13.2 Drinking Water Laboratory Testing

Key Takeaways

  • Electrometric pH testing utilizes a glass measuring electrode and an Ag/AgCl reference junction with Automatic Temperature Compensation (ATC); valid calibrations require a three-point curve (pH 4.0, 7.0, 10.0) with an electrode slope between 95% and 105% (56.2 to 62.1 mV/pH at 25°C).
  • Turbidity compliance under the Surface Water Treatment Rule (SWTR) requires Combined Filter Effluent (CFE) to remain ≤0.3 NTU in at least 95% of monthly measurements and never exceed 1.0 NTU; Individual Filter Effluent (IFE) triggers mandate recording profiles when exceeding 0.5 or 1.0 NTU.
  • DPD colorimetry (515–530 nm) detects free chlorine instantly, whereas total chlorine requires potassium iodide (KI); chlorine levels above 4–5 mg/L cause bleaching of the pink Wurster dye into a colorless imine, resulting in false-low readings that require serial dilution.
  • Alkalinity titration uses 0.02 N H2SO4 to a pH 4.5 endpoint (mg/L as CaCO3), while total hardness titration uses 0.01 M EDTA at pH 10 with Eriochrome Black T indicator, shifting sharply from wine red to pure cobalt blue.
  • Microbiological examination employs Membrane Filtration (m-Endo agar at 35°C producing golden-green metallic sheen colonies), Multiple-Tube Fermentation (LTB presumptive, BGLB confirmed, EC+MUG fecal), or modern enzyme substrates (Colilert: ONPG turning yellow for total coliform, MUG fluorescing blue under 365 nm UV for E. coli).
Last updated: September 2026

13.2 Drinking Water Laboratory Testing

Core Principle: Safe drinking water purveyors in New Jersey operate under stringent statutory mandates enforced by the NJDEP Bureau of Safe Drinking Water. Water operators must master standardized physical, chemical, and bacteriological analytical procedures to verify finished water disinfection, optimize chemical coagulation, control corrosivity, and prevent waterborne pathogen outbreaks.


1. Electrometric pH Testing & Calibration Dynamics

pH represents the negative base-10 logarithm of hydrogen ion activity ($pH = -\log_{10}[H^+]$). In water treatment, pH governs chemical coagulation chemistry, calcium carbonate precipitation, corrosion indices, and hypochlorous acid disinfection kinetics ($HOCl \rightleftharpoons H^+ + OCl^-$).

                       COMBINATION pH ELECTRODE ARCHITECTURE
  ┌────────────────────────────────────────────────────────────────────────┐
  │ Coaxial Shielded Cable ──► High-Impedance Voltmeter (mV)               │
  │                                                                        │
  │ ┌───────────────────────────────┐    ┌───────────────────────────────┐ │
  │ │    Glass Measuring System     │    │   Reference Half-Cell System  │ │
  │ │ - Ag/AgCl Internal Element    │    │ - Ag/AgCl Reference Element   │ │
  │ │ - Buffered Internal Solution  │    │ - Saturated KCl Salt Solution │ │
  │ │   (pH 7.0 Buffer)             │    │ - Porous Ceramic Liquid       │ │
  │ │ - Thin Hydrated Glass Bulbed  │    │   Junction (Wick)             │ │
  │ │   Membrane (~100 nm Gel Layer)│    │                               │ │
  │ └───────────────┬───────────────┘    └───────────────┬───────────────┘ │
  └─────────────────┼────────────────────────────────────┼─────────────────┘
                    ▼                                    ▼
          Phase Potential ($E_g$)              Stable Reference ($E_{ref}$)
                    │                                    │
                    └─────────────────┬──────────────────┘
                                      ▼
                  Net Potential: $E_{cell} = E_g - E_{ref}$

The Nernst Equation & Temperature Compensation

The electrical potential developed across the hydrated glass bulb membrane is governed by the Nernst Equation:

