11.2 Georgia Rule 240-4: Cleansing, Disinfection, Storage, and Prohibited Implements

Key Takeaways

  • Georgia's current cleansing, disinfection, storage, hand-washing, and prohibited-implement rules sit in Chapter 240-4, especially Rules 240-4-.03, 240-4-.04, and 240-4-.05; Chapter 240-5 is apprentices, not the sanitation chapter.
  • After cleaning with warm soapy water, multi-use implements require complete immersion for 10 minutes in an EPA-registered disinfectant; the solution must be changed daily and tools cannot be stored in disinfectant.
  • Ultraviolet lights, dishwashers, and steam devices are not acceptable methods of cleansing and disinfection; an autoclave meeting ANSI/AAMI ST55:2010 or ISO 17665 is an acceptable sanitizing method.
  • Clean tools belong in a closed airtight drawer or container labeled Clean Implements; soiled tools belong in a separate closed container labeled Dirty Implements.
  • Rule 240-4-.03 prohibits credo blades, razor-type callus shavers, cutters, scrapers, graters, and anything that abrades or removes layers of skin; products may not be double-dipped.
Last updated: September 2026

Chapter 240-4 is sanitation; Chapter 240-5 is apprentices

Georgia candidates lose easy points by studying the wrong chapter number. On https://rules.sos.ga.gov/gac/240-4, Chapter 240-4 is titled FACILITY REQUIREMENTS and contains the health, safety, cleansing, disinfection, and storage rules:

  • 240-4-.01 Requirements of Operating a Facility
  • 240-4-.02 Posting of Licenses, Rules, Signs and Inspection Reports
  • 240-4-.03 General Health and Safety
  • 240-4-.04 Cleansing and Disinfection of Implements and Equipment
  • 240-4-.05 Storage of Implements and Equipment

Chapter 240-5 on the same rules site is APPRENTICES. Do not treat 240-5 as the disinfection chapter. Rule 240-4-.02(4) on the page opened for this chapter still says facilities shall comply with sanitation, health, and disinfectant rules "in Chapter 240-5." Treat that sentence as a leftover cross-reference on the posted text. The numbered procedures inspectors actually walk — hand washing, credo-blade ban, 10-minute EPA immersion, Clean/Dirty labels, no double-dip — are 240-4-.03 through 240-4-.05.

Rule 240-4-.04(1) is the blanket prohibition: the use of any article that is not properly cleansed and disinfected on any client or subject is prohibited.

Hands, clothing, sharps, and the credo-blade ban

Rule 240-4-.03(2) requires hand washing before and between providing services to each client. An anti-bacterial soap is recommended to sanitize the hands and the exposed portions of arms before services and after smoking, drinking, eating, and using restrooms. That is a between-every-client duty, not a morning-only sink trip.

Practitioners shall wear appropriate protective clothing and/or footwear during clinical services to prevent occupational exposure; the rule points to OSHA for garments that may include clinical jackets, gloves, and similar outer garments (240-4-.03(3)). Clients must wear appropriate clothing and footwear, or the shop may provide a covering (240-4-.03(4)).

Pets shall not be allowed in facilities, with the exception of guide and assistance dogs for qualifying clients as permitted by Title 30 of the Georgia Code (240-4-.03(5)). Waste must be removed daily. Garbage is stored in a covered, washable container and shall not be left in the establishment overnight. Work stations must be free from food and soiled dishes (240-4-.03(6)).

Sharps such as razor blades go in a sharps container: stored upright, with OSHA biohazard labels, manufacturer's safety-flap lid in place, and properly disposed of when three-fourths (3/4) full (240-4-.03(7)).

Rule 240-4-.03(8) is the implement ban every Georgia practical and theory candidate should be able to recite. In order to reduce the chance of injury or infection, licensees shall not use any implement designed to remove layers of skin such as:

  • razor-type callus shavers designed and intended to cut growths of skin such as corns or calluses;
  • credo blades;
  • cutters;
  • scrapers;
  • graters;
  • anything that abrades or removes layers of skin.

A Duluth facialist who keeps a credo blade "only for heel callus add-ons" is already outside the rule. The ban is the implement type, not the body part.

Multi-use versus single-use

Rule 240-4-.04 splits the world into items you disinfect and items you throw away.

Multi-use tools, implements, and equipment constructed of nonporous materials such as metal, glass, or plastic, for use on more than one client, include (the rule's list is not exclusive): nail clippers; cuticle nippers and pushers; scissors and shears; reusable nail forms; manicure and pedicure bowls; foot files; glass; metal and fiberglass files; metal drill bits; tweezers; comedone extractors; brushes; nylon-bristle dusting brushes that can be disinfected; combs; clips; reusable pencil sharpeners; reusable gloves; and other metal or non-porous tools.

Single-use items cannot be disinfected and should not be reused. They are discarded after one use. The rule lists buffers; emery boards; nail files; sleeves and sanders for electric files; orange-wood or birch-wood sticks; wooden applicator sticks or spatulas; porous foot files; pedicure slippers and toe separators; disposable gloves; paraffin liners; cotton balls, strips, or swabs; neck strips and muslin strips; or any item that cannot be disinfected.

Esthetic stations still live this split: a metal comedone extractor is multi-use. A wooden wax stick is single-use. A muslin strip is single-use. A nylon makeup brush that can be disinfected is multi-use; a disposable mascara wand is not.

