1.4 Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges

Key Takeaways

  • Analytical reagent (ACS) grade chemicals are suitable for most clinical analyses, while technical or commercial grade chemicals must never be used to prepare reagents.

  • CLSI clinical laboratory reagent water has a resistivity of at least 10 MΩ·cm at 25 °C, under 500 ppb total organic carbon, and under 10 CFU/mL of bacteria.

  • Deionization removes dissolved ions but not organics or bacteria, so it is combined with reverse osmosis, carbon, filtration, or UV oxidation.

  • A TC pipette such as the Sahli pipette must be rinsed out with diluent, while a blow-out pipette is marked by an etched or frosted band near the top.

  • Relative centrifugal force equals 1.118 × 10^-5 × rotor radius in centimeters × rpm squared, so procedures should specify RCF rather than rpm.

Last updated: October 2026

Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges

The Clinical Chemistry blueprint assigns 5% to reagent preparation and laboratory mathematics. Reagent quality sets the ceiling on analytical quality: an assay run with the wrong chemical grade, contaminated water, or a misread pipette will fail quality control no matter how good the analyzer is. This section covers chemical grades and standards, laboratory water, solution preparation with hydrates, glassware and pipettes, balances and desiccants, and centrifuge calculations. Dilution, molarity, and normality calculations are worked in the quality control and laboratory mathematics section.


1. Chemical Grades and Reference Materials

GradePurity and Use
Ultrapure (spectrograde, HPLC grade, nanograde)Extra purification steps for sensitive techniques such as chromatography, atomic absorption, and fluorometry
Analytical reagent (AR) / ACS reagent gradeMeets American Chemical Society specifications; impurity limits are stated on the label; suitable for most laboratory analyses
USP / NF gradeMeets United States Pharmacopeia or National Formulary standards for drugs; free of harmful impurities but may not be pure enough for analysis
Chemically pure (CP) / pure gradeImpurity limits are not stated; not recommended for clinical analysis without verification
Technical or commercial gradeFor industrial and cleaning use only; never for reagent preparation
  • A primary standard is a highly purified chemical that can be weighed directly to make a solution of exactly known concentration. It must be stable, non-hygroscopic, and of known composition.
  • A secondary standard has its concentration determined by comparison with a primary standard.
  • Standard reference materials (SRMs) from the US National Institute of Standards and Technology (NIST) and other certified reference materials give calibration traceability (see the calibration hierarchy in the quality control section).
  • Every reagent container is labeled with its name, concentration, date received, date opened or prepared, expiration date, storage conditions, hazard warnings, and the preparer's initials.

2. Laboratory Water

Water Quality Categories

The Clinical and Laboratory Standards Institute (CLSI) replaced the older Type I, II, and III classification with categories based on intended use:

  • Clinical laboratory reagent water (CLRW): The default water for reagent preparation and most testing. Typical specifications are resistivity of at least 10 MΩ·cm at 25 °C, total organic carbon below 500 ppb, microbial content below 10 CFU/mL, and filtration through a 0.22 µm filter.
  • Special reagent water (SRW): Water with extra specifications for a particular method, such as low organics for HPLC or low trace metals for metal analysis.
  • Instrument feed water: Water meeting an analyzer manufacturer's specifications.
  • Water for autoclaves and washers, and commercially packaged purified water, have their own specifications.

Older texts still cited by reviewers use Type I water (highest purity, at least 10 MΩ·cm; for trace metals, enzymes, electrolytes, and buffers), Type II (at least 1 MΩ·cm; general chemistry, hematology, and microbiology media), and Type III (at least 0.1 MΩ·cm; glassware washing).

Purification Methods

MethodWhat It RemovesLimitations
DistillationMost ions, organics, microorganisms, and pyrogensVolatile impurities and dissolved gases (CO2, chlorine) carry over; energy intensive
Deionization (ion exchange)Dissolved ions (cation and anion resins exchange them for H+ and OH-)Does not remove organics, bacteria, or particles; resins can grow bacteria
Reverse osmosisAbout 90% to 99% of ions, organics, particles, and microorganismsOften used as pretreatment; membranes foul over time
Activated carbonOrganic compounds and chlorineUsed as a pretreatment step
Ultrafiltration and 0.22 µm filtrationParticles, bacteria, and (ultrafiltration) pyrogensFinal polishing steps
Ultraviolet oxidationBreaks down organic compounds and kills bacteriaUsed with other steps

Resistivity is monitored continuously or daily, and bacterial counts are checked on a schedule. Falling resistivity means the deionizer cartridges are exhausted.


3. Preparing Solutions and Hydrates

  • Percent solutions: w/v (grams per 100 mL), v/v (mL per 100 mL), and w/w (grams per 100 g). Physiologic saline is 0.85% to 0.9% NaCl (w/v), or 8.5 to 9 g/L.
  • Molarity is moles of solute per liter of solution. Normality is equivalents per liter. Molality is moles per kilogram of solvent, which does not change with temperature.
  • Hydrates: When the available chemical is a hydrate, use the hydrate's molecular weight, or scale an anhydrous mass by the ratio of molecular weights:

g of hydrate=g of anhydrous compound×MW of hydrateMW of anhydrous compound\text{g of hydrate} = \text{g of anhydrous compound} \times \frac{\text{MW of hydrate}}{\text{MW of anhydrous compound}}

  • Worked example: Prepare 500 mL of 0.1 M copper sulfate from CuSO4·5H2O (MW 249.7). Grams = 0.1 mol/L × 0.5 L × 249.7 g/mol = 12.48 g. If you used the anhydrous MW (159.6) by mistake, you would weigh only 7.98 g and the solution would be about 36% too weak.
  • Terms: Hygroscopic substances absorb moisture from the air; deliquescent substances absorb so much that they dissolve (for example, sodium hydroxide pellets); efflorescent substances lose their water of hydration to the air. Store hygroscopic chemicals and primary standards in a desiccator with a drying agent such as silica gel or anhydrous calcium chloride.

