Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation

Key Takeaways

  • The modified Hughson-Westlake threshold search procedure follows the ascending 'down 10 dB, up 5 dB' rule, defining auditory threshold as the lowest intensity where a response occurs in at least 50% of presentations on ascending runs (minimum 2 out of 3, or 3 out of 5 ascending presentations).

  • Standard pure-tone test order begins at 1000 Hz, proceeds through high frequencies (2000, 3000, 4000, 6000, 8000 Hz), retests 1000 Hz as an internal reliability check (must agree within ±5 dB\pm 5\,\text{dB}), and concludes with low frequencies (500, 250, 125 Hz); inter-octaves (750, 1500, 3000, 6000 Hz) must be tested whenever adjacent octaves differ by ≥20 dB\ge 20\,\text{dB}.

  • Insert earphones (ER-3A/5A) provide markedly higher interaural attenuation (about 50−70 dB50-70\,\text{dB}, frequency dependent, vs. 40 dB40\,\text{dB}), eliminate collapsed ear canal artifacts, minimize the occlusion effect, and enhance test-retest reliability compared to supra-aural headphones (TDH-39).

  • Bone conduction testing evaluates sensorineural sensitivity from 250 to 4000 Hz; mastoid placement provides greater dynamic range due to lower RETSPLs by ≈10−15 dB\approx 10-15\,\text{dB} than forehead placement, but bone conduction interaural attenuation is 0 dB0\,\text{dB} across all cranial sites.

  • Audiometric classification establishes degree (normal −10-10 to 15 dB HL15\,\text{dB HL} in children / 25 dB HL25\,\text{dB HL} in adults through profound ≥91 dB HL\ge 91\,\text{dB HL}) and type: sensorineural (AC = BC within 10 dB10\,\text{dB}, abnormal thresholds), conductive (normal BC≤25 dB HL\text{BC} \le 25\,\text{dB HL}, abnormal AC, ABG≥15 dB\text{ABG} \ge 15\,\text{dB}), or mixed (abnormal BC and AC with ABG≥15 dB\text{ABG} \ge 15\,\text{dB}), while accounting for low-frequency vibrotactile artifacts at 250 Hz (≥40−45 dB HL\ge 40-45\,\text{dB HL}) and 500 Hz (≥55−60 dB HL\ge 55-60\,\text{dB HL}).

Last updated: October 2026

Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation

Behavioral pure-tone audiometry is the gold standard diagnostic battery in audiological practice. It establishes auditory sensitivity across the speech frequency spectrum, differentiates conductive from sensorineural and mixed hearing loss, quantifies impairment severity, guides medical and surgical management, and serves as the baseline for rehabilitative audiologic interventions.


The Modified Hughson-Westlake Threshold Search Protocol

Clinical pure-tone audiometry utilizes the modified Hughson-Westlake procedure (Carhart & Jerger, 1959; adapted from Hughson & Westlake, 1944), standardized under ANSI/ASA S3.21 and ISO 8253-1. This psychophysical method uses an ascending approach to determine behavioral threshold.

Psychophysical Rationale of the Ascending Technique

In classic psychophysics, thresholds determined via descending runs are consistently lower (more sensitive) than those obtained via ascending runs due to stimulus perseveration, central expectation, and auditory persistence. The modified Hughson-Westlake protocol deliberately employs an ascending search for three critical clinical reasons:

  • It prevents auditory fatigue and tone adaptation, especially in patients with retrocochlear or sensory pathology.
  • It establishes a well-defined psychoacoustic criterion: the patient responds to a sound that emerges from complete inaudibility, minimizing false-positive guessing.
  • It provides high test-retest reliability across clinical sessions (typically within ±5 dB\pm 5\,\text{dB}).

The "Down 10 dB, Up 5 dB" Algorithm

  1. Familiarization / Initial Presentation:
    • Present a pure tone at 1000 Hz1000\,\text{Hz} at an intensity of 30 dB HL30\,\text{dB HL}.
    • If the patient responds clearly, threshold search begins immediately.
    • If no response occurs, increase the presentation level to 50 dB HL50\,\text{dB HL}.
    • If still inaudible, increase the intensity in successive 10 dB10\,\text{dB} steps until a definite behavioral response is obtained (ASHA, 2005).
  2. The Descent ("Down 10 dB"):
    • Following every positive response, attenuate the stimulus by 10 dB10\,\text{dB} and present the tone again.
    • Continue stepping down by 10 dB10\,\text{dB} until the patient fails to respond.
  3. The Ascent ("Up 5 dB"):
    • Following every absent response (failure to respond), increase the stimulus level by 5 dB5\,\text{dB} and present the tone.
    • Continue stepping up by 5 dB5\,\text{dB} until the patient responds.
  4. Ascending Run Definition:
    • Only responses obtained on an ascending run (moving from inaudible to audible in 5 dB5\,\text{dB} increments) are counted toward threshold determination. Responses on the initial descent do not count.

Definition of Behavioral Auditory Threshold

The clinical auditory threshold is formally defined as: Threshold=The lowest hearing level (in dB HL) at which a response occurs in ≥50% of ascending presentations\text{Threshold} = \text{The lowest hearing level (in dB HL) at which a response occurs in } \ge 50\% \text{ of ascending presentations} Clinically, this criterion requires a minimum of 2 responses out of 3 presentations (or 3 responses out of 5 presentations) at that specific intensity on ascending sweeps.

Stimulus Characteristics & Presentation Timing

  • Tone Duration: Stimulus duration should be standardized between 1.01.0 and 2.0 seconds2.0\,\text{seconds} (typically 1.0−1.5 s1.0-1.5\,\text{s}). Tone presentations shorter than 200 ms200\,\text{ms} suffer from temporal integration roll-off, artificially elevating thresholds because the auditory system integrates acoustic energy over ≈200−300 ms\approx 200-300\,\text{ms}. Presentations longer than 2.5 s2.5\,\text{s} induce auditory adaptation.
  • Pulsed Tones vs. Continuous Tones: Pulsed tones (2–3 pulses per second with 50% duty cycle50\%\,\text{duty cycle} and 20−50 ms20-50\,\text{ms} rise/fall times) are strongly preferred over continuous tones. Pulsed tones reduce patient listening fatigue, enhance signal detection in background physiologic noise, and prevent confusion in patients experiencing subjective high-frequency tinnitus.
  • Inter-Stimulus Intervals: Clinicians must randomize the interval between presentations (1.5−4.0 seconds1.5-4.0\,\text{seconds}). Regular or rhythmic stimulus delivery leads to false-positive rhythmic conditioning, where the patient presses the response button in anticipation of the cadence.

Test Frequency Sequence & Inter-Octave Rules

Standard Testing Sequence

Pure-tone air conduction testing must adhere to a standardized sequence to maximize reliability and identify participant inconsistency:

  1. Initial Frequency (1000 Hz1000\,\text{Hz}): Testing begins at 1000 Hz1000\,\text{Hz} because human hearing exhibits superior test-retest reliability, low susceptibility to physiologic background noise, and central placement within the speech-frequency spectrum.
  2. High-Frequency Progression: Proceed sequentially in ascending octaves: 2000 Hz2000\,\text{Hz}, 3000 Hz3000\,\text{Hz}, 4000 Hz4000\,\text{Hz}, 6000 Hz6000\,\text{Hz}, and 8000 Hz8000\,\text{Hz}.
  3. Mandatory 1000 Hz Retest (Internal Reliability Check): Before evaluating low frequencies, retest threshold at 1000 Hz1000\,\text{Hz} in the first ear tested.
    • Acceptable Reliability: The retest threshold must agree with the initial threshold within ±5 dB\pm 5\,\text{dB}.
    • Recording Rule: When the two thresholds agree within ±5 dB\pm 5\,\text{dB}, record the lower (better) one. If they differ by more than 5 dB5\,\text{dB}, ASHA (2005) allows the lower threshold to be accepted but calls for retesting at least one other frequency; also reinstruct the patient and check transducer placement.
  4. Low-Frequency Progression: Proceed in descending order: 500 Hz500\,\text{Hz}, 250 Hz250\,\text{Hz}, and (when indicated) 125 Hz125\,\text{Hz}.
  5. Contralateral Ear Sequence: Test the second ear starting at 1000 Hz1000\,\text{Hz}, proceeding to high frequencies, and finishing with low frequencies. The 1000 Hz1000\,\text{Hz} retest is not required in the second ear unless responses appear erratic or questionable.

Inter-Octave Testing Rule

Standard octave frequencies include 250,500,1000,2000,4000, and 8000 Hz250, 500, 1000, 2000, 4000,\,\text{and } 8000\,\text{Hz}. Inter-octave testing (750,1500,3000, and 6000 Hz750, 1500, 3000,\,\text{and } 6000\,\text{Hz}) is clinically mandatory under the following conditions:

  • The ≥20 dB\ge 20\,\text{dB} Rule: Whenever a threshold difference of 20 dB20\,\text{dB} or greater is observed between two adjacent octave frequencies (e.g., 1000 Hz=20 dB HL1000\,\text{Hz} = 20\,\text{dB HL} and 2000 Hz=45 dB HL2000\,\text{Hz} = 45\,\text{dB HL}), the intermediate inter-octave frequency (1500 Hz1500\,\text{Hz}) must be tested.
  • Hearing Aid Fittings: Testing 1500 Hz1500\,\text{Hz} and 3000 Hz3000\,\text{Hz} is standard when prescribing modern multi-channel hearing aids to ensure accurate targets.
  • Noise-Induced Hearing Loss & Ototoxicity: Testing 3000 Hz3000\,\text{Hz} and 6000 Hz6000\,\text{Hz} is routine in occupational noise surveillance (OSHA/NIOSH standards) and ototoxicity monitoring protocols because acoustic trauma notches and drug toxicity frequently emerge between 30003000 and 6000 Hz6000\,\text{Hz} before manifesting at standard octaves.

Transducer Selection & Acoustic Coupling Mechanics

The selection of acoustic transducers dictates calibration standards, ambient noise attenuation, collapsed canal vulnerability, and cross-hearing thresholds.

Insert Earphones (Etymotic ER-3A / ER-5C / 3C)

Insert earphones couple to the patient's external auditory canal via disposable expandable foam or silicon eartips.

  • Interaural Attenuation (IA): Provides much greater interaural attenuation than supra-aural earphones (conservatively about 70 dB70\,\text{dB} at 1000 Hz1000\,\text{Hz} and below and about 50 dB50\,\text{dB} above 1000 Hz1000\,\text{Hz} with deep insertion). This greatly reduces how often masking is needed in asymmetrical losses.
  • Canal Collapse Prevention: Deep insertion mechanically stents the cartilaginous external meatus, completely preventing collapsed ear canal artifacts.
  • Occlusion Effect Reduction: Deep insertion into the bony canal dramatically minimizes the occlusion effect (0 dB0\,\text{dB} at ≥500 Hz\ge 500\,\text{Hz}).
  • Ambient Noise Attenuation: Superior passive low-frequency ambient noise rejection compared to supra-aural earphones, facilitating valid testing in non-ideal sound suites.
  • Contraindications: Obstructing cerumen impaction, active otorrhea/exudate, canal atresia or severe stenosis, acute otitis externa, or unhealed post-surgical incisions.

Supra-Aural Headphones (Telephonics TDH-39 / TDH-49 / TDH-50)

Supra-aural earphones consist of dynamic transducers housed in MX-41/AR cushions resting directly on the pinnae.

  • Interaural Attenuation (IA): Conservative standard IA=40 dB\text{IA} = 40\,\text{dB} across all frequencies.
  • Ear Canal Collapse: Cushion headband tension compresses the pinna against the mastoid, causing the posterior cartilaginous meatal wall to collapse over the canal opening in elderly individuals (loss of cartilage elasticity) and infants. This produces an artificial, pseudo-conductive air-bone gap of 15−35 dB15-35\,\text{dB} localized to 1000−4000 Hz1000-4000\,\text{Hz}.
  • Sanitation & Occlusion: Higher occlusion effect (30 dB30\,\text{dB} at 250 Hz, 20 dB20\,\text{dB} at 500 Hz) and lower ambient noise attenuation.

Circumaural Headphones (Sennheiser HDA 200 / HDA 300)

Circumaural earphones feature large earcups with cushions that completely encircle the pinnae, resting against the cranium.

  • Application: Standard for extended high-frequency audiometry (8000−20000 Hz8000-20000\,\text{Hz}) used in ototoxicity surveillance, cisplatin/aminoglycoside monitoring, and advanced tinnitus evaluations.

Bone Conduction Transducers (Radioear B-71 / B-81)

Bone vibrators convert electrical signals into mechanical vibrations to directly stimulate the temporal bones and cochlear fluids.

  • Frequency Limits: Testing is restricted to 250−4000 Hz250-4000\,\text{Hz}. Frequencies above 4000 Hz4000\,\text{Hz} are excluded due to severe harmonic distortion and physical output limits of magnetic transducers.
  • Interaural Attenuation: Bone conduction IA=0 dB\text{IA} = 0\,\text{dB} across the entire skull. Contralateral masking is required to isolate ear-specific bone conduction whenever an unmasked air-bone gap exceeds 10 dB10\,\text{dB}.
Transducer TypeAcoustic Coupler (Calibration)Conservative Interaural AttenuationPrimary Clinical AdvantagesDiagnostic Vulnerabilities / Traps
Insert Earphones (ER-3A / ER-5C)2-cc Coupler (HA-1, HA-2) / Occluded Ear Simulatorabout 70 dB70\,\text{dB} (250-1000 Hz), about 50 dB50\,\text{dB} (above 1000 Hz)Eliminates collapsed canals; high IA; minimal occlusion effectIncompatible with active otorrhea, canal atresia, or impacted cerumen
Supra-Aural (TDH-39 / TDH-49)NBS-9A 6-cc Coupler (ANSI S3.6)40 dB40\,\text{dB} across all frequenciesEasy placement on deformed pinnae; standard clinical familiarizationEar canal collapse mimics high-frequency conductive loss; low IA (40 dB40\,\text{dB})
Circumaural (HDA 200 / HDA 300)Artificial Ear (IEC 60318-1 / 60318-2)40−50 dB40-50\,\text{dB}Extended high frequencies (8-20 kHz); high ambient noise rejectionBulky; headband placement critical; coupler resonance calibration
Bone Vibrator (Radioear B-71 / B-81)Mechanical Coupler (Artificial Mastoid, ANSI S3.13)0 dB0\,\text{dB} across all frequenciesDirect evaluation of cochlear sensorineural reserveIA=0 dB\text{IA} = 0\,\text{dB}; low-frequency tactile artifacts; harmonic distortion at high outputs

Bone Conduction Mechanics: Mastoid vs. Forehead Placement

Bone conduction sound energy reaches the organ of Corti via three physical pathways:

  1. Compressional / Distortional Bone Conduction: Skull vibrations alternate between compression and expansion, distorting the bony otic capsule and cyclically compressing the inner ear perilymph. Because the scala vestibuli has greater volume than the scala tympani and the round window is 20 times more compliant than the oval window, fluid displacement creates a traveling wave along the basilar membrane.
  2. Inertial Bone Conduction: The ossicular chain is suspended elastically within the middle ear cavity. Cranial vibration moves the temporal bone around the ossicles; due to inertia, the ossicles lag behind, generating relative stapedial footplate displacement at the oval window.
  3. Osseotympanic Bone Conduction: Cranial vibration vibrates the cartilaginous and bony walls of the external auditory meatus, radiating acoustic sound pressure into the canal. In an unoccluded ear, this energy escapes into the room; in an occluded ear, it drives the tympanic membrane.

Mastoid Placement vs. Forehead Placement

Clinical CharacteristicMastoid Placement (Radioear B-71)Forehead Placement (Radioear B-71)
Anatomical SitePost-auricular mastoid process (without touching pinna)Midline frontal bone (2-3 cm above brow)
RETSPL / SensitivityMore sensitive: requires ≈10−15 dB\approx 10-15\,\text{dB} less physical force for thresholdLess sensitive: requires ≈10−15 dB\approx 10-15\,\text{dB} higher output for threshold
Maximum Output LimitsGreater dynamic range before amplifier clipping and distortionReduced dynamic range; output ceiling reached early in severe losses
Test-Retest VariabilitySlightly higher due to differences in mastoid cellularity and positioningLowest test-retest variability; flat bony surface prevents slippage
Ear-Canal RadiationMild acoustic radiation from mastoid into canal at high frequenciesMinimal acoustic radiation
Interaural Attenuation0 dB0\,\text{dB} (stimulates both cochleae simultaneously)0 dB0\,\text{dB} (stimulates both cochleae simultaneously)

Clinical Recommendation: Mastoid placement is universally favored in diagnostic practice because its lower RETSPLs provide 10−15 dB10-15\,\text{dB} greater testable dynamic range, allowing accurate evaluation of severe sensorineural thresholds before reaching transducer output limits.


Pure-Tone Averages (PTAs) & Clinical Applications

Pure-tone averages synthesize discrete thresholds into a single numerical index for speech cross-checks, disability ratings, and medical classifications.

1. Standard 3-Frequency Pure-Tone Average (PTA)

The standard PTA is the arithmetic mean of air-conduction thresholds at 500 Hz500\,\text{Hz}, 1000 Hz1000\,\text{Hz}, and 2000 Hz2000\,\text{Hz}: PTA3=AC500+AC1000+AC20003\text{PTA}_{3} = \frac{\text{AC}_{500} + \text{AC}_{1000} + \text{AC}_{2000}}{3}

  • Speech Recognition Threshold (SRT) Cross-Check: The standard PTA must agree with the measured SRT within ±6−8 dB\pm 6-8\,\text{dB}. A discrepancy >10−12 dB> 10-12\,\text{dB} suggests non-organic hearing loss (pseudohypacusis), auditory neuropathy, or cognitive/linguistic barriers.

2. Two-Frequency Fletcher Average (Best-Two PTA)

The Fletcher average is calculated by taking the arithmetic mean of the two best (lowest threshold) frequencies among 500 Hz500\,\text{Hz}, 1000 Hz1000\,\text{Hz}, and 2000 Hz2000\,\text{Hz}: PTAFletcher=Best Threshold1+Best Threshold22\text{PTA}_{\text{Fletcher}} = \frac{\text{Best Threshold}_1 + \text{Best Threshold}_2}{2}

  • Clinical Indication: Utilized in steeply sloping, precipitous, or ski-slope high-frequency hearing loss, where thresholds at 2000 Hz2000\,\text{Hz} drop precipitously by ≥20−30 dB\ge 20-30\,\text{dB} relative to 500500 and 1000 Hz1000\,\text{Hz}. In such patients, the standard 3-frequency PTA is skewed artificially poor, whereas the Fletcher average aligns accurately with the SRT.

