8.1 Metrology Foundations & Environmental Controls
Key Takeaways
- Metrology is the science of measurement and its application, providing the quantitative foundation for quality assurance, dimensional inspection, and process control.
- ISO 1 establishes the international standard reference temperature for geometrical product specification and dimensional inspection at exactly 20°C (68°F).
- Linear thermal expansion is governed by ΔL = L · α · ΔT; differential thermal expansion occurs when measuring instruments and workpieces possess unequal coefficients of thermal expansion (CTE).
- Abbe's Principle mandates that the line of measurement must be collinear with the line of dimension to avoid first-order angular tilt errors, explaining why micrometers are Abbe-compliant while calipers suffer Abbe offset errors.
- Cosine error occurs when an indicator stylus or measuring probe is inclined at an angle θ to the dimension vector, causing the instrument to misreport displacement according to ΔL = L · (1 - cos θ).
8.1 Metrology Foundations & Environmental Controls
The Science of Metrology: Definitions and Core Purpose
In quality assurance and manufacturing excellence, metrology is formally defined as the science of measurement and its application. Metrology encompasses all theoretical and practical aspects of measurement, regardless of the measurement uncertainty or field of application. For an ASQ Certified Quality Technician (CQT), metrology provides the empirical bedrock upon which all product verification, process capability evaluation, and statistical process control decisions rest. Without rigorous metrological controls, inspection data is meaningless, tolerance verification is untrustworthy, and product compliance cannot be demonstrated.
Metrology is traditionally divided into three distinct branches:
- Scientific (Fundamental) Metrology: The establishment, development, and maintenance of fundamental measurement units, realization of SI base standards, and research into novel measurement methods (conducted by national laboratories such as NIST).
- Industrial (Applied) Metrology: The application of measurement science to manufacturing processes, quality testing, dimensional verification, and calibration of production measuring tools on the shop floor.
- Legal Metrology: Regulatory requirements and statutory enforcement governing measuring instruments and measurements that impact commercial transactions, trade weights, environmental safety, and public health.
True Value, Error, and the Metrology Dilemma
A foundational tenet of metrology is that the true value of a physical measurand is an idealized concept that can never be known with infinite precision. Every physical measurement yields a measured value that contains an inherent measurement error (the difference between measured value and reference value). Measurement errors consist of:
- Systematic Errors (Bias): Predictable, repeatable deviations in one direction caused by identifiable factors such as improper zeroing, mechanical scale bias, thermal expansion, or kinematic Abbe offsets. Systematic errors can be calculated, compensated, or calibrated out.
- Random Errors (Precision / Noise): Unpredictable fluctuations caused by microscopic environmental drafts, mechanical friction, visual parallax, or electronic thermal noise. Random errors cannot be eliminated individually, but their dispersion can be quantified statistically using the standard deviation of repeated readings.
The International Standard Reference Temperature: ISO 1
Dimensional measurements depend directly on the physical temperature of both the measuring instrument and the component being evaluated. Because virtually all engineering materials expand when heated and contract when cooled, the length of an object is not a fixed static number; it is a function of its thermal state.
To establish global uniformity in manufacturing and trade, the International Organization for Standardization issued ISO 1 (Geometrical product specifications (GPS) — Standard reference temperature for geometrical product specification and verification). ISO 1 defines the international standard reference temperature as:
Practical Significance on Engineering Drawings
Unless an engineering drawing or procurement specification explicitly states an alternative reference temperature (which is extraordinarily rare), every dimensional tolerance, nominal dimension, and geometric dimensioning and tolerancing (GD&T) feature control frame applies strictly at 20°C (68°F). If a shaft is specified as $50.000 \pm 0.005\text{ mm}$, that tolerance band defines the part's physical boundary only when the shaft is thermally stabilized at 20°C.
[!IMPORTANT] The ISO 1 Mandate: In metrology laboratories and receiving inspection areas, all critical measurements must either be performed at $20^\circ\text{C}$ after adequate thermal soaking, or mathematically compensated back to $20^\circ\text{C}$ using thermal expansion equations. Performing precision inspections at uncontrolled room temperatures (such as $28^\circ\text{C} / 82.4^\circ\text{F}$) without thermal compensation will lead to false rejections of conforming parts or false acceptances of defective product.
