12.2 Temperature Measurement, Humidity, Gas Analysis and SI Units

Key Takeaways

  • Thermistors are semiconductors whose resistance usually falls non-linearly as temperature rises, whereas thermocouples generate a voltage at the junction of two dissimilar metals (the Seebeck effect).

  • Accurate core temperature sites include the pulmonary artery, distal oesophagus, nasopharynx and tympanic membrane (contact probe); bladder and rectal temperatures lag during rapid changes.

  • Absolute humidity is the mass of water vapour per volume of gas, about 44 mg/L when air is fully saturated at 37 °C; heat and moisture exchangers typically deliver about 25-35 mg/L.

  • Infrared analysers measure carbon dioxide, nitrous oxide and volatile agents because molecules with two or more different atoms absorb infrared light; oxygen does not absorb infrared and is measured by paramagnetic or fuel-cell analysers.

  • Paramagnetic oxygen analysis exploits the attraction of oxygen to a magnetic field (unpaired outer electrons) and responds quickly enough for breath-by-breath measurement, unlike the slower galvanic fuel cell.

Last updated: October 2026

12.2 Temperature Measurement, Humidity, Gas Analysis and SI Units

Temperature Measurement

Units and Scales

The SI unit of temperature is the kelvin (K). A change of 1 K equals a change of 1 °C, and T(K)=T(°C)+273.15T(\text{K}) = T(°\text{C}) + 273.15. The triple point of water (273.16 K) is the reference point. Absolute zero is 0 K.

Types of Thermometer

DevicePrincipleFeatures
Liquid expansion (mercury, alcohol)Liquid expands with temperatureSlow; mercury now largely withdrawn
Bimetallic stripTwo metals with different expansion coefficients bendUsed in dial thermometers and variable-bypass vaporizer temperature compensation
Resistance thermometerResistance of a platinum wire rises almost linearly with temperatureAccurate; used as a reference
ThermistorSemiconductor (metal oxide) whose resistance usually falls non-linearly with temperatureSmall, rapid response, cheap; used in pulmonary artery catheters for thermodilution
ThermocoupleJunction of two dissimilar metals (for example copper and constantan) generates a voltage proportional to temperature difference (Seebeck effect)Needs a reference junction; small and robust
Infrared (tympanic, skin)Measures emitted infrared radiationRapid; tympanic readings depend on technique

Resistance thermometers and thermistors are usually used in a Wheatstone bridge circuit to measure small changes in resistance accurately.

Core Temperature Sites

  • Pulmonary artery: the reference standard for core temperature.
  • Distal oesophagus: probe in the lower third, behind the heart; accurate and practical during general anaesthesia.
  • Nasopharynx: close to the internal carotid artery; reflects brain temperature but affected by leaks around the airway.
  • Tympanic membrane: contact thermistors are accurate; infrared devices are variable.
  • Bladder and rectum: acceptable but lag behind rapid changes such as rewarming after bypass.
  • Skin and axilla: reflect peripheral temperature and underestimate core temperature.

Monitoring is recommended during anaesthesia lasting more than about 30 minutes, and active warming should keep core temperature above 36.0 °C (for example NICE guidance on inadvertent perioperative hypothermia).

Humidity

TermDefinition
Absolute humidityMass of water vapour per unit volume of gas (mg/L or g/m³)
Relative humidityActual water vapour mass divided by the mass required to saturate the gas at that temperature (%)
Dew pointTemperature at which gas becomes saturated and water condenses
Saturated vapour pressure of waterAbout 6.3 kPa (47 mmHg) at 37 °C

Fully saturated gas at 37 °C contains about 44 mg/L of water; at 20 °C it holds only about 17 mg/L. Cooling saturated gas therefore causes condensation (rain-out) in breathing circuits.

Why Humidification Matters

Normal breathing warms and humidifies inspired air in the upper airway so that gas reaching the isothermic saturation boundary (just below the carina) is fully saturated at 37 °C. Dry medical gases delivered through a tracheal tube bypass the nose, causing heat and water loss, thickened secretions, impaired mucociliary clearance, atelectasis and airway damage.

Measuring Humidity

  • Hair hygrometer: a hair lengthens as humidity rises; works between about 30% and 90% relative humidity.
  • Wet and dry bulb hygrometer: evaporation cools the wet bulb; the temperature difference indicates relative humidity.
  • Regnault's hygrometer: cooling a polished surface until dew forms gives the dew point.
  • Electronic sensors: humidity changes the capacitance or resistance of a sensing material.

