3.1 Montage, 10-20 System & Electrode Application
Key Takeaways
The American Academy of Sleep Medicine (AASM) recommended electroencephalogram (EEG) montage uses F4-M1, C4-M1, and O2-M1 with F3-M2, C3-M2, and O1-M2 as backups.
Electrode sites are placed using the International 10-20 System, where measurements are taken as 10% or 20% of the distance between bony landmarks (nasion, inion, and preauricular points).
Impedances should be balanced and at or below 5 kilohms (5 kΩ) for EEG, electrooculogram (EOG), and electromyogram (EMG) electrodes before recording.
Gold cup or silver/silver-chloride (Ag/AgCl) electrodes are applied with conductive paste or collodion; reference (M1, M2) electrodes are placed on the mastoid processes.
A clean, low-impedance, mechanically secure electrode application is the single biggest determinant of a scoreable, artifact-free study.
Why the Montage Matters for the RPSGT Exam
Sleep Study Preparation and Performance is the largest domain on the Registered Polysomnographic Technologist (RPSGT) examination at 27.3% of scored items per the Board of Registered Polysomnographic Technologists (BRPT) exam blueprint (tied with Treatment and Intervention, also 27.3%). More questions test how you derive a montage, measure head sites, and achieve clean impedances than test any single scoring rule. The RPSGT exam is a 175-item computer-based test (150 scored plus 25 unscored pretest items) delivered in 180 minutes, with a scaled passing score of 350 on a 200-500 scale.
A study can only be scored as accurately as it was acquired, so the BRPT expects you to build a recording that meets AASM technical specifications before the patient ever falls asleep.
A montage is the specific set of derivations (electrode pairs) displayed on the recording. Each channel shows the voltage difference between an exploring electrode and a reference electrode. Because PSG channels are differential amplifiers, a clean reference is as important as the exploring site, and unbalanced impedances destroy common-mode rejection.
The Recommended PSG EEG Montage
The AASM recommends three frontal, central, and occipital EEG derivations referenced to the contralateral mastoid:
| Derivation | Exploring | Reference | Captures |
|---|---|---|---|
| F4-M1 | Right frontal | Left mastoid | K-complexes, slow-wave activity |
| C4-M1 | Right central | Left mastoid | Sleep spindles, vertex sharp waves |
| O2-M1 | Right occipital | Left mastoid | Posterior dominant (alpha) rhythm |
Backup derivations F3-M2, C3-M2, and O1-M2 mirror the left hemisphere so the study stays scoreable if a primary electrode fails mid-night. Mastoid references are labeled M1 (left) and M2 (right); the older 10-20 labels for these sites are A1 and A2. The AASM also specifies a minimum EEG sampling rate of 200 Hz (desirable 500 Hz) so spindle and slow-wave morphology survive digitization.
Worked Example
Suppose C4 pops free at 1:30 a.m. and cannot be re-secured without waking the patient. You switch staging to the C3-M2 backup derivation. Because the backup mirrors the same region on the opposite hemisphere, spindle detection and arousal scoring continue uninterrupted — provided you logged the change with a timestamp so the scorer knows which channel was active. This is exactly why the AASM mandates a full backup chain rather than a single C-channel.
The International 10-20 System
The International 10-20 System is a standardized method for placing scalp electrodes so recordings are reproducible between technologists, nights, and labs. The name refers to the spacing: each electrode sits at 10% or 20% of the total measured distance between defined skull landmarks. Because every site is a percentage of the individual's head, the system self-scales to infants and adults alike.
Bony Landmarks
- Nasion — the depression at the bridge of the nose, between the eyes
- Inion — the bony bump at the back of the skull (external occipital protuberance)
- Preauricular points — the depressions just in front of each ear canal
- Vertex (Cz) — the true top of the head, found where the front-to-back and ear-to-ear lines cross
Naming Convention
- Letters indicate the brain region: F = frontal, C = central, O = occipital, P = parietal, T = temporal, Fp = frontal pole
- Odd numbers (1, 3, 5) are on the left hemisphere; even numbers (2, 4, 6) are on the right
- z denotes a midline electrode (Fz, Cz, Pz)
Core Measurement Steps
- Measure the anterior-posterior distance from nasion to inion over the vertex.