Thermal Expansion Mechanics and Differential Errors
Thermal expansion is the physical tendency of matter to change its shape, area, and volume in response to a change in temperature. In dimensional metrology, quality technicians focus predominantly on linear thermal expansion.
The Linear Thermal Expansion Equation
For small temperature deviations around room temperature, the change in length $\Delta L$ of a physical body is linearly proportional to its nominal length $L$, its material-specific Coefficient of Thermal Expansion (CTE) denoted by $\alpha$, and the temperature deviation $\Delta T$ from the standard $20^\circ\text{C}$ reference: Where:
- $\Delta L$ = Change in length (in millimeters or inches)
- $L$ = Nominal length of the dimension at $20^\circ\text{C}$ (in millimeters or inches)
- $\alpha$ = Linear Coefficient of Thermal Expansion (expressed in units of $10^{-6}/^\circ\text{C}$ or $10^{-6}/^\circ\text{F}$, commonly called parts per million per degree or $\mu\text{m}/(\text{m}\cdot^\circ\text{C})$)
- $\Delta T = (T - 20^\circ\text{C})$ = Temperature deviation from the standard $20^\circ\text{C}$ baseline
Coefficients of Thermal Expansion for Common Engineering Materials
| Material | CTE (Metric: $\alpha$ in $10^{-6}/^\circ\text{C}$) | CTE (Imperial: $\alpha$ in $10^{-6}/^\circ\text{F}$) | Metrological Significance |
|---|---|---|---|
| Structural / Tool Steel | $\approx 11.5$ | $\approx 6.4$ | Baseline for gage blocks, micrometer frames, machine tool ways |
| Aluminum Alloys (6061 / 7075) | $\approx 23.0$ | $\approx 12.8$ | Expands twice as fast as steel; high differential error risk |
| Yellow Brass / Bronze | $\approx 19.0$ | $\approx 10.5$ | Moderate-high expansion; commonly used for bushings and fittings |
| Cast Iron (Gray / Ductile) | $\approx 10.5$ | $\approx 5.8$ | Very close to tool steel; stable machine bases and surface plates |
| Granite (Black Jinan / Charcoal) | $\approx 6.0$ | $\approx 3.3$ | Low thermal expansion; superior stability for precision surface plates |
| Tungsten Carbide (WC) | $\approx 4.5$ | $\approx 2.5$ | Extremely hard; very low expansion; used for premium wear gage blocks |
| Ceramic Zirconia ($\text{ZrO}_2$) | $\approx 9.2$ | $\approx 5.1$ | Close to steel CTE; immune to corrosion; premium gage blocks |
| Invar (Nickel-Iron Alloy) | $\approx 1.2$ | $\approx 0.67$ | Near-zero thermal expansion; used for master reference scale standards |
Differential Thermal Expansion Errors
In practical shop inspection, a frequent misconception is: "If my steel micrometer and the part are both sitting on the same workbench at 26°C, they are at the same temperature, so thermal error cancels out!"
This statement is only true if the gage and the workpiece are made of the exact same material with identical CTEs (for example, measuring a tool steel gage block with a tool steel micrometer). If the materials differ, a severe differential thermal expansion error occurs:
Consider measuring an aluminum aerospace component with a steel micrometer at $28^\circ\text{C}$ ($82.4^\circ\text{F}$):
- The aluminum expands at $\alpha_{\text{Al}} \approx 23.0 \times 10^{-6}/^\circ\text{C}$.
- The steel micrometer frame and spindle expand at $\alpha_{\text{Steel}} \approx 11.5 \times 10^{-6}/^\circ\text{C}$.
- The aluminum expands at double the rate of the steel instrument! As a result, the steel micrometer severely under-measures the true thermal expansion of the aluminum, or conversely, if the part is machined to nominal at $28^\circ\text{C}$, it will shrink significantly below the lower tolerance limit once it cools to the standard $20^\circ\text{C}$ operating temperature.
Thermal Soaking (Soak Time) and Temperature Equalization
When components or gages are brought into an environmentally controlled metrology lab from a hot manufacturing plant or cold shipping dock, they must undergo thermal soaking before inspection. Thermal soaking is the process of allowing an object to reach complete thermal equilibrium with the laboratory environment ($20^\circ\text{C}$) throughout its entire cross-section.
- Thermal mass and conductivity: Heavy steel castings or dense granite blocks take substantially longer to equalize than thin sheet metal stampings. As a rule of thumb, precision inspection standards require a minimum of 1 to 2 hours of soak time per 25 mm (1 inch) of material thickness.
- Operator body heat transfer: A technician's hand radiates heat at approximately $37^\circ\text{C}$ ($98.6^\circ\text{F}$). Bare-hand contact with a 100 mm steel gage block for just 30 seconds can transfer sufficient thermal energy to expand the block by $1$ to $2\text{ }\mu\text{m}$ ($40$ to $80\text{ }\mu\text{in}$), completely consuming its calibration tolerance. Technicians must handle precision standards using insulated thermal pads, wooden tongs, or lint-free thermal inspection gloves.
Metrology Laboratory Environmental Controls
High-precision dimensional measurements cannot be sustained on an open shop floor subject to ambient drafts, direct sunlight, forklift vibrations, and seasonal humidity fluctuations. Precision measurement requires a controlled metrology laboratory environment.
+-----------------------------------------------------------------------------------+
| PRECISION METROLOGY LABORATORY ENVELOPE |
| |
| [HVAC System: 20°C ± 0.5°C] [Humidity Control: 40% - 50% RH] |
| Continuous laminar air diffusion Prevents corrosion (>55%) & ESD (<30%) |
| |
| [Positive Air Pressure (+15 Pa)] [HEPA Air Filtration: ISO Class 7 / 8] |
| Air sweeps outward at doors Traps abrasive airborne particulate swarf |
| |
| [Vibration Isolation Trench] [Pneumatic Air Tables] |
| Decoupled structural slab Damps low-frequency floor vibrations |
+-----------------------------------------------------------------------------------+
1. Temperature Stability Tiers
Metrology laboratories are categorized into precision tiers based on temperature tolerance bands:
- Reference / Primary Calibration Labs (National / Research): $20^\circ\text{C} \pm 0.1^\circ\text{C}$ ($68^\circ\text{F} \pm 0.2^\circ\text{F}$).
- Standard Gage Calibration Labs (Gage Cribs / QA Labs): $20^\circ\text{C} \pm 0.5^\circ\text{C}$ ($68^\circ\text{F} \pm 1.0^\circ\text{F}$).
- Shop-Floor Precision Inspection Enclosures: $20^\circ\text{C} \pm 1.0^\circ\text{C}$ to $\pm 2.0^\circ\text{C}$. Air must be delivered through high-volume, low-velocity porous fabric diffusers to eliminate high-speed thermal drafts across surface plates.
2. Relative Humidity (RH) Control (40% to 50%)
Relative humidity must be strictly controlled within the narrow target band of 40% to 50% RH:
- Consequences of High Humidity ($>50% - 55%\text{ RH}$): Moisture condenses onto polished steel surfaces, causing microscopic galvanic and atmospheric oxidation (rusting and pitting) of expensive gage blocks, master setting rings, and machine ways.
- Consequences of Low Humidity ($<30% - 35%\text{ RH}$): Dry air dramatically accelerates electrostatic discharge (ESD) generation. Static charges attract airborne dust particles onto optical flats and gage blocks, distort electronic indicator signals, and present severe latch-up risks to sensitive digital metrology scales and CMM touch probes.
3. Air Cleanliness and Positive Pressure
Metrology laboratories must maintain particulate control conforming to ISO 14644-1 Class 7 or Class 8 (formerly Federal Standard 209E Class 10,000 or 100,000). A single particle of airborne grinding dust ($10\text{ }\mu\text{m}$ to $25\text{ }\mu\text{m}$ diameter) trapped between a gage block and a surface plate can induce an angular error of several arc-seconds, completely invalidating height gage setups.
- Positive Air Pressure: The laboratory must maintain positive static pressure (typically $+10$ to $+15\text{ Pascals}$) relative to the surrounding plant floor. When doors open, clean air sweeps outward, preventing the ingress of airborne oil mist, grinding swarf, and dust.
4. Vibration Isolation
Mechanical vibrations from nearby stamping presses, CNC milling centers, overhead bridge cranes, and forklift traffic transmit through concrete building slabs, causing indicator needle flutter, laser interferometer beam jitter, and Coordinate Measuring Machine (CMM) probe triggering errors.
- Structural Foundation Decoupling: Precision CMMs and primary surface plates are anchored to dedicated concrete inertia slabs (weighing tens of tons) that are physically decoupled from the plant floor by a perimeter expansion joint filled with vibration-damping mastic or cork.
