9.8 Hematology Instrumentation: Microscopy, Counters, Analyzers & Centrifuges
Key Takeaways
- Total magnification is objective power multiplied by ocular power; immersion oil preserves the numerical aperture of the 100x objective, and the aperture diaphragm is set by Kohler illumination to about 70 to 80 percent of that numerical aperture and is never used to control brightness.
- Phase contrast is used for manual platelet and body fluid chamber counts on unstained cells, and polarized light with a first-order red compensator identifies synovial crystals.
- Histograms are single-parameter distributions, scatter plots are two-parameter dot plots generating the five-part differential, and digital imaging pre-classifies cells but always requires technologist verification and prior validation.
- Photo-optical coagulation detection is vulnerable to lipemia, icterus, and hemolysis; electromechanical steel-ball detection is the alternative for optically interfering specimens, and platelet-poor plasma must contain fewer than 10 x 10^9/L platelets.
- Relative centrifugal force is RCF = 1.118 x 10^-5 x r x (rpm)^2, so rpm and x g are not interchangeable; rotors are balanced by weight, sealed cups are used for aerosol-generating spins, and timers and speeds are verified and documented.
Hematology Instrumentation: Microscopy, Counters, Analyzers & Centrifuges
The Laboratory Operations content area names six instrument groups: microscopes, cell counters, differential analyzers (histograms, digital imaging, scatter plots), coagulation analyzers, point-of-care analyzers, and centrifuges. Sections 8.1 and 8.2 covered cell-counting and flow cytometric measurement principles; this section covers the instruments as equipment, including the optics, maintenance, and physical calculations the examination asks about.
1. Microscopy
The Brightfield Microscope
Light travels from the illuminator through the condenser, which focuses it on the specimen, then through the objective, which forms the primary magnified image, and finally through the ocular (usually 10x), which magnifies it again. Total magnification = objective power x ocular power, so a 100x oil objective with a 10x ocular gives 1,000x.
- Numerical aperture (NA) describes the light-gathering cone of an objective. Resolution improves as NA increases: $d = \dfrac{0.61\lambda}{NA}$, where a smaller $d$ means finer detail is distinguishable.
- Immersion oil has a refractive index (about 1.515) close to that of glass. It prevents the light rays from bending away as they cross the air gap, preserving the high numerical aperture of the 100x objective. Without oil, the 100x objective cannot resolve the chromatin detail needed for a differential.
- Kohler illumination is the standard alignment procedure: focus the specimen, close the field diaphragm, focus and center it with the condenser height and centering screws, then reopen it just beyond the field of view and set the aperture (iris) diaphragm to roughly 70 to 80 percent of the objective's numerical aperture. Closing the aperture diaphragm too far increases contrast but destroys resolution and creates diffraction halos; the aperture diaphragm is never used to control brightness, which is adjusted with the light intensity control or neutral density filters.
Other Microscopy Modes
| Mode | Principle | Hematology Use |
|---|---|---|
| Phase contrast | Converts small differences in refractive index into amplitude differences | Manual platelet counts and body fluid chamber counts on unstained, living cells |
| Polarizing with a first-order red compensator | Detects birefringent material and determines the sign of birefringence | Synovial fluid crystals (monosodium urate versus calcium pyrophosphate) |
| Darkfield | Only obliquely scattered light enters the objective | Spirochetes; rarely used in hematology |
| Fluorescence | Excitation at one wavelength, emission at a longer one | Fluorescent spot G6PD screening, FISH, immunofluorescence |
Care. Clean oil from objectives after each use with lens paper and an approved lens cleaner, never with xylene on modern cemented optics or with ordinary tissue, which scratches. Carry the microscope with one hand on the arm and one under the base, store with the low-power objective in position and the stage lowered, and cover it when idle.
2. Cell Counters and Differential Analyzers
Cell counters operate by electrical impedance, optical light scatter, or a combination, as detailed in Section 8.1. As equipment, their daily requirements are background counts within acceptance limits, aperture and flow-cell cleaning, verification that carryover is within specification, and startup checks before patient testing.
- Coincidence is the simultaneous passage of two cells through the sensing zone, counted as one; hydrodynamic focusing and electronic coincidence correction address it.
- Aperture obstruction by a clot, fibrin, or debris produces erratic low counts with a characteristic pulse-height abnormality and is cleared by the instrument's burn or backflush cycle.
- Carryover is checked with a high-concentration specimen followed by two low-concentration specimens.
