13.2 CIED Interrogation Workflow, Programmer Operation & Peri-Procedural Reprogramming

Key Takeaways

  • Task B6 requires interrogating cardiac implantable electronic devices as a distinct intra-procedural objective, separate from implant assistance.
  • Programmers are manufacturer-specific and communicate through an inductive wand or radiofrequency telemetry; there is no universal programmer.
  • Pacing dependence is determined by decrementing the pacing rate to observe an underlying rhythm, not by assuming the mode implies dependence.
  • Pacemaker-dependent patients are reprogrammed to an asynchronous mode before electrocautery or MRI, and ICD tachytherapy is suspended.
  • Every device reprogrammed for a procedure must be restored and re-interrogated afterward with thresholds, sensing, and impedances documented.
Last updated: September 2026

13.2 CIED Interrogation Workflow, Programmer Operation & Peri-Procedural Reprogramming

CCI lists "Interrogate Cardiac Implantable Electronic Devices (CIEDs)" as task B6, and separately lists CIED programmer function and operation(s), basic CIED programming, and basic CIED function in the knowledge list. In an EP lab a device is interrogated before nearly every case in a device patient, again after any external shock or cautery exposure, and always before discharge.


1. The Programmer and Telemetry

A programmer is a manufacturer-specific computer that communicates with an implanted device. There is no universal programmer: an Abbott device requires an Abbott programmer, a Medtronic device a Medtronic programmer, and so on. Identifying the manufacturer is therefore the first step in any device encounter.

Identification methodDetail
Patient identification cardManufacturer, model, serial number, implant date, implanting physician
Chest radiographRadiopaque alphanumeric identifier stamped in the header; distinctive generator shape and lead architecture
Institutional or manufacturer registryDevice tracking database lookup
Manufacturer support line24-hour technical support for every major manufacturer
Trial-and-error interrogationAttempting each available programmer — the slowest option, used only when nothing else identifies the device

Telemetry is either inductive, requiring a wand held directly over the generator, or radiofrequency (wireless), which works over a few metres. Inductive telemetry is more robust in an electrically noisy lab; radiofrequency telemetry is convenient but can drop out. Sterile-field interrogation during a procedure uses a wand inside a sterile sleeve.


2. The Systematic Interrogation Sequence

  1. Identify manufacturer, model, serial number, implant date, and indication.
  2. Establish telemetry and save the pre-existing ("initial") programmed settings before changing anything.
  3. Battery status — voltage, estimated longevity, and replacement indicators.
  4. Lead measurements — impedance, sensed amplitude, and pacing threshold for each lead.
  5. Programmed parameters — mode, rates, AV delays, outputs, sensitivities, refractory periods, and (for an ICD) detection zones and therapies.
  6. Diagnostics — percentage paced per chamber, rate histograms, mode switch episodes, arrhythmia counters, heart failure diagnostics.
  7. Stored episodes — electrograms and markers for arrhythmias and delivered therapies.
  8. Determine pacing dependence.
  9. Reprogram as clinically required.
  10. Print or save the final report to the medical record, and restore original settings when the procedural need has passed.

Battery status terminology

IndicatorMeaning
BOL — beginning of lifeFull battery
ERI / RRT — elective replacement indicator / recommended replacement timeSchedule replacement; typically ~3 months of normal function remain
EOL / EOS — end of life / end of serviceUrgent; device may cease reliable therapy

At ERI, devices typically change behavior in a manufacturer-specific way — often reverting to a fixed VVI rate, disabling rate response, and lengthening the magnet rate — and those changes are themselves the clue on an ECG.

Normal measured values

ParameterNormalAbnormal suggests
Pacing lead impedance~300-1,200 Ω (varies by lead)> 2,000 Ω or an abrupt rise = conductor fracture / loose set screw; < 200 Ω or an abrupt fall = insulation breach
High-voltage (shock) coil impedance~20-100 ΩOut of range suggests coil or connection failure
Atrial sensed P wave> 1.5-2.0 mVUndersensing risk
Ventricular sensed R wave> 5 mVUndersensing risk; ICD needs adequate R waves for VF detection
Atrial threshold< 1.5 V at 0.5 msExit block, dislodgement, perforation
Ventricular threshold< 1.0-1.5 V at 0.5 msSame

Trends matter more than single values. A ventricular threshold of 1.2 V is unremarkable in isolation but alarming if it was 0.5 V three months ago. Interrogation reports plot these trends, and reading the trend is the skill being tested.


