12.2 Temperature Measurement, Humidity, Gas Analysis and SI Units
Key Takeaways
Thermistors are semiconductors whose resistance usually falls non-linearly as temperature rises, whereas thermocouples generate a voltage at the junction of two dissimilar metals (the Seebeck effect).
Accurate core temperature sites include the pulmonary artery, distal oesophagus, nasopharynx and tympanic membrane (contact probe); bladder and rectal temperatures lag during rapid changes.
Absolute humidity is the mass of water vapour per volume of gas, about 44 mg/L when air is fully saturated at 37 °C; heat and moisture exchangers typically deliver about 25-35 mg/L.
Infrared analysers measure carbon dioxide, nitrous oxide and volatile agents because molecules with two or more different atoms absorb infrared light; oxygen does not absorb infrared and is measured by paramagnetic or fuel-cell analysers.
Paramagnetic oxygen analysis exploits the attraction of oxygen to a magnetic field (unpaired outer electrons) and responds quickly enough for breath-by-breath measurement, unlike the slower galvanic fuel cell.
12.2 Temperature Measurement, Humidity, Gas Analysis and SI Units
Temperature Measurement
Units and Scales
The SI unit of temperature is the kelvin (K). A change of 1 K equals a change of 1 °C, and . The triple point of water (273.16 K) is the reference point. Absolute zero is 0 K.
Types of Thermometer
| Device | Principle | Features |
|---|---|---|
| Liquid expansion (mercury, alcohol) | Liquid expands with temperature | Slow; mercury now largely withdrawn |
| Bimetallic strip | Two metals with different expansion coefficients bend | Used in dial thermometers and variable-bypass vaporizer temperature compensation |
| Resistance thermometer | Resistance of a platinum wire rises almost linearly with temperature | Accurate; used as a reference |
| Thermistor | Semiconductor (metal oxide) whose resistance usually falls non-linearly with temperature | Small, rapid response, cheap; used in pulmonary artery catheters for thermodilution |
| Thermocouple | Junction of two dissimilar metals (for example copper and constantan) generates a voltage proportional to temperature difference (Seebeck effect) | Needs a reference junction; small and robust |
| Infrared (tympanic, skin) | Measures emitted infrared radiation | Rapid; tympanic readings depend on technique |
Resistance thermometers and thermistors are usually used in a Wheatstone bridge circuit to measure small changes in resistance accurately.
Core Temperature Sites
- Pulmonary artery: the reference standard for core temperature.
- Distal oesophagus: probe in the lower third, behind the heart; accurate and practical during general anaesthesia.
- Nasopharynx: close to the internal carotid artery; reflects brain temperature but affected by leaks around the airway.
- Tympanic membrane: contact thermistors are accurate; infrared devices are variable.
- Bladder and rectum: acceptable but lag behind rapid changes such as rewarming after bypass.
- Skin and axilla: reflect peripheral temperature and underestimate core temperature.
Monitoring is recommended during anaesthesia lasting more than about 30 minutes, and active warming should keep core temperature above 36.0 °C (for example NICE guidance on inadvertent perioperative hypothermia).
Humidity
| Term | Definition |
|---|---|
| Absolute humidity | Mass of water vapour per unit volume of gas (mg/L or g/m³) |
| Relative humidity | Actual water vapour mass divided by the mass required to saturate the gas at that temperature (%) |
| Dew point | Temperature at which gas becomes saturated and water condenses |
| Saturated vapour pressure of water | About 6.3 kPa (47 mmHg) at 37 °C |
Fully saturated gas at 37 °C contains about 44 mg/L of water; at 20 °C it holds only about 17 mg/L. Cooling saturated gas therefore causes condensation (rain-out) in breathing circuits.
Why Humidification Matters
Normal breathing warms and humidifies inspired air in the upper airway so that gas reaching the isothermic saturation boundary (just below the carina) is fully saturated at 37 °C. Dry medical gases delivered through a tracheal tube bypass the nose, causing heat and water loss, thickened secretions, impaired mucociliary clearance, atelectasis and airway damage.
Measuring Humidity
- Hair hygrometer: a hair lengthens as humidity rises; works between about 30% and 90% relative humidity.
- Wet and dry bulb hygrometer: evaporation cools the wet bulb; the temperature difference indicates relative humidity.
- Regnault's hygrometer: cooling a polished surface until dew forms gives the dew point.
- Electronic sensors: humidity changes the capacitance or resistance of a sensing material.
