Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation
Key Takeaways
The modified Hughson-Westlake threshold search procedure follows the ascending 'down 10 dB, up 5 dB' rule, defining auditory threshold as the lowest intensity where a response occurs in at least 50% of presentations on ascending runs (minimum 2 out of 3, or 3 out of 5 ascending presentations).
Standard pure-tone test order begins at 1000 Hz, proceeds through high frequencies (2000, 3000, 4000, 6000, 8000 Hz), retests 1000 Hz as an internal reliability check (must agree within ), and concludes with low frequencies (500, 250, 125 Hz); inter-octaves (750, 1500, 3000, 6000 Hz) must be tested whenever adjacent octaves differ by .
Insert earphones (ER-3A/5A) provide markedly higher interaural attenuation (about , frequency dependent, vs. ), eliminate collapsed ear canal artifacts, minimize the occlusion effect, and enhance test-retest reliability compared to supra-aural headphones (TDH-39).
Bone conduction testing evaluates sensorineural sensitivity from 250 to 4000 Hz; mastoid placement provides greater dynamic range due to lower RETSPLs by than forehead placement, but bone conduction interaural attenuation is across all cranial sites.
Audiometric classification establishes degree (normal to in children / in adults through profound ) and type: sensorineural (AC = BC within , abnormal thresholds), conductive (normal , abnormal AC, ), or mixed (abnormal BC and AC with ), while accounting for low-frequency vibrotactile artifacts at 250 Hz () and 500 Hz ().
Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation
Behavioral pure-tone audiometry is the gold standard diagnostic battery in audiological practice. It establishes auditory sensitivity across the speech frequency spectrum, differentiates conductive from sensorineural and mixed hearing loss, quantifies impairment severity, guides medical and surgical management, and serves as the baseline for rehabilitative audiologic interventions.
The Modified Hughson-Westlake Threshold Search Protocol
Clinical pure-tone audiometry utilizes the modified Hughson-Westlake procedure (Carhart & Jerger, 1959; adapted from Hughson & Westlake, 1944), standardized under ANSI/ASA S3.21 and ISO 8253-1. This psychophysical method uses an ascending approach to determine behavioral threshold.
Psychophysical Rationale of the Ascending Technique
In classic psychophysics, thresholds determined via descending runs are consistently lower (more sensitive) than those obtained via ascending runs due to stimulus perseveration, central expectation, and auditory persistence. The modified Hughson-Westlake protocol deliberately employs an ascending search for three critical clinical reasons:
- It prevents auditory fatigue and tone adaptation, especially in patients with retrocochlear or sensory pathology.
- It establishes a well-defined psychoacoustic criterion: the patient responds to a sound that emerges from complete inaudibility, minimizing false-positive guessing.
- It provides high test-retest reliability across clinical sessions (typically within ).
The "Down 10 dB, Up 5 dB" Algorithm
- Familiarization / Initial Presentation:
- Present a pure tone at at an intensity of .
- If the patient responds clearly, threshold search begins immediately.
- If no response occurs, increase the presentation level to .
- If still inaudible, increase the intensity in successive steps until a definite behavioral response is obtained (ASHA, 2005).
- The Descent ("Down 10 dB"):
- Following every positive response, attenuate the stimulus by and present the tone again.
- Continue stepping down by until the patient fails to respond.
- The Ascent ("Up 5 dB"):
- Following every absent response (failure to respond), increase the stimulus level by and present the tone.
- Continue stepping up by until the patient responds.
- Ascending Run Definition:
- Only responses obtained on an ascending run (moving from inaudible to audible in increments) are counted toward threshold determination. Responses on the initial descent do not count.
Definition of Behavioral Auditory Threshold
The clinical auditory threshold is formally defined as: Clinically, this criterion requires a minimum of 2 responses out of 3 presentations (or 3 responses out of 5 presentations) at that specific intensity on ascending sweeps.
Stimulus Characteristics & Presentation Timing
- Tone Duration: Stimulus duration should be standardized between and (typically ). Tone presentations shorter than suffer from temporal integration roll-off, artificially elevating thresholds because the auditory system integrates acoustic energy over . Presentations longer than induce auditory adaptation.
- Pulsed Tones vs. Continuous Tones: Pulsed tones (2–3 pulses per second with and rise/fall times) are strongly preferred over continuous tones. Pulsed tones reduce patient listening fatigue, enhance signal detection in background physiologic noise, and prevent confusion in patients experiencing subjective high-frequency tinnitus.