3. Determining Pacing Dependence

Pacing dependence cannot be inferred from the mode or from the percentage paced. A patient can be 100% paced and still have a stable underlying escape rhythm; another can be 40% paced and have no escape at all.

The determination is made by decrementing the programmed lower rate (or briefly programming to VVI at 30 bpm) and observing whether an intrinsic rhythm emerges, with the patient monitored, pads on, and the programmer ready to restore pacing immediately. A patient with no escape rhythm, or an escape below about 30-40 bpm that is hemodynamically inadequate, is pacemaker-dependent.

That single determination drives every subsequent peri-procedural decision.


4. Peri-Procedural Reprogramming

SituationPacemakerICD
Electrocautery (monopolar), pacing-dependentAsynchronous: VOO or DOOAsynchronous pacing plus suspend tachytherapy
Electrocautery, not pacing-dependentOften no change; monitorSuspend tachytherapy
External cardioversion / defibrillationPads ≥ 8 cm from the generator, ideally anteroposterior; interrogate afterwardSame; interrogate afterward
MRI (conditional system)Enable MRI mode, asynchronous if dependentEnable MRI mode, tachytherapy off
Radiation therapyDose limits to the generator; relocate the device if it lies in the treatment fieldSame, plus therapy suspension during fractions if indicated
Lithotripsy, RF ablation, TENS, electroconvulsive therapyCase-specific; asynchronous if dependentSuspend tachytherapy

Suspending tachytherapy is accomplished either by programming detection off (preferred, because it is precise and documented) or by applying a magnet over the ICD. The magnet's effect must be understood exactly, because the two device types behave differently:

Magnet over a...Effect
PacemakerAsynchronous pacing (VOO/DOO/AOO) at a manufacturer-specific magnet rate; the magnet rate itself indicates battery status
ICDSuspends tachyarrhythmia detection and therapy only. Bradycardia pacing does not change and does not become asynchronous

A pacemaker-dependent ICD patient undergoing electrocautery therefore cannot be managed with a magnet alone: the magnet stops inappropriate shocks but leaves the bradycardia pacing susceptible to oversensing and inhibition. That patient requires programming.

The mandatory closing step: any device reprogrammed for a procedure is restored and re-interrogated before the patient leaves monitored care, with thresholds, sensing, and impedances documented and compared with the pre-procedure values. Leaving a device asynchronous, in MRI mode, or with therapies off is a well-documented and entirely preventable adverse event, and it is the reason the patient stays on continuous monitoring until restoration is confirmed.


5. Special Intra-Procedural Situations

Radiofrequency ablation in a device patient. RF current can be oversensed, and lesions near a lead can raise thresholds. Practice includes suspending tachytherapy in an ICD, programming asynchronously in a dependent patient, keeping the RF dispersive pad positioned so the current vector does not cross the generator or leads, and interrogating fully afterward.

Cryoablation near a lead can cause transient threshold rise and undersensing during the freeze, which typically recovers on rewarming but must be watched in a dependent patient.

External cardioversion in a device patient uses pads at least 8 cm from the generator, preferably in an anteroposterior orientation, at the lowest effective energy. Post-shock interrogation checks for parameter reset, threshold change, and impedance change.

Power-on reset occurs when strong interference causes the device to revert to factory backup settings — typically a fixed-rate VVI mode with nominal outputs, and with tachytherapy off in an ICD. It is recognized on interrogation as a set of parameters nobody programmed, along with a reset event marker in the diagnostics, and it requires full reprogramming to the patient's intended settings.

Test Your Knowledge

A patient with a dual-chamber ICD requires monopolar electrocautery for an unrelated procedure and is pacemaker-dependent. A colleague proposes taping a magnet over the device for the duration. Why is this inadequate?

A
B
C
D
Test Your Knowledge

Interrogation of a pacemaker before a scheduled ablation shows the patient is paced 98 percent of the time in DDD mode. What does this establish about pacing dependence?

A
B
C
D
Test Your Knowledge

Post-procedure interrogation of a pacemaker exposed to prolonged monopolar electrocautery reveals VVI mode at 65 beats per minute with nominal outputs and rate response off, although the patient was implanted and programmed as DDDR at 60 with a rate-adaptive sensor. Diagnostics show a reset event marker. What has occurred?

A
B
C
D