The 10-minute EPA immersion, daily solution, no soaking storage

Rule 240-4-.04 states the sequence after each client for multi-use tools:

  1. Clean with warm soapy water to remove all hair and debris.
  2. Disinfect by complete immersion. Disinfectant containers must be large enough for complete immersion (enough solution to cover all surfaces of the item) for 10 minutes in a disinfectant that indicates on its label that it has been registered with the EPA as a disinfectant. The disinfecting solution must be free of debris and contaminants at all times and must be changed daily.
  3. After immersion, rinse thoroughly with clean warm water and blot dry.
  4. Store following Rule 240-4-.05. Tools cannot be stored in disinfectant.

Memorize the order: wash, then immerse 10 minutes, then rinse and dry, then store dry in a labeled clean container. Leaving tweezers overnight in the jar is a storage violation even if the liquid is EPA-registered. A cloudy, hair-flecked jar fails "free of debris" even if the clock said 10 minutes at 8 a.m.

An autoclave that meets or exceeds ANSI/AAMI ST55:2010 or ISO 17665 standards for producing high-temperature steam (121°C to 134°C) is an acceptable method of sanitizing implements (240-4-.04(c)). That is the steam exception with a standard number on it.

Ultraviolet lights, dishwashers, and steam devices are not acceptable methods of cleansing and disinfection (240-4-.04(d)). A UV cabinet may be used later for storage only, and only after proper cleansing and disinfection; the cabinet itself must still be labeled "Clean Implements", and the rule repeats that ultraviolet lights are not acceptable methods of cleansing and disinfection (240-4-.05(1)(b)).

MethodGeorgia Rule 240-4 status
Warm soapy water cleaning, then 10-minute complete immersion in EPA-registered disinfectant, solution changed dailyRequired sequence for multi-use tools
Storing tools in the disinfectant jar between clientsProhibited
Autoclave meeting ANSI/AAMI ST55:2010 or ISO 17665 (121°C to 134°C)Acceptable sanitizing method
UV light as the disinfection stepNot acceptable
Dishwasher as the disinfection stepNot acceptable
Steam device (other than a qualifying autoclave) as the disinfection stepNot acceptable

Towels and linens contaminated with blood or bodily fluids must be washed separately with an EPA-registered disinfectant along with laundry detergent per the manufacturer's instructions (240-4-.04(5)). Sinks and shampoo bowls must be thoroughly cleansed and disinfected with an EPA-registered disinfectant and kept free of hair or other debris (240-4-.04(6)).

Pedicure-equipment procedures in 240-4-.04(4) (chelating soap, 15-minute disinfection after each client, end-of-day circulation, weekly bleach procedure, and a log book in the pedicure area) apply to any facility that holds that equipment. An esthetician in a mixed shop still needs to know the rule exists even when the personal service is a facial.

Storage labels and the no-double-dip rule

Rule 240-4-.05 is the label chapter inspectors can read from the doorway.

  • All disinfected tools and implements shall be stored in a sanitary manner in a clean, closed, airtight drawer or container labeled "Clean Implements".
  • Soiled and dirty tools and implements must be stored in a separate, closed and covered container labeled "Dirty Implements." They shall not be used again until properly cleansed and disinfected and shall not be stored together with clean implements.
  • A hazardous-waste container for blood-contaminated disposable implements is labeled with a biohazard label and/or the words "Contaminated Implements".

Clean towels stay in a closed cabinet, container, or closet (restroom hand towels are the stated exception). Used towels go in a closed container until laundered. Bloody towels go in a separate closed container labeled biohazard and/or "contaminated linen."

Wax, creams, lotions, and other products must be kept in sanitary, closed containers, uncontaminated and free of debris. Products removed from containers must be removed with a stick, spatula, or spoon that is disposable or that can be disinfectedno double dipping, and no product use or storage that contaminates products through contact with the client (240-4-.05(3)).

Inspector-style sanitation walk

Walk a Decatur treatment room the way a Board inspector walks it. Hands were washed in the client's sight before gloves. The extractor sat in a covered Clean Implements box, not in the disinfectant. After the extraction, the extractor went into Dirty Implements, not back into the clean box. Wax was lifted with a fresh wooden stick; the stick never returned to the pot after touching skin. Muslin was discarded. No credo blade was in the drawer. The EPA disinfectant label faced out, the jar was full enough to cover every surface of a tweezer, and the shop could say when today's solution was mixed. A UV box on the counter, if present, was labeled for storage of already-disinfected tools, not advertised as the sterilizer. That walk is Rule 240-4 in one room.

Test Your Knowledge

After a Georgia esthetician washes a metal comedone extractor with warm soapy water, what disinfection step does Rule 240-4-.04 require before the tool may be stored as clean?

A
B
C
D
Test Your Knowledge

Which statement matches Rule 240-4-.04 on ultraviolet lights, dishwashers, steam devices, and autoclaves?

A
B
C
D
Test Your Knowledge

During a facial add-on, a Georgia licensee reaches for a credo blade to reduce a client's heel callus. What does Rule 240-4-.03 say?

A
B
C
D
Test Your Knowledge

How must clean and dirty multi-use implements be stored under Rule 240-4-.05, and how must wax be removed from a pot?

A
B
C
D