- Inter-Stimulus Intervals: Clinicians must randomize the interval between presentations (). Regular or rhythmic stimulus delivery leads to false-positive rhythmic conditioning, where the patient presses the response button in anticipation of the cadence.
Test Frequency Sequence & Inter-Octave Rules
Standard Testing Sequence
Pure-tone air conduction testing must adhere to a standardized sequence to maximize reliability and identify participant inconsistency:
- Initial Frequency (): Testing begins at because human hearing exhibits superior test-retest reliability, low susceptibility to physiologic background noise, and central placement within the speech-frequency spectrum.
- High-Frequency Progression: Proceed sequentially in ascending octaves: , , , , and .
- Mandatory 1000 Hz Retest (Internal Reliability Check): Before evaluating low frequencies, retest threshold at in the first ear tested.
- Acceptable Reliability: The retest threshold must agree with the initial threshold within .
- Recording Rule: When the two thresholds agree within , record the lower (better) one. If they differ by more than , ASHA (2005) allows the lower threshold to be accepted but calls for retesting at least one other frequency; also reinstruct the patient and check transducer placement.
- Low-Frequency Progression: Proceed in descending order: , , and (when indicated) .
- Contralateral Ear Sequence: Test the second ear starting at , proceeding to high frequencies, and finishing with low frequencies. The retest is not required in the second ear unless responses appear erratic or questionable.
Inter-Octave Testing Rule
Standard octave frequencies include . Inter-octave testing () is clinically mandatory under the following conditions:
- The Rule: Whenever a threshold difference of or greater is observed between two adjacent octave frequencies (e.g., and ), the intermediate inter-octave frequency () must be tested.
- Hearing Aid Fittings: Testing and is standard when prescribing modern multi-channel hearing aids to ensure accurate targets.
- Noise-Induced Hearing Loss & Ototoxicity: Testing and is routine in occupational noise surveillance (OSHA/NIOSH standards) and ototoxicity monitoring protocols because acoustic trauma notches and drug toxicity frequently emerge between and before manifesting at standard octaves.
Transducer Selection & Acoustic Coupling Mechanics
The selection of acoustic transducers dictates calibration standards, ambient noise attenuation, collapsed canal vulnerability, and cross-hearing thresholds.
Insert Earphones (Etymotic ER-3A / ER-5C / 3C)
Insert earphones couple to the patient's external auditory canal via disposable expandable foam or silicon eartips.
- Interaural Attenuation (IA): Provides much greater interaural attenuation than supra-aural earphones (conservatively about at and below and about above with deep insertion). This greatly reduces how often masking is needed in asymmetrical losses.
- Canal Collapse Prevention: Deep insertion mechanically stents the cartilaginous external meatus, completely preventing collapsed ear canal artifacts.
- Occlusion Effect Reduction: Deep insertion into the bony canal dramatically minimizes the occlusion effect ( at ).
- Ambient Noise Attenuation: Superior passive low-frequency ambient noise rejection compared to supra-aural earphones, facilitating valid testing in non-ideal sound suites.
- Contraindications: Obstructing cerumen impaction, active otorrhea/exudate, canal atresia or severe stenosis, acute otitis externa, or unhealed post-surgical incisions.
Supra-Aural Headphones (Telephonics TDH-39 / TDH-49 / TDH-50)
Supra-aural earphones consist of dynamic transducers housed in MX-41/AR cushions resting directly on the pinnae.
- Interaural Attenuation (IA): Conservative standard across all frequencies.
- Ear Canal Collapse: Cushion headband tension compresses the pinna against the mastoid, causing the posterior cartilaginous meatal wall to collapse over the canal opening in elderly individuals (loss of cartilage elasticity) and infants. This produces an artificial, pseudo-conductive air-bone gap of localized to .
- Sanitation & Occlusion: Higher occlusion effect ( at 250 Hz, at 500 Hz) and lower ambient noise attenuation.
Circumaural Headphones (Sennheiser HDA 200 / HDA 300)
Circumaural earphones feature large earcups with cushions that completely encircle the pinnae, resting against the cranium.
- Application: Standard for extended high-frequency audiometry () used in ototoxicity surveillance, cisplatin/aminoglycoside monitoring, and advanced tinnitus evaluations.
