3.1 Physical Principles of Dual-Energy X-Ray Absorptiometry

Key Takeaways

  • For a uniform absorber and a narrow monoenergetic beam, the transmitted intensity decreases exponentially with thickness.

  • Photoelectric absorption and scattering contribute to attenuation. Their relative contributions vary with photon energy and the material.

  • Two measurements with different material attenuation ratios can separate two model components. Real DXA also uses spectral calibration and a soft-tissue model.

  • K-edge filtration and rapid tube-potential switching produce different energy spectra. Tube potential and photon energy are distinct quantities.

Last updated: November 2026

Describe what the detector measures

Dual-energy X-ray absorptiometry (DXA) measures the transmission of two different X-ray spectra through the patient. Bone mineral and soft tissue change the transmitted signal differently as photon energy changes. The scanner uses those differences and its calibration to estimate bone mineral mass within the projected image.

For a narrow monoenergetic beam passing through a uniform material with negligible detected scatter, the Beer–Lambert model is

I=I0e−μx.I=I_0e^{-\mu x}.

Here I0I_0 is the incident intensity, II is the transmitted intensity, μ\mu is the linear attenuation coefficient in cm⁻¹ and xx is path length in cm. The exponent is dimensionless. Writing the mass attenuation coefficient as μ/ρ\mu/\rho in cm²/g and the areal mass as M=ρxM=\rho x in g/cm² gives I=I0e−(μ/ρ)MI=I_0e^{-(\mu/\rho)M}. Attenuation includes photons absorbed and photons scattered out of the primary beam.

Photoelectric absorption removes the incident photon and ejects a bound electron. Its contribution is sensitive to elemental composition and increases at lower energies away from absorption edges. This helps distinguish mineral-rich bone from soft tissue. Compton scattering changes the direction and energy of a photon and transfers energy to an electron. Its contribution per unit mass relates to the number of electrons per gram and becomes relatively more important as energy rises in the diagnostic range. Coherent scattering can also contribute. Actual attenuation coefficients depend on photon energy and tissue composition. A single atomic-number power or fixed bone-to-soft-tissue contrast ratio cannot predict both spectra.

Solve a simplified two-basis measurement

Let AL=ln⁡(I0,L/IL)A_L=\ln(I_{0,L}/I_L) and AH=ln⁡(I0,H/IH)A_H=\ln(I_{0,H}/I_H) be the logarithmic attenuations at low and high energies. Let bb and ss stand for the bone and soft-tissue mass attenuation coefficients at each energy. In a two-basis model,

AL=bLMB+sLMS,AH=bHMB+sHMS.\begin{aligned} A_L&=b_LM_B+s_LM_S,\\ A_H&=b_HM_B+s_HM_S. \end{aligned}

The unknowns MBM_B and MSM_S are the projected bone and soft-tissue mass per area. Multiply the second equation by RS=sL/sHR_S=s_L/s_H and subtract to cancel the soft-tissue term.

MB=AL−RSAHbL−RSbH.M_B=\frac{A_L-R_SA_H}{b_L-R_Sb_H}.

The denominator must differ from zero. If both materials have the same low-to-high attenuation ratio, the measurements cannot separate the two contributions. As a worked example, use illustrative coefficients bL=0.80b_L=0.80, bH=0.35b_H=0.35, sL=0.25s_L=0.25, sH=0.20s_H=0.20 in cm²/g, with measured log attenuations of 1.14 and 0.68. Then RS=1.25R_S=1.25 and the estimated bone areal mass is (1.14−1.25×0.68)/(0.80−1.25×0.35)=0.80(1.14-1.25\times0.68)/(0.80-1.25\times0.35)=0.80 g/cm². Substituting gives MS=2.0M_S=2.0 g/cm².

Commercial beams contain distributions of energies. Lower-energy photons are preferentially attenuated along the path, changing the transmitted spectrum. The ideal equations describe the separation principle. Commercial systems account for spectra and beam hardening with instrument-specific calibration. They also need assumptions about the soft tissue overlying bone. Two energy measurements cannot independently solve three unknown areal masses for mineral, fat and lean tissue in the same pixel. Estimated composition in nearby soft tissue helps model the overlying tissue. Nonuniform fat distribution and major body-composition change can bias measured BMD, with magnitude and direction depending on the site and setup.

Distinguish tube potential and photon energy

MethodHow the two spectra are obtained
K-edge filtrationA filter changes transmission sharply near its K absorption edge. An energy-sensitive detection scheme separates the low- and high-energy components from a filtered spectrum. FDA device specifications list a 100 kV tube and samarium filter for the Lunar iDXA and a 76 kV tube and cerium filter for Lunar Astra.
Rapid tube-potential switchingThe tube alternates between two settings and acquisition is synchronized to them. A measurement study of Hologic Discovery and Horizon scanners lists 100 and 140 kV for its equipment. Verify the actual model rather than assuming all scanners use those settings.

A tube potential in kilovolts describes electron acceleration and bounds the maximum bremsstrahlung photon energy in keV. For example, a 100 kV tube can produce photons up to about 100 keV. The beam includes a range of lower energies. Neither the tube voltage nor a filter's K edge alone specifies the average photon energy. Beam filtration changes the energy distribution and therefore its penetrating ability. A quality-control measurement such as half-value layer describes penetration by measuring the thickness of a specified material needed to halve air kerma under defined conditions. Compare the same beam and acquisition settings when using that measurement.

IAEA DXA physics · NIST photon cross sections · FDA GE device specifications · Hologic measurement study · Fat-layering study

Test Your Knowledge

Which interaction most helps explain why composition-dependent contrast between bone mineral and soft tissue is pronounced toward the lower end of the DXA photon spectrum?

A

Photoelectric absorption, which varies strongly with energy and elemental composition.

B

Pair production, which predominates at the low end of DXA spectra.

C

Compton scattering, which depends primarily on mass and is independent of photon energy.

D

Radiofrequency resonance of hydrogen nuclei.

Test Your Knowledge

In a two-basis attenuation model, why must the bone and soft-tissue low-to-high attenuation ratios differ?

A

This forces both X-ray spectra to be monoenergetic.

B

The bone and soft-tissue terms cannot be separated if the two equations are proportional.

C

Otherwise the scanner measures organ equivalent dose instead of density.

D

A difference is required only when the two spectra are produced with a K-edge filter.

Test Your Knowledge

The Lunar iDXA specification lists a 100 kV tube and a samarium K-edge filter. What follows from that tube setting?

A

Both recorded energy groups consist solely of 100 keV photons.

B

Samarium emits a single 100 keV line that replaces the entire incident spectrum.

C

The tube alternates between 100 and 140 kV on this specified model.

D

The bremsstrahlung endpoint is about 100 keV. Filtration and detection distinguish portions of the spectrum.

Sections you finish are checked off in the contents.