7.4 Calibration, Verification, Validation & Maintenance of Equipment

Key Takeaways

  • Calibration adjusts an instrument against a traceable standard; verification checks that it still reads within tolerance; validation proves that a process (sterilization, depyrogenation, cleaning) consistently achieves its result; maintenance keeps equipment within its performance limits.

  • USP <797> requires PECs and classified rooms to be certified initially and at least every 6 months, and a CVE, BSC or CACI used for presterilization steps is also certified at least every 6 months.

  • Temperature, humidity and pressure monitoring devices are calibrated or verified as the manufacturer recommends, or every 12 months if no interval is given.

  • Steam and dry heat sterilization are verified with biological indicators with every run or load, and depyrogenation cycles with endotoxin challenge vials initially and at least annually.

  • When equipment is found out of tolerance, it is taken out of service and repaired or recalibrated, and the CSPs made since the last acceptable check are assessed. All of this is documented, as USP <797> Section 20 requires.

Last updated: September 2026

7.4 Calibration, Verification, Validation & Maintenance of Equipment

Every measurement a compounding pharmacist relies on comes from equipment: particle counts, pressure readings, delivered volumes, weights, incubation temperatures, sterilizer cycles. If that equipment is off, the data are wrong, and a CSP can look acceptable when it is not. The BCSCP outline (task 1A8) expects you to design and supervise the program that keeps every instrument trustworthy.

Four terms that exams mix up

TermMeaningExample
CalibrationComparing an instrument with a traceable standard and adjusting it, documenting the before and after valuesA service technician calibrates a balance with certified weights
VerificationChecking that an instrument or process still performs within tolerance, without adjusting itDaily balance check with a test weight; daily ACD accuracy check
ValidationDocumented evidence that a process consistently produces the intended resultProving an autoclave cycle sterilizes a defined load; proving a depyrogenation cycle gives a 3-log endotoxin reduction
MaintenancePreventive and corrective work that keeps equipment within its limitsReplacing ACD pump tubing on schedule; replacing a HEPA filter

For sterilizers and similar equipment, validation is often organized as installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ). IQ shows the equipment is installed correctly, OQ that it works across its operating ranges, and PQ that it performs the real process on real loads.

Master table of checks and intervals

Equipment or processRequirementSource
PECs and classified roomsCertified initially, then at least every 6 months: airflow, HEPA leak test, total particle count under dynamic conditions, smoke pattern tests. Also after redesign, construction, PEC replacement or relocation, or changes affecting airflow. The designated person reviews every reportUSP <797> 5
CVE, BSC or CACI used for presterilization weighingCertified at least every 6 monthsUSP <797