16.3 Artifacts, Markers & Annotation

Key Takeaways

  • RTR.6.5 requires accurate laterality markers and annotations; physical radiopaque markers remain best practice, and digital labels can be wrong, flipped, or copied forward.
  • Never release an image when side identification is uncertain—wrong-side documentation is a critical safety event.
  • RTR.6.6 artifacts include motion, grid cutoff, clothing/foreign objects, CR/DR processing defects, double exposure, scatter fog, moiré, and detector hardware faults.
  • Correct the root cause (geometry, collimation, gowning, detector hygiene, exposure time) rather than only windowing over the problem.
  • If an artifact obscures the region of interest or laterality is compromised, reject/correct under ALARA and escalate equipment failures through QC/service pathways.
Last updated: July 2026

16.3 Artifacts, Markers & Annotation

Quick Answer: Correct laterality markers and accurate annotation are patient-safety requirements (RTR.6.5). Artifacts (RTR.6.6)—motion, grid cutoff, foreign objects/clothing, CR/DR processing problems, double exposure, scatter fog—must be recognized, explained, and corrected. Digital tools make annotation errors easier, not rarer. If an artifact obscures the clinical region or laterality is uncertain, do not release until fixed or properly escalated.

Image analysis is incomplete if you only judge mAs and rotation. Legal identification and artifactual densities decide whether an image is usable and whether the wrong side of a patient could be treated.

Markers & Annotation Accuracy (RTR.6.5)

Physical lead markers remain the gold standard habit

Best practice in Canadian departments still emphasizes radiopaque side markers placed in the primary beam (outside essential anatomy when possible) at exposure:

RuleWhy it matters
Correct side (R / L)Wrong-side surgery and treatment risk
Correct patient contact sideMarker on the side being imaged, not the opposite cassette edge by habit alone
Not over critical anatomyMarker over fracture line or lung apex can force a repeat
Visible after collimationTight collimation that clips the marker creates incomplete legal image
Supplemental markers as protocolErect/supine, weight-bearing, inspiration/expiration, minutes post-contrast, mobile

Digital annotation risks

Electronic annotation is useful for add-on labels (portable, upright, post-void) but creates unique failure modes:

  1. Post-exposure side labels applied to the wrong side after the fact
  2. Copy-forward annotations from a prior study on the wrong patient
  3. Image flip/rotate after annotation so the “R” no longer matches anatomy
  4. Missing annotation on stitched or reformatted secondaries
  5. Over-reliance on default laterality from the worklist without visual confirmation

Exam principle: A digital “R” is not automatically trustworthy if the physical marker is absent and anatomy is ambiguous (e.g., AP supine abdomen, unilateral extremity without landmarks). When laterality is uncertain, investigate before release.

Annotation content that supports care

  • Projection (PA/AP/lateral/oblique) when not obvious
  • Patient position (erect, semi-erect, supine, decubitus with side down)
  • Behavioral/technique notes required by protocol (inspiration/expiration)
  • Time-sensitive series labels (pre/post void, post-reduction)

Avoid cluttering the image with redundant text that covers pathology.

Artifacts (RTR.6.6): Recognition Table & Corrective Response

ArtifactAppearance / cuesCommon causesCorrective response
MotionBlur, ghosting of edges, repeated contoursLong exposure, patient pain/tremor, kids, respiration, equipment vibrationShorten exposure time (↑ mA), immobilize, coach breath-hold, supports; pediatric distraction; check tube/stand locks
Grid cutoff (lateral decentering)Overall or unilateral density loss; underexposed look across fieldCR angled across lead strips; grid upside-down; off-level; wrong focal range SIDRe-center, match SID to grid focal distance, verify grid orientation, use appropriate ratio
Grid cutoff (upside-down focused grid)Severe density loss laterally both sides or characteristic patternFocused grid invertedFlip grid to correct orientation; repeat
Grid linesFine parallel linesStationary grid visible; wrong grid use; moving grid failureUse moving Bucky when indicated; check mechanism; accept only if lines do not hide ROI and protocol allows
Clothing / buttons / zippersOpaque shapes, fabrics, sparklesIncomplete gowningRemove radiopaque objects from FOV; re-gown; explain to patient
Jewelry / hair pins / ECG leadsHigh-density discrete objectsNot removed/relocatedRemove or reposition leads outside ROI when safe
Foreign objects on IR/detectorDust, cracks, scratches, glueDirty plate, damaged detectorClean per manufacturer; take QC flat field; remove from service if hardware defect
Double exposureSuperimposed anatomies, dark/overexposed combined imageCassette reused (CR); wrong detector assignment; procedural errorReject; QC process; re-image once with clear detector workflow
Scatter fogReduced contrast, gray veilPoor collimation, no grid when needed, high kVp large field, open back of cassette (film era concepts still tested), patient scatterCollimate, grid when indicated, optimize kVp, control field size
CR plate reader artifactsLines, skip, plate erasure incompletenessReader rollers, incomplete erase, light leakReprocess/erase plate; service if persistent; do not keep sending bad plates
DR processing / flat-field issuesCoring, dropout, structured noise, dead pixels clustersCalibration overdue, dropped detector, software LUT errorsRecalibrate; service; alternate detector; do not “fix” hardware failure with edge enhance
Moire / aliasingWavy interference patternGrid frequency vs sampling; electronic interferenceAdjust grid/processing; change orientation if protocol allows; service evaluation
Saturation / clippingPure white/black loss of data in regionsExtreme overexposure or processingCannot recover lost data; repeat with corrected technique
Quantum mottleGrainy noiseUnderexposureRepeat with increased exposure as needed—not just windowing
Pajama folds / skin foldsLinear densitiesSoft tissueReposition, smooth skin, note portable limitations
Cast / splint / backboardUniform or structured opacityTrauma immobilizationIncrease exposure appropriately; annotate; do not remove unstable immobilization without clinical order