E=E0(2.303RTF)×pHE = E^0 - \left( \frac{2.303 R T}{F} \right) \times pH

  • Where $R$ = universal gas constant ($8.314 \text{ J}/\text{mol}\cdot\text{K}$), $T$ = absolute temperature in Kelvin ($K = ^\circ\text{C} + 273.15$), and $F$ = Faraday constant ($96,485 \text{ C}/\text{mol}$).
  • Theoretical Nernstian Slope at 25°C (298.15 K): Slope=2.303×8.314×298.1596,485=0.05916 V/pH=59.16 mV/pH unit\text{Slope} = \frac{2.303 \times 8.314 \times 298.15}{96,485} = \mathbf{0.05916 \text{ V/pH}} = \mathbf{59.16 \text{ mV/pH unit}}
  • Role of Automatic Temperature Compensation (ATC): The ATC probe measures sample temperature and dynamically corrects the slope factor ($\frac{2.303RT}{F}$) applied to raw millivolt conversion. At 0°C, the Nernst slope drops to 54.20 mV/pH; at 50°C, it rises to 64.12 mV/pH. Note: ATC adjusts the meter's mathematical slope conversion—it does not alter the actual temperature-dependent chemical equilibrium of the sample.

Three-Point Buffer Calibration & Slope Verification

+----------------------+-----------------------------+---------------------------------------------+
| Standard Buffer      | Primary Chemical Matrix     | Analytical Role in Meter Calibration        |
+----------------------+-----------------------------+---------------------------------------------+
| pH 7.00 (Zero Point) | Potassium phosphate /       | Establishes electrical zero (isopotential   |
|                      | Sodium hydroxide            | point, ideally 0.0 mV ± 30 mV at pH 7.00).  |
+----------------------+-----------------------------+---------------------------------------------+
| pH 4.00 (Acid Range) | Potassium hydrogen          | Calibrates acid slope for raw water,        |
|                      | phthalate (KHP)             | coagulant addition, and alum dosing.        |
+----------------------+-----------------------------+---------------------------------------------+
| pH 10.00 (Base Range)| Sodium bicarbonate /        | Calibrates basic slope for lime softening,  |
|                      | Sodium carbonate            | caustic soda addition, and corrosion control|
+----------------------+-----------------------------+---------------------------------------------+
  • Electrode Slope Verification Formula: After calibrating with pH 7.00 buffer and measuring pH 4.00 buffer, the instrument calculates the operational slope percentage: % Slope=ΔmV MeasuredΔmV Theoretical×100=mVpH4mVpH73×59.16×100\% \text{ Slope} = \frac{\Delta \text{mV Measured}}{\Delta \text{mV Theoretical}} \times 100 = \frac{|\text{mV}_{pH4} - \text{mV}_{pH7}|}{3 \times 59.16} \times 100
  • Acceptance Criteria: A clean, compliant electrode must maintain an operational slope between 95% and 105% (corresponding to 56.2 to 62.1 mV per pH unit at 25°C). If the slope drops below 95%, the electrode exhibits sluggish response due to a fouled glass membrane, crystallized KCl salt clogging the ceramic junction, or depleted reference electrolyte.
  • Electrode Maintenance Rules: Clean fouled glass bulbs using 0.1 N hydrochloric acid ($HCl$) for mineral scaling or mild non-ionic detergent for oils. Store electrodes strictly in 3M or 4M KCl storage solution. Never store a pH electrode in deionized (DI) water; osmotic pressure leaches mobile lithium and potassium ions from the hydrated glass gel layer, permanently destroying electrode sensitivity.

2. Turbidity Measurement & Surface Water Treatment Rule (SWTR) Standards

Turbidity is not a direct measurement of suspended mass, but an optical property that quantifies the scattering and absorption of incident light by suspended colloidal particles (clay, silt, organic tannins, algae, and microorganisms).

                      NEPHELOMETRIC TURBIDITY MEASUREMENT (90°)
             Incident Light Beam (Tungsten Lamp, 2200K - 3000K)
       ─────────────────────────────────────────────────────────────► Transmitted Beam
                                   │
                                   │ 90° Scattered Light Path
                                   ▼
                        ┌───────────────────────┐
                        │ Photodiode Detector   │ ──► Generates current proportional
                        │ Positioned at 90°     │     to scattered photons (NTU)
                        └───────────────────────┘

Nephelometric Method (EPA Method 180.1)

  1. 90-Degree Scattered Light Detection: A stabilized tungsten filament lamp passes a light beam through the sample vial. A photoelectric detector is mounted at a strict 90-degree ($\pm 30^\circ$) angle relative to the light path, measuring light scattered by suspended particulates in Nephelometric Turbidity Units (NTU).
  2. Calibration Standards:
    • Primary Calibration Standard: Formazin polymer suspension (synthesized by reacting hydrazine sulfate $[(NH_2)_2 \cdot H_2SO_4]$ with hexamethylenetetramine $[(CH_2)_6N_4]$ to produce a 4,000 NTU stock).
    • Secondary Standards: Sealed sub-micron styrene divinylbenzene copolymer beads (AMCO Clear) or sealed silicone oil suspensions utilized for daily calibration verification.

