ANSI S3.22 Electroacoustic Analysis & Quality Control Standards

Key Takeaways

  • ANSI S3.22 provides standardized, reproducible electroacoustic test parameters in a 2-cc acoustic coupler within an anechoic sound box, enabling objective verification of manufacturer specifications, pre-dispensing quality control, and repair integrity.

  • Acoustic coupler selection is strictly transducer- and coupling-dependent: the HA-1 coupler utilizes a direct silicone putty seal for custom ITE/ITC devices and slim-tube/RIC custom tips, whereas the HA-2 coupler incorporates a rigid brass neck and 25 mm25\,\text{mm} of #13 tubing for standard BTE instruments.

  • Full-On Gain (FOG) electroacoustic metrics include Maximum OSPL90 (tolerance: ≤spec+3 dB\le \text{spec} + 3\,\text{dB}), High-Frequency Average OSPL90 (HFA-OSPL90 at 1000, 1600, and 2500 Hz; tolerance: ±4 dB\pm 4\,\text{dB}), and HFA-FOG with a 50 dB SPL50\,\text{dB SPL} input (tolerance: ±5 dB\pm 5\,\text{dB}).

  • The Reference Test Setting (RTS) defines a linear, non-saturating operating gain control position derived as RTS=HFA-OSPL90−77 dB\text{RTS} = \text{HFA-OSPL90} - 77\,\text{dB}; if an instrument's HFA-FOG is less than this calculated target, Full-On Gain serves as the RTS.

  • Quality control verification requires precise adherence to ANSI tolerances: Total Harmonic Distortion (≤spec+3%\le \text{spec} + 3\% at 500, 800, and 1600 Hz), Equivalent Input Noise (coupler noise minus HFA gain at RTS, ≤spec+3 dB\le \text{spec} + 3\,\text{dB}, typically <25−28 dB SPL<25-28\,\text{dB SPL}), battery current (≤spec+20%\le \text{spec} + 20\%), and induction coil telecoil response (HFA-SPLITS within ±6 dB\pm 6\,\text{dB}).

Last updated: October 2026

ANSI S3.22 Electroacoustic Analysis & Quality Control Standards

Electroacoustic analysis (EAA) of hearing aids provides an objective, standardized bench evaluation of hearing aid performance independent of the human ear canal. Governing all commercial hearing aids in the United States under Food and Drug Administration (FDA) regulations, the ANSI S3.22 standard ("Specification of Hearing Aid Characteristics") establishes rigorous measurement protocols, standardized acoustic couplers, and tolerances. Audiologists must perform electroacoustic quality control on newly received devices prior to patient delivery, following manufacturer repairs, and whenever troubleshooting acoustic distortion or intermittent function.


Purpose and Regulatory Mandate of ANSI S3.22

While real-ear probe-microphone measurements assess the acoustic output delivered to an individual's unique ear canal, ANSI S3.22 electroacoustic analysis serves fundamentally different clinical objectives:

  • Manufacturer Specification Compliance: Verifies that a newly manufactured or repaired hearing aid performs within the published engineering specifications filed with the FDA.
  • Pre-Fitting Quality Control: Ensures that an instrument is defect-free before the patient arrives for their fitting appointment, eliminating fitting delays caused by dead receivers, excessive internal noise, or harmonic distortion.
  • Repair Verification: Objectively determines whether a manufacturer repair successfully resolved an internal failure or introduced new transducer artifacts.
  • Diagnostic Troubleshooting: Objectively differentiates patient perceptual complaints (e.g., loudness intolerance, speech muffling) from true hardware degradation (e.g., failed internal dampers, cracked receiver suspensions, moisture damage).

