3.1 SI Measurement, Scientific Notation, and Error Analysis

Key Takeaways

  • The SI has seven base units and uses fixed defining constants.

  • Derived units and conversion factors help track dimensions. For example, the pascal is a newton per square meter.

  • Accuracy concerns agreement with the quantity measured. Precision concerns agreement among repeated values under specified conditions.

  • Scientific notation makes significant trailing zeros explicit. Keep guard digits until final rounding.

  • Absolute percent error is the magnitude of a discrepancy relative to a nonzero stated reference.

Last updated: November 2026

3.1 SI Measurement, Scientific Notation, and Error Analysis

Measurements are easier to compare when they use common units and clearly report how the values were obtained and rounded.


The International System of Units (SI)

The International System of Units (Système International d'Unités, or SI) is the globally adopted metric framework maintained by the General Conference on Weights and Measures (CGPM). The modern SI fixes seven defining constants. The table connects each base unit to the constant most directly associated with its definition. There are also derived units, and some quantities are dimensionless.

Physical QuantityBase Unit NameSymbolDefining Physical Constant / Standard
LengthMeterm\text{m}Distance light travels in a vacuum in 1/299,792,4581 / 299,792,458 of a second (cc).
MassKilogramkg\text{kg}Defined via the Planck constant hh (6.62607015×10−34 kg⋅m2/s6.62607015 \times 10^{-34} \text{ kg}\cdot\text{m}^2/\text{s}). (Note: the base unit is the kilogram, not the gram.)
TimeSeconds\text{s}Unperturbed ground-state hyperfine transition frequency of cesium-133 (ΔνCs=9,192,631,770 Hz\Delta \nu_{\text{Cs}} = 9,192,631,770 \text{ Hz}).
Thermodynamic TemperatureKelvinK\text{K}Defined via the Boltzmann constant kk (1.380649×10−23 J/K1.380649 \times 10^{-23} \text{ J/K}). (Note: no degree symbol is used with kelvin.)
Amount of SubstanceMolemol\text{mol}Exactly 6.02214076×10236.02214076 \times 10^{23} elementary entities (Avogadro's constant NAN_A).
Electric CurrentAmpereA\text{A}Defined via the elementary electric charge ee (1.602176634×10−19 C1.602176634 \times 10^{-19} \text{ C}).
Luminous IntensityCandelacd\text{cd}Luminous efficacy KcdK_{\rm cd} of monochromatic radiation at 540×1012 Hz540 \times 10^{12}\ \mathrm{Hz}, fixed at 683 lm/W683\ \mathrm{lm/W}.

Metric Prefixes and Exponential Scaling

Metric prefixes modify SI units by factors of ten, enabling scientists to describe phenomena spanning dozens of orders of magnitude—from subatomic distances to galactic scales—without unwieldy strings of zeros.

PrefixSymbolExponential MultiplierDecimal EquivalentLaboratory / Scientific Example
Giga-G\text{G}10910^91,000,000,0001,000,000,000Regional electrical generation (gigawatts, GW\text{GW}); computer processor clocks (gigahertz, GHz\text{GHz})
Mega-M\text{M}10610^61,000,0001,000,000Radio frequencies (megahertz, MHz\text{MHz}); energy (megajoules, MJ\text{MJ})
Kilo-k\text{k}10310^31,0001,000Macroscopic mass measurements (kilograms, kg\text{kg}); geographic distances (kilometers, km\text{km})
Deci-d\text{d}10−110^{-1}0.10.1Clinical fluid measurements in human physiology (deciliters, dL\text{dL})
Centi-c\text{c}10−210^{-2}0.010.01Standard benchtop glassware scale markings (centimeters, cm\text{cm})
Milli-m\text{m}10−310^{-3}0.0010.001Precision analytical reagent masses (milligrams, mg\text{mg}); pipette liquid volumes (milliliters, mL\text{mL})
Micro-μ\mu10−610^{-6}0.0000010.000001Eukaryotic cell diameters and bacterial dimensions (micrometers, μm\mu\text{m})
Nano-n\text{n}10−910^{-9}0.0000000010.000000001Optical wavelengths and nanoscale dimensions

Derived Units and Dimensional Equivalence

Derived units are formed from base units according to the relationships among the quantities (F=maF = ma, W=FdW = Fd, P=F/AP = F/A, etc.).