SWTR & Enhanced SWTR Compliance Thresholds

Under the Safe Drinking Water Act Surface Water Treatment Rule (SWTR) and LT2ESWTR, turbidity monitoring serves as the primary operational surrogate for the physical removal of chlorine-resistant encysted protozoan parasites: Cryptosporidium parvum oocysts (4–6 µm) and Giardia lamblia cysts (8–12 µm).

+-----------------------+-----------------------------------------------------------------------------+
| Regulatory Threshold  | Statutory Requirement & Operational Mandates                                |
+-----------------------+-----------------------------------------------------------------------------+
| Combined Filter       | Must be **≤ 0.3 NTU** in at least **95% of all measurements** recorded      |
| Effluent (CFE) - 95%  | each calendar month (continuous monitoring or grab every 4 hours).          |
+-----------------------+-----------------------------------------------------------------------------+
| CFE Maximum Limit     | Must **NEVER exceed 1.0 NTU** at any instantaneous moment. A single reading |
| (Instantaneous Cap)   | > 1.0 NTU requires immediate operational investigation and state reporting. |
+-----------------------+-----------------------------------------------------------------------------+
| Individual Filter     | Trigger 1: If IFE > 0.5 NTU after 4 hours of continuous post-backwash run,  |
| Effluent (IFE)        | the utility must produce a filter profile or state reason.                  |
| Action Triggers       | Trigger 2: If IFE > 1.0 NTU in two consecutive readings taken 15 minutes   |
|                       | apart, must produce profile and report within 10 days to NJDEP.             |
|                       | Trigger 3: If IFE > 2.0 NTU in two consecutive readings taken 15 minutes   |
|                       | apart in 2 consecutive months, utility must arrange a third-party review.   |
+-----------------------+-----------------------------------------------------------------------------+

3. Chlorine Residual Analysis: DPD Colorimetric Method

Chlorine is the principal chemical disinfectant utilized in New Jersey water distribution networks. To guarantee biological safety, facilities must maintain a measurable disinfectant residual throughout the distribution system (minimum 0.2 mg/L free chlorine under standard rules).

                     CHLORINE FRACTIONATION BREAKDOWN
                         [ Total Chlorine Residual ]
                                      │
             ┌────────────────────────┴────────────────────────┐
             ▼                                                 ▼
   [ Free Available Chlorine ]                       [ Combined Chlorine Residual ]
   - Instantaneous DPD reaction                      - Monochloramine ($NH_2Cl$)
   - Hypochlorous acid ($HOCl$)                      - Dichloramine ($NHCl_2$)
   - Hypochlorite ion ($OCl^-$)                      - Nitrogen trichloride ($NCl_3$)
                                                     - Requires Potassium Iodide ($KI$)

DPD Spectrophotometric Chemistry (Standard Methods 4500-Cl G)

  1. Free Available Chlorine ($HOCl + OCl^-$): In a neutral, phosphate-buffered matrix, free chlorine reacts instantaneously with $N,N\text{-diethyl-}p\text{-phenylenediamine}$ (DPD) indicator. Free chlorine oxidizes the amine group, forming a vivid pink/magenta semiquinoid dye known as Wurster dye. Absorbance is measured photometrically at 515 nm to 530 nm within 1 minute of reagent addition.
  2. Total Chlorine ($Free + Combined$): After reading free chlorine, a catalytic quantity of potassium iodide ($KI$) is introduced. Combined chloramines oxidize iodide ions to elemental iodine ($I_2$), which subsequently oxidizes additional DPD to Wurster dye. Total chlorine is recorded after a 2-minute reaction period.
  3. Combined Chlorine Mathematical Determination: Combined Chlorine Residual (mg/L)=Total ChlorineFree Chlorine\text{Combined Chlorine Residual (mg/L)} = \text{Total Chlorine} - \text{Free Chlorine}

