7.1 Radiographic Examination Techniques & Image Quality Indicators (Article 2)

Key Takeaways

  • ASME Section V Article 2 mandates standard radiographic examination procedures, specifying radiation source selection, geometric unsharpness limits, Image Quality Indicator (IQI) sensitivity, and transmitted film density.

  • Industrial gamma sources are selected based on material thickness: Selenium-75 (0.15 to 1.18 in. / 4 to 30 mm steel), Iridium-192 (0.25 to 2.5 in. / 6 to 64 mm steel; half-life 73.8 days), and Cobalt-60 (1.5 to 7.0 in. / 38 to 178 mm steel; half-life 5.27 years).

  • Image Quality Indicators (IQIs) must be placed on the source side of the component; film-side placement is permitted only when the source side is physically inaccessible, requiring a lead letter 'F' marker adjacent to the IQI.

  • Transmitted radiographic film density must be 1.8 minimum for X-ray and 2.0 minimum for gamma sources, up to a maximum of 4.0 for single-film viewing; composite viewing requires a minimum density of 2.6, and density across the area of interest cannot vary by more than -15% or +30% from the IQI density.

  • Maximum allowable geometric unsharpness (Ug = F * d / D) per Table T-274.2 is 0.020 in. (0.51 mm) for material thickness under 2 in., 0.030 in. for 2 to 3 in., 0.040 in. for 3 to 4 in., and 0.070 in. for thicknesses exceeding 4 in.

Last updated: August 2026

Radiographic Examination Techniques & Image Quality Indicators (Article 2)

Radiographic Examination (RT) is one of the foundational volumetric nondestructive examination methods referenced by API 510 and ASME Section VIII, Division 1. By recording differential radiation absorption on photographic film or digital detector arrays, radiography allows inspectors to detect internal volumetric flaws (porosity, slag inclusions) and planar defects (incomplete penetration, lack of fusion, cracks) in pressure vessel shells, heads, and nozzle welds.

ASME Boiler and Pressure Vessel Code (BPVC) Section V, Article 2 establishes the mandatory requirements for conducting radiographic examination, verifying image quality, controlling optical density, and maintaining viewing environments. An API 510 Inspector must master Article 2 rules to audit radiographic reports, verify radiograph quality, and interpret code compliance.


1. Fundamentals of Radiographic Examination & ASME Section V Scope

Radiography relies on the attenuation of high-energy electromagnetic radiation (photons) as they pass through matter. Dense materials (or thicker sections) absorb more photons, resulting in fewer photons reaching the film (producing lighter regions). Voids, cracks, or thinner sections absorb fewer photons, allowing more radiation to expose the film emulsion (producing darker regions after chemical development).

+-----------------------------------------------------------------------------+
|                     BASIC RADIOGRAPHIC EXPOSURE GEOMETRY                    |
|                                                                             |
|                      [ RADIATION SOURCE (Focal Spot F) ]                    |
|                                     |                                       |
|                                     |                                       |
|                                     | Source-to-Object Distance (D)         |
|                                     |                                       |
|                                     v                                       |
|   ========================[ SOURCE SIDE IQI ]=============================  |
|   \\\\\\\\\\\\\\\\\\\\  WELD REINFORCEMENT  //////////////////////////////  |
|   ------------------------------------------------------------------------  |
|                                                                             |
|                              BASE METAL THICKNESS                           |
|                                      (t)                                    |
|                                                                             |
|   ------------------------------------------------------------------------  |
|   ////////////////////////  WELD ROOT PASS  \\\\\\\\\\\\\\\\\\\\\\\\\\\\\\  |
|   ========================================================================  |
|                                     | Object-to-Film Distance (d)           |
|                                     v                                       |
|   -----------------------[ FILM-SIDE LEAD 'F' ]---------------------------  |
|   [========================= RADIOGRAPHIC FILM ==========================]  |
|   [------------------------- LEAD SCREEN BACKING ------------------------]  |
|   [========================= LEAD LETTER 'B' ============================]  |
+-----------------------------------------------------------------------------+

Code Scope & Responsibilities (T-210 / T-220)

  • ASME Section V provides the methodology (how to perform the examination and verify image quality).
  • ASME Section VIII, Division 1 (UW-51 & UW-52) and API 510 provide the acceptance criteria (what flaw sizes are rejectable).
  • Written procedures are mandatory for all radiographic examinations governed by Article 2.

