NBRC RPFT Exam Guide 2026: PFT Exam Format, High Cut Score, Domains, and Free Practice
Video
A spirometer is being calibrated using a 3-liter calibration syringe. The spirometer reads 2.88 liters on three consecutive trials. What is the most appropriate action?
Instrumentation And Equipment
Question · Medium
A current 2026 RPFT exam guide for pulmonary function technologists: NBRC PFT exam format, eligibility paths, domain weights, high-yield procedures, quality control traps, and free practice questions.
Blog
What maximum permissible error must a spirometer meet under the 2019 ATS/ERS standard, and what tolerance do you actually apply at the bench?
Instrumentation: Set Up, Maintain, Calibrate
Flashcard
Which of the following is the primary purpose of a body plethysmograph in pulmonary function testing?
Question · Easy
Introduction & Exam Overview
1.1 Current Exam Facts
Study guide
During a nitrogen washout test for FRC measurement, which of the following would most likely cause an erroneously high result?
Nursing & Healthcare Exam Prep Podcast
Listen on Spotify
Podcast
What accuracy must a 3-L calibration syringe hold, and why should you avoid gripping its barrel?
Which gas is used as the tracer in the single-breath diffusion capacity (DLCO) test?
A PFT lab is performing quality control on a blood gas analyzer. The control sample results fall outside the ±2 standard deviation (SD) range. What should the technologist do FIRST?
Why do gas analyzer circuits use absorbers such as Drierite or Nafion tubing?
A helium dilution test for measuring FRC yields a significantly lower value than a body plethysmography test performed on the same patient. Which condition best explains this discrepancy?
Procedures
1.2 Content Outline Domains & Study Plan
When performing spirometry, a patient's FVC maneuver shows a plateau on the volume-time curve but the expiratory time is only 4 seconds. According to ATS/ERS standards, what should the technologist do?
Question · Hard
A patient with severe kyphoscoliosis cannot stand for a stadiometer height. What do you measure instead?
During a cardiopulmonary exercise test (CPET), the patient's ventilatory equivalent for oxygen (VE/VO2) reaches its nadir and then begins to rise while VE/VCO2 continues to decrease. This point represents the:
Which of the following is the correct breath-hold time for a standard single-breath DLCO test according to ATS/ERS guidelines?
When must the spirometry system perform a zero-flow setting?
A patient's arterial blood gas results show: pH 7.28, PaCO2 55 mmHg, HCO3 26 mEq/L, PaO2 58 mmHg. How should these results be interpreted?
Data Management
Spirometry Instrumentation
2.1 Spirometer Types, Setup & Maintenance
A flow-volume loop shows a fixed flattening of both the inspiratory and expiratory limbs. This pattern is most consistent with:
Daily verification passes at low flow but fails at high flow. What does that pattern point to?
Instrumentation: Troubleshoot
What is the normal alveolar-arterial oxygen gradient (A-a gradient) for a healthy young adult breathing room air at sea level?
A pneumotachograph measures airflow by detecting:
During a nitrogen washout the expired N2 concentration suddenly climbs instead of falling. What do you check?
When performing a methacholine challenge test, a positive result is defined as a provocative concentration (or dose) that causes the FEV1 to decrease by at least:
2.2 Spirometer Calibration & Linearity
A patient performs a bronchodilator response test. Pre-bronchodilator FEV1 is 2.50 L, post-bronchodilator FEV1 is 2.85 L, and predicted FEV1 is 3.00 L. According to the ATS/ERS 2021 interpretive strategies, is this a significant bronchodilator response?
In closed-circuit helium dilution, the helium concentration drops abruptly and never plateaus. What has happened, and which way is FRC wrong?
Which of the following conditions would cause an INCREASED diffusion capacity (DLCO)?
During quality control of a body plethysmograph, the technologist should verify the box volume using a:
A blood gas analyzer passes its one-point calibration but fails the two-point check. What is the difference between them?
A patient's spirometry results show: FEV1 65% predicted, FVC 90% predicted, FEV1/FVC ratio 55%. These results are most consistent with:
2.3 Spirometer Troubleshooting
When converting a gas volume from ATPS (ambient temperature and pressure, saturated) to BTPS (body temperature and pressure, saturated), the volume will:
Plethysmograph box pressure drifts steadily during quiet breathing before the shutter closes. What is the usual cause?
A patient scheduled for pulmonary function testing reports using a short-acting bronchodilator (albuterol) 2 hours before the appointment. What should the technologist recommend?
During a maximum voluntary ventilation (MVV) test, a patient achieves a value of 45 L/min. The predicted value is 120 L/min, and the FEV1 is 1.8 L. What is the estimated MVV using the FEV1 × 40 formula, and how does it compare?
A smoke-inhalation patient looks cyanotic but the pulse oximeter reads SpO2 98%. Why, and what measures the truth?
Which of the following electrode systems is used in a Clark electrode to measure PaO2?
Blood Gas & Gas Analyzers
3.1 Blood Gas Analyzer Setup, Maintain, Calibrate
A patient's 6-minute walk test (6MWT) results show a distance of 280 meters with SpO2 dropping from 96% to 84% during the test. The most appropriate interpretation is:
You suspect the CO analyzer in a DLCO system is nonlinear. How do you demonstrate it?
When measuring lung compliance, which of the following describes static compliance?
A patient with suspected occupational asthma is referred for serial peak expiratory flow (PEF) monitoring. The recommended monitoring protocol typically requires PEF measurements at least:
State the full daily spirometer verification procedure required by the 2019 standard.
Instrumentation: Quality Control
The Severinghaus electrode in a blood gas analyzer measures: