4.1 CLIA Regulations, Quality Control & Lab Safety

Key Takeaways

  • The Clinical Laboratory Improvement Amendments of 1988 (CLIA '88) establish federal quality standards for all diagnostic testing performed on human specimens across four regulatory tiers: Waived, Provider-Performed Microscopy Procedures (PPMP), Moderate Complexity, and High Complexity.
  • CLIA-waived tests are simple laboratory examinations with an insignificant risk of an erroneous result; facilities performing them must obtain a Certificate of Waiver (CoW) and follow manufacturer package insert instructions strictly without modification.
  • Quality Control (QC) protocols require verifying internal procedural controls with every test run and performing external liquid controls (positive and negative) with each new lot number, new shipment, storage temperature excursion, or unexpected patient result.
  • Quality Assurance (QA) programs oversee the entire testing workflow—pre-analytical, analytical, and post-analytical phases—and mandate participation in external proficiency testing (PT) programs three times per year for non-waived laboratories.
  • OSHA laboratory standards require discarding blood-saturated soft waste in biohazard bags, replacing rigid sharps containers when 2/3 to 3/4 full, maintaining Safety Data Sheets (SDS) under the Chemical Hygiene Plan, and flushing emergency eyewash stations weekly for 15 minutes of continuous tepid water (60°F to 100°F).
Last updated: August 2026

4.1 CLIA Regulations, Quality Control & Lab Safety

Clinical laboratory testing plays a pivotal role in modern medicine, influencing approximately 70% of all medical diagnoses and therapeutic decisions. In the ambulatory care environment, Certified Medical Assistants (CMAs) routinely collect diagnostic specimens, process samples, operate automated analyzers, perform point-of-care screening, and document laboratory results. To guarantee that every patient result is accurate, reliable, reproducible, and timely, clinical testing is strictly governed by federal regulations, robust quality management systems, and stringent occupational safety standards.


1. The CLIA '88 Regulatory Framework

Congress enacted the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88) in response to nationwide investigations revealing significant diagnostic errors in clinical laboratories—most notably misread Pap smears and erroneous automated blood screenings. CLIA establishes uniform federal quality standards for all laboratory testing performed on specimens derived from the human body for the purpose of health assessment, disease diagnosis, prevention, or treatment.

+-----------------------------------------------------------------------------------+
|                    CLIA '88 FEDERAL TRI-AGENCY OVERSIGHT                          |
+-----------------------------------------------------------------------------------+
|                                                                                   |
|  [Food and Drug Administration]    [Centers for Medicare & Medicaid Services]    |
|             (FDA)                                    (CMS)                        |
|   - Categorizes test complexity             - Issues Laboratory Certificates      |
|   - Evaluates manufacturer 510(k)           - Inspects laboratory facilities      |
|     and CLIA waiver applications            - Collects biennial user fees         |
|   - Clears home & POC test devices          - Enforces regulatory sanctions       |
|                                                                                   |
|                     [Centers for Disease Control & Prevention]                    |
|                                       (CDC)                                       |
|                      - Conducts laboratory research & studies                     |
|                      - Develops technical standards & guidelines                  |
|                      - Evaluates proficiency testing (PT) programs                |
+-----------------------------------------------------------------------------------+

The Tri-Agency Governance Model

CLIA '88 is administered through the coordinated efforts of three federal entities:

  1. Food and Drug Administration (FDA): Categorizes commercial diagnostic test systems into specific complexity tiers based on a 7-point scoring rubric (evaluating operator skill, reagent preparation, operational steps, calibration requirements, troubleshooting, and interpretation difficulty). The FDA reviews manufacturer premarket approval applications and grants CLIA waivers.
  2. Centers for Medicare & Medicaid Services (CMS): Exercises primary operational and financial oversight. CMS issues laboratory certificates, assesses and collects biennial certification fees, conducts unannounced on-site laboratory inspections, monitors compliance, and imposes enforcement sanctions (e.g., fines, suspension of testing authority, revocation of laboratory operating certificates).
  3. Centers for Disease Control and Prevention (CDC): Provides ongoing scientific, epidemiological, and technical consultation. The CDC develops clinical laboratory standards, monitors laboratory medicine practices, and evaluates external proficiency testing programs.

Universal Scope of CLIA

CLIA applies to all facilities that perform testing on human specimens, regardless of setting or test volume. Whether testing occurs in a 1,000-bed tertiary academic medical center, a physician office laboratory (POL), an urgent care center, an ambulatory surgical suite, a public health clinic, or a mobile health van, the facility must hold a valid CLIA certificate appropriate for the highest complexity level of testing performed on-site.


2. The Four CLIA Diagnostic Test Complexity Tiers

CLIA categorizes laboratory testing into four distinct regulatory tiers based on the potential risk of public harm from an erroneous result, the technical knowledge required to perform the assay, and the degree of operator judgment involved in interpreting the data.

+---------------------------------------------------------------------------------------+
|                         CLIA '88 TEST COMPLEXITY HIERARCHY                            |
|                                                                                       |
|   +-------------------------------------------------------------------------------+   |
|   | 1. WAIVED TESTS (Certificate of Waiver - CoW)                                 |   |
|   |    - Simple, lowest risk of error; cleared for home/POC use; CMAs perform.    |   |
|   +-------------------------------------------------------------------------------+   |
|   | 2. PROVIDER-PERFORMED MICROSCOPY PROCEDURES (PPMP)                            |   |
|   |    - Moderate subcategory; labile specimens; MD, DO, PA, NP personally perform.   |   |
|   +-------------------------------------------------------------------------------+   |
|   | 3. MODERATE COMPLEXITY TESTS                                                  |   |
|   |    - Automated analyzers, urine sediment; requires QA, QC, and PT programs.   |   |
|   +-------------------------------------------------------------------------------+   |
|   | 4. HIGH COMPLEXITY TESTS                                                      |   |
|   |    - Molecular PCR, cytology, histopathology, crossmatching; specialized lab. |   |
|   +-------------------------------------------------------------------------------+   |
+---------------------------------------------------------------------------------------+

1. Waived Tests (Certificate of Waiver - CoW)

  • Definition: Simple diagnostic laboratory examinations and procedures that have been cleared by the FDA for home use or that employ methodologies so simple and accurate as to render the likelihood of erroneous results negligible, posing no reasonable harm to the patient if performed incorrectly.
  • Certificate Required: Certificate of Waiver (CoW), renewed biennially through CMS.
  • Personnel Requirements: No specific federal education or credentialing mandates. Certified Medical Assistants, phlebotomists, and other trained clinical staff routinely perform waived assays under the supervision of the laboratory director.
  • Mandatory Compliance Standard: The testing facility MUST strictly follow the manufacturer's package insert instructions without any deviation or modification. If a laboratory alters any parameter (e.g., using a different buffer volume, shortening incubation time, or applying the test to an off-label specimen type), the test immediately loses its waived status and is reclassified as High Complexity, subjecting the facility to severe regulatory violations.
  • Common Waived Assays:
    • Urine pregnancy tests (visual lateral flow hCG)
    • Rapid Group A Streptococcus antigen swabs
    • Rapid Influenza A and B screening
    • Rapid Respiratory Syncytial Virus (RSV) antigen
    • Rapid Infectious Mononucleosis heterophile antibody
    • Capillary blood glucose using automated glucometers
    • CLIA-waived automated Hemoglobin A1c analyzers
    • Chemical urine reagent dipstick testing (visual or automated reader)
    • Fecal occult blood testing (gFOBT and FIT)
    • CLIA-waived prothrombin time (PT/INR) point-of-care coagulation
    • Rapid COVID-19 (SARS-CoV-2) antigen lateral flow assays
    • Urine drug screening cups (visual immunoassay)
    • Microhematocrit centrifugation using capillary tubes

2. Provider-Performed Microscopy Procedures (PPMP)

  • Definition: A specialized subcategory of moderate-complexity testing that meets four specific criteria:
    1. The examination is performed personally by a licensed healthcare provider (physician, physician assistant, nurse practitioner, certified nurse midwife, or dentist).
    2. The procedure is performed during the course of a patient encounter in a clinic or outpatient setting.
    3. The examination utilizes a brightfield or phase-contrast microscope.
    4. The specimen is inherently labile and cannot be stored or transported without rapid cellular degradation.
  • Permitted PPMP Tests:
    • Wet mounts of vaginal, cervical, or skin specimens (evaluating Trichomonas vaginalis, clue cells for bacterial vaginosis, or motile bacteria)
    • Potassium hydroxide (KOH) preparations of skin scrapings, nail clippings, or vaginal discharge (detecting fungal hyphae, pseudohyphae, and budding yeast of dermatophytes or Candida)
    • Pinworm examinations using cellulose adhesive tape applied to the perianal folds (identifying Enterobius vermicularis ova)
    • Microscopic examination of urine sediment (identifying cells, casts, and crystals)
    • Post-coital examinations of cervical mucus (evaluating sperm motility)
    • Nasal smears for granulocytes/eosinophils (allergic rhinitis evaluation)
    • Fecal leukocyte examinations (differentiating inflammatory from non-inflammatory diarrhea)
  • Medical Assistant Role: CMAs assist by preparing the patient, labeling slides, setting up microscope illumination, applying reagents (e.g., 10% KOH solution, normal saline), and sanitizing equipment. In most states, CMAs do not perform diagnostic microscopic interpretation.

3. Moderate Complexity Tests

  • Scope: Represents approximately 75% of all non-waived diagnostic assays. These assays involve automated analytical systems with minimal manual operator steps but require clinical judgment, standard operating procedures, and automated data validation.
  • Examples: Automated complete blood count (CBC) hematology analyzers with automated 5-part differentials, automated multi-channel chemistry panels, automated immunoassays, urine sediment microscopic analysis performed by dedicated lab technologists, and manual Gram stains.
  • Requirements: Facilities must hold a Certificate of Compliance or Certificate of Accreditation, appoint a qualified Laboratory Director and Technical Consultant, employ qualified testing personnel (minimum high school diploma plus documented clinical training or an associate degree in medical laboratory technology), conduct daily two-level quality control, maintain calibration verification, and participate in external proficiency testing.

4. High Complexity Tests

  • Scope: Highly sophisticated, multi-step analytical procedures requiring specialized manual pipetting, complex molecular biochemistry, extensive scientific interpretation, and profound operator judgment.
  • Examples: Clinical cytogenetics (karyotyping), surgical histopathology (frozen section biopsy analysis), diagnostic cytopathology (Pap smear microscopic screening), molecular polymerase chain reaction (PCR) assays for viral loads, flow cytometry, Western blot testing, and manual blood bank compatibility crossmatching.
  • Requirements: Stringent personnel standards (testing personnel must hold at least an associate or bachelor's degree in medical laboratory science; directors must hold an MD/DO or doctoral degree with board certification), extensive method validation, daily multi-level QC, comprehensive QA plans, and bi-annual CMS inspections.

3. Quality Control (QC): Internal vs. External Systems

Quality Control (QC) refers to the operational techniques and activities used to monitor the precision and accuracy of a specific analytical test system, instrument, or reagent lot before reporting patient results.

+-----------------------------------------------------------------------------------+
|                         QUALITY CONTROL (QC) TAXONOMY                             |
+-----------------------------------------------------------------------------------+
|                                                                                   |
|  1. INTERNAL (PROCEDURAL) CONTROLS        2. EXTERNAL (LIQUID) CONTROLS           |
|  - Built into the testing device/strip.   - Manufactured liquid samples of        |
|  - Confirms adequate sample volume,         known target concentrations.          |
|    wicking, and reagent integrity.        - Positive Control: Yields known high   |
|  - Must appear on EVERY patient run.        or reactive positive result.          |
|  - Example: "Control (C)" line on a       - Negative Control: Yields known zero   |
|    rapid strep or COVID-19 strip.           or non-reactive negative result.      |
+-----------------------------------------------------------------------------------+

Internal (Procedural) Controls

Internal controls are built directly into the manufacturing architecture of lateral flow strips, reagent cartridges, and test cassettes.

  • Mechanism: When a liquid specimen flows across the membrane via capillary action, it mobilizes conjugated antibodies. A secondary control line contains anti-species antibodies that capture unbound tracer conjugate, producing a distinct colored Control ("C") band.
  • What Internal Controls Prove:
    1. An adequate volume of liquid specimen was applied to the sample well.
    2. Complete capillary migration and wicking across the nitrocellulose membrane occurred.
    3. The chemical reagents embedded within the test strip are biologically active.
  • Action on Failure: If the Control ("C") line does not appear—even if a prominent Test ("T") line is visible—the test is INVALID. The medical assistant must never interpret or report an invalid result. The test must be repeated immediately using a fresh test device from the kit.

External Quality Controls

External controls are manufactured positive and negative control solutions that mimic human clinical specimens with known analyte concentrations.

  • Positive Control: Formulated with a standardized concentration of the target antigen, analyte, or hormone (e.g., inactivated Group A Streptococcus antigen or purified hCG) to confirm that the assay reliably produces a reactive/positive result.
  • Negative Control: Formulated without the target analyte (or using sterile buffer) to verify that the test does not generate false-positive reactions.

Mandatory Triggers for Running External Controls

Under CLIA and accreditation standards, medical assistants must perform external quality control under the following mandatory clinical circumstances:

  1. Opening a New Test Kit Lot: Whenever a new lot number of test kits, cartridges, or reagent strips is placed into service.
  2. Receiving a New Shipment: Whenever a new shipment of an existing lot arrives (to verify reagents were not exposed to extreme temperatures or physical degradation during transit).
  3. Storage Temperature Excursions: When daily monitoring reveals that room, refrigerator, or incubator storage temperatures deviated outside the manufacturer's specified storage range.
  4. Questionable or Inconsistent Patient Results: When a patient's test result contradicts their clinical presentation or when multiple consecutive patient results yield the identical unexpected outcome.
  5. Periodic Regulatory Schedules: Daily, weekly, or monthly as dictated by the facility's written Standard Operating Procedures (SOPs) or manufacturer guidelines.
  6. Following Instrument Maintenance or Software Updates: After performing preventative maintenance, bulb replacements, or firmware updates on automated benchtop analyzers.

The Golden Rule of Quality Control: If external quality controls fail (i.e., the positive control yields a negative result, or the negative control yields a positive result), NO PATIENT TESTING MAY OCCUR AND NO PATIENT RESULTS MAY BE REPORTED. The medical assistant must quarantine the lot, document the failure in the QC log, investigate potential causes (expired reagents, improper reconstitution, incorrect storage temperature), re-test with new control vials, and notify the laboratory supervisor if the issue persists.

Maintenance Logs & Temperature Monitoring

Every clinical laboratory must maintain comprehensive equipment maintenance logs:

  • Refrigerators (Reagent & Specimen Storage): Must maintain temperatures between 2°C and 8°C (36°F to 46°F). Temperatures must be checked and recorded daily on a physical or digital log sheet.
  • Freezers: Standard laboratory freezers must remain at -20°C (-4°F) or colder; ultra-low freezers must maintain -70°C to -80°C.
  • Incubators & Water Baths: Maintained at physiological body temperature 37°C ± 1°C (98.6°F).
  • Ambient Room Temperature: Monitored daily (20°C to 25°C / 68°F to 77°F) to ensure dry strip reagents and lateral flow kits do not degrade from ambient heat or humidity.

4. Quality Assurance (QA) Programs & Proficiency Testing

While Quality Control focuses on the analytical accuracy of individual tests, Quality Assurance (QA) is an overarching, comprehensive clinical management system designed to monitor, evaluate, and optimize the entire laboratory testing process to ensure continuous high-quality patient care.

+-----------------------------------------------------------------------------------------+
|                        THE THREE PHASES OF QUALITY ASSURANCE                            |
+-----------------------------------------------------------------------------------------+
|                                                                                         |
|  1. PRE-ANALYTICAL PHASE (60–70% of all laboratory errors occur here!)                  |
|     - Patient identification using 2 unique identifiers.                                |
|     - Patient fasting & dietary preparation verification.                               |
|     - Proper specimen collection technique & sterile container selection.               |
|     - Correct phlebotomy order of draw & proper tube inversion.                         |
|     - Immediate, error-free bedside labeling in the presence of the patient.            |
|     - Timely transport, centrifuging, and temperature-controlled storage.               |
|                                                                                         |
|  2. ANALYTICAL PHASE (Testing execution)                                                |
|     - Strict adherence to manufacturer package insert instructions.                     |
|     - Daily equipment calibration, temperature logging, and maintenance.                |
|     - Verification of internal and external QC before patient testing.                  |
|     - Proper reagent storage and expiration date verification.                          |
|                                                                                         |
|  3. POST-ANALYTICAL PHASE (Result reporting and management)                             |
|     - Accurate transcription and EHR entry of quantitative/qualitative data.            |
|     - Immediate verbal notification and read-back of CRITICAL (panic) values to provider.|
|     - Proper documentation of critical call logs (date, time, caller, recipient).       |
|     - Secure, HIPAA-compliant archiving of laboratory requisitions and logs.            |
+-----------------------------------------------------------------------------------------+

Critical Laboratory Values (Panic Values)

A critical value (or panic value) is an analytical test result that falls so far outside the reference range that it indicates an immediate, life-threatening danger to the patient requiring urgent clinical intervention. Examples include a capillary blood glucose $<40\text{ mg/dL}$ or $>500\text{ mg/dL}$, a potassium $<2.5\text{ mEq/L}$ or $>6.0\text{ mEq/L}$, or a critical PT/INR $>5.0$.

  • Protocol for Critical Value Reporting:
    1. Immediately verify the test result (repeat the assay or check for gross specimen hemolysis/clotting).
    2. Verbally notify the ordering licensed provider immediately via direct telephone or in-person communication.
    3. Request the provider to "read back" the patient name, date of birth, analyte, and specific numeric value to ensure zero auditory misinterpretation.
    4. Document the critical communication in the EHR: recording the exact date, timestamp, name of the provider notified, test result reported, and provider read-back confirmation.

External Proficiency Testing (PT)

Proficiency Testing is an external quality assessment mechanism required under CLIA for all moderate- and high-complexity laboratories (and recommended for waived facilities):

  • Mechanism: An approved external accrediting organization (e.g., College of American Pathologists [CAP], American Proficiency Institute [API]) sends "blind" unknown specimens to the clinic three times per year (events), typically containing 5 challenge specimens per analyte.
  • Execution Rule: Staff must test proficiency samples in the exact same manner as routine patient specimens, using the same equipment, reagents, and personnel. Collaborating with other laboratories or sending PT samples out is strictly prohibited and constitutes federal fraud.
  • Grading: Laboratories must maintain a minimum score of 80% per testing event to maintain CLIA certification for that specialty.

5. Laboratory Safety & OSHA Standards

Clinical laboratories contain biological, chemical, physical, and ergonomic hazards. The Occupational Safety and Health Administration (OSHA) enforces strict federal mandates to safeguard healthcare personnel against occupational exposures.

+---------------------------------------------------------------------------------------+
|                                 OSHA HAZARD SIGNAGE                                   |
|                                                                                       |
|          BIOHAZARD WARNING                          NFPA 704 CHEMICAL DIAMOND         |
|               (####)                                           [RED]                  |
|             (########)                                      Flammability              |
|           (############)                                       (0-4)                  |
|         (####)  ||  (####)                         [BLUE]                 [YELLOW]    |
|                 ||                              Health Hazard            Instability  |
|          Fluorescent Orange                         (0-4)                   (0-4)     |
|               or Red                                         [WHITE]                  |
|                                                              Special                  |
|                                                           (W, OX, ACID)               |
+---------------------------------------------------------------------------------------+

Biohazard Waste Management

  • Regulated Medical Waste (Biohazard Bags): Red or fluorescent orange plastic bags bearing the universal biohazard symbol. Reserved strictly for items saturated, dripping, or caked with liquid blood or potentially infectious bodily fluids, bulk blood, microbiologic cultures, and pathological tissues. Non-saturated gauze, lightly stained exam paper, and clean wrappers belong in regular trash to control disposal costs.
  • Biohazard Sharps Containers: Rigid, puncture-resistant, leak-proof polypropylene containers labeled with the biohazard symbol. Used for all used needles, lancets, scalpel blades, glass capillary tubes, broken glassware, and glass slides.
    • Replacement Threshold: Sharps containers must be closed, locked, and replaced when filled to the fill line (2/3 to 3/4 full). Never force, compress, or overfill a sharps container.

OSHA Chemical Hygiene Plan & Safety Data Sheets (SDS)

Under the OSHA Hazard Communication Standard (29 CFR 1910.1200), every facility working with chemicals must maintain a written Chemical Hygiene Plan (CHP) and an accessible repository of Safety Data Sheets (SDS):

  • SDS Organization: SDS documents are standardized into 16 mandatory sections detailing chemical identity, manufacturer contact, hazardous ingredients, health hazards, first-aid measures, fire-fighting procedures, accidental release measures, handling and storage, exposure controls/PPE, physical/chemical properties, stability/reactivity, and toxicological information.
  • Accessibility: SDS files must be immediately accessible to all employees 24/7 in physical binders or unhindered digital databases without requiring passwords or supervisor permission.
  • NFPA 704 Hazard Diamond: Uses a color-coded numerical rating from 0 (minimal hazard) to 4 (severe hazard):
    • Blue (Left): Health Hazard
    • Red (Top): Flammability Hazard
    • Yellow (Right): Instability / Reactivity Hazard
    • White (Bottom): Special Hazards (e.g., W with a slash = reacts violently with water; OX = oxidizer; ACID = strong acid; COR = corrosive).