Derived QuantityUnit NameSymbolBase Unit EquivalenceGoverning Physical Equation
ForceNewtonN\text{N}kg⋅m/s2\text{kg}\cdot\text{m}/\text{s}^2Newton’s second law: Fnet=ma\mathbf{F}_{\rm net}=m\mathbf a
Energy, Work, HeatJouleJ\text{J}N⋅m=kg⋅m2/s2\text{N}\cdot\text{m} = \text{kg}\cdot\text{m}^2/\text{s}^2Work for constant force: W=F⋅ΔrW=\mathbf{F}\cdot\Delta\mathbf{r}
PowerWattW\text{W}J/s=kg⋅m2/s3\text{J}/\text{s} = \text{kg}\cdot\text{m}^2/\text{s}^3Average power: Pav=W/ΔtP_{\rm av}=W/\Delta t
Pressure, StressPascalPa\text{Pa}N/m2=kg/(m⋅s2)\text{N}/\text{m}^2 = \text{kg}/(\text{m}\cdot\text{s}^2)Average pressure: p=F⊥/Ap=F_\perp/A
FrequencyHertzHz\text{Hz}1/s=s−11/\text{s} = \text{s}^{-1}Frequency: f=1/Tf = 1 / T
Electric PotentialVoltV\text{V}J/C=kg⋅m2/(A⋅s3)\text{J}/\text{C} = \text{kg}\cdot\text{m}^2/(\text{A}\cdot\text{s}^3)Potential Difference: V=W/qV = W / q

Understanding base unit equivalence allows candidates to verify formulas dimensionally. An equation relating the same quantity must be dimensionally consistent when both sides use compatible units. Dimensional agreement alone cannot show that a formula is correct.


Scientific Notation and Dimensional Analysis

Scientific Notation

Normalized scientific notation expresses a nonzero number as a coefficient times an integer power of ten:

M×10nwhere 1≤∣M∣<10 and n∈ZM \times 10^n \quad \text{where } 1 \le |M| < 10 \text{ and } n \in \mathbb{Z}
  • Magnitudes of at least 10 have positive exponents: 149,600,000 km=1.496×108 km149,600,000\text{ km} = 1.496 \times 10^8\text{ km} (Earth-Sun distance).
  • Nonzero magnitudes below 1 have negative exponents: 0.000000155 m=1.55×10−7 m0.000000155\text{ m} = 1.55 \times 10^{-7}\text{ m}.
  • When multiplying: multiply coefficients and add exponents: (2.0×104)×(3.0×10−6)=6.0×10−2(2.0 \times 10^4) \times (3.0 \times 10^{-6}) = 6.0 \times 10^{-2}.
  • When dividing: divide coefficients and subtract exponents: (8.0×105)/(2.0×102)=4.0×103(8.0 \times 10^5) / (2.0 \times 10^2) = 4.0 \times 10^3.

Dimensional Analysis (Unit Factor Method)

Dimensional analysis treats units as algebraic quantities that can be multiplied, divided, and canceled. A conversion factor is a ratio equal to 1, formed from an established mathematical equality (e.g., 1 km=1000 m1\text{ km} = 1000\text{ m}, so 1000 m1 km=1\frac{1000\text{ m}}{1\text{ km}} = 1).

Worked Example 1: Velocity Conversion

Convert 72.0 km/h72.0\text{ km/h} to m/s\text{m/s}:

72.0 km/h×(1000 m1 km)×(1 h3600 s)=72.0×10003600 m/s=20.0 m/s72.0\text{ km/h} \times \left(\frac{1000\text{ m}}{1\text{ km}}\right) \times \left(\frac{1\text{ h}}{3600\text{ s}}\right) = \frac{72.0 \times 1000}{3600}\text{ m/s} = 20.0\text{ m/s}

Worked Example 2: Multidimensional Density Conversion

Convert a mineral density of 2.70 g/cm32.70\text{ g/cm}^3 to kg/m3\text{kg/m}^3. Remember that cubic units require cubing the conversion factor:

2.70 g/cm3×(1 kg1000 g)×(100 cm1 m)3=2.70×(11000)×1,000,000 kg/m3=2.70×103 kg/m32.70\text{ g/cm}^3 \times \left(\frac{1\text{ kg}}{1000\text{ g}}\right) \times \left(\frac{100\text{ cm}}{1\text{ m}}\right)^3 = 2.70 \times \left(\frac{1}{1000}\right) \times 1,000,000\text{ kg/m}^3 = 2.70 \times 10^3\text{ kg/m}^3

Precision versus Accuracy

Accuracy concerns closeness to the value of the quantity being measured. A certified reference can help assess closeness, with its own stated uncertainty. Precision concerns agreement among repeated measurements under specified conditions. Precision under the same procedure, operator, instrument and short-term conditions is repeatability. A closely grouped set of readings may still have a substantial bias.

Tight cluster near reference         Tight cluster away from reference
Little scatter; small difference    Little scatter; large difference

Wide scatter centered near reference Wide scatter away from reference
Mean may be close; readings vary     Readings vary and mean differs

Worked Example: Comparing Balance Readings

Use 50.000 g as the assigned reference value in this example. A reference measurement can have uncertainty even with three reported decimal places.

  • Balance A: 47.121, 47.120, 47.122 and 47.121 g. The mean is 47.121 g, and the readings span 0.002 g. They cluster closely but have a mean 2.879 g below the reference. This pattern suggests bias under the tested conditions. Additional reference checks can help identify its cause.
  • Balance B: 49.8, 50.3, 49.7 and 50.2 g. The mean is 50.0 g and the range is 0.6 g. The mean is close to the reference, with more scatter than the Balance A set. A small mean difference does not make each reading precise.

