6.4 A-scan biometry & IOL calculation basics

Key Takeaways

  • Biometry measures axial length, keratometry, and anterior chamber depth to select the correct IOL power before cataract surgery.

  • Contact (applanation) A-scan can indent the cornea and falsely shorten axial length; immersion A-scan avoids corneal contact and is more accurate.

  • Optical biometry is non-contact. IOLMaster 500 uses partial-coherence interferometry and IOLMaster 700 uses swept-source OCT. Ultrasound can help when optical measurement fails.

  • IOL formulas include SRK/T, Barrett Universal II, Holladay, Hoffer Q, and Haigis; the regression SRK is conceptually P = A - 2.5L - 0.9K.

  • Axial-length errors can significantly affect the refractive result. The effect varies with eye length and formula, so review repeatability and artifact.

Last updated: November 2026

Why biometry matters

Biometry is the measurement of the eye's anatomy needed to choose the correct intraocular lens (IOL) power before cataract surgery. The goal is to implant a lens that delivers the surgeon's target refraction, usually emmetropia or a small planned myopia. Core measurements include axial length (AL) and keratometry (K). Some formulas also use measured anterior chamber depth (ACD), lens thickness, or other variables. Accurate values are what separate a happy postoperative patient from a costly refractive surprise, so the technician's care at this step directly shapes the surgical result.

A-scan ultrasound biometry

A-scan (amplitude-scan) ultrasound sends a 10 MHz sound pulse along the visual axis and measures the time echoes take to return from each ocular interface: cornea, anterior and posterior lens surfaces, and retina. The machine converts travel time into distance using assumed sound velocities for the cornea, aqueous, lens, and vitreous, so a clean series of tall, steeply rising spikes is essential for a trustworthy axial length.

Contact versus immersion

There are two ultrasound techniques:

  • Contact (applanation): the probe touches the cornea directly. It is fast and simple, but the probe can indent the cornea, compressing the globe and falsely shortening the axial length. Even a fraction of a millimeter of indentation shifts the calculated IOL power. Approach gently along the visual axis, barely touch, take several readings, and discard the short outliers.
  • Immersion: a saline-filled scleral shell couples the probe to the eye without touching the cornea. Because there is no corneal compression, immersion is more accurate and reproducible and is preferred when precision matters, especially in very short or very long eyes.

Review repeatability and the instrument’s quality indicators. On ultrasound, confirm the expected retinal echo and probe alignment. On optical instruments, inspect fixation, signal quality and the device-specific scan checks.

Reading the waveform

A valid axial-length trace shows a clean sequence of spikes: a corneal spike, then the anterior and posterior lens spikes, then a tall retinal spike, followed by lower scleral and orbital echoes. The retinal spike must rise steeply from the baseline; a stepped or gradually sloping spike often means the beam is off-axis or reflecting from the optic nerve rather than the fovea, which lengthens the reading. Ask the patient to fixate on the probe's internal light and average several concordant scans. The machine assumes fixed sound velocities for each ocular segment, so a silicone-oil-filled or pseudophakic eye needs the correct velocity mode selected, otherwise the axial length will be systematically wrong.

Optical biometry

Optical biometry uses light to measure the eye. For example, the IOLMaster 500 uses partial-coherence interferometry, whereas the IOLMaster 700 uses swept-source OCT. Both are non-contact.

  • Non-contact: nothing touches the cornea, so there is no indentation error and negligible infection risk.
  • Higher resolution and reproducibility than ultrasound, measuring to the fovea and matching the true visual axis.
  • Captures AL, K, ACD, white-to-white, and lens thickness in a single acquisition.

Optical readings can fail with dense cataract or other media opacity. Swept-source devices penetrate many cataracts that older optical systems cannot measure. When reliable optical biometry remains unobtainable, ultrasound A-scan is an alternative.

A-scan versus optical biometry

FeatureA-scan ultrasoundOptical biometry
Energy sourceSound (10 MHz)Light (technology varies by model)
Corneal contactContact indents; immersion is non-contactNon-contact
Main error sourceCorneal compression (contact method)Dense media blocking light
ReproducibilityLowerHigher
Works in dense cataractOften usableDepends on opacity and device

From measurements to IOL power

An IOL formula estimates implant power from the required biometry and lens constants. Modern choices include SRK/T, Barrett Universal II, Holladay 1 and 2, Hoffer Q, and Haigis. Barrett Universal II is widely favored for accuracy across a broad range of eye lengths. The older regression SRK formula is worth knowing conceptually:

P = A - 2.5L - 0.9K

Here P is the IOL power, A is the lens constant unique to each IOL model, L is axial length in millimeters, and K is average keratometry in diopters. The A-constant encodes the effective lens position for that specific implant, which is why entering the correct lens model is as important as the measurements.

The other inputs

Keratometry supplies the corneal power in two meridians; a difference between them is corneal astigmatism, which guides toric IOL planning. Take K readings before any drops that dry the cornea, and repeat if the mires look distorted. Anterior chamber depth and, in newer formulas, lens thickness and white-to-white help predict the effective lens position (ELP), the depth at which the implant will finally sit. Because the surgeon cannot know ELP until after surgery, formulas estimate it from these measurements, and this estimation is the second-largest error source after axial length.

Why axial length dominates

Axial length strongly affects predicted implant power. A contact probe indenting the cornea can shorten the measurement and lead to selection of an overly powerful IOL, all else equal. Inspect the traces and repeat an unexpected measurement using the clinical protocol.

Test Your Knowledge

A cataract patient's contact applanation A-scan reads 1.2 mm shorter than an immersion scan taken minutes later. What most likely explains the difference?

A

The immersion scleral shell stretched and lengthened the globe

B

The contact probe indented the cornea, falsely shortening the axial length

C

Optical biometry principles always make ultrasound read longer

D

The A-constant for the IOL was entered incorrectly

Test Your Knowledge

All other inputs being equal, what happens if contact A-scan underestimates axial length and the calculation uses that measurement?

A

It predicts too little IOL power

B

It changes only the toric axis

C

It has no effect on estimated IOL power

D

It predicts too much IOL power

Test Your Knowledge

The IOLMaster cannot obtain an axial length on a patient with a dense brunescent cataract. What is the appropriate next step?

A

Use A-scan ultrasound biometry as the fallback measurement

B

Proceed using an average estimated axial length

C

Cancel the cataract surgery entirely

D

Increase the A-constant to compensate

Sections you finish are checked off in the contents.