Interferences & The High-Chlorine Bleaching Trap

  • Oxidized Manganese ($Mn^{4+}$ / Permanganate): Directly oxidizes DPD to magenta, generating a severe false-positive free chlorine reading. Operators must run a blank with sodium arsenite ($NaAsO_2$) or thioacetamide to chemically reduce chlorine while leaving manganese intact, subtracting the interference.
  • The Bleaching Phenomenon (Concentrations > 4.0 to 5.0 mg/L):
    • When free chlorine concentrations exceed 4.0 to 5.0 mg/L (common during main disinfection, storage tank shock-treatment, or break repairs), the massive excess of hypochlorite oxidizes the pink Wurster intermediate into a colorless, uncharged imine compound.
    • The sample vial will flash faint pink for a split second and then turn completely crystal clear!
    • Catastrophic Error: Unsuspecting operators place the clear vial into the colorimeter and report "0.00 mg/L residual," mistakenly assuming the chlorinator failed when the water actually contains a dangerous overdose (e.g., 25 mg/L).
    • Corrective Action: When testing disinfected mains, perform a volumetric serial dilution (e.g., 1:10 dilution: 10 mL sample diluted to 100 mL with chlorine demand-free deionized water), re-test with DPD, and multiply the reading by the dilution factor (10×).

4. Alkalinity, Hardness Titrations & Jar Testing

Alkalinity Titration (Standard Methods 2320 B)

Alkalinity measures the acid-neutralizing buffering capacity of water, comprising bicarbonate ($HCO_3^-$), carbonate ($CO_3^{2-}$), and hydroxide ($OH^-$) ions.

                               ALKALINITY TITRATION PH CURVE
    pH
    11 ┌────────────────────────────────────────────────────────┐
    10 │                                                        │
     9 │                                                        │
     8 ├─────────────────* [ Phenolphthalein Endpoint: pH 8.3 ] │ ──► P-Alkalinity
     7 │                  ╲                                     │     (All OH- + 1/2 CO3^2-)
     6 │                   ╲                                    │
     5 │                    ╲                                   │
     4 ├─────────────────────* [ Methyl Orange Endpoint: pH 4.5]│ ──► Total (M) Alkalinity
     3 │                      ╲                                 │     (All Bicarbonate, Carbonate, OH-)
     2 └───────────────────────┴────────────────────────────────┘
       0                        Volume of 0.0200 N H2SO4 Added
  1. Titrant: Standardized 0.0200 N sulfuric acid ($H_2SO_4$).
  2. Phenolphthalein Alkalinity ($P$-Alkalinity): Titrated to pH 8.3 using phenolphthalein indicator (pink to clear), measuring all hydroxide and half of carbonate.
  3. Total Alkalinity ($T$-Alkalinity or $M$-Alkalinity): Titrated to pH 4.5 using bromocresol green-methyl red or methyl orange indicator (blue-green to light pink), measuring all bicarbonate, carbonate, and hydroxide.
  4. Calculation Formula: Alkalinity (mg/L as CaCO3)=A×N×50,000mL of sample\text{Alkalinity (mg/L as } CaCO_3) = \frac{A \times N \times 50,000}{\text{mL of sample}}
    • Where $A = \text{mL of } H_2SO_4$ titrant used, and $N = \text{normality of acid (0.0200 N)}$.
    • When using a standard 100 mL sample aliquot, the formula simplifies to: Alkalinity=A×0.0200×50,000100=A×10\text{Alkalinity} = \frac{A \times 0.0200 \times 50,000}{100} = \mathbf{A \times 10}

Total Hardness Titration (Standard Methods 2340 C)

Hardness is the total concentration of polyvalent metallic cations, predominantly calcium ($Ca^{2+}$) and magnesium ($Mg^{2+}$).

  1. Titration Mechanics: Complexometric titration using 0.0100 M EDTA (ethylenediaminetetraacetic acid disodium salt) buffered to pH 10.0 $\pm$ 0.1 using an ammonium chloride/ammonium hydroxide ($NH_4Cl/NH_4OH$) buffer.
  2. Indicator Dynamics: Eriochrome Black T (EBT) or Calmagite forms a wine-red chelate complex with free magnesium ions. As EDTA is added, it binds calcium first, then extracts magnesium from the indicator. At the exact equivalence point, the free uncomplexed indicator turns a sharp, distinct cobalt blue.
  3. Calculation Formula: Hardness (mg/L as CaCO3)=mL EDTA×B×1,000mL of sample\text{Hardness (mg/L as } CaCO_3) = \frac{\text{mL EDTA} \times B \times 1,000}{\text{mL of sample}}
    • Where $B = \text{mg } CaCO_3 \text{ equivalent to 1.00 mL EDTA titrant (typically 1.00 mg)}$.