Bone Conduction Transducers (Radioear B-71 / B-81)
Bone vibrators convert electrical signals into mechanical vibrations to directly stimulate the temporal bones and cochlear fluids.
- Frequency Limits: Testing is restricted to . Frequencies above are excluded due to severe harmonic distortion and physical output limits of magnetic transducers.
- Interaural Attenuation: Bone conduction across the entire skull. Contralateral masking is required to isolate ear-specific bone conduction whenever an unmasked air-bone gap exceeds .
| Transducer Type | Acoustic Coupler (Calibration) | Conservative Interaural Attenuation | Primary Clinical Advantages | Diagnostic Vulnerabilities / Traps |
|---|---|---|---|---|
| Insert Earphones (ER-3A / ER-5C) | 2-cc Coupler (HA-1, HA-2) / Occluded Ear Simulator | about (250-1000 Hz), about (above 1000 Hz) | Eliminates collapsed canals; high IA; minimal occlusion effect | Incompatible with active otorrhea, canal atresia, or impacted cerumen |
| Supra-Aural (TDH-39 / TDH-49) | NBS-9A 6-cc Coupler (ANSI S3.6) | across all frequencies | Easy placement on deformed pinnae; standard clinical familiarization | Ear canal collapse mimics high-frequency conductive loss; low IA () |
| Circumaural (HDA 200 / HDA 300) | Artificial Ear (IEC 60318-1 / 60318-2) | Extended high frequencies (8-20 kHz); high ambient noise rejection | Bulky; headband placement critical; coupler resonance calibration | |
| Bone Vibrator (Radioear B-71 / B-81) | Mechanical Coupler (Artificial Mastoid, ANSI S3.13) | across all frequencies | Direct evaluation of cochlear sensorineural reserve | ; low-frequency tactile artifacts; harmonic distortion at high outputs |
Bone Conduction Mechanics: Mastoid vs. Forehead Placement
Bone conduction sound energy reaches the organ of Corti via three physical pathways:
- Compressional / Distortional Bone Conduction: Skull vibrations alternate between compression and expansion, distorting the bony otic capsule and cyclically compressing the inner ear perilymph. Because the scala vestibuli has greater volume than the scala tympani and the round window is 20 times more compliant than the oval window, fluid displacement creates a traveling wave along the basilar membrane.
- Inertial Bone Conduction: The ossicular chain is suspended elastically within the middle ear cavity. Cranial vibration moves the temporal bone around the ossicles; due to inertia, the ossicles lag behind, generating relative stapedial footplate displacement at the oval window.
- Osseotympanic Bone Conduction: Cranial vibration vibrates the cartilaginous and bony walls of the external auditory meatus, radiating acoustic sound pressure into the canal. In an unoccluded ear, this energy escapes into the room; in an occluded ear, it drives the tympanic membrane.
Mastoid Placement vs. Forehead Placement
| Clinical Characteristic | Mastoid Placement (Radioear B-71) | Forehead Placement (Radioear B-71) |
|---|---|---|
| Anatomical Site | Post-auricular mastoid process (without touching pinna) | Midline frontal bone (2-3 cm above brow) |
| RETSPL / Sensitivity | More sensitive: requires less physical force for threshold | Less sensitive: requires higher output for threshold |
| Maximum Output Limits | Greater dynamic range before amplifier clipping and distortion | Reduced dynamic range; output ceiling reached early in severe losses |
| Test-Retest Variability | Slightly higher due to differences in mastoid cellularity and positioning | Lowest test-retest variability; flat bony surface prevents slippage |
| Ear-Canal Radiation | Mild acoustic radiation from mastoid into canal at high frequencies | Minimal acoustic radiation |
| Interaural Attenuation | (stimulates both cochleae simultaneously) | (stimulates both cochleae simultaneously) |
Clinical Recommendation: Mastoid placement is universally favored in diagnostic practice because its lower RETSPLs provide greater testable dynamic range, allowing accurate evaluation of severe sensorineural thresholds before reaching transducer output limits.
Pure-Tone Averages (PTAs) & Clinical Applications
Pure-tone averages synthesize discrete thresholds into a single numerical index for speech cross-checks, disability ratings, and medical classifications.
1. Standard 3-Frequency Pure-Tone Average (PTA)
The standard PTA is the arithmetic mean of air-conduction thresholds at , , and :
- Speech Recognition Threshold (SRT) Cross-Check: The standard PTA must agree with the measured SRT within . A discrepancy suggests non-organic hearing loss (pseudohypacusis), auditory neuropathy, or cognitive/linguistic barriers.