2. Industrial Radiation Sources: X-Ray Machines vs. Gamma Radioisotopes

Industrial radiography utilizes two primary radiation sources: electrically generated X-radiation and isotope-generated Gamma radiation.

+-----------------------------------------------------------------------------+
|                      INDUSTRIAL RADIATION SOURCE MATRIX                     |
|                                                                             |
|   [X-RAY GENERATORS]                     [GAMMA RADIOISOTOPES]              |
|   - Continuous energy spectrum           - Discrete monoenergetic lines     |
|   - Adjustable kV and mA                 - Fixed photon energies per isotope|
|   - Switched on/off electrically         - Emits continuously (decay)       |
|   - Higher contrast / sensitivity        - Highly portable field cameras    |
|   - Heavy, requires field power          - Requires no power source         |
+-----------------------------------------------------------------------------+

Gamma Radioisotope Characteristics (Article 2 & API 577)

RadioisotopeChemical SymbolAverage Photon EnergyHalf-Life (t1/2t_{1/2})Practical Steel Thickness RangeKey Inspection Applications
Selenium-7575Se^{75}\text{Se}≈0.22 MeV\approx 0.22\text{ MeV} (range 0.066–0.40 MeV0.066\text{--}0.40\text{ MeV})119.8 days119.8\text{ days}0.15 in. to 1.18 in.(4 to 30 mm)0.15\text{ in.} \text{ to } 1.18\text{ in.} (4\text{ to } 30\text{ mm})Thin-wall piping, small-diameter vessel nozzles; high radiographic contrast close to X-ray.
Iridium-192192Ir^{192}\text{Ir}≈0.38 MeV\approx 0.38\text{ MeV} (range 0.20–0.61 MeV0.20\text{--}0.61\text{ MeV})73.8 days73.8\text{ days}0.25 in. to 2.50 in.(6 to 64 mm)0.25\text{ in.} \text{ to } 2.50\text{ in.} (6\text{ to } 64\text{ mm})Standard refinery workhorse; used for vessel girth seams, shell welds, and piping.
Cobalt-6060Co^{60}\text{Co}1.25 MeV1.25\text{ MeV} (1.17 and 1.33 MeV1.17\text{ and } 1.33\text{ MeV})5.27 years5.27\text{ years}1.50 in. to 7.00 in.(38 to 178 mm)1.50\text{ in.} \text{ to } 7.00\text{ in.} (38\text{ to } 178\text{ mm})Heavy-wall reactor vessels, thick forged flanges, heavy castings; highly penetrating.

Note

Isotope Decay & Exposure Calculations: Radioisotopes decay according to A(t)=A0⋅e−λt=A0⋅(0.5)t/t1/2A(t) = A_0 \cdot e^{-\lambda t} = A_0 \cdot (0.5)^{t / t_{1/2}}. Because radioactive sources decay continuously, exposure times must be recalculated daily using decay charts or source calculators to achieve target radiographic film densities.


3. Image Quality Indicators (IQIs) / Penetrameters

An Image Quality Indicator (IQI), historically termed a penetrameter, is a standardized gauge placed on the part during radiography to verify radiographic technique sensitivity, contrast, and resolution. An IQI does not measure flaw sizes—it proves that the radiograph has sufficient definition to reveal flaws of a specified magnitude.

ASME Section V Article 2 recognizes two standard IQI designs:

  1. Hole-Type IQIs (Plaques per Table T-233.1 / ASTM E1025)
  2. Wire-Type IQIs (Per Table T-233.2 / ASTM E747)
+-----------------------------------------------------------------------------+
|                        HOLE-TYPE VS. WIRE-TYPE IQIs                         |
|                                                                             |
|   [HOLE-TYPE IQI (PLAQUE)]                    [WIRE-TYPE IQI (ASTM E747)]   |
|   +-------------------------------+          +---------------------------+  |
|   |  25         4T   1T   2T      |          |  [1] [2] [3] [4] [5] [6]  |  |
|   |  ---       (O)  (o)  (O)      |          |   |   |   |   |   |   |   |  |
|   |  AISI                         |          |   |   |   |   |   |   |   |  |
|   +-------------------------------+          |  SET A  ASTM  DIN  FE     |  |
|   - Plaque thickness T = 2% of nominal       +---------------------------+  |
|   - 1T Hole dia = 1 x T                      - Set of 6 encapsulated     |  |
|   - 2T Hole dia = 2 x T (Essential Hole)       wires of increasing dia   |  |
|   - 4T Hole dia = 4 x T                      - Essential Wire must be    |  |
|   - Material alloy matches base metal          clearly visible on film   |  |
+-----------------------------------------------------------------------------+

Hole-Type IQI Mechanics

  • A rectangular metallic plaque whose thickness TplaqueT_{\text{plaque}} is nominally 2% of the material thickness being radiographed.
  • Contains three precision drilled holes:
    • 1T1T Hole: Diameter equals 1×Tplaque1 \times T_{\text{plaque}}
    • 2T2T Hole: Diameter equals 2×Tplaque2 \times T_{\text{plaque}} (Standard Essential Hole)
    • 4T4T Hole: Diameter equals 4×Tplaque4 \times T_{\text{plaque}}
  • Standard radiographic quality level is 2-2T2\text{-}2T (2% plaque thickness, 2T2T essential hole must be visible on the processed radiograph).
  • Higher sensitivity levels include 2-1T2\text{-}1T (2% plaque, 1T1T hole visible) and 1-1T1\text{-}1T (1% plaque, 1T1T hole visible).

Wire-Type IQI Sets (ASTM E747 / ASME Table T-233.2)

Wire IQIs consist of six parallel wires of increasing diameter encased in transparent plastic. ASTM E747 groups wires into four standard sets:

Wire SetEncapsulated Wire NumbersWire Diameters (inches)Wire Diameters (mm)Typical Thickness Application
Set A1, 2, 3, 4, 5, 60.0032, 0.004, 0.005, 0.0063, 0.008, 0.0100.08, 0.10, 0.13, 0.16, 0.20, 0.25Up to 0.25 in.(6.4 mm)0.25\text{ in.} (6.4\text{ mm})
Set B6, 7, 8, 9, 10, 110.010, 0.0125, 0.016, 0.020, 0.025, 0.0320.25, 0.32, 0.40, 0.51, 0.64, 0.81>0.25 in. to 0.75 in.(6.4 to 19 mm)> 0.25\text{ in.} \text{ to } 0.75\text{ in.} (6.4\text{ to } 19\text{ mm})
Set C11, 12, 13, 14, 15, 160.032, 0.040, 0.050, 0.063, 0.080, 0.1000.81, 1.02, 1.27, 1.60, 2.03, 2.54>0.75 in. to 2.00 in.(19 to 51 mm)> 0.75\text{ in.} \text{ to } 2.00\text{ in.} (19\text{ to } 51\text{ mm})
Set D16, 17, 18, 19, 20, 210.100, 0.126, 0.160, 0.200, 0.250, 0.3202.54, 3.20, 4.06, 5.08, 6.35, 8.13>2.00 in. to 8.00 in.(51 to 203 mm)> 2.00\text{ in.} \text{ to } 8.00\text{ in.} (51\text{ to } 203\text{ mm})

4. IQI Selection, Placement Rules, Shims & Lead Markers

Nominal Thickness Determination for IQI Selection (T-276.1)

  • For welds with reinforcement, the governing thickness used to select the IQI is the nominal base metal thickness plus allowable weld reinforcement (both sides, crown plus root).
  • Backing rings or strips are included in the thickness calculation if they remain in place during radiography.
Governing Thickness = Nominal Plate Thickness + Internal Reinforcement + External Reinforcement

IQI Placement Rules (T-277.1)

  1. Source-Side Placement (Mandatory Default): The IQI must be placed on the source side of the component being examined.
  2. Film-Side Placement Exception: When physical accessibility prevents source-side placement (such as an enclosed vessel with no internal access and a panoramic central source is impossible):
    • The IQI may be placed on the film side.
    • A lead letter "F" (at least 1/4 in./6 mm1/4\text{ in.} / 6\text{ mm} high) must be placed adjacent to or on the IQI.
    • The essential hole or wire requirement is typically one size smaller (more stringent) to compensate for the reduced geometric unsharpness on the film side.
  3. Orientation: Wire IQIs must be laid perpendicular to the weld axis so that the wires cross the weld reinforcement and HAZ.