3. High-Frequency Pure-Tone Average (HF-PTA)

The high-frequency PTA averages thresholds at 1000 Hz1000\,\text{Hz}, 2000 Hz2000\,\text{Hz}, and 4000 Hz4000\,\text{Hz} (or 1000,2000,3000 Hz1000, 2000, 3000\,\text{Hz} in industrial audiometry): HF-PTA=AC1000+AC2000+AC40003\text{HF-PTA} = \frac{\text{AC}_{1000} + \text{AC}_{2000} + \text{AC}_{4000}}{3}

  • Clinical Indication: Predicts speech perception difficulty in competing background noise and monitors noise-induced hearing loss, presbycusis, and ototoxic deterioration.

Audiometric Interpretation: Degree, Configuration & Type

Degree of Hearing Loss (ANSI / Clark / Goodman Scales)

Hearing loss degree is defined by pure-tone air-conduction thresholds across the speech frequency spectrum.

Degree CategoryPediatric Threshold Range (dB HL)Adult Threshold Range (dB HL)Functional Speech & Communication Impact
Normal−10 to 15 dB HL-10\,\text{to } 15\,\text{dB HL}−10 to 25 dB HL-10\,\text{to } 25\,\text{dB HL}Complete access to conversational speech cues in quiet and noise.
Slight / Minimal16 to 25 dB HL16\,\text{to } 25\,\text{dB HL}Classified as NormalSubtle speech morphology errors; misses unstressed words and faint consonants in classroom noise.
Mild26 to 40 dB HL26\,\text{to } 40\,\text{dB HL}26 to 40 dB HL26\,\text{to } 40\,\text{dB HL}Misses 25−40%25-40\% of speech signal; difficulty hearing faint or distant voices; struggles in noisy environments.
Moderate41 to 55 dB HL41\,\text{to } 55\,\text{dB HL}41 to 55 dB HL41\,\text{to } 55\,\text{dB HL}Misses 50−75%50-75\% of spoken conversational speech at normal vocal levels (60−65 dB SPL60-65\,\text{dB SPL}).
Moderately Severe56 to 70 dB HL56\,\text{to } 70\,\text{dB HL}56 to 70 dB HL56\,\text{to } 70\,\text{dB HL}Conversational speech inaudible without amplification; requires loud shouting; severe conversational disruption.
Severe71 to 90 dB HL71\,\text{to } 90\,\text{dB HL}71 to 90 dB HL71\,\text{to } 90\,\text{dB HL}Cannot hear conversational speech; perceives only loud shouts or amplified environmental sounds.
Profound≥91 dB HL\ge 91\,\text{dB HL}≥91 dB HL\ge 91\,\text{dB HL}Reliant on visual speech cues, ASL, cochlear implants; sound perceived primarily as tactile vibration.

Pediatric Imperative: The pediatric cutoff for normal hearing is strictly 15 dB HL15\,\text{dB HL}. Thresholds between 1616 and 25 dB HL25\,\text{dB HL} in children represent minimal/slight hearing loss, which compromises phonological awareness, academic performance, and reading comprehension.

Audiogram Configurations

  • Flat: Thresholds do not vary by more than 5−10 dB5-10\,\text{dB} across octaves.
  • Sloping: Thresholds worsen gradually across octave frequencies (5−10 dB/octave5-10\,\text{dB/octave}). Characteristic of sensory presbycusis and metabolic hearing loss.
  • Precipitous / Steeply Sloping: Thresholds drop sharply by ≥20 dB\ge 20\,\text{dB} per octave. Classic for acoustic trauma, ototoxicity, or apical-to-basal lesions.
  • Rising: Low-frequency thresholds are poorer than high-frequency thresholds. Classic for early Ménière's disease (endolymphatic hydrops) or middle-ear stiffness pathologies (otosclerosis, Eustachian tube dysfunction).
  • Cookie-Bite / Trough-Shaped: Mid-frequency thresholds (500−2000 Hz500-2000\,\text{Hz}) are abnormal, with preserved low and high frequencies. Often hereditary, for example autosomal dominant TECTA-related loss (DFNA8/12).
  • Reverse Cookie-Bite / Saucer: Mid-frequency hearing is preserved, while low and high frequencies are impaired.
  • Acoustic Notch: Sharp drop in thresholds at 3000 Hz3000\,\text{Hz}, 4000 Hz4000\,\text{Hz}, or 6000 Hz6000\,\text{Hz}, with significant recovery at 8000 Hz8000\,\text{Hz}. Pathognomonic for noise-induced hearing loss (NIHL), driven by outer ear acoustic resonance (2.5−3.5 kHz2.5-3.5\,\text{kHz}) focusing high sound pressure on the 4 kHz cochlear partition.
  • Corner Audiogram: Responses obtained only at the lowest frequencies (250−500 Hz250-500\,\text{Hz}) at the upper limits of the audiometer (90−110 dB HL90-110\,\text{dB HL}), with no response at all remaining frequencies.