- Pneumatic Isolation Tables: Optical comparators, roundness testers, and surface roughness profilometers sit on pneumatic air-spring isolation tables that filter out high- and low-frequency floor vibrations.
Abbe's Principle of Measurement
Formulated in 1890 by German physicist and optical engineer Ernst Abbe, Abbe's Principle is the single most important kinematic rule in dimensional instrument design.
[!IMPORTANT] Abbe's Principle: The measuring instrument's reference scale must be collinear with (lie along the same straight line as) the line of dimension / feature being measured. In other words, the measurement axis and the reference scale axis must coincide.
ABBE PRINCIPLE COMPLIANCE VS. VIOLATION
MICROMETER (Abbe Compliant):
[Anvil] ======= [Workpiece] ======= [Spindle] ------ [Internal Scale Screw]
<----------------- Single Collinear Line of Measurement ------------------>
Offset distance h = 0 --> No first-order angular tilt error
CALIPER (Abbe Non-Compliant / Abbe Violation):
[Fixed Jaw] [Moving Jaw]
| | <-- Workpiece clamped out at jaw tips
+--- h (Offset) --+ Offset distance h = 30 to 50 mm!
| |
============ [Main Beam Reference Scale] ============>
Any mechanical play or canting angle θ produces first-order Abbe Error: ε = h * tan(θ)
Kinematics of Abbe Error (Abbe Offset)
When an instrument violates Abbe's principle, the line of measurement is separated from the reference scale by a perpendicular distance known as the Abbe offset ($h$). If the moving measuring carriage undergoes any angular tilt or rotation (canting angle $\theta$, expressed in radians) due to mechanical play, gib looseness, jaw wear, or uneven measuring pressure, a first-order Abbe error ($\epsilon_{\text{Abbe}}$) is generated: Where:
- $\epsilon_{\text{Abbe}}$ = Measurement error
- $h$ = Abbe offset distance (perpendicular distance between reference scale and measuring contact point)
- $\theta$ = Angular tilt or canting of the measuring carriage (in radians)
Caliper Abbe Offset vs. Micrometer Abbe Compliance
This kinematic principle explains why a standard outside micrometer has a fundamental accuracy ten times superior to a vernier, dial, or digital caliper:
- The Caliper Violates Abbe's Principle: On a caliper, the measuring jaws extend perpendicular to the beam by an offset distance $h$ (typically $30\text{ mm}$ to $50\text{ mm}$ for standard jaws, and up to $100\text{ mm}$ or more for long-jaw calipers). The reference scale is etched along the main beam. When a technician clamps a workpiece at the tip of the caliper jaws, any play in the sliding jaw carriage causes the jaw to tip forward or backward by an angle $\theta$. Because $h$ is large, this angular tilt is multiplied directly by $h$, generating substantial first-order measurement errors (often $0.025\text{ mm}$ to $0.050\text{ mm}$ / $0.001\text{ in}$ to $0.002\text{ in}$).
- The Micrometer Complies with Abbe's Principle: On a standard external micrometer, the anvil, the workpiece, the measuring spindle, and the internal calibrated lead screw lie along the exact same straight line ($h = 0$). Because the Abbe offset is zero, any microscopic tilt or deformation of the spindle carriage generates only a second-order cosine error, which is mathematically negligible (on the order of millionths of an inch). Consequently, micrometers provide inherently superior precision and repeatability.
Cosine Error in Indicator Setups
Cosine error is a geometrical measurement error caused by angular misalignment between the operational axis of a measuring probe or indicator stylus and the true vector of dimensional displacement.
Cosine error is most frequently encountered when using dial test indicators (DTIs) (lever-type indicators with swiveling contact styli), electronic height gage probes, and Coordinate Measuring Machine (CMM) touch triggers.
COSINE ERROR IN LEVER-TYPE DIAL TEST INDICATOR (DTI)
[Indicator Dial Body]
|
| (Pivot)
/
/ <-- Stylus contact arm
/ θ (Angle of inclination)
Surface of Part =========*====================
|<-- True vertical motion L_true
Apparent reading on dial: L_reading = L_true / cos(θ)
True displacement: L_true = L_reading * cos(θ)
Cosine Error: ΔL = L_reading * (1 - cos θ)
The Mathematical Formulation of Cosine Error
When a lever-type dial test indicator's contact stylus is set at an inclination angle $\theta$ relative to the surface being measured (instead of lying parallel to the surface, where contact movement is perpendicular to the pivot arm):
- The physical movement of the surface displaces the stylus tip.