2. Two-Frequency Fletcher Average (Best-Two PTA)
The Fletcher average is calculated by taking the arithmetic mean of the two best (lowest threshold) frequencies among , , and :
- Clinical Indication: Utilized in steeply sloping, precipitous, or ski-slope high-frequency hearing loss, where thresholds at drop precipitously by relative to and . In such patients, the standard 3-frequency PTA is skewed artificially poor, whereas the Fletcher average aligns accurately with the SRT.
3. High-Frequency Pure-Tone Average (HF-PTA)
The high-frequency PTA averages thresholds at , , and (or in industrial audiometry):
- Clinical Indication: Predicts speech perception difficulty in competing background noise and monitors noise-induced hearing loss, presbycusis, and ototoxic deterioration.
Audiometric Interpretation: Degree, Configuration & Type
Degree of Hearing Loss (ANSI / Clark / Goodman Scales)
Hearing loss degree is defined by pure-tone air-conduction thresholds across the speech frequency spectrum.
| Degree Category | Pediatric Threshold Range (dB HL) | Adult Threshold Range (dB HL) | Functional Speech & Communication Impact |
|---|---|---|---|
| Normal | Complete access to conversational speech cues in quiet and noise. | ||
| Slight / Minimal | Classified as Normal | Subtle speech morphology errors; misses unstressed words and faint consonants in classroom noise. | |
| Mild | Misses of speech signal; difficulty hearing faint or distant voices; struggles in noisy environments. | ||
| Moderate | Misses of spoken conversational speech at normal vocal levels (). | ||
| Moderately Severe | Conversational speech inaudible without amplification; requires loud shouting; severe conversational disruption. | ||
| Severe | Cannot hear conversational speech; perceives only loud shouts or amplified environmental sounds. | ||
| Profound | Reliant on visual speech cues, ASL, cochlear implants; sound perceived primarily as tactile vibration. |
Pediatric Imperative: The pediatric cutoff for normal hearing is strictly . Thresholds between and in children represent minimal/slight hearing loss, which compromises phonological awareness, academic performance, and reading comprehension.
Audiogram Configurations
- Flat: Thresholds do not vary by more than across octaves.
- Sloping: Thresholds worsen gradually across octave frequencies (). Characteristic of sensory presbycusis and metabolic hearing loss.
- Precipitous / Steeply Sloping: Thresholds drop sharply by per octave. Classic for acoustic trauma, ototoxicity, or apical-to-basal lesions.
- Rising: Low-frequency thresholds are poorer than high-frequency thresholds. Classic for early Ménière's disease (endolymphatic hydrops) or middle-ear stiffness pathologies (otosclerosis, Eustachian tube dysfunction).
- Cookie-Bite / Trough-Shaped: Mid-frequency thresholds () are abnormal, with preserved low and high frequencies. Often hereditary, for example autosomal dominant TECTA-related loss (DFNA8/12).
- Reverse Cookie-Bite / Saucer: Mid-frequency hearing is preserved, while low and high frequencies are impaired.
- Acoustic Notch: Sharp drop in thresholds at , , or , with significant recovery at . Pathognomonic for noise-induced hearing loss (NIHL), driven by outer ear acoustic resonance () focusing high sound pressure on the 4 kHz cochlear partition.
- Corner Audiogram: Responses obtained only at the lowest frequencies () at the upper limits of the audiometer (), with no response at all remaining frequencies.
Type of Hearing Loss
The relationship between air-conduction (AC) and bone-conduction (BC) thresholds determines the pathology type:
-
Sensorineural Hearing Loss (SNHL):
- Air-conduction thresholds: Abnormal ( in adults, in children).
- Bone-conduction thresholds: Abnormal ( in adults, in children).
- Air-Bone Gap (ABG): Absent or clinically insignificant ().
- Pathophysiology: Outer and middle ear transmission is intact. Deficit resides in the cochlea (sensory, outer/inner hair cells) or VIII cranial nerve (neural).
-
Conductive Hearing Loss (CHL):
- Air-conduction thresholds: Abnormal ( in adults, in children).
- Bone-conduction thresholds: Completely normal ( in adults, in children).
- Air-Bone Gap (ABG): Clinically significant ().