Type of Hearing Loss

The relationship between air-conduction (AC) and bone-conduction (BC) thresholds determines the pathology type:

  1. Sensorineural Hearing Loss (SNHL):

    • Air-conduction thresholds: Abnormal (>25 dB HL> 25\,\text{dB HL} in adults, >15 dB HL> 15\,\text{dB HL} in children).
    • Bone-conduction thresholds: Abnormal (>25 dB HL> 25\,\text{dB HL} in adults, >15 dB HL> 15\,\text{dB HL} in children).
    • Air-Bone Gap (ABG): Absent or clinically insignificant (ABG≤10 dB\text{ABG} \le 10\,\text{dB}).
    • Pathophysiology: Outer and middle ear transmission is intact. Deficit resides in the cochlea (sensory, outer/inner hair cells) or VIII cranial nerve (neural).
  2. Conductive Hearing Loss (CHL):

    • Air-conduction thresholds: Abnormal (>25 dB HL> 25\,\text{dB HL} in adults, >15 dB HL> 15\,\text{dB HL} in children).
    • Bone-conduction thresholds: Completely normal (BC≤25 dB HL\text{BC} \le 25\,\text{dB HL} in adults, ≤15 dB HL\le 15\,\text{dB HL} in children).
    • Air-Bone Gap (ABG): Clinically significant (ABG≥15 dB\text{ABG} \ge 15\,\text{dB}).
    • Pathophysiology: Cochlear and retrocochlear structures are intact. Obstruction or mechanical lesion in the external canal, tympanic membrane, or ossicular chain (cerumen, otitis media, otosclerosis, ossicular discontinuity).
  3. Mixed Hearing Loss (MHL):

    • Air-conduction thresholds: Abnormal (>25 dB HL> 25\,\text{dB HL} in adults, >15 dB HL> 15\,\text{dB HL} in children).
    • Bone-conduction thresholds: Abnormal (>25 dB HL> 25\,\text{dB HL} in adults, >15 dB HL> 15\,\text{dB HL} in children).
    • Air-Bone Gap (ABG): Clinically significant (ABG≥15 dB\text{ABG} \ge 15\,\text{dB}).
    • Pathophysiology: Concurrent presence of sensorineural damage in the inner ear/nerve and conductive pathology in the outer/middle ear (e.g., presbycusis complicated by otitis media with effusion, or advanced otosclerosis with cochlear involvement).

Diagnostic Pitfalls & Vibrotactile Artifacts

Low-Frequency Vibrotactile Responses

At high stimulus intensities, bone vibrators (and occasionally supra-aural headphones) generate strong physical mechanical vibrations that activate somatosensory mechanoreceptors (cutaneous Pacinian and Meissner corpuscles) in the periosteum and skin, rather than stimulating the auditory hair cells.

  • Vibrotactile Threshold Limits:
    • 250 Hz250\,\text{Hz}: Tactile responses occur at ≥40−45 dB HL\ge 40-45\,\text{dB HL}.
    • 500 Hz500\,\text{Hz}: Tactile responses occur at ≥55−60 dB HL\ge 55-60\,\text{dB HL}.
    • 1000 Hz1000\,\text{Hz}: Tactile responses occur at ≥70−75 dB HL\ge 70-75\,\text{dB HL}.

