patient has severe emphysema with bullae. Which lung volume method do you select, and why?
Body plethysmography. It measures all compressible thoracic gas, including gas in bullae and poorly communicating units, because it works from Boyle's law rather than from washing a tracer in or out. Gas dilution and washout only reach ventilated lung, so they systematically underestimate FRC and TLC in severe obstruction.
Which bronchial provocation agents act directly on airway smooth muscle, and which act indirectly?
Methacholine acts directly by stimulating muscarinic receptors on smooth muscle. Exercise, eucapnic voluntary hyperventilation, and mannitol act indirectly by provoking mediator release from airway cells. Indirect challenges are less sensitive but more specific for active airway inflammation, so they map better onto a clinical asthma diagnosis.
When would you order MIP, MEP, and cough peak flow rather than more spirometry?
When you need to separate weak respiratory muscles from stiff lungs or chest wall, as in suspected neuromuscular disease, unexplained restriction with a normal TLC pattern, or before decisions about ventilatory support. MIP reflects inspiratory strength, MEP expiratory strength, and cough peak flow the ability to clear secretions.
How do you choose between a six-minute walk test and an incremental shuttle walk test?
The 6MWT is self-paced, so the patient sets an intensity they can sustain and it reflects functional capacity in daily life. The shuttle walk is externally paced by an audio signal that speeds up in stages, so it pushes toward a symptom-limited maximum and behaves more like a graded exercise test. Neither replaces a monitored cardiopulmonary exercise test when you need to explain why exercise is limited.
Why must bronchodilators be withheld before bronchodilator responsiveness and methacholine testing?
Residual drug blunts the measured response and can turn a truly responsive patient into a negative result, or push a methacholine PC20 falsely high. Withholding intervals follow the drug's duration of action, so short-acting agents need the shortest hold and long-acting agents the longest. Always record what the patient actually took and when, because that determines whether the study is interpretable.
During a monitored exercise test, which findings require you to stop the test immediately?
Chest pain suggestive of ischemia or ischemic ECG changes; complex ectopy or second- or third-degree heart block; a systolic pressure fall of more than 20 mmHg from the highest value reached; hypertension above 250 systolic or 120 diastolic; SpO2 of 80% or less with signs of severe hypoxemia; sudden pallor, loss of coordination, dizziness, confusion, or signs of respiratory failure (ATS/ACCP CPET statement). Fatigue and dyspnea are normal end points, not termination criteria.
What is back-extrapolated volume, and what is its acceptability limit?
BEV is the volume already exhaled before the extrapolated true start of the forced effort, and it quantifies a hesitant or slow start. It must be no more than 5% of the FVC or 0.100 L, whichever is greater. An excessive BEV means the timing zero is wrong, which inflates FEV1 because the first, fastest part of the breath is not counted against the clock.
Name the three end of forced expiration (EOFE) indicators; a maneuver needs only one.
First, a plateau: less than 0.025 L change in volume over at least 1 second. Second, the patient has tried to exhale for at least 15 seconds. Third, the measured FVC is within the repeatability tolerance of, or exceeds, the largest FVC already observed in that testing set, which covers patients who cannot sustain exhalation. The 2019 standard renamed the old end-of-test criteria to EOFE.
Give the inspiration and breath-hold requirements for an acceptable single-breath DLCO.
The patient must inhale at least 90% of the largest VC from the same session, fast enough that 85% of the test gas is in within 4 seconds, and the calculated breath-hold must last 10 seconds plus or minus 2 (8 to 12 s) with no Valsalva or Mueller effort. A shallower inspiration, between 85 and 90% of VC, can still be accepted only when its alveolar volume matches the largest VA from acceptable trials to within 200 mL or 5%, whichever is larger.
What are the sampling and spacing rules for DLCO trials?
Sample collection must be completed within 4 seconds of the start of exhalation; on rapid gas analyzer systems the virtual sample starts once dead-space washout is complete. Allow at least 4 minutes between trials on classical systems so test gas clears the lungs, or on RGA systems until end-exhalation tracer is at most 2% of the test gas concentration. Retained CO from the prior trial lowers the next DLCO.
When is a multiple-breath nitrogen washout complete, and what follows immediately?
When the end-tidal tracer gas concentration falls below one fortieth of its starting concentration, roughly 2% N2, for at least three consecutive tidal breaths. The patient then performs a linked maneuver, an inspiratory capacity followed by an expiratory vital capacity, so RV and TLC are derived from the same continuous recording as FRC rather than from a separate spirometry session.
Describe correct closed-shutter panting for FRCpleth, including the frequency.
Support the cheeks and floor of the mouth with both hands, breathe quietly until a stable end-expiratory level is reached (usually 3 to 10 breaths), then close the shutter and pant at 0.5 to 1.0 Hz, that is 30 to 60 per minute. Up to about 1.5 Hz (90 per minute) is acceptable when spirometry shows no or minimal obstruction. Unsupported cheeks act as a compliant chamber and cause overestimation.
