13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

Key Takeaways

  • Failure to capture shows a pacing spike with no depolarization; failure to output shows no spike at all where one was expected.
  • Undersensing produces inappropriate pacing spikes that fall too early, while oversensing produces inappropriate pauses with no spike at all.
  • A higher programmed sensitivity number means a less sensitive device, so raising the millivolt value corrects oversensing and lowering it corrects undersensing.
  • Hyperkalemia, antiarrhythmic drugs, myocardial infarction at the lead tip, and acidosis all raise the capture threshold and can cause acute loss of capture.
  • Temporary transvenous pacing output is set at two to three times threshold, with sensitivity set to reliably sense the intrinsic R wave in demand mode.
Last updated: September 2026

13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

CCI lists "Evaluation of temporary and permanent pacing malfunctions", "CIED troubleshooting", and "Temporary pacemaker operation" as three separate knowledge areas. Troubleshooting is a small, closed set of problems, and the fastest route to the answer is a fixed framework.


1. The Four Malfunctions

MalfunctionECG appearanceMeaning
Failure to captureSpike present, no depolarization followsThe stimulus reached the tissue but did not depolarize it
Failure to outputNo spike where one was expectedThe device did not deliver a stimulus
UndersensingInappropriate spikes appearing too soon after an intrinsic beatThe device did not see an intrinsic event
OversensingInappropriate pauses with no spikeThe device saw something that was not an intrinsic cardiac event and inhibited

The two most reliable discriminators: failure to capture has a spike, failure to output does not; and undersensing adds spikes, oversensing removes them.

Failure to capture — causes

CategoryExamples
LeadDislodgement, micro-dislodgement, perforation, poor contact
Threshold riseHyperkalemia, acidosis, antiarrhythmic drugs (class IC especially), myocardial infarction at the lead tip, exit block from fibrosis
DeviceOutput programmed too low, battery depletion, header connection problem
CircuitConductor fracture with intermittent contact, loose set screw

Hyperkalemia is the classic acute answer because it raises the resting membrane potential, inactivates sodium channels, and both raises threshold and widens the QRS at the same time.

Failure to output — causes

Complete conductor fracture (open circuit, very high impedance), a loose or disconnected set screw, battery at end of life, oversensing that inhibits output (which is really a sensing problem presenting as no output), or the device being programmed to a rate below the intrinsic rate so no output was ever expected — the last of which is not a malfunction at all.

Undersensing — causes

Sensitivity set too low (numerically too high a millivolt threshold), lead dislodgement, insulation or conductor failure, a low-amplitude intrinsic signal (as in the substrate change of infarction), a signal falling within a refractory or blanking period, or a change in the intrinsic rhythm's morphology such as new bundle branch block or ventricular ectopy.

A specific and testable point: an intrinsic beat that falls inside a programmed refractory period is not undersensing. It is normal, designed behavior, and correcting it means adjusting the refractory period, not the sensitivity.

Oversensing — causes

SourceTypical signature
T-wave oversensingCounted rate is exactly double the true rate
Myopotential oversensingPectoral muscle activity or deep breathing; more common with unipolar sensing
Far-field R-wave sensing on the atrial channelVentricular events counted as atrial events
Lead fracture / insulation failure noiseNon-physiologic, very short intervals; impedance abnormality
CrosstalkVentricular channel senses the atrial output
Electromagnetic interferenceCorrelates with an external source

2. The Sensitivity Convention

This is the single most frequently inverted concept in device troubleshooting.

Sensitivity is programmed as a voltage threshold in millivolts, and the device only counts signals larger than that threshold.

Programmed valueDevice behaviorCorrects
Lower number (e.g. 0.5 mV)More sensitive — detects smaller signalsUndersensing
Higher number (e.g. 5.0 mV)Less sensitive — ignores smaller signalsOversensing

So the counterintuitive rule is: to make a device more sensitive, program a smaller number. A patient with T-wave oversensing needs the sensitivity value raised (made numerically larger, so the smaller T wave is ignored); a patient undersensing a small R wave needs it lowered.

Sensing configuration matters too. Bipolar sensing uses the closely spaced tip and ring, producing a small antenna that rejects far-field signals and myopotentials. Unipolar sensing uses the tip against the generator can, producing a large antenna that is far more vulnerable to myopotential and electromagnetic oversensing. Converting a unipolar system to bipolar is a standard fix for myopotential oversensing.