Motion Artifact — Special Emphasis

Motion is still one of the most common avoidable repeats.

Types:

  • Voluntary — breathing, movement (coaching, supports, clear instructions)
  • Involuntary — tremor, spasm, cardiac (minimize time, stabilize)
  • Equipment — unlocked tube arm, moving table, grid mechanism

Pediatric tip: immobilization devices and caregiver assistance per policy beat multiple failed “please hold still” exposures. ALARA prefers one good immobilized image over three blurred ones.

Grid Cutoff Patterns Worth Memorizing

ErrorTypical density pattern
Off-level gridOverall decrease in exposure across image
Lateral decenteringMore density loss toward one side
Upside-down focused gridSevere bilateral lateral underexposure
Wrong SID outside focal rangeDensity loss toward periphery

Do not compensate solely by raising mAs without fixing geometry—you increase dose and may still fail.

Foreign Objects & Clothing: Workflow Discipline

Build a pre-exposure checklist into muscle memory:

  1. Gown / remove artifacts from region
  2. Empty pockets
  3. Move oxygen tubing, NG slack, monitor cables when safe
  4. Explain why metal must move (“it can hide a fracture”)
  5. Document non-removable objects (jewelry that cannot come off, permanent medical devices)

Piercings and religious items: use culturally safe communication; if item cannot be removed, annotate and collimate/angle when possible without losing required anatomy.

Double Exposure & Workflow Mix-ups

Modern DR reduces classic double exposure but it still appears in stems and in CR-heavy sites:

  • Two body parts on one image
  • “Ghost” of previous anatomy
  • Unexpected darkness

Response: reject image, secure correct detector/plate, re-identify patient, single controlled re-exposure, report system issues if persistent.

Scatter Fog vs Pathology Veil

A gray, low-contrast image may be:

  • Scatter fog (collimation/grid problem)
  • Correct technique on a large soft-tissue patient
  • True pathologic veil (edema, effusion layering)

Fix what you control first (collimation, grid, kVp, field size), reassess, then consider pathology recognition (16.2).

Corrective Response Hierarchy

  1. Identify artifact type (do not guess randomly)
  2. Decide impact: Does it obscure ROI or create wrong-side risk?
  3. Correct cause (not only symptoms)
  4. Repeat or add image under ALARA if needed (16.4)
  5. Escalate equipment problems to service/QC
  6. Document when non-removable artifacts remain

CAMRT-Style Stems

Expect items such as:

  • “Density loss on both lateral margins after using a focused grid—most likely?” → upside-down grid
  • “Electronic L marker placed after exposure on a hand without physical marker—risk?” → wrong laterality annotation
  • “Wavy pattern over lumbar spine with stationary grid—what is it?” → moiré/grid-line interference
  • “Blurred trabeculae with correct EI—cause?” → motion, not mAs

Master tables above until recognition is automatic under exam timing.

Test Your Knowledge

A focused grid produces an image with severe loss of density on both lateral sides of the field. What is the most likely cause?

A
B
C
D
Test Your Knowledge

Which practice best reduces wrong-side errors in digital radiography?

A
B
C
D
Test Your Knowledge

Trabecular detail is blurred on a wrist image, yet the exposure indicator is within acceptable range and there is no grid cutoff. What is the most likely problem?

A
B
C
D
Test Your Knowledge

A CR cassette is accidentally exposed twice with different projections. What is the appropriate technologist response?

A
B
C
D