Test Chamber Instrumentation & Coupler Acoustics

Electroacoustic analysis is conducted inside an anechoic sound-attenuating test box to prevent ambient room noise from corrupting low-frequency measurements. The test system comprises three primary electroacoustic components:

  1. Calibrated Reference Loudspeaker: Delivers controlled acoustic signals across the audio bandwidth (100100 to 10,000 Hz10,000\,\text{Hz}).
  2. Reference Regulating Microphone: Positioned adjacent to the hearing aid's microphone sound inlet; continuously samples sound field intensity to dynamically adjust (level) loudspeaker output, maintaining an exact input level (e.g., 50,60,70,50, 60, 70, or 90 dB SPL90\,\text{dB SPL}).
  3. 2-cc Coupler & Coupler Measurement Microphone: Simulates the nominal acoustic impedance of an average adult ear canal occluded by an earmold. The coupler terminates in a precision laboratory-grade measurement microphone.

Standardized 2-cc Coupler Types

Selecting the correct 2-cc coupler is critical to prevent severe acoustic artifacts in measured gain and output:

Coupler TypeMechanical ConfigurationClinical Indication / Transducer Application
HA-1 CouplerDirect connection; features an open aperture where the device nozzle is sealed directly using acoustic putty.Custom hearing aids (ITE, ITC, CIC, IIC); Receiver-in-the-Canal (RIC) custom molds or soft eartips; thin-tube custom earpieces.
HA-2 CouplerFeatures a 25 mm25\,\text{mm} rigid neck of standard #13 acoustic tubing (1.93 mm1.93\,\text{mm} ID) terminating in a brass coupling nipple.Standard Behind-the-Ear (BTE) hearing aids connected to traditional acoustic earmolds via standard #13 tubing.

Clinical Warning: Sealing a standard BTE instrument into an HA-1 coupler with putty without the standardized 25 mm25\,\text{mm} length of #13 tubing abolishes the natural acoustic quarter-wave resonance of the tubing, generating massive high-frequency measurement errors that violate ANSI testing protocols.


Maximum Output & Full-On Gain (FOG) Metrics

ANSI S3.22 specifies two primary measurements with the hearing aid volume/gain control set to Full-On and all adaptive digital features (noise reduction, feedback suppression, directionality) disabled in specialized "test mode."

1. OSPL90 (Output Sound Pressure Level with 90 dB SPL90\,\text{dB SPL} Input)

The OSPL90 curve evaluates the absolute maximum acoustic saturation output the instrument can generate across frequency when driven by an intense 90 dB SPL90\,\text{dB SPL} swept pure tone:

  • Maximum OSPL90: The highest sound pressure level recorded at any single frequency along the entire swept curve.
    • ANSI Tolerance: Measured Maximum OSPL90 must not exceed the manufacturer's specification by more than +3 dB+3\,\text{dB}.
    • Clinical Significance: This is a critical patient safety parameter; an instrument exceeding its maximum output specification risks delivering sound pressures that exceed the patient's loudness discomfort levels, causing acoustic trauma.
  • High-Frequency Average OSPL90 (HFA-OSPL90): The linear mathematical average of OSPL90 sound pressure levels at 1000 Hz1000\,\text{Hz}, 1600 Hz1600\,\text{Hz}, and 2500 Hz2500\,\text{Hz}: HFA-OSPL90=OSPL901000+OSPL901600+OSPL9025003\text{HFA-OSPL90} = \frac{\text{OSPL90}_{1000} + \text{OSPL90}_{1600} + \text{OSPL90}_{2500}}{3}
    • ANSI Tolerance: Measured HFA-OSPL90 must fall within ±4 dB\pm 4\,\text{dB} of the manufacturer's published specification.
    • Special Purpose Average (SPA): If the hearing aid's frequency response does not span the standard HFA frequencies, the manufacturer specifies three alternative frequencies separated by at least two-thirds of an octave.