Differential analyzers report through three display formats. Histograms are single-parameter frequency distributions (red cell volume, platelet volume, and the impedance three-part leukocyte distribution). Scatter plots (scattergrams) are two-parameter dot plots that create the five-part differential and flag blasts, immature granulocytes, and atypical lymphoid cells. Digital imaging systems photograph the stained monolayer, use a neural network to pre-classify each cell, and present sorted galleries on screen. Digital imaging does not replace the technologist: pre-classification must be verified and reclassified by a qualified operator, and the system must be validated against manual differentials before use. Its advantages are standardization, remote review, and permanent image storage for competency assessment and consultation.
3. Coagulation Analyzers
| Detection Method | Principle | Notes and Interferences |
|---|---|---|
| Photo-optical (turbidometric) | Measures increasing light scatter or decreasing transmittance as fibrin forms | Convenient and widely used, but lipemia, icterus, and hemolysis shift the optical baseline and can cause premature or missed endpoints |
| Electromechanical / magnetic steel ball | Detects the change in movement of a steel ball as viscosity rises | Unaffected by optical interference, so it is the method of choice for lipemic, icteric, or hemolyzed specimens |
| Chromogenic (synthetic substrate) | Enzyme cleaves a substrate releasing para-nitroaniline measured at 405 nm | Anti-Xa, antithrombin, protein C activity |
| Immunologic (latex agglutination, ELISA) | Antibody-coated particles agglutinate with the analyte | D-dimer, von Willebrand factor antigen |
| Nephelometric | Measures scattered light at an angle | Some fibrinogen and antigen assays |
Coagulation analyzers require a validated platelet-poor plasma input, defined as a residual platelet count below 10 x 10^9/L, achieved by centrifugation at approximately 1,500 x g for 15 minutes (a double spin is used for specimens to be frozen for lupus anticoagulant testing).
4. Point-of-Care Analyzers
Handheld and benchtop point-of-care devices use the same underlying chemistries in miniature: reflectance photometry for hemoglobin, electrochemical or optical strip endpoints for INR and activated clotting time, and cartridge electrode arrays for blood gases. Their operating requirements are covered in Section 9.4; as equipment, the recurring issues are cartridge and strip storage temperature, lot-to-lot verification, battery and connectivity status, and documented correlation with the central laboratory method.
5. Centrifuges
Types
| Type | Configuration | Hematology Application |
|---|---|---|
| Fixed-angle | Tubes held at a constant angle, typically 25 to 40 degrees | General serum and plasma separation; fast |
| Swinging-bucket (horizontal) | Tubes swing to horizontal during the spin | Coagulation platelet-poor plasma and gel-barrier tubes, because the sediment packs flat |
| Microhematocrit | Very high speed, small capillary tubes | Spun hematocrit, typically 10,000 to 15,000 x g for 5 minutes |
| Cytocentrifuge (Cytospin) | Low speed with an absorbent filter card | Body fluid differentials, concentrating cells into a monolayer with preserved morphology |
Relative Centrifugal Force
Protocols specify RCF in x g, not revolutions per minute, because the force depends on the rotor radius as well as the speed:
where $r$ is the rotating radius in centimetres. Two centrifuges running at the same rpm with different rotor radii deliver different forces, which is why a procedure copied between instruments without recalculating rpm produces incorrectly separated plasma.
Safe and Correct Operation
- Balance opposing tube positions by weight, not by appearance; an unbalanced rotor damages bearings and can fail catastrophically. Use a water-filled balance tube of equal weight when an odd number of specimens is spun.
- Never open the lid until the rotor has stopped completely, and use sealed safety cups or a sealed rotor for any specimen that may aerosolize; a tube that breaks in an open rotor generates an aerosol that persists for minutes.
- Spin tubes stoppered to prevent evaporation, aerosol, and pH change.
- Maintenance and verification: check the timer and speed (tachometer or optical tachometer) at defined intervals, inspect the rotor and buckets for corrosion and cracks, clean and disinfect the chamber, and document all of it. Under-centrifugation is a common source of residual platelets in coagulation plasma, which neutralizes heparin and falsely shortens heparin-sensitive assays.
While examining a peripheral smear at 1,000x, a technologist closes the aperture (iris) diaphragm nearly shut to make the cells appear darker and easier to see. What is the effect?
A coagulation laboratory receives a grossly lipemic specimen for a PT and APTT. Which analyzer detection method should be used and why?
A procedure specifies centrifugation at 1,500 x g for 15 minutes. A technologist enters 1,500 rpm on a centrifuge with a rotor radius of 15 cm. What is the consequence?
Which statement about digital imaging (automated cell locator) systems for the differential is correct?
You've completed this section
Continue exploring other exams