Jar Testing Procedure for Coagulation Optimization

Jar testing simulates full-scale conventional water treatment (rapid mix, flocculation, and gravity sedimentation) using a 6-gang variable-speed paddle stirrer with 2-liter square B-Ker (Gator) jars.

+-----------------------+-----------------------------+--------------------+-----------------------------+
| Treatment Phase       | Paddle Stirrer Speed (RPM)  | Duration / Time    | Hydraulic & Chemical Purpose|
+-----------------------+-----------------------------+--------------------+-----------------------------+
| 1. Flash / Rapid Mix  | 100 to 120 RPM              | 1 to 2 Minutes     | Disperses coagulant (alum,  |
|                       | (High shear: G > 300 s⁻¹)   |                    | ferric) to neutralize charge|
+-----------------------+-----------------------------+--------------------+-----------------------------+
| 2. Slow Flocculation  | 20 to 30 RPM                | 15 to 20 Minutes   | Promotes inter-particle     |
|                       | (Tapered: G = 20 - 50 s⁻¹)  |                    | collision without floc shear|
+-----------------------+-----------------------------+--------------------+-----------------------------+
| 3. Quiescent Settling | 0 RPM                       | 15 to 30 Minutes   | Evaluates gravity settling  |
|                       | (Paddles stopped)           |                    | rate and pin-floc carryover |
+-----------------------+-----------------------------+--------------------+-----------------------------+
  • Supernatant Analytical Evaluation: After settling, sample supernatant 2 inches below the liquid surface. Analyze for: (1) settled turbidity, (2) filtered turbidity (through Whatman 40 or 0.45 µm filter paper to simulate dual-media filtration), (3) residual dissolved aluminum or iron, (4) finished pH, and (5) $UV_{254}$ absorbance for organic DBP precursor removal. The jar with the lowest filtered turbidity and acceptable coagulant cost identifies the optimal plant chemical dose.

5. Microbiological Examination of Drinking Water

Drinking water is tested for the Coliform group as an indicator of fecal contamination and treatment barrier failure.

+-----------------------------------------------------------------------------------------------------+
| SCIENTIFIC DEFINITION OF THE COLIFORM BACTERIOLOGICAL GROUP                                        |
| All aerobic and facultatively anaerobic, Gram-negative, non-spore-forming, rod-shaped bacteria     |
| that ferment lactose with gas and acid formation within 48 hours at 35.0°C ± 0.5°C.                |
+-----------------------------------------------------------------------------------------------------+

Analytical Methodologies Comparison

+-----------------------+-----------------------------+--------------------+-----------------------------+
| Analytical Method     | Growth Media & Conditions   | Positive Result    | Technical Advantages & Uses |
+-----------------------+-----------------------------+--------------------+-----------------------------+
| Membrane Filtration   | 100 mL sample filtered on   | Dark red colonies  | Direct colony enumeration   |
| (MF / SM 9222 B)      | 0.45 µm grid filter; m-Endo | with golden-green  | (CFU/100 mL); rapid 24-hr   |
|                       | broth/agar, 35°C, 24±2 hr.  | metallic sheen.    | results; fails on turbid H2O|
+-----------------------+-----------------------------+--------------------+-----------------------------+
| Multiple-Tube         | 3-Phase Fermentation:       | Gas in Durham tube | Statistical Most Probable   |
| Fermentation (MTF /   | 1. LTB (Presumptive 35°C)   | and turbidity in   | Number (MPN); handles high  |
| Most Probable Number) | 2. BGLB (Confirmed 35°C)    | both presumptive & | turbidity and sludges; takes|
|                       | 3. EC+MUG (Fecal/E. coli)   | confirmed phases.  | 48 to 72 hours to complete. |
+-----------------------+-----------------------------+--------------------+-----------------------------+
| Enzyme Substrate /    | Defined Substrate (DST)     | Yellow = Total     | Simultaneous detection of   |
| Presence-Absence      | with ONPG and MUG;          | Coliform;          | Total Coliform and E. coli; |
| (Colilert / SM 9223B) | incubated at 35°C, 24 hr.   | Blue Fluorescence  | zero confirmation tubes;    |
|                       |                             | under 365nm UV =   | standard for SDWA compliance|
|                       |                             | E. coli.           |                             |
+-----------------------+-----------------------------+--------------------+-----------------------------+

