9.4 Method Evaluation, Diagnostic Statistics & Point-of-Care Testing

Key Takeaways

  • Diagnostic sensitivity is TP/(TP+FN) and rules disease OUT when negative; specificity is TN/(TN+FP) and rules disease IN when positive.
  • Sensitivity and specificity are intrinsic to the test, but positive and negative predictive values depend on disease PREVALENCE, so a good assay used for screening produces mostly false positives.
  • Before reporting nonwaived results, CLIA requires verification of manufacturer claims for accuracy, precision, reportable range, and reference intervals; a laboratory-modified or laboratory-developed test must additionally ESTABLISH analytical sensitivity and specificity.
  • The analytical measurement range is the directly measured linear interval; the clinically reportable range extends it by validated dilution.
  • IQCP is a voluntary CLIA alternative for NONWAIVED testing requiring a risk assessment, a written quality control plan, and ongoing quality assessment; it never applies to waived testing and never removes proficiency testing or competency obligations.
Last updated: August 2026

Method Evaluation, Diagnostic Statistics & Point-of-Care Testing

Laboratory Operations includes two topics that are frequently confused: the analytical performance characteristics that describe how well an instrument measures, and the diagnostic statistics that describe how well a result predicts disease. Point-of-care testing is where both meet under a distinct regulatory model.


1. Diagnostic Statistics: The Two-by-Two Table

Disease PresentDisease Absent
Test PositiveTrue positive (TP)False positive (FP)
Test NegativeFalse negative (FN)True negative (TN)
MeasureFormulaPlain Meaning
Diagnostic sensitivity$\dfrac{TP}{TP+FN}\times 100$Of those WITH disease, the proportion the test detects. A highly sensitive test rules disease out when negative.
Diagnostic specificity$\dfrac{TN}{TN+FP}\times 100$Of those WITHOUT disease, the proportion the test correctly calls negative. A highly specific test rules disease in when positive.
Positive predictive value$\dfrac{TP}{TP+FP}\times 100$Of those testing positive, the proportion that truly have disease
Negative predictive value$\dfrac{TN}{TN+FN}\times 100$Of those testing negative, the proportion that truly do not
Efficiency$\dfrac{TP+TN}{\text{Total}}\times 100$Overall proportion correctly classified

The single most tested principle: sensitivity and specificity are intrinsic to the test and do not change with the population. Predictive values depend on prevalence. The same assay used to screen a healthy population produces a much lower positive predictive value than when it is used in a symptomatic population, because most positives in a low-prevalence setting are false positives. This is why a D-dimer with 95% sensitivity is used to EXCLUDE venous thromboembolism in low-probability patients but never to confirm it.

Worked example. A flow cytometry screen for a PNH clone is run on 1,000 patients, 50 of whom truly have PNH. It detects 48 of them and calls 38 of the 950 unaffected patients positive. Sensitivity = 48/50 = 96%. Specificity = 912/950 = 96%. Positive predictive value = 48/(48+38) = 56%. Despite excellent sensitivity and specificity, fewer than 6 in 10 positives are real, because the disease is uncommon.


2. Analytical Performance Characteristics

CharacteristicDefinitionHow It Is Established
Analytical sensitivity (limit of detection)The lowest concentration reliably distinguished from zeroReplicate analysis of blank and low-level material
Analytical specificityFreedom from interference by substances other than the analyteInterference studies with hemolysate, bilirubin, lipid, drugs
PrecisionReproducibility, expressed as within-run and between-run CVRepeated measurement of controls over at least 20 days
Accuracy / biasAgreement with a reference method or reference materialMethod comparison against a comparative method, regression analysis
Linearity and analytical measurement rangeThe interval over which results are directly proportional to concentrationDilution series spanning the claimed range
Clinically reportable rangeThe AMR extended by permitted dilution or concentrationAMR multiplied by the validated dilution factor
CarryoverContamination of one specimen by the preceding oneHigh-low-low sequence testing
Reference interval verificationConfirming that a published interval applies to this populationAnalyzing a modest number of healthy local subjects (commonly 20) and confirming acceptable agreement

For nonwaived tests, CLIA requires the laboratory to verify manufacturer performance claims for accuracy, precision, reportable range, and reference intervals before reporting patient results. If the laboratory modifies an FDA-cleared method or develops its own, it must additionally establish analytical sensitivity, analytical specificity, and other performance characteristics.


3. Point-of-Care Testing

Point-of-care testing is performed near the patient by staff who are usually not laboratory-trained, which shifts the quality burden from the instrument to the management program.

Hematology and Coagulation Point-of-Care Devices

Device CategoryTypical Measurement PrincipleCommon Pitfalls
Portable hemoglobinometerReflectance or transmission photometry on a microcuvettePoor capillary technique, milking the finger, interstitial fluid dilution; first drop must be wiped away
Point-of-care INR meterCapillary whole blood clotting on a test strip with an electrochemical or optical endpointNot validated for lupus anticoagulant patients or for direct oral anticoagulants; INR agreement degrades above about 4.0
Activated clotting time (ACT)Whole blood clotting after contact activation, used for high-dose heparin during bypass and catheterizationInsensitive at low heparin doses; hemodilution and hypothermia prolong it
Point-of-care platelet function analyzersShear-induced platelet adhesion or aggregation on a cartridgeDependent on platelet count and hematocrit; low values in thrombocytopenia or anemia are not necessarily drug effects
Handheld blood gas and electrolyte cartridgesMicro-electrode arraysCartridge storage temperature and air bubbles

Regulatory Model

  • CLIA complexity. Waived tests carry the lowest regulatory burden but still require adherence to manufacturer instructions and a CLIA certificate of waiver. Moderate and high complexity point-of-care tests carry the same personnel, quality control, proficiency testing, and competency obligations as the main laboratory.
  • Individualized Quality Control Plan (IQCP). An optional CLIA-recognized alternative for nonwaived testing that permits a reduced control frequency only after a documented risk assessment covering specimen, environment, reagent, testing personnel, and test system, plus a written quality control plan and ongoing quality assessment. IQCP does not apply to waived testing.
  • Program requirements include a designated point-of-care coordinator, documented operator training and competency, lot-to-lot verification, connectivity so results reach the permanent record, and correlation studies between point-of-care devices and the central laboratory whenever both report the same analyte.
Test Your Knowledge

A screening assay has a diagnostic sensitivity of 96% and a specificity of 96%. It is applied first in a specialty clinic where disease prevalence is 40%, and then in a general screening population where prevalence is 1%. What happens to the performance measures?

A
B
C
D
Test Your Knowledge

A new hematology analyzer is being brought into service and the laboratory needs to define the interval over which reported results are directly proportional to concentration. Which performance characteristic is being established, and how?

A
B
C
D
Test Your Knowledge

Which statement about the Individualized Quality Control Plan (IQCP) is correct?

A
B
C
D
Test Your Knowledge

A point-of-care INR meter reports 2.4 on a patient whose plasma INR from the central laboratory is 4.6. The patient has a known lupus anticoagulant. What is the best explanation?

A
B
C
D