2. Full-On Gain (FOG) at 50 dB SPL50\,\text{dB SPL} Input

Full-On Gain evaluates the total amplification capacity of the device when driven by a low-level input (50 dB SPL50\,\text{dB SPL}) to ensure the amplifier operates linearly without saturating:

  • High-Frequency Average Full-On Gain (HFA-FOG): The average gain across 1000 Hz1000\,\text{Hz}, 1600 Hz1600\,\text{Hz}, and 2500 Hz2500\,\text{Hz} with input at 50 dB SPL50\,\text{dB SPL} and gain control at maximum: HFA-FOG=FOG1000+FOG1600+FOG25003\text{HFA-FOG} = \frac{\text{FOG}_{1000} + \text{FOG}_{1600} + \text{FOG}_{2500}}{3}
    • ANSI Tolerance: Measured HFA-FOG must fall within ±5 dB\pm 5\,\text{dB} of the manufacturer's published specification.

Reference Test Setting (RTS) & Reference Test Gain (RTG)

Modern hearing aids cannot be tested for distortion, internal noise, or bandwidth at Full-On Gain because a moderate input signal (60 dB SPL60\,\text{dB SPL}) would drive the amplifier into deep non-linear saturation. The Reference Test Setting (RTS) defines a standardized gain control position that models typical conversational listening without saturation.

Mathematical Formulation of RTS

The Reference Test Setting is calculated directly from the measured HFA-OSPL90: RTS Target Gain=HFA-OSPL90−77 dB\text{RTS Target Gain} = \text{HFA-OSPL90} - 77\,\text{dB}

Step-by-Step Procedure for Setting RTS:

  1. Determine the measured HFA-OSPL90 from the initial 90 dB SPL90\,\text{dB SPL} sweep.
  2. Calculate the target output for a 60 dB SPL60\,\text{dB SPL} input: Target Coupler Output=60 dB+RTS Target Gain=60 dB+(HFA-OSPL90−77 dB)=HFA-OSPL90−17 dB\text{Target Coupler Output} = 60\,\text{dB} + \text{RTS Target Gain} = 60\,\text{dB} + (\text{HFA-OSPL90} - 77\,\text{dB}) = \text{HFA-OSPL90} - 17\,\text{dB}
  3. Present a 60 dB SPL60\,\text{dB SPL} input tone (or speech-shaped noise) at the HFA frequencies (1000,1600,2500 Hz1000, 1600, 2500\,\text{Hz}).
  4. Adjust the hearing aid gain control downward until the measured HFA output matches HFA-OSPL90−17 dB\text{HFA-OSPL90} - 17\,\text{dB} within ±1.5 dB\pm 1.5\,\text{dB}.
  5. The RTS Exception Rule: If the hearing aid's Full-On Gain (HFA-FOG) is less than (HFA-OSPL90−77 dB)(\text{HFA-OSPL90} - 77\,\text{dB}), the volume control is left at Full-On Gain (i.e., RTS=FOG\text{RTS} = \text{FOG}). In mild-gain hearing aids, full-on gain does not provide enough amplification to saturate the amplifier at 60 dB SPL60\,\text{dB SPL}.

Frequency Response Curve & Frequency Bandwidth (f1f_1 & f2f_2)

With the hearing aid set to RTS, a 60 dB SPL60\,\text{dB SPL} input signal (or 50 dB SPL50\,\text{dB SPL} if specified) is presented to record the Reference Test Frequency Response Curve:

  1. Determine HFA Response: Calculate the average sound pressure level across 1000,1600,1000, 1600, and 2500 Hz2500\,\text{Hz} on the frequency response curve.
  2. Establish the Response Limit Line: Subtract 20 dB20\,\text{dB} from the calculated HFA response: Response Limit=HFA Response−20 dB\text{Response Limit} = \text{HFA Response} - 20\,\text{dB} Draw a horizontal line across the frequency chart at this exact sound pressure level.
  3. Identify Cutoff Frequencies:
    • Low-Frequency Cutoff (f1f_1): The lowest frequency at which the frequency response curve crosses the 20 dB20\,\text{dB} response limit line.
    • High-Frequency Cutoff (f2f_2): The highest frequency at which the frequency response curve crosses the 20 dB20\,\text{dB} response limit line.
  4. Frequency Range: Stated as f1f_1 to f2f_2. ANSI S3.22 reports this range for information only; it carries no tolerance.
    • Response-Curve Tolerance: Instead, the frequency response curve must stay within ±4 dB\pm 4\,\text{dB} of specification in the low band (from the higher of 1.25f11.25 f_1 or 200 Hz200\,\text{Hz} up to 2000 Hz2000\,\text{Hz}) and within ±6 dB\pm 6\,\text{dB} in the high band (from 2000 Hz2000\,\text{Hz} to the lower of 4000 Hz4000\,\text{Hz} or 0.8f20.8 f_2).