Use of Shims Under IQIs (T-277.3)

When an IQI is placed on the base metal adjacent to a weld that has reinforcement, the base metal is thinner than the weld joint. To prevent the IQI from experiencing an artificially high exposure:

  • A shim of radiographically similar material must be placed underneath the IQI.
  • The thickness of the shim must equal the total weld reinforcement thickness, ensuring the optical density through the IQI matches the optical density through the weld area of interest.
+-----------------------------------------------------------------------------+
|                        SHIM PLACEMENT UNDER HOLE IQI                        |
|                                                                             |
|               [ HOLE IQI ]                                                  |
|             +--------------+                                                |
|             |  20  (o)(O)  |                                                |
|             +--------------+      /\\\\\\\\ WELD CROWN REINFORCEMENT        |
|             | SHIM (t=t_r) |     /  \\\\\\\                                 |
|     +-------+--------------+----+    \\\\\\+--------------------------+      |
|     |                           |     \\\\\|                          |      |
|     |      BASE METAL           |      \\\\|      BASE METAL          |      |
|     |      THICKNESS            |       \\\|      THICKNESS           |      |
|     |                           |        \\|                          |      |
+-----+---------------------------+---------+--------------------------+------+ 
|  Shim Thickness (t_shim) = Total Weld Reinforcement Thickness (t_reinforce) |
+-----------------------------------------------------------------------------+

Backscatter Radiation & Lead Letter "B" Marker (T-223 / T-284)

Radiation scattering off walls, floors, or equipment behind the film holder creates non-image-forming fog that degrades radiograph contrast.

  • A lead letter "B" (minimum 1/2 in./13 mm1/2\text{ in.} / 13\text{ mm} high and 1/16 in./1.5 mm1/16\text{ in.} / 1.5\text{ mm} thick) must be affixed to the back of each film cassette/holder.
  • Acceptance Rule: If a light (low-density) image of the letter "B" appears on a darker background of the processed radiograph, protection from backscatter is insufficient, and the radiograph must be rejected and re-shot.
  • If a dark image of "B" appears, or no image appears, backscatter shielding is acceptable.

5. Radiographic Film Density & Calibration Requirements (T-282)

Transmitted Film Optical Density (DD) is a quantitative measurement of film blackening, defined as:

D=log⁡10(I0It)D = \log_{10}\left(\frac{I_0}{I_t}\right)

Where I0I_0 is incident light intensity and ItI_t is transmitted light intensity. A density of 2.02.0 transmits 1/1001/100 (1%) of incident light; a density of 4.04.0 transmits 1/10,0001/10,000 (0.01%) of incident light.

+-----------------------------------------------------------------------------+
|                   ASME SECTION V ARTICLE 2 FILM DENSITY LIMITS              |
|                                                                             |
|   [X-RAY RADIOGRAPHS]              [GAMMA RADIOGRAPHS]                      |
|   - Single film viewing:           - Single film viewing:                   |
|     1.8 Minimum                      2.0 Minimum                            |
|     4.0 Maximum                      4.0 Maximum                            |
|                                                                             |
|   [COMPOSITE VIEWING (DOUBLE FILM)]                                         |
|   - Composite Density: 2.6 Minimum to 4.0 Maximum                           |
|   - Individual Film Density in Composite: 1.3 Minimum Each                  |
+-----------------------------------------------------------------------------+

Comprehensive Optical Density Thresholds (ASME T-282.1)

Radiation SourceSingle Film Minimum DensitySingle Film Maximum DensityComposite Film Minimum DensityComposite Film Maximum Density
X-Ray Sources1.84.02.6 (each film ≥1.3\ge 1.3)4.0
Gamma Sources (Ir-192, Co-60, Se-75)2.04.02.6 (each film ≥1.3\ge 1.3)4.0

Density Variation Limits (T-282.2)

The optical density within the area of interest (the weld deposit and adjacent heat-affected zone) must remain consistent with the density measured through the IQI:

  • Allowable Density Range: The density in the area of interest shall not vary by more than -15% to +30% from the density measured through the essential hole of the hole-type IQI or adjacent to the essential wire of the wire-type IQI.