Humidification Devices

DeviceTypical outputNotes
Heat and moisture exchanger (HME)About 25-35 mg/LPassive; adds dead space (about 30-90 mL) and resistance; may include a bacterial and viral filter (HMEF)
Heated water bath humidifierUp to about 44 mg/LActive; risk of scalding, infection and overhydration; condensation in tubing
Nebulisers (jet, ultrasonic)Can exceed saturation with dropletsUsed for drug delivery; ultrasonic devices produce very small droplets

Gas Analysis

Infrared Absorption Spectroscopy

Molecules containing two or more different atoms (carbon dioxide, nitrous oxide, volatile agents, water vapour) absorb infrared radiation at characteristic wavelengths. Carbon dioxide is measured at about 4.3 μm\mu\text{m}; volatile agents are identified using wavelengths around 8-12 μm\mu\text{m}. The amount absorbed follows the Beer-Lambert law.

  • Side-stream analysers aspirate about 50-200 mL/min from the breathing system, which must be returned or scavenged; there is a short delay.
  • Main-stream analysers place the sensor at the airway, giving an immediate signal but adding weight and dead space.
  • Collision broadening: nitrous oxide broadens the carbon dioxide absorption band, causing slight over-reading of carbon dioxide unless corrected.
  • Oxygen, nitrogen and other symmetrical diatomic gases do not absorb infrared, so other methods are needed.

Oxygen Analysis

MethodPrincipleResponse timeNotes
ParamagneticOxygen has unpaired electrons and is attracted into a magnetic field; modern analysers use a pressure difference in a switched magnetic fieldRapid (breath-by-breath)Standard in anaesthetic monitors
Galvanic fuel cellOxygen is reduced at a cathode, generating a current proportional to oxygen partial pressureSlow (about 20-30 seconds)Used in anaesthetic machines; limited lifespan; no external power needed
Polarographic (Clark) electrodeOxygen reduced at a platinum cathode with an applied voltage; current proportional to PO2P\text{O}_2ModerateUsed in blood gas analysers

Other Methods

  • Piezoelectric analysis: a quartz crystal coated with oil absorbs volatile agent, changing its resonant frequency; it cannot identify which agent is present.
  • Mass spectrometry: separates ionised gas molecules by mass-to-charge ratio; accurate for multiple gases but expensive.
  • Raman spectroscopy: measures scattered light shifted in wavelength by gas molecules.
  • Chemiluminescence is used to measure nitric oxide.

SI Units and Conversions

QuantitySI unitCommon conversions
PressurePascal (Pa = N/m²)1 kPa = 7.5 mmHg = 10.2 cmH₂O; 1 atmosphere = 101.325 kPa = 760 mmHg; 1 bar = 100 kPa
ForceNewton (N = kg·m/s²)
EnergyJoule (J)1 calorie = 4.18 J
PowerWatt (W = J/s)
Amount of substanceMole (mol)
TemperatureKelvin (K)
Luminous intensityCandela (cd)

The seven SI base units are the metre, kilogram, second, ampere, kelvin, mole and candela. All others, including the pascal, newton, joule, volt and watt, are derived units.

Tip

A quick pressure check: 1 kPa is about 7.5 mmHg, so a normal PaCO2P_a\text{CO}_2 of 5.3 kPa is about 40 mmHg and a PaO2P_a\text{O}_2 of 13 kPa is about 98 mmHg.

Test Your Knowledge

Which analyser is suitable for breath-by-breath measurement of inspired and expired oxygen in an anaesthetic monitor?

A

An infrared analyser using the 4.3 micrometre absorption band

B

A galvanic fuel cell, which responds within about 1 second

C

A paramagnetic analyser, which responds rapidly

D

A piezoelectric quartz crystal coated with oil

Test Your Knowledge

Which statement about temperature measurement is correct?

A

A thermocouple uses a semiconductor whose resistance falls as temperature rises

B

Bladder temperature reflects core temperature most accurately during rapid rewarming after cardiopulmonary bypass

C

Axillary temperature is a reliable measure of core temperature during anaesthesia and surgery

D

A thermocouple generates a temperature-dependent voltage at a junction of two dissimilar metals

Test Your Knowledge

Gas fully saturated with water vapour at 37 °C contains about how much water per litre, and how does a heat and moisture exchanger compare?

A

About 44 mg/L at 37 °C; a heat and moisture exchanger typically delivers about 25-35 mg/L

B

About 17 mg/L at 37 °C; a heat and moisture exchanger delivers more than 60 mg/L

C

About 100 mg/L at 37 °C; a heat and moisture exchanger provides complete saturation at body temperature

D

About 6.3 mg/L at 37 °C, matching the saturated vapour pressure in kilopascals

Sections you finish are checked off in the contents.