The Enzyme Substrate Reaction Mechanism (Colilert DST Chemistry)

Modern compliance testing predominantly utilizes enzyme substrate liquid media (Colilert) due to its specificity and operational simplicity:

  1. Total Coliform Detection ($eta$-D-galactosidase): Total coliform bacteria metabolize the nutrient indicator ONPG (ortho-nitrophenyl-$\beta$-D-galactopyranoside) using the intracellular enzyme $\beta$-D-galactosidase. Hydrolysis cleaves the colorless ONPG molecule, releasing $ortho$-nitrophenol, turning the sample vivid yellow.
  2. E. coli Detection ($\beta$-glucuronidase): Escherichia coli uniquely synthesizes the target enzyme $\beta$-glucuronidase, which hydrolyzes the fluorogenic indicator MUG (4-methylumbelliferyl-$\beta$-D-glucuronide). Hydrolysis releases 4-methylumbelliferone, which fluoresces bright blue under a 365 nm long-wave ultraviolet (UV) lamp.

6. Practical Operational Scenarios & Exam Traps

Practical Operational Scenario

A water distribution technician collects a routine monthly compliance bacteriological sample from an authorized designated tap. After 24 hours of incubation at 35.0°C in a certified incubator, the Colilert vessel is examined: the liquid exhibits a vibrant yellow color and fluoresces bright blue under a 365 nm UV viewer.

  • Diagnostic Investigation:
    1. The yellow color confirms the presence of Total Coliform bacteria (positive $\beta$-galactosidase reaction on ONPG).
    2. The bright blue fluorescence confirms the presence of Escherichia coli (positive $\beta$-glucuronidase reaction on MUG).
    3. Under the Revised Total Coliform Rule (RTCR), an E. coli-positive result represents an acute public health violation.
  • Immediate Remediation Protocol:
    1. The operator must notify the NJDEP Bureau of Safe Drinking Water within 24 hours of receiving the lab result.
    2. The utility must collect repeat compliance samples within 24 hours: at the original positive tap, within 5 service connections upstream, and within 5 service connections downstream.
    3. If repeat samples confirm E. coli, an immediate Boil Water Advisory (BWA) must be issued to consumers in coordination with the NJDEP and local health departments.

Critical Exam Traps

  • Trap 1: Bleaching of DPD. If high chlorine water flashes pink and clears, never report "zero residual." Dilute the sample with deionized demand-free water and re-read.
  • Trap 2: Membrane Filtration Sheen Colony Appearance. Coliforms on m-Endo media appear dark red with a distinctive golden-green metallic sheen. Colonies that are pink, clear, or lack the metallic sheen are non-coliforms.
  • Trap 3: Combined Filter Effluent vs. Individual Filter Effluent. The SWTR CFE limit is $\le 0.3 \text{ NTU}$ in 95% of readings, with a maximum cap of 1.0 NTU. Do not confuse this with IFE action triggers (profiles required if IFE exceeds 0.5 or 1.0 NTU).
  • Trap 4: pH Buffer Sequence. Always calibrate pH meters starting with pH 7.00 buffer first (to establish zero mV baseline), followed by pH 4.00 or pH 10.00 to establish the slope.
Test Your Knowledge

A distribution system operator is testing the free chlorine residual of a newly installed and disinfected water main using the standard DPD colorimetric method. Upon adding the DPD reagent powder pillow to the sample vial, the solution flashes a faint pink for less than a second and then immediately turns completely clear and colorless. The operator places the vial into the spectrophotometer and records a reading of 0.00 mg/L. What physical or chemical phenomenon has occurred, and what corrective action must the operator take?

A
B
C
D
Test Your Knowledge

A laboratory technician is conducting a routine three-point calibration on an electrometric benchtop pH meter using standard buffer solutions of pH 4.00, 7.00, and 10.00 at 25°C. When evaluating the calibration curve, what is the required electrode slope percentage range, and what millivolt (mV) response per pH unit represents acceptable Nernstian performance?

A
B
C
D
Test Your Knowledge

A water treatment facility uses the chromogenic/fluorogenic enzyme substrate method (Colilert) to analyze 100 mL potable distribution compliance samples for coliform bacteria. After 24 hours of incubation at 35°C, a sample vessel turns distinctly yellow, and when placed under a 365 nm long-wave ultraviolet light, it fluoresces bright blue. What biochemical reactions have taken place, and what does this result indicate?

A
B
C
D