Densitometer Calibration Protocols (T-262)

  1. National Standard Step Tablet: Densitometers must be calibrated against a certified national standard step tablet (NIST-traceable) at least every 90 days.
  2. Periodic Verification: Calibration verification must be performed at the beginning of each shift, after 8 hours of continuous operation, or whenever the instrument is energized/relocated.
  3. Verification Tolerance: The densitometer readings must match the step tablet values within ±0.05\pm 0.05 density units.

6. Geometric Unsharpness (UgU_g) Physics & Code Limits (T-274.2)

Geometric Unsharpness (UgU_g) is the penumbral blurring of edges on a radiograph caused by the physical dimensions of the radiation source (focal spot or isotope pellet diameter) and the geometry of the exposure setup.

+-----------------------------------------------------------------------------+
|                        GEOMETRIC UNSHARPNESS FORMULA                        |
|                                                                             |
|                                  F * d                                      |
|                           Ug = ---------                                    |
|                                    D                                        |
|                                                                             |
|   Where:                                                                    |
|   Ug = Geometric Unsharpness (inches or mm)                                 |
|   F  = Source focal spot size or physical isotope dimensions (in. or mm)    |
|   d  = Object-to-Film Distance (OFD) (source-side of object to film)        |
|   D  = Source-to-Object Distance (SOD) (source to source-side of object)    |
|                                                                             |
|   Note: Source-to-Film Distance (SFD) = D + d                               |
+-----------------------------------------------------------------------------+

Maximum Allowable Geometric Unsharpness (ASME Table T-274.2)

Material Thickness Range (tt)Maximum Allowable UgU_g (inches)Maximum Allowable UgU_g (mm)
Under 2.0 in.(50 mm)2.0\text{ in.} (50\text{ mm})0.020 in.0.020\text{ in.}0.51 mm0.51\text{ mm}
2.0 in. to 3.0 in.(50 to 75 mm)2.0\text{ in.} \text{ to } 3.0\text{ in.} (50\text{ to } 75\text{ mm})0.030 in.0.030\text{ in.}0.76 mm0.76\text{ mm}
>3.0 in. to 4.0 in.(75 to 100 mm)> 3.0\text{ in.} \text{ to } 4.0\text{ in.} (75\text{ to } 100\text{ mm})0.040 in.0.040\text{ in.}1.02 mm1.02\text{ mm}
Greater than 4.0 in.(100 mm)4.0\text{ in.} (100\text{ mm})0.070 in.0.070\text{ in.}1.78 mm1.78\text{ mm}

Calculation Example:

An inspector audits a shot on a 1.50 in.1.50\text{ in.} thick shell plate using an Iridium-192 source with a focal size F=0.150 in.F = 0.150\text{ in.}. The source-to-object distance D=20 in.D = 20\text{ in.} and object-to-film distance d=1.65 in.d = 1.65\text{ in.} (plate thickness plus standoff).

Ug=F⋅dD=0.150 in.×1.65 in.20 in.=0.247520=0.0124 in.U_g = \frac{F \cdot d}{D} = \frac{0.150\text{ in.} \times 1.65\text{ in.}}{20\text{ in.}} = \frac{0.2475}{20} = \mathbf{0.0124\text{ in.}}
  • Evaluation: For thickness <2.0 in.< 2.0\text{ in.}, maximum allowable Ug=0.020 in.U_g = 0.020\text{ in.}. Since 0.0124 in.≤0.020 in.0.0124\text{ in.} \le 0.020\text{ in.}, the geometric setup is acceptable.

7. Radiographic Viewing Facilities, Densitometers & Interpretation Protocols

Radiographic film interpretation requires controlled optical environments to ensure subtle flaw indications are visible:

+-----------------------------------------------------------------------------+
|                     RADIOGRAPH VIEWING ENVIRONMENT RULES                    |
|                                                                             |
|   [VIEWING ILLUMINATOR]                      [AMBIENT DARKROOM ENVIRONMENT] |
|   - Variable intensity transmitted light     - Subdued lighting             |
|   - Minimum light intensity:                 - Eliminate reflections from   |
|     1000 cd/m2 or 100 foot-candles             film surface                 |
|   - Masking frames to block glare            - Minimum 5-10 min eye         |
|     around film edges                          adaptation time              |
+-----------------------------------------------------------------------------+

Key Viewing & Interpretation Standards (T-290)

  1. Illuminator Capacity: Illuminators must supply variable-intensity transmitted light capable of clearly reading optical densities up to 4.0.
  2. Masking: High-intensity light escaping around the outer perimeter of the film creates peripheral glare that severely impairs visual acuity. Masking plates or diaphragms must be used to block all light outside the film border.
  3. Eye Dark Adaptation: Interpreters entering the darkroom or viewing area from bright daylight must allow sufficient time (typically 5 to 10 minutes5\text{ to } 10\text{ minutes}) for rhodopsin regeneration and dark adaptation before evaluating radiographs.