Emergency Eyewash Station Protocol

OSHA and ANSI standard Z358.1 mandate emergency eyewash stations wherever caustic or corrosive chemicals (e.g., 10% bleach, acid reagents, formalin) are utilized:

  • Location: Must be located within 10 seconds of travel time (approximately 55 feet) from the chemical hazard, along an unobstructed path.
  • Weekly Flushing Requirement: Eyewash stations must be activated and flushed weekly to clear sediment, rust, and stagnant water biofilm from the plumbing lines, verifying operational water flow.
  • Emergency Exposure Protocol: If a chemical splashes into the eyes or face:
    1. Immediately hold eyelids open with fingers and position eyes directly over the eyewash nozzles.
    2. Flush both eyes continuously for a minimum of 15 full minutes.
    3. Water temperature must be tepid (60°F to 100°F / 15.6°C to 37.8°C) to prevent thermal shock and chemical acceleration.
    4. Seek immediate medical evaluation and file an OSHA incident report.

CLIA Test Complexity Tiers & Quality Control Protocols

Complexity TierFederal Certificate RequiredAuthorized Testing PersonnelRepresentative Diagnostic AssaysMandatory Quality Control & Regulatory Oversight
WaivedCertificate of Waiver (CoW)Certified Medical Assistants (CMAs), phlebotomists, trained clinical staff under supervisionRapid Strep A, Rapid Flu A/B, Rapid RSV, Urine hCG, Capillary Glucose, HbA1c POC, Urine 10-dipstick, Guaiac/FIT FOBT, PT/INR POCStrictly follow manufacturer package insert without modification; verify built-in internal controls; perform external positive/negative controls per lot/shipment/excursion.
Provider-Performed Microscopy (PPMP)Certificate of PPMPPhysicians (MD/DO), Physician Assistants (PA), Nurse Practitioners (NP), Midwives, DentistsVaginal wet mounts (Trichomonas/clue cells), KOH prep (fungal hyphae), Pinworm tape prep, Microscopic urine sediment, Fecal leukocytesPerformed personally by provider during patient visit on labile specimens; brightfield/phase microscopy; annual microscope maintenance and calibration verification.
Moderate ComplexityCertificate of Compliance or Certificate of AccreditationMedical Laboratory Technicians (MLT / Associate degree) or higher; specialized clinical trainingAutomated CBC with 5-part differential, automated clinical chemistry panels, automated immunoassays, manual Gram stain, manual urine microscopyLaboratory Director oversight; biennial on-site CMS/accreditation inspection; daily 2-level QC; calibration verification every 6 months; 3x/year external Proficiency Testing.
High ComplexityCertificate of Compliance or Certificate of AccreditationMedical Laboratory Scientists (MLS / Bachelor's degree), Cytotechnologists, Pathologists (MD/DO/PhD)Histopathology biopsy review, Pap smear cytology screening, Molecular viral load PCR, Flow cytometry, Western blot, Manual blood crossmatchingStringent personnel credentials; extensive validation of test accuracy and precision; daily multi-level QC; comprehensive QA program; 3x/year external Proficiency Testing.
Test Your Knowledge

A Certified Medical Assistant opens a newly delivered shipment of rapid Strep A test kits that share the same manufacturer lot number as the kits currently in use. Under CLIA quality control standards, what action must the medical assistant perform before using the new test kits on patients?

A
B
C
D
Test Your Knowledge

According to OSHA and ANSI standards, which operational protocol must be maintained for emergency eyewash stations located in ambulatory clinical laboratories?

A
B
C
D
Test Your Knowledge

A physician office laboratory holds a CLIA Certificate of Waiver (CoW). The clinic provider asks the medical assistant to modify a waived rapid COVID-19 antigen test protocol by using half the recommended extraction buffer to make the remaining buffer last longer. How must the medical assistant respond regarding CLIA regulations?

A
B
C
D