What makes a six-minute walk test standardized and repeatable?
A flat, straight hard-surfaced course of fixed length with turnaround markers, the patient walking alone at their own pace while the technologist stays put and uses only scripted encouragement at set intervals, and recording of baseline and end SpO2, heart rate, and Borg dyspnea and fatigue scores. Walking beside or ahead of the patient paces them and invalidates comparison with prior tests.
What do you confirm and control before and after a radial arterial puncture?
Before: adequate collateral circulation, traditionally with a modified Allen test, plus anticoagulant status and site condition. Use a dry heparinized syringe, because liquid heparin dilutes the sample and lowers PCO2. After: apply firm direct pressure for several minutes, longer if the patient is anticoagulated, expel air bubbles immediately, and analyze promptly or ice the sample.
State the repeatability criteria for adult spirometry and the minimum number of maneuvers.
The difference between the largest and next largest FVC must be no more than 0.150 L, and the same for FEV1. An adequate test needs a minimum of three acceptable maneuvers with two of them meeting repeatability. Report the largest FVC, largest FEV1, and largest peak flow, taken from any acceptable maneuvers, not necessarily the same one.
How do the spirometry criteria change for a child aged 6 years or younger?
Repeatability loosens to 0.100 L or 10% of the highest value, whichever is greater, for both FVC and FEV1. Cough and glottic closure are judged over the first 0.75 seconds rather than the first second, because FEV0.75 is often the reportable index when a preschool child cannot sustain a full second of forced expiration.
Two DLCO trials read 40.0 and 37.0 mL/min/mmHg. Are they repeatable under the 2017 standard?
No. The 2017 criterion is two acceptable measurements within 2 mL/min/mmHg (0.67 mmol/min/kPa) of each other, an absolute limit only. This pair differs by 3.0 and fails. The 2005 alternative of 3 units or 10% was dropped, so applying a percentage rule would wrongly pass these and report an average built from discordant data.
How many plethysmographic FRC values do you need, how closely must they agree, and what do you report?
At least three acceptable TGV (FRCpleth) values agreeing within 5%, calculated as the difference between the highest and lowest divided by the mean. Report the average of those values, not the best single effort. Airway resistance must not be measured during the same maneuver as lung volumes, because the two require different panting frequencies.
What is the agreement requirement for FRC measured by washout or helium dilution?
Aim for at least two technically acceptable trials whose FRC values agree within 10% of their mean (highest minus lowest, divided by the mean, no more than 0.10), and report the mean. The tolerance is wider than the 5% used for plethysmography because a steady tidal breathing pattern is harder to sustain than a short panting maneuver. Wait at least twice the washout or dilution time between attempts.
Why compare FIVC with FVC, and what is the limit?
After the forced expiration the patient should inhale rapidly back to full inflation. This inspiratory limb is strongly recommended, though its absence alone does not make the maneuver unacceptable. If the FIVC exceeds the FVC, FIVC minus FVC must be no more than 0.100 L or 5% of FVC, whichever is greater. A larger gap means the patient did not start the forced expiration from full inflation, so the FVC and FEV1 are understated.
How does the timing of a cough or glottic closure change which spirometry value you must reject?
A cough in the first second makes FEV1 unacceptable but leaves FVC usable. Glottic closure in the first second invalidates both FEV1 and FVC. Glottic closure, early inspiration, or coming off the mouthpiece after the first second invalidates FVC but leaves FEV1 usable. Since 2019, FEV1 and FVC are graded separately, so you may report a graded FVC from a maneuver whose FEV1 you rejected, and the reverse.
What are the GLI-2012 equations, and how is the lower limit of normal defined?
GLI-2012 are all-age multi-ethnic spirometry reference equations built with the LMS method, giving a predicted value and a z-score for each index. The LLN is the fifth percentile, which is a z-score of about -1.645. Values above that z-score are normal. Because the LLN adjusts for age, height, and sex, it avoids the age bias that a flat 80% of predicted introduces.
What is BTPS correction and why is it applied?
Body temperature, ambient pressure, saturated with water vapor. Gas leaving the lungs cools and loses water vapor in the circuit, so it occupies less volume at the sensor than it did in the chest. The BTPS factor scales measured volume back to conditions inside the patient. Skipping it, or using a wrong ambient temperature, systematically understates every reported volume and flow.
Which adjustments should be considered for DLCO, and what is KCO?
Consider adjusting for hemoglobin, carboxyhemoglobin, and altitude or inspired oxygen tension, since each changes how much CO the blood can take up independently of the alveolar-capillary membrane. KCO is DLCO divided by alveolar volume, the transfer coefficient. Reporting KCO alongside DLCO separates a small accessible lung from genuinely impaired gas transfer per unit of lung.
How is bronchodilator responsiveness calculated under the 2022 interpretive standard?