8. Common Exam Traps & Real-World Pitfalls in Radiographic Examination

Warning

Trap 1: The Backscatter 'B' Image - Light vs. Dark: A favorite API 510 question asks: "During interpretation, a dark image of the letter 'B' appears on the film. Is the radiograph acceptable?"

  • Rule: Only a LIGHT image of the letter 'B' on a darker background indicates excessive backscatter and causes REJECTION.
  • A dark image of 'B' is simply an overexposure artifact or secondary reflection and is ACCEPTABLE unless it obscures the weld area.

Important

Trap 2: Minimum Density Confusion (X-Ray vs. Gamma): Do not mix up the minimum density limits: 1.8 for X-ray versus 2.0 for Gamma sources. Both share the same single-film maximum limit of 4.0.

Tip

Trap 3: IQI Selection Governing Thickness: When selecting an IQI for a welded joint, always include the weld reinforcement thickness added to the nominal plate thickness. If base metal is 0.500 in.0.500\text{ in.} with 0.125 in.0.125\text{ in.} reinforcement on both crown and root, the governing thickness is 0.500+0.125+0.125=0.750 in.0.500 + 0.125 + 0.125 = 0.750\text{ in.}, which changes the required essential wire or hole designation!

Loading diagram...
ASME Section V Article 2 Radiographic Quality & Acceptance Workflow
Test Your Knowledge

An API 510 inspector reviews an industrial gamma radiograph taken with an Iridium-192 source on a 1.0-inch thick pressure vessel shell weld. When examining the film on an illuminator, a distinct LIGHT image of the lead letter 'B' appears against the darker background of the radiograph. What action does ASME Section V Article 2 require?

A

The radiograph must be rejected because a light image of the letter 'B' indicates that unacceptable backscatter radiation reached the film cassette.

B

The radiograph is acceptable because the appearance of the letter 'B' confirms that backscatter shielding was successfully deployed.

C

The radiograph is acceptable provided the transmitted film optical density through the weld is at least 1.8.

D

The radiograph is acceptable only if the essential wire on the wire-type IQI is clearly resolved.

Test Your Knowledge

What are the minimum and maximum transmitted film optical density limits permitted by ASME Section V Article 2 for a single radiograph exposed using an Iridium-192 gamma ray source?

A

Minimum density 1.8, maximum density 3.5

B

Minimum density 2.0, maximum density 4.0

C

Minimum density 1.5, maximum density 4.0

D

Minimum density 2.6, maximum density 4.5

Test Your Knowledge

When radiographing a vessel circumferential seam where the inside of the vessel is physically inaccessible, an inspector must place the wire-type IQI on the film side. Which of the following is MANDATORY under ASME Section V Article 2?

A

A lead letter 'S' must be affixed to the source side of the weld.

B

The radiograph density must be increased to a minimum of 3.0.

C

A lead letter 'F' must be placed adjacent to or on the IQI on the film side.

D

The object-to-film distance must be increased by at least 25%.

Test Your Knowledge

An exposure setup uses an X-ray tube with a focal spot size (F) of 0.120 in. (3.0 mm). The distance from the radiation source to the source-side of a 1.25-inch thick vessel shell (D) is 24.0 in., and the distance from the source-side of the shell to the film (d) is 1.50 in. What is the calculated geometric unsharpness (Ug), and is it acceptable per ASME Section V Table T-274.2?

A

Ug = 0.030 in.; Rejected because maximum allowable Ug for material under 2.0 in. is 0.020 in.

B

Ug = 0.015 in.; Rejected because maximum allowable Ug for material under 2.0 in. is 0.010 in.

C

Ug = 0.025 in.; Acceptable because maximum allowable Ug for material under 2.0 in. is 0.030 in.

D

Ug = 0.0075 in.; Acceptable because maximum allowable Ug for material under 2.0 in. is 0.020 in.

Sections you finish are checked off in the contents.