6.1 Tonometry & pachymetry
Key Takeaways
Goldmann applanation tonometry (GAT) is the gold standard and applies the Imbert-Fick principle (P = F/A), using a 3.06 mm diameter applanation zone to estimate IOP.
For GAT you instill anesthetic plus fluorescein, view through the cobalt-blue filter, and align the inner edges of the two green mires; multiply the dial reading by 10 to get mmHg.
Normal IOP is 10–21 mmHg; alternatives include non-contact (air-puff), iCare rebound (no anesthetic), and the handheld Tono-Pen.
Errors: too much fluorescein or lid/globe pressure reads falsely high, too little reads low, and astigmatism over ~3 D distorts the mires.
Thin corneas tend to bias Goldmann measurements low and thicker corneas high. Record CCT with IOP and consider other corneal properties.
Measuring Intraocular Pressure
Tonometry is the measurement of intraocular pressure (IOP), the fluid pressure inside the eye produced by the balance between aqueous humor production and its outflow. A typical adult IOP reference interval is 10 to 21 mmHg. Glaucoma can occur within it, and elevated pressure is a major modifiable risk factor. As a COT you will perform tonometry many times each day, so mastering the principles, instruments, and error sources is essential for accurate, repeatable numbers that the physician can trust.
Goldmann Applanation Tonometry (GAT)
Goldmann applanation tonometry is the gold standard. It mounts on the slit lamp and works on the Imbert-Fick principle, which states that the pressure inside a sphere equals the force needed to flatten a portion of its surface divided by the flattened area (P = F/A). Goldmann chose a flattening diameter of 3.06 mm because at exactly that diameter the cornea's structural rigidity and tear film surface tension tend to offset each other, allowing an estimate of IOP.
Procedure: instill a topical anesthetic (such as proparacaine) combined with fluorescein dye, then view the cornea through the cobalt-blue filter. The doubling prism splits the illuminated tear meniscus into two glowing green semicircles called mires. Turn the calibrated dial until the inner edges of the two mires just touch, forming a smooth S-shape as they pulse gently. Read the number on the dial at that endpoint and multiply it by 10 to obtain the IOP in mmHg — a dial reading of 1.6 equals 16 mmHg.
Common errors to avoid:
- Too much fluorescein produces thick mires and a falsely high reading; too little dye gives thin mires and a falsely low reading.
- Pressure on the globe from the eyelids, the patient squeezing, or the examiner holding the lids can raise the reading artificially.
- Corneal astigmatism greater than about 3 diopters distorts the mires into unequal shapes. For compatible Haag-Streit prisms, align the prism graduation for the flattest corneal meridian with the 43° mark, or follow the manufacturer’s two-measurement method at perpendicular orientations.
- Repeated applications deplete the tear film and can abrade the epithelium, so limit re-checks.
Other Tonometry Methods
- Non-contact (air-puff) tonometry flattens the cornea with a calibrated pulse of air. It needs no anesthetic or dye, which makes it useful for screening and for staff who do not perform contact tonometry, but it is more affected by corneal properties and is less accurate at higher pressures.
- iCare rebound tonometry bounces a tiny disposable probe off the cornea and measures its deceleration. It requires no anesthetic, is very fast, and works well in children and anxious patients.
- The Tono-Pen is a handheld electronic applanation device that averages several taps and reports a mean with a reliability percentage. It needs anesthetic and a disposable latex-free tip cover, and it excels at the bedside, in wheelchairs, and on scarred or irregular corneas.
Pachymetry and Corneal Thickness
Pachymetry measures central corneal thickness (CCT), normally about 540–560 µm. CCT matters because applanation tonometry assumes an average thickness, so any deviation biases the reading:
- A thin cornea underestimates the true IOP — the number looks reassuringly low while the real pressure may be higher.
- A thick cornea overestimates the true IOP — the number looks alarmingly high while the real pressure may be normal.
This is why CCT is a critical part of evaluating ocular hypertension and glaucoma risk. The Ocular Hypertension Treatment Study established thin CCT as an independent predictor of progression to glaucoma. Ultrasound pachymetry uses a small contact probe (with anesthetic) held perpendicular to the central cornea, while optical and OCT-based pachymetry are non-contact. Record the measured IOP and CCT so the ophthalmologist can interpret them together. There is no generally accepted correction formula for individual eyes.
Practical Technique and Safety
Seat the patient at the slit lamp with the forehead firmly against the band and the chin in the rest, and instruct them to look straight ahead, breathe normally, and avoid squeezing or holding their breath, all of which transiently raise IOP. Set the cobalt-blue illumination at low-to-medium magnification with the beam angled about 60°. Advance the prism until it just contacts the cornea; over-indenting flattens too much and inflates the reading, while an incomplete touch reads low. Take the measurement quickly to avoid drying the cornea, and if the patient blinks, re-wet with a fresh drop rather than pressing harder.
Check GAT calibration at the checkpoints and intervals in the model’s instructions. If a check fails, take the instrument out of service and report it for qualified inspection rather than adjusting patient readings. Between patients, disinfect a reusable prism exactly per manufacturer instructions — for example a compatible validated sodium hypochlorite procedure followed by rinsing and drying. A quick isopropyl alcohol wipe or a 3% hydrogen peroxide soak is unreliable against adenovirus. A new single-use prism or tip is another option for compatible equipment. Never touch the tip surface that contacts the eye, and always record which eye was measured and the time of day, since IOP has diurnal variation.
Tonometry Methods at a Glance
| Method | Contact? | Anesthetic + dye | Notes |
|---|---|---|---|
| Goldmann (GAT) | Yes | Yes | Gold standard; align 2 mires, dial ×10 |
| Non-contact (air-puff) | No | No | Screening; less accurate at high IOP |
| iCare rebound | Yes (probe) | No | Fast; great for kids/anxious patients |
| Tono-Pen | Yes | Anesthetic only | Handheld; irregular corneas, bedside |
When performing Goldmann applanation tonometry, what is the correct endpoint the technician looks for before reading the dial?
The outer edges of the two green mires are separated by one mire width
The inner edges of the two green mires just touch, forming an S-shape
A single continuous green ring fills the entire cobalt-blue field
The mires disappear completely as the prism seats on the cornea
A patient's central corneal thickness measures 610 µm. How does this likely affect the Goldmann IOP reading relative to true intraocular pressure?
It has no effect because Goldmann corrects for thickness automatically
The reading overestimates the true IOP
The reading underestimates the true IOP
It makes the mires impossible to align
Which tonometry instrument requires NO topical anesthetic and uses a small disposable probe that rebounds off the cornea, making it well suited to children?
iCare rebound tonometer
Tono-Pen
Goldmann applanation tonometer
Ultrasound pachymeter
Sections you finish are checked off in the contents.