Rules for Significant Figures

Significant figures show which digits are reported in a rounded measurement. They do not give its full uncertainty. For an analog scale, an estimate between divisions often determines the final recorded digit. A digital readout does not necessarily measure accurately to its last digit.

Identification Rules

  1. Non-zero digits are always significant:
    • 38.4 g38.4\text{ g} has 3 significant figures.
  2. Captive zeros between non-zero digits are always significant:
    • 4008 km4008\text{ km} has 4 significant figures; 3.05 s3.05\text{ s} has 3 significant figures.
  3. Leading zeros appearing before the first non-zero digit are never significant; they are mere place-holders:
    • 0.0042 m0.0042\text{ m} has only 2 significant figures (44 and 22). In scientific notation, 4.2×10−3 m4.2 \times 10^{-3}\text{ m} makes this evident.
  4. Trailing zeros after a displayed decimal point can express measured precision. A whole number with ending zeros can be ambiguous:
    • 120.0 mL120.0\text{ mL} has 4 significant figures (the decimal confirms precision to tenths).
    • 50. s50.\text{ s} has 2 significant figures (the explicit decimal point indicates precision to the ones place).
    • 1500 m1500\text{ m} leaves the status of the zeros unclear. Use 1.5×1031.5\times10^3 for 2, 1.50×1031.50\times10^3 for 3 or 1.500×1031.500\times10^3 for 4 significant figures.
  5. Exact numbers (counted objects such as 12 test tubes, or defined conversion factors such as 1000 m=1 km1000\text{ m} = 1\text{ km}) do not impose a rounding limit on the calculation.

Classroom Rounding Rules

Keep extra digits during intermediate steps and round the final answer. If measurement uncertainties are available, explicit uncertainty analysis is more informative than these classroom conventions.

  • Addition and Subtraction (Decimal Place Rule): The final result must be rounded to match the coarsest reported place value after expressing quantities in common units: 142.1 g+0.345 g+12.08 g=154.525 g→154.5 g142.1\text{ g} + 0.345\text{ g} + 12.08\text{ g} = 154.525\text{ g} \rightarrow 154.5\text{ g} (Since 142.1142.1 has only one decimal place, the answer rounds to the tenths place.)

  • Multiplication and Division (Significant Figures Rule): The final result must be rounded to possess the same number of significant figures as the originating factor with the fewest significant figures: Area=14.25 cm (4 sig figs)×3.2 cm (2 sig figs)=45.6 cm2→46 cm2\text{Area} = 14.25\text{ cm (4 sig figs)} \times 3.2\text{ cm (2 sig figs)} = 45.6\text{ cm}^2 \rightarrow 46\text{ cm}^2


Error Calculations and Analysis

Measurement error is the measured value minus a reference value. In practice the reference is often an estimate with uncertainty. Keep the sign when the direction matters. An absolute error is the magnitude of that difference. A commonly used classroom absolute percent error is

∣measured value−reference value∣∣reference value∣×100%.\frac{|\text{measured value}-\text{reference value}|}{|\text{reference value}|}\times100\%.

Use compatible units and a nonzero reference. For a zero reference, report a difference or an uncertainty in the original unit instead. A small percent error for one reading does not establish precision or characterize the entire measuring process.

A systematic component remains constant or changes predictably under repeated conditions. Independent random errors cause scatter, so averaging can reduce the uncertainty in the mean without removing a systematic offset. Sampling variation can also arise from choosing different subsets of a population. It is distinguished from variation in repeated measurements of one sample.

Worked Example: Estimating Gravitational Acceleration

A pendulum experiment gives 9.54 m/s29.54\ \mathrm{m/s^2}, and the stated reference for the exercise is 9.81 m/s29.81\ \mathrm{m/s^2}. The absolute error is 0.27 m/s20.27\ \mathrm{m/s^2}. Dividing 0.270.27 by 9.819.81 and multiplying by 100%100\% gives about 2.75%2.75\%, or 2.8%2.8\% rounded to two significant figures. This is agreement with the stated reference in the example. Local gravitational acceleration varies with location.

Test Your Knowledge

Mass readings are 105.2 g, 8.455 g and 0.08 g. Using the standard classroom place-value rule, what total should be reported?

A

113.735 g

B

113.74 g

C

114 g

D

113.7 g

Test Your Knowledge

Which expression gives one pascal entirely in SI base units?

A

kg / (m·s²)

B

kg·m / s²

C

kg·m² / s²

D

kg / (m²·s)

Test Your Knowledge

A balance gives 22.411, 22.410, 22.412 and 22.411 g for a reference mass assigned 25.000 g. Which description compares these results with the reference?

A

Small scatter and close agreement with the reference

B

Large scatter and close agreement with the reference

C

Small scatter and poor agreement with the reference

D

Large scatter and poor agreement with the reference