- Because the stylus sweeps through an inclined arc, the indicator's internal mechanical gearing registers an apparent displacement ($L_{\text{reading}}$) that is larger than the true vertical displacement ($L_{\text{true}}$): The resulting cosine error magnitude is:
Table of Cosine Correction Factors and Error Magnitudes
| Stylus Inclination Angle ($\theta$) | $\cos\theta$ | Correction Multiplier | Measurement Error Percentage ($1 - \cos\theta$) |
|---|---|---|---|
| $0^\circ$ (Ideal / Parallel) | $1.00000$ | $1.0000$ | $0.00%$ (Zero error) |
| $5^\circ$ | $0.99619$ | $0.9962$ | $0.38%$ (Negligible for general shop work) |
| $10^\circ$ | $0.98481$ | $0.9848$ | $1.52%$ (Practical shop threshold) |
| $15^\circ$ | $0.96593$ | $0.9659$ | $3.41%$ (Must apply correction) |
| $20^\circ$ | $0.93969$ | $0.9397$ | $6.03%$ (Severe error; exceeds typical gage R&R) |
| $30^\circ$ | $0.86603$ | $0.8660$ | $13.40%$ (Gross measurement failure) |
| $45^\circ$ | $0.70711$ | $0.7071$ | $29.29%$ (Unusable setup) |
Practical Rules for Quality Technicians
- Keep the Stylus as Parallel as Possible: When mounting a dial test indicator on a height gage or magnetic stand, align the stylus arm so that the angle $\theta$ between the stylus and the part surface is less than $10^\circ$. Below $10^\circ$, the cosine error is less than $1.5%$, which is often within acceptable gage uncertainty bands.
- Apply Mathematical Correction when Angle is Unavoidable: When measuring deep recesses or undercuts where clearance requires the stylus to be angled steeply (e.g., at $\theta = 30^\circ$), the technician must multiply the dial reading by $\cos 30^\circ = 0.8660$ to determine the true dimensional deviation.
Step-by-Step Worked Numerical Examples
Worked Example 1: Differential Thermal Expansion of an Aluminum Housing
Scenario: A quality technician is inspecting a precision CNC-machined aluminum housing ($6061\text{-T6}$, CTE $\alpha_{\text{Al}} = 23.0 \times 10^{-6}/^\circ\text{C}$) on the plant floor during a hot summer shift where the ambient temperature is $30^\circ\text{C}$ ($86^\circ\text{F}$). The drawing specifies a critical bore center-to-center length of $L = 300.000\text{ mm} \pm 0.025\text{ mm}$ at $20^\circ\text{C}$. The technician measures the feature using a large vernier height gage with a hardened tool steel column (CTE $\alpha_{\text{Steel}} = 11.5 \times 10^{-6}/^\circ\text{C}$). Both the aluminum part and the steel height gage have stabilized at the plant ambient temperature of $30^\circ\text{C}$. The height gage displays a reading of $300.040\text{ mm}$.
Step A: Calculate the Temperature Deviation ($\Delta T$)
Step B: Calculate Individual Thermal Expansions
- Expansion of the Aluminum Housing:
- Expansion of the Steel Height Gage Column:
Step C: Calculate the Differential Thermal Expansion Error
Step D: Calculate the True Length at Standard $20^\circ\text{C}$
Because the aluminum expanded by $0.0690\text{ mm}$ due to heat, its true physical length when cooled to $20^\circ\text{C}$ is: Since the height gage itself expanded by $0.0345\text{ mm}$, the actual physical length of the part at $30^\circ\text{C}$ was: Now, cooling the part to $20^\circ\text{C}$:
Quality Evaluation:
- At $30^\circ\text{C}$, the uncompensated reading was $300.040\text{ mm}$, which exceeded the upper specification limit ($300.025\text{ mm}$) by $0.015\text{ mm}$! An uncalibrated inspector would have falsely scrapped or rejected this part.
- When correctly compensated back to standard $20^\circ\text{C}$, the true length is $300.0055\text{ mm}$, which falls comfortably within the $\pm 0.025\text{ mm}$ tolerance band! This demonstrates how critical thermal metrology is to avoiding costly scrap.