3. Electromagnetic Interference

SourceRiskPractical guidance
Monopolar electrocauteryHigh — oversensing, inhibition, inappropriate shocks, power-on resetBipolar cautery, short bursts, asynchronous programming, therapy suspension
MRIHigh if the system is not conditionalConditional systems only, MRI mode, monitored scan
External defibrillationHighPads ≥ 8 cm from the generator, anteroposterior orientation, interrogate afterward
Radiation therapyCumulative dose damage and single-event upsetsDose limits; relocate the generator if it is in the field
Radiofrequency ablationModerateProgram asynchronously if dependent, suspend tachytherapy, position the dispersive pad away from the device
TENS units, electrolysis, electroconvulsive therapyModerateCase-specific evaluation
Airport / retail security gatesLowWalk through at a normal pace; do not linger
Cell phonesLowKeep ≥ 15 cm (6 inches) from the generator; use the opposite ear
Consumer devices with strong magnetsLow but realKeep certain phones, phone cases, wireless chargers, and smart-watch bands ≥ 15 cm from the generator
Household appliances, microwaves, induction hobsNegligible to lowNormal use is safe; keep induction cooktops at arm's length
Arc welding, large industrial motors, high-power radio transmittersHighOccupational assessment required

4. Temporary Pacing

Temporary pacing appears in the knowledge list explicitly and in task D8 as "emergency pacing."

Modalities

ModalitySpeedNotes
TranscutaneousFastestPads on the chest; painful, requires sedation; verify mechanical capture with a pulse, not just electrical capture on the monitor
TransvenousMinutesBalloon-tipped or standard wire to the RV apex; reliable and tolerated
EpicardialPost-cardiac-surgeryWires already in place
Through an existing EP catheterImmediate in the labConnect the RV diagnostic catheter to the stimulator or a temporary generator

Operating a temporary transvenous generator

Three controls, each with a rule:

  1. Rate — set above the intrinsic rate to achieve consistent pacing, typically 60-80 bpm, or higher for overdrive suppression of a pause-dependent arrhythmia.
  2. Output (mA) — determine the threshold by decrementing until capture is lost, then set the output at 2 to 3 times threshold. A typical acute ventricular threshold is under 1 mA, so a typical setting is 2-5 mA.
  3. Sensitivity (mV) — in demand (VVI) mode, set low enough (a smaller number) to sense the intrinsic R wave reliably. Turning the dial fully to asynchronous disables sensing and paces at a fixed rate regardless of intrinsic activity — appropriate only for specific situations such as cautery interference, and hazardous otherwise because of R-on-T risk.

Safety practice: insulate all exposed wire terminals (a glove over the connector block is the classic method) so that stray current cannot reach the endocardium — a microshock hazard at as little as 10-100 µA. Secure the wire, obtain a chest radiograph to confirm position, check thresholds at least daily because they rise over the first days, and monitor continuously.

Complications: RV perforation (thresholds rise, a pericardial rub or effusion appears, and pacing may capture the diaphragm), lead dislodgement with loss of capture, infection at the access site, arrhythmia during placement, and thrombosis. Thresholds that rise steadily over hours to days are expected; an abrupt rise with new diaphragmatic pacing suggests perforation.

Test Your Knowledge

A patient with a VVI pacemaker programmed to a sensitivity of 4.0 mV shows intermittent pacing spikes falling shortly after clearly visible intrinsic QRS complexes. Interrogation shows a sensed R wave amplitude of 3.2 mV and a stable lead impedance of 510 ohms. What is the malfunction and the correct programming change?

A
B
C
D
Test Your Knowledge

A dialysis patient with a permanent pacemaker presents with pacing spikes that are not followed by ventricular depolarization, a widened intrinsic QRS, and peaked T waves. Lead impedance and the chest radiograph are unchanged from six months ago. What is the most likely cause?

A
B
C
D
Test Your Knowledge

After placing a temporary transvenous pacing wire in the right ventricle, the specialist finds the capture threshold is 0.8 mA. What output should be programmed, and what additional safety step is required at the connector block?

A
B
C
D