- Mark Fpz at 10%, Fz at 30%, Cz at 50%, Pz at 70%, and Oz at 90% along that line.
- Measure the transverse distance between the two preauricular points through Cz; true Cz is where both measurements intersect.
- Measure the circumference and place the temporal/lateral chain at the matching 10% and 20% increments.
Using a flexible tape and a skin-marking pencil, the technologist must measure rather than estimate. A misplaced central electrode shifts spindle amplitude; a misplaced occipital electrode weakens the posterior dominant rhythm needed to identify wake and stage N1, both of which propagate into sleep-stage scoring errors.
Common Traps
- Eyeballing Cz instead of crossing both measured lines — the most common placement error.
- Reversing odd/even numbering (left vs. right) — mislabels the whole montage.
- Forgetting that F4 references M1 (contralateral), not the same-side mastoid; same-side referencing reduces the captured signal and invites ECG contamination from the nearby carotid.
Electrode Types, Application & Impedance
Electrode Types
- Gold cup electrodes — durable, low-noise, used for EEG/EOG; filled with conductive paste
- Silver/silver-chloride (Ag/AgCl) — stable low-offset potential, excellent for slow signals such as DC-coupled recordings
- Adhesive snap (disposable) — common for electrocardiogram (ECG) and limb electromyogram (EMG)
Application Technique
- Locate and mark the site using 10-20 measurements.
- Prep the skin with a mild abrasive prep gel to remove dead skin and oils, lowering impedance at the electrode-skin interface.
- Apply conductive paste in the cup, place the electrode, and secure with a gauze square or collodion (a fast-drying adhesive used for long studies and pediatrics).
- Strain-relieve lead wires and bundle them so head movement does not pop electrodes.
Impedance Targets
Impedance is the opposition to alternating current at the electrode-skin interface, measured in kilohms. High or unequal impedances cause 60 Hz (mains) artifact, electrode pop, and signal attenuation.
| Electrode group | Target impedance |
|---|---|
| EEG, EOG, EMG | At or below 5 kΩ |
| Paired electrodes | Balanced (similar values) for common-mode rejection |
| ECG / limb (less critical) | Low and stable, though slightly higher tolerated |
The AASM standard is concrete: re-prep and reapply any electrode that exceeds 5 kΩ rather than accepting a noisy channel. Equally important, impedances must be balanced — a 4 kΩ exploring electrode paired with an 18 kΩ reference produces 60 Hz contamination even though one electrode is technically in range, because the mismatch defeats the amplifier's common-mode rejection.
Worked Scenario
During impedance check the C4 site reads 12 kΩ. You re-abrade the scalp, add fresh paste, and re-seat the cup; it now reads 3 kΩ, matching its M1 reference at 3 kΩ. The pair is now both low and balanced. Always resolve impedance problems before lights out — fixing a popped occipital electrode at 2 a.m. risks waking the patient and losing data, and the BRPT explicitly tests the principle that prevention beats mid-night repair.
What is the AASM-recommended primary EEG derivation set for routine polysomnography?
Fp1-F7, F7-T3, T3-T5
Fz-Cz, Cz-Pz, Pz-Oz
C3-A1, C4-A2 only
F4-M1, C4-M1, O2-M1
In the International 10-20 System, which two landmarks define the anterior-posterior measurement line used to place midline electrodes?
The nasion and the inion
The two preauricular points
The vertex and the nasion
The mastoid processes
Before lights out, an EEG electrode reads an impedance of 18 kΩ while its paired reference reads 4 kΩ. What is the correct action?
Begin recording; impedance does not affect EEG quality
Reduce the EEG sensitivity to hide the artifact
Switch off the 60 Hz notch filter and start the study
Re-prep the skin and reapply the high-impedance electrode to bring it to ≤5 kΩ and balance the pair
Sections you finish are checked off in the contents.