Humidification Devices
| Device | Typical output | Notes |
|---|---|---|
| Heat and moisture exchanger (HME) | About 25-35 mg/L | Passive; adds dead space (about 30-90 mL) and resistance; may include a bacterial and viral filter (HMEF) |
| Heated water bath humidifier | Up to about 44 mg/L | Active; risk of scalding, infection and overhydration; condensation in tubing |
| Nebulisers (jet, ultrasonic) | Can exceed saturation with droplets | Used for drug delivery; ultrasonic devices produce very small droplets |
Gas Analysis
Infrared Absorption Spectroscopy
Molecules containing two or more different atoms (carbon dioxide, nitrous oxide, volatile agents, water vapour) absorb infrared radiation at characteristic wavelengths. Carbon dioxide is measured at about 4.3 ; volatile agents are identified using wavelengths around 8-12 . The amount absorbed follows the Beer-Lambert law.
- Side-stream analysers aspirate about 50-200 mL/min from the breathing system, which must be returned or scavenged; there is a short delay.
- Main-stream analysers place the sensor at the airway, giving an immediate signal but adding weight and dead space.
- Collision broadening: nitrous oxide broadens the carbon dioxide absorption band, causing slight over-reading of carbon dioxide unless corrected.
- Oxygen, nitrogen and other symmetrical diatomic gases do not absorb infrared, so other methods are needed.
Oxygen Analysis
| Method | Principle | Response time | Notes |
|---|---|---|---|
| Paramagnetic | Oxygen has unpaired electrons and is attracted into a magnetic field; modern analysers use a pressure difference in a switched magnetic field | Rapid (breath-by-breath) | Standard in anaesthetic monitors |
| Galvanic fuel cell | Oxygen is reduced at a cathode, generating a current proportional to oxygen partial pressure | Slow (about 20-30 seconds) | Used in anaesthetic machines; limited lifespan; no external power needed |
| Polarographic (Clark) electrode | Oxygen reduced at a platinum cathode with an applied voltage; current proportional to | Moderate | Used in blood gas analysers |
Other Methods
- Piezoelectric analysis: a quartz crystal coated with oil absorbs volatile agent, changing its resonant frequency; it cannot identify which agent is present.
- Mass spectrometry: separates ionised gas molecules by mass-to-charge ratio; accurate for multiple gases but expensive.
- Raman spectroscopy: measures scattered light shifted in wavelength by gas molecules.
- Chemiluminescence is used to measure nitric oxide.
SI Units and Conversions
| Quantity | SI unit | Common conversions |
|---|---|---|
| Pressure | Pascal (Pa = N/m²) | 1 kPa = 7.5 mmHg = 10.2 cmH₂O; 1 atmosphere = 101.325 kPa = 760 mmHg; 1 bar = 100 kPa |
| Force | Newton (N = kg·m/s²) | |
| Energy | Joule (J) | 1 calorie = 4.18 J |
| Power | Watt (W = J/s) | |
| Amount of substance | Mole (mol) | |
| Temperature | Kelvin (K) | |
| Luminous intensity | Candela (cd) |
The seven SI base units are the metre, kilogram, second, ampere, kelvin, mole and candela. All others, including the pascal, newton, joule, volt and watt, are derived units.
Tip
A quick pressure check: 1 kPa is about 7.5 mmHg, so a normal of 5.3 kPa is about 40 mmHg and a of 13 kPa is about 98 mmHg.
Which analyser is suitable for breath-by-breath measurement of inspired and expired oxygen in an anaesthetic monitor?
An infrared analyser using the 4.3 micrometre absorption band
A galvanic fuel cell, which responds within about 1 second
A paramagnetic analyser, which responds rapidly
A piezoelectric quartz crystal coated with oil
Which statement about temperature measurement is correct?
A thermocouple uses a semiconductor whose resistance falls as temperature rises
Bladder temperature reflects core temperature most accurately during rapid rewarming after cardiopulmonary bypass
Axillary temperature is a reliable measure of core temperature during anaesthesia and surgery
A thermocouple generates a temperature-dependent voltage at a junction of two dissimilar metals
Gas fully saturated with water vapour at 37 °C contains about how much water per litre, and how does a heat and moisture exchanger compare?
About 44 mg/L at 37 °C; a heat and moisture exchanger typically delivers about 25-35 mg/L
About 17 mg/L at 37 °C; a heat and moisture exchanger delivers more than 60 mg/L
About 100 mg/L at 37 °C; a heat and moisture exchanger provides complete saturation at body temperature
About 6.3 mg/L at 37 °C, matching the saturated vapour pressure in kilopascals
Sections you finish are checked off in the contents.