The Severe Diagnostic Trap: A patient with bilateral profound sensorineural hearing loss (true hearing threshold >100 dB HL> 100\,\text{dB HL}) is tested with unmasked bone conduction. At 250 Hz250\,\text{Hz}, the patient feels the vibrator buzzing at 45 dB HL45\,\text{dB HL} and presses the response button. If the audiologist plots 45 dB HL45\,\text{dB HL} as a true hearing threshold, it creates a false 55 dB55\,\text{dB} air-bone gap (100 dB AC−45 dB BC100\,\text{dB AC} - 45\,\text{dB BC}), misdiagnosing profound SNHL as a mixed hearing loss. The clinician must explicitly ask the patient: "Did you hear a tone in your ear, or did you only feel a vibrating or buzzing sensation on your head?" Tactile responses must be documented on the audiogram with a tactile symbol ("V" or "T") or annotated as "no response at maximum output."

Ear Canal Collapse

Canal collapse occurs when pressure from supra-aural headphone cushions pinches the cartilaginous meatus closed. This creates an artificial conductive component of 15−30 dB15-30\,\text{dB} across 1000−4000 Hz1000-4000\,\text{Hz}. Always verify suspicious high-frequency air-bone gaps by switching to insert earphones, which physically stent the canal open.

Loading diagram...
Modified Hughson-Westlake Threshold Search & Frequency Testing Algorithm
Test Your Knowledge

During pure-tone threshold testing using the modified Hughson-Westlake procedure, a patient responds at 40 dB HL on an ascending run. After stepping down 10 dB to 30 dB HL, the patient fails to respond. The clinician increases the level to 35 dB HL (no response), and then to 40 dB HL, where the patient responds again. According to clinical threshold definitions, what action should the clinician take next?

A

Accept 40 dB HL as the threshold, as 2 out of 2 ascending presentations at this level elicited a response.

B

Decrease by 10 dB to 30 dB HL and then test at 35 dB HL to confirm whether 35 dB HL is the true threshold.

C

Increase by 5 dB to 45 dB HL to test for suprathreshold auditory fatigue.

D

Reject the run and switch immediately to a descending method of limits.

Test Your Knowledge

A 74-year-old patient is evaluated using TDH-49 supra-aural headphones. Audiometry reveals bilateral normal bone conduction thresholds (10-15 dB HL at all frequencies), but air conduction thresholds at 2000 Hz, 3000 Hz, and 4000 Hz are 40 dB HL bilaterally, yielding a 25-30 dB high-frequency air-bone gap. Tympanometry reveals normal Type A tympanograms with robust acoustic reflexes. What is the most appropriate next clinical step?

A

Refer the patient immediately for an otolaryngology surgical consultation for bilateral ossicular chain reconstruction.

B

Re-evaluate air-conduction thresholds using insert earphones to rule out collapsed external auditory meatuses.

C

Diagnose bilateral mixed hearing loss and recommend immediate binaural bone-conduction hearing aids.

D

Perform masked bone conduction at 6000 Hz and 8000 Hz to assess high-frequency sensorineural reserve.

Test Your Knowledge

A patient with suspected profound bilateral hearing loss is tested with an unmasked Radioear B-71 bone vibrator on the right mastoid. The patient signals a response at 45 dB HL at 250 Hz and at 55 dB HL at 500 Hz. However, air conduction thresholds at both frequencies exceed 100 dB HL bilaterally, and otoacoustic emissions and acoustic reflexes are absent. What is the most accurate interpretation of these bone conduction responses?

A

The patient exhibits a profound conductive hearing loss that requires urgent surgical ossicular reconstruction.

B

The responses represent true cochlear thresholds of the non-test left ear via cross-hearing.

C

They are likely vibrotactile responses detected through the skin rather than true auditory thresholds.

D

The responses demonstrate an organic acoustic notch resulting from acute high-intensity impulse noise exposure.