Total Harmonic Distortion (THD)

Harmonic distortion occurs when non-linear amplifier or transducer behavior generates unwanted acoustic energy at integer multiples of the input fundamental frequency (2f,3f,4f2f, 3f, 4f):

  • Measurement Parameters: Measured with the hearing aid positioned at RTS:
    • At 500 Hz500\,\text{Hz} using a 70 dB SPL70\,\text{dB SPL} input.
    • At 800 Hz800\,\text{Hz} using a 70 dB SPL70\,\text{dB SPL} input.
    • At 1600 Hz1600\,\text{Hz} using a 65 dB SPL65\,\text{dB SPL} input.
  • The 12 dB12\,\text{dB} Rule Exemption: If the frequency response curve displays a sharp peak at 2f2f (the second harmonic) that rises 12 dB12\,\text{dB} or more above the response at the fundamental frequency ff, harmonic distortion at that test frequency is exempt from compliance testing. Acoustic resonant peaks artificially inflate measured distortion percentages through acoustic resonance rather than true transducer non-linearity.
  • ANSI Tolerance: Measured THD at each frequency must not exceed the manufacturer's published specification by more than +3%+3\% (e.g., if the manufacturer specification is 2%2\%, a measured distortion up to 5%5\% passes; 5.1%5.1\% fails).

Internal Circuit Noise: Equivalent Input Noise (EIN)

Equivalent Input Noise (EIN) quantifies the internal thermal and electrical circuit noise generated by the microphone and pre-amplifier, expressed as the equivalent acoustic input level required to generate that noise at the output:

Mathematical Formulation

EIN=Lnoise−HFA gain at RTS\text{EIN} = L_{\text{noise}} - \text{HFA gain at RTS} where:

  • LnoiseL_{\text{noise}} is the HFA sound pressure level measured in the 2-cc coupler with the input signal turned off while the hearing aid is at RTS.
  • The HFA gain is measured at RTS with a 50 dB SPL50\,\text{dB SPL} input (the HFA coupler output minus 50 dB).

ANSI Tolerance & Clinical Rejection Criteria

  • ANSI Tolerance: Measured EIN must not exceed the manufacturer's published specification by more than +3 dB SPL+3\,\text{dB SPL}.
  • Clinical Threshold: High-performing modern digital hearing aids exhibit an EIN<25−28 dB SPL\text{EIN} < 25-28\,\text{dB SPL}. When an instrument's EIN exceeds 3030 to 32 dB SPL32\,\text{dB SPL}, patients with normal or near-normal low-frequency thresholds will perceive an audible, distracting internal "hissing," "frying," or "rushing water" circuit noise, resulting in immediate device rejection.