Worked Example 2: Abbe Offset Error on a Long-Jaw Caliper
Scenario: A technician uses a heavy-duty shop caliper with long jaws ($h = 80\text{ mm}$ from beam scale to contact tips) to inspect the outside diameter of a machined shaft. Due to mechanical wear in the sliding jaw gibs and firm thumb clamping pressure, the moving jaw tilts by an angular deflection of $\theta = 0.06^\circ$. What is the resulting Abbe error?
- Convert Angular Deflection to Radians:
- Calculate First-Order Abbe Error:
Evaluation: An Abbe error of $0.084\text{ mm}$ ($3.3\text{ thousandths of an inch}$) completely swamps typical precision tolerances (which are often $\pm 0.013\text{ mm}$ / $\pm 0.0005\text{ in}$). This proves why calipers must never be used for final product buyoff on tight tolerances.
Worked Example 3: Cosine Error on a Dial Test Indicator Setup
Scenario: An inspector sets up a lever-type dial test indicator on a surface plate stand to verify the parallelism of a ground steel plate. Due to clamping interference, the indicator stylus is angled at $\theta = 25^\circ$ relative to the surface plane. During the sweep across the plate, the dial needle indicates a total indicator reading (TIR) of $0.0040\text{ in}$ ($0.1016\text{ mm}$). What is the true height deviation of the plate, and what is the absolute cosine error?
- Calculate True Height Deviation ($L_{\text{true}}$):
- Calculate Absolute Cosine Error:
Evaluation: The indicator exaggerated the surface out-of-parallelism by nearly $10%$. If the parallelism specification was $0.0038\text{ in}$ max, the raw dial reading of $0.0040\text{ in}$ would have falsely caused a conforming plate to be rejected.
Technician Inspection Scenarios & Common Exam Traps
Real-World Shop Scenario: Machining Inconel Turbine Rings
A precision aerospace machining cell turns Inconel 718 rings ($L = 500\text{ mm}$, CTE $\approx 13.0 \times 10^{-6}/^\circ\text{C}$). The cutting fluid heats the parts to $42^\circ\text{C}$ during turning. An operator unloads a freshly turned ring and immediately checks the outside diameter with a shop micrometer that has been sitting at room temperature ($22^\circ\text{C}$). The unsoaked part measures $0.130\text{ mm}$ larger than nominal. An untrained operator would immediately adjust the CNC offset to cut an additional $0.130\text{ mm}$ of stock. Two hours later, after the part cools to $20^\circ\text{C}$, the ring shrinks below the minimum specification limit, resulting in a $$12,000$ scrapped forging. The technician must institute a mandatory thermal soaking protocol before any dimensional buyoff.
Common Exam Traps for CQT Candidates
- Exam Trap 1: Assuming Same Temperature Means Zero Thermal Error: Exam questions often describe a steel gage measuring an aluminum or brass component where both are at $28^\circ\text{C}$. Candidates mistakenly assume zero error because "both parts are at equilibrium." Differential thermal expansion occurs because their CTE values differ!
- Exam Trap 2: Believing Calipers Comply with Abbe's Principle: Questions frequently ask which tool complies with Abbe's principle: caliper, height gage, depth gage, or micrometer. The micrometer is the only common hand tool that complies with Abbe's principle; calipers and height gages violate it.
- Exam Trap 3: Confusing Relative Humidity Limits: Remembering why 40%–50% RH is mandated: if humidity drops below 30%, electrostatic discharge (ESD) and airborne dust increase; if it rises above 55%, steel gages and standards rust and corrode.
- Exam Trap 4: Reversing the Cosine Formula: Dial test indicators over-report displacement when angled ($L_{\text{true}} = L_{\text{reading}} \cdot \cos\theta$). Do not divide the reading by cosine when solving for true part displacement.
According to ISO 1, what is the international standard reference temperature at which all engineering drawing dimensions, geometric tolerances, and gage calibrations are defined?
Why does a standard external micrometer achieve inherently higher dimensional accuracy and lower measurement uncertainty than a standard slide caliper?
A quality technician sets up a lever-type dial test indicator (DTI) on a height gage stand to measure a step height on a precision ground fixture. Because of physical clearance constraints, the indicator stylus contact arm is inclined at an angle of 20° relative to the horizontal surface plane. If the indicator dial registers an indicated displacement of 0.0050 inches, what is the true vertical step height, and what is the percentage measurement error induced by the cosine effect?