Take post-bronchodilator minus pre-bronchodilator, divide by the predicted value, and multiply by 100. A change greater than 10% of predicted in FEV1 or FVC is a significant response. Expressing the change against predicted rather than against the patient's own baseline removes the bias that made the older 12% plus 200 mL rule easier to meet in small people with low baselines.
TLC by plethysmography is 7.10 L and TLC by helium dilution is 5.40 L in the same patient. What does the gap mean?
About 1.7 L of trapped or poorly communicating gas. Plethysmography measures all compressible thoracic gas while dilution only reaches units the tracer can enter, so the difference estimates trapped volume and is a hallmark of severe obstruction and bullous disease. A large gap validates both numbers rather than invalidating either; report the method with every volume.
What must stay constant before you call a change on serial pulmonary function testing real?
The measurement method, the equipment, and the reference equation set, and ideally the operator and time of day. Changing from dilution to plethysmography, or from one reference set to another, shifts values on its own. A real change must also exceed the combined biological and measurement variability for that index, which is wider for DLCO than for FEV1.
The afternoon calibration verification fails. What is the status of the patient results collected since the last good check?
They are suspect and cannot be released as valid. Document the failure, correct or recalibrate the device, repeat the verification, and recall affected patients for retesting. This is why the quality control log must record findings, repairs, adjustments, and software updates with dates: it defines exactly which window of patient data is in question.
What defines airflow obstruction under the 2022 ERS/ATS interpretive standard?
FEV1 divided by the largest vital capacity below the fifth percentile, that is below its LLN. A fixed ratio such as 0.70 is not the recommended criterion: it over-diagnoses obstruction in older adults, whose ratio falls normally with age, and under-diagnoses it in young adults. Confirm hyperinflation or air trapping with lung volumes rather than inferring it from the ratio.
Spirometry shows a low FVC with a normal FEV1/FVC ratio. Can you report restriction?
Not from spirometry alone; that pattern is only possible restriction. Restriction requires TLC below the fifth percentile with a normal FEV1/VC. When both FEV1/VC and TLC fall below the fifth percentile the pattern is mixed. A low FVC with normal TLC is instead the non-specific pattern, often seen with poor effort, obesity, or air trapping.
Give the three-level severity grading from the 2022 interpretive standard.
Grade by z-score, applied to the relevant index. Above -1.645 is normal, -1.65 to -2.5 is mild, -2.51 to -4.0 is moderate, and below -4.1 is severe (the standard's own cut points). This replaced the five-level scheme built on percent-predicted FEV1 cut points of 70, 60, 50, and 35 percent, which graded the same physiology differently at different ages.
Spirometry and lung volumes are normal but DLCO is clearly below the LLN. What does that combination suggest?
Impaired gas transfer without a ventilatory defect. Consider pulmonary vascular disease such as pulmonary hypertension or chronic thromboembolic disease, early interstitial lung disease, or early emphysema before airflow obstruction appears. Also rule out the non-pulmonary causes first: anemia lowers DLCO and a raised carboxyhemoglobin level lowers it too, and both are correctable on the report.
Frequently Asked Questions
Is the RPFT a different exam from the CPFT?
No. The NBRC gives one Pulmonary Function Technology (PFT) Examination and scores it against two cut scores. Reaching the low cut score awards the CPFT credential; reaching the high cut score awards the RPFT credential. You do not choose between two separate tests.
How many questions are on the PFT Examination and how long is it?
The exam has 115 multiple-choice items: 100 scored and 15 pretest items that do not count. You get 2 hours. Pretest items are not identified, so treat every question as if it counts.
Does the NBRC publish RPFT pass rates?
Yes. The NBRC posts annual Examinations in Review statistics. For 2025, 68.8% of new candidates reached the low (CPFT) cut score and 31.5% reached the high (RPFT) cut score. Repeat candidates reached 54.2% and 15.7% respectively.
Who is eligible to sit for the PFT Examination?
You must be at least 18 and meet one of four routes: an associate degree or higher from a CoARC-accredited respiratory therapy program; the NBRC CRT or RRT credential; 62 semester hours of regionally accredited college credit that includes biology and mathematics; or a current CPFT credential. The NBRC does not require chemistry coursework or documented clinical experience.
Can I take the PFT Examination at home?
No. The NBRC discontinued live remote proctoring effective December 31, 2024. Since January 1, 2025 all NBRC examinations are delivered in person at PSI assessment centers, which operate Monday through Saturday at more than 300 US locations.
How long does the RPFT credential last?
Each NBRC credential is issued for five years from the end of the month it was awarded. You maintain it through the Credential Maintenance Program using one of three options: complete assessments and any continuing education they trigger, retake the PFT Examination in your final credential year, or pass an NBRC examination you have not taken before.
Explore More NBRC Respiratory Care Credentials
Continue into nearby exams from the same family. Each card keeps practice questions, study guides, flashcards, videos, and articles in one place.
More From This Family
Videos and articles for deeper review.