4. Glassware, Plasticware, and Pipettes

Glass and Plastic Types

  • Borosilicate glass (such as Pyrex and Kimax) resists heat and most chemicals and is the standard for laboratory glassware.
  • Low-actinic (amber or red) glassware protects light-sensitive analytes such as bilirubin and vitamin A.
  • Soda-lime (flint) glass is inexpensive but less resistant to heat and chemicals. Boron-free alkali-resistant glass is used for strong alkaline solutions.
  • Plastics: Polypropylene and polyethylene resist most chemicals; polystyrene is clear but cracks with many solvents; polycarbonate is strong; fluorocarbon (Teflon) is nearly inert.
  • Class A volumetric glassware meets the strictest tolerance standards (traceable to NIST) and is used for preparing standards. Graduated cylinders are much less accurate than volumetric flasks.

Pipette Terminology

  • TD (to deliver): Calibrated to deliver the stated volume when drained.
  • TC (to contain): Holds the stated volume; the contents must be rinsed out completely with diluent. The Sahli hemoglobin pipette is a TC pipette.
  • Blow-out: Marked by an etched or frosted band (or double ring) near the top; the last drop is expelled.
  • Self-draining: The last drop is allowed to remain in the tip.

Pipette Types

PipetteDescriptionUse
Volumetric (transfer)Single bulb, one calibration mark, TD, self-drainingMost accurate fixed volumes; standards and controls
Ostwald-FolinBulb close to the delivery tip; blow-outViscous fluids such as whole blood and serum
Mohr (measuring)Graduated, but the graduations stop above the tip; not blow-outDelivering a volume between two marks
Serological (measuring)Graduated to the tip; usually blow-outDelivering variable volumes; less precise than volumetric
Air-displacement micropipetteAdjustable volume with disposable tipsRoutine microliter volumes
Positive-displacement micropipettePiston in direct contact with the liquidViscous, dense, or volatile liquids
  • Read the bottom of the meniscus at eye level for clear liquids. Hold pipettes vertically while drawing and delivering.
  • Reverse pipetting (aspirate past the first stop, dispense to the first stop) improves accuracy with viscous or foaming liquids.
  • Pipettes are calibrated gravimetrically by weighing delivered distilled water and correcting for its density at the measured temperature.
  • Glassware is cleaned with laboratory detergent, rinsed thoroughly with tap water, then rinsed with reagent-grade water. Glassware for trace-metal work is soaked in dilute nitric acid. Chromic acid (dichromate) cleaning solution is now avoided because hexavalent chromium is toxic and carcinogenic.

5. Balances and Centrifuges

  • Analytical balances weigh to 0.0001 g (0.1 mg) inside a draft shield. They are leveled, kept away from vibration and air currents, and checked with certified weights. Top-loading balances are used for less exacting work.
  • Relative centrifugal force (RCF) depends on both speed and rotor radius:

RCF (×g)=1.118×10−5×r (cm)×(rpm)2\text{RCF } (\times g) = 1.118 \times 10^{-5} \times r \text{ (cm)} \times (\text{rpm})^2

  • Worked example: A rotor radius of 10 cm at 3,000 rpm gives RCF = 1.118 × 10^-5 × 10 × 9,000,000 ≈ 1,006 × g. Procedures should state RCF, not rpm, because the same rpm produces different forces in rotors of different radii.
  • Centrifuges are balanced with tubes of equal mass opposite each other, the lid is kept closed until the rotor stops, and speed is verified with a tachometer as part of maintenance.
Test Your Knowledge

A medical technologist must prepare 250 mL of a 0.2 M magnesium sulfate solution, but only MgSO4·7H2O (MW 246.5) is available. The anhydrous MgSO4 has an MW of 120.4. How many grams of the hydrate should be weighed?

A

6.02 g

B

12.33 g

C

24.65 g

D

49.30 g

Test Your Knowledge

Which pipette has graduations that stop above the tip, so the liquid is delivered between two marks and is not blown out?

A

Serological pipette

B

Volumetric transfer pipette

C

Ostwald-Folin pipette

D

Mohr pipette

Test Your Knowledge

The resistivity reading on a laboratory water system falls from 15 MΩ·cm to 0.8 MΩ·cm over two weeks. What is the most likely cause, and what is the effect on the water?

A

The deionization resin cartridges are exhausted, so the water now contains more dissolved ions

B

Bacteria have grown in the storage tank, which raises the water's resistivity

C

The 0.22 µm final filter is clogged, so particles are now passing into the water

D

The UV lamp has failed, so organic carbon has increased and resistivity has risen