- Pathophysiology: Cochlear and retrocochlear structures are intact. Obstruction or mechanical lesion in the external canal, tympanic membrane, or ossicular chain (cerumen, otitis media, otosclerosis, ossicular discontinuity).
-
Mixed Hearing Loss (MHL):
- Air-conduction thresholds: Abnormal ( in adults, in children).
- Bone-conduction thresholds: Abnormal ( in adults, in children).
- Air-Bone Gap (ABG): Clinically significant ().
- Pathophysiology: Concurrent presence of sensorineural damage in the inner ear/nerve and conductive pathology in the outer/middle ear (e.g., presbycusis complicated by otitis media with effusion, or advanced otosclerosis with cochlear involvement).
Diagnostic Pitfalls & Vibrotactile Artifacts
Low-Frequency Vibrotactile Responses
At high stimulus intensities, bone vibrators (and occasionally supra-aural headphones) generate strong physical mechanical vibrations that activate somatosensory mechanoreceptors (cutaneous Pacinian and Meissner corpuscles) in the periosteum and skin, rather than stimulating the auditory hair cells.
- Vibrotactile Threshold Limits:
- : Tactile responses occur at .
- : Tactile responses occur at .
- : Tactile responses occur at .
The Severe Diagnostic Trap: A patient with bilateral profound sensorineural hearing loss (true hearing threshold ) is tested with unmasked bone conduction. At , the patient feels the vibrator buzzing at and presses the response button. If the audiologist plots as a true hearing threshold, it creates a false air-bone gap (), misdiagnosing profound SNHL as a mixed hearing loss. The clinician must explicitly ask the patient: "Did you hear a tone in your ear, or did you only feel a vibrating or buzzing sensation on your head?" Tactile responses must be documented on the audiogram with a tactile symbol ("V" or "T") or annotated as "no response at maximum output."
Ear Canal Collapse
Canal collapse occurs when pressure from supra-aural headphone cushions pinches the cartilaginous meatus closed. This creates an artificial conductive component of across . Always verify suspicious high-frequency air-bone gaps by switching to insert earphones, which physically stent the canal open.
During pure-tone threshold testing using the modified Hughson-Westlake procedure, a patient responds at 40 dB HL on an ascending run. After stepping down 10 dB to 30 dB HL, the patient fails to respond. The clinician increases the level to 35 dB HL (no response), and then to 40 dB HL, where the patient responds again. According to clinical threshold definitions, what action should the clinician take next?
Accept 40 dB HL as the threshold, as 2 out of 2 ascending presentations at this level elicited a response.
Decrease by 10 dB to 30 dB HL and then test at 35 dB HL to confirm whether 35 dB HL is the true threshold.
Increase by 5 dB to 45 dB HL to test for suprathreshold auditory fatigue.
Reject the run and switch immediately to a descending method of limits.
A 74-year-old patient is evaluated using TDH-49 supra-aural headphones. Audiometry reveals bilateral normal bone conduction thresholds (10-15 dB HL at all frequencies), but air conduction thresholds at 2000 Hz, 3000 Hz, and 4000 Hz are 40 dB HL bilaterally, yielding a 25-30 dB high-frequency air-bone gap. Tympanometry reveals normal Type A tympanograms with robust acoustic reflexes. What is the most appropriate next clinical step?
Refer the patient immediately for an otolaryngology surgical consultation for bilateral ossicular chain reconstruction.
Re-evaluate air-conduction thresholds using insert earphones to rule out collapsed external auditory meatuses.
Diagnose bilateral mixed hearing loss and recommend immediate binaural bone-conduction hearing aids.
Perform masked bone conduction at 6000 Hz and 8000 Hz to assess high-frequency sensorineural reserve.
A patient with suspected profound bilateral hearing loss is tested with an unmasked Radioear B-71 bone vibrator on the right mastoid. The patient signals a response at 45 dB HL at 250 Hz and at 55 dB HL at 500 Hz. However, air conduction thresholds at both frequencies exceed 100 dB HL bilaterally, and otoacoustic emissions and acoustic reflexes are absent. What is the most accurate interpretation of these bone conduction responses?
The patient exhibits a profound conductive hearing loss that requires urgent surgical ossicular reconstruction.
The responses represent true cochlear thresholds of the non-test left ear via cross-hearing.
They are likely vibrotactile responses detected through the skin rather than true auditory thresholds.
The responses demonstrate an organic acoustic notch resulting from acute high-intensity impulse noise exposure.