Battery Drain, Telecoil & Directional Quality Verification

Battery Current Drain Test

  • Measurement Setup: The hearing aid's physical battery is replaced with a calibrated "battery simulator pill" connected to a regulated DC power supply delivering nominal operating voltage (1.30 V1.30\,\text{V} for Zinc-Air; 3.80 V3.80\,\text{V} for rechargeable Lithium-Ion).
  • Protocol: Evaluated at RTS using a 1000 Hz1000\,\text{Hz} pure tone at 65 dB SPL65\,\text{dB SPL}.
  • ANSI Tolerance: Measured current drain (in milliamperes, mA\text{mA}) must not exceed the manufacturer's specification by more than +20%+20\%.
  • Battery Life Estimation Formula: Battery Operating Hours=Battery Capacity (mAh)Measured Current Drain (mA)\text{Battery Operating Hours} = \frac{\text{Battery Capacity (mAh)}}{\text{Measured Current Drain (mA)}} Clinical Example: A Size 312 Zinc-Air battery with a nominal capacity of 160 mAh160\,\text{mAh} powered in a hearing aid with a measured current drain of 1.6 mA1.6\,\text{mA} yields an expected battery life of 1601.6=100 hours\frac{160}{1.6} = 100\,\text{hours} (approximately 6−7 days6-7\,\text{days} at 15 hours/day15\,\text{hours/day}).

Telecoil / Induction Coil Verification (SPLITS)

  • Simulator Field: The hearing aid is switched to telecoil mode and placed inside a standardized Telephone Magnetic Field Simulator (TMFS) producing a magnetic field strength of 31.6 mA/m31.6\,\text{mA/m}.
  • SPLITS (Coupler Sound Pressure Level for an Inductive Telephone Simulator): Measured across frequency with the gain control at the reference test setting (RTS).
  • HFA-SPLITS: Average of SPLITS at 1000,1600,1000, 1600, and 2500 Hz2500\,\text{Hz}.
    • ANSI Tolerance: Measured HFA-SPLITS must fall within ±6 dB\pm 6\,\text{dB} of manufacturer specification.
  • Relative Simulated Equivalent Telephone Sensitivity (RSETS): Compares telecoil output with microphone output: RSETS=HFA-SPLITS−(RTG+60 dB)\text{RSETS} = \text{HFA-SPLITS} - (\text{RTG} + 60\,\text{dB}). A value near 0 dB0\,\text{dB} means the telecoil sounds about as loud as the microphone does for typical speech.

Directional Microphone Test Box Verification

Directional microphone verification compares frequency sweeps presented at 0∘0^\circ azimuth (front) versus 180∘180^\circ azimuth (rear) inside the test box:

  • Front-to-Back Ratio (FBR): The difference in output between 0∘0^\circ and 180∘180^\circ, typically providing 1515 to 25 dB25\,\text{dB} of low-to-mid frequency attenuation for sounds originating from the rear.
  • Clinical Artifact: Physical occlusion of the rear microphone port by debris, wax, or skin oil collapses the directional polar pattern into an omnidirectional mode, eliminating background noise suppression.

ANSI S3.22 Tolerance Summary Reference Table

Electroacoustic ParameterInput Level & ConditionsMeasurement SettingANSI S3.22 Tolerance Limit
Maximum OSPL9090 dB SPL90\,\text{dB SPL} swept pure toneFull-On Gain≤Manufacturer Spec+3 dB\le \text{Manufacturer Spec} + 3\,\text{dB}
HFA-OSPL9090 dB SPL90\,\text{dB SPL} (1000,1600,2500 Hz1000, 1600, 2500\,\text{Hz})Full-On Gain±4 dB\pm 4\,\text{dB} of Manufacturer Spec
HFA-FOG50 dB SPL50\,\text{dB SPL} (1000,1600,2500 Hz1000, 1600, 2500\,\text{Hz})Full-On Gain±5 dB\pm 5\,\text{dB} of Manufacturer Spec
Reference Test Setting (RTS)Formula: HFA-OSPL90−77 dB\text{HFA-OSPL90} - 77\,\text{dB}Gain Control AdjustedOutput within ±1.5 dB\pm 1.5\,\text{dB} of target
Frequency Range (f1f_1 to f2f_2)60 dB SPL60\,\text{dB SPL} input (at RTS)RTSInformational only (no tolerance)
Frequency Response Curve60 dB SPL60\,\text{dB SPL} input (at RTS)RTS±4 dB\pm 4\,\text{dB} low band; ±6 dB\pm 6\,\text{dB} high band
Harmonic Distortion (THD)500/800 Hz500/800\,\text{Hz} at 70 dB70\,\text{dB}; 1600 Hz1600\,\text{Hz} at 65 dB65\,\text{dB}RTS≤Manufacturer Spec+3%\le \text{Manufacturer Spec} + 3\%
Equivalent Input Noise (EIN)Ambient sound off (0 dB SPL0\,\text{dB SPL})RTS (noise minus HFA gain at RTS)≤Manufacturer Spec+3 dB SPL\le \text{Manufacturer Spec} + 3\,\text{dB SPL}
Battery Current Drain1000 Hz1000\,\text{Hz} at 65 dB SPL65\,\text{dB SPL}RTS≤Manufacturer Spec+20%\le \text{Manufacturer Spec} + 20\%
HFA-SPLITS (Telecoil)31.6 mA/m31.6\,\text{mA/m} magnetic fieldRTS, Telecoil Mode±6 dB\pm 6\,\text{dB} of Manufacturer Spec
Loading diagram...
ANSI S3.22 Electroacoustic Analysis & Quality Control Decision Flowchart
Test Your Knowledge

An audiologist performs an ANSI S3.22 electroacoustic analysis on a newly received behind-the-ear hearing aid. The measured HFA-OSPL90 is 122 dB SPL122\,\text{dB SPL}, and the measured HFA-FOG is 56 dB56\,\text{dB}. What is the calculated Reference Test Setting (RTS) target gain, and what target output must be achieved in the 2-cc coupler when presenting a 60 dB SPL60\,\text{dB SPL} input?

A

RTS Target Gain is 56 dB56\,\text{dB}; Target Coupler Output is 116 dB SPL116\,\text{dB SPL}.

B

RTS Target Gain is 62 dB62\,\text{dB}; Target Coupler Output is 122 dB SPL122\,\text{dB SPL}.

C

RTS Target Gain is 35 dB35\,\text{dB}; Target Coupler Output is 95 dB SPL95\,\text{dB SPL}.

D

RTS Target Gain is 45 dB45\,\text{dB}; Target Coupler Output is 105 dB SPL105\,\text{dB SPL}.

Test Your Knowledge

Which acoustic coupler configuration is required by ANSI S3.22 when performing quality control electroacoustic analysis on a custom In-the-Ear (ITE) hearing aid?

A

An HA-1 coupler, in which the hearing aid canal tip is sealed directly into the coupler aperture using acoustic putty.

B

An HA-2 coupler featuring a 25 mm25\,\text{mm} length of #13 plastic tubing connecting the receiver to the measurement microphone.

C

An HA-4 coupler connected to a high-impedance artificial ear simulator.

D

An occluded ear simulator (IEC 60318-4) fitted with an unvented pediatric ear tip.

Test Your Knowledge

An audiologist tests a repaired hearing aid against the manufacturer's published ANSI S3.22 specification sheet. The specification sheet indicates an Equivalent Input Noise (EIN) of 26 dB SPL26\,\text{dB SPL} and a Total Harmonic Distortion (THD) of 2.0%2.0\% at 800 Hz800\,\text{Hz}. Which of the following measured test box results indicates that the hearing aid fails to meet ANSI S3.22 compliance?

A

Measured EIN is 28.5 dB SPL28.5\,\text{dB SPL}; measured THD at 800 Hz800\,\text{Hz} is 4.5%4.5\%.

B

Measured EIN is 30.5 dB SPL30.5\,\text{dB SPL}; measured THD at 800 Hz800\,\text{Hz} is 3.5%3.5\%.

C

Measured EIN is 27.0 dB SPL27.0\,\text{dB SPL}; measured THD at 800 Hz800\,\text{Hz} is 5.0%5.0\%.

D

Measured EIN is 29.0 dB SPL29.0\,\text{dB SPL}; measured THD at 800 Hz800\,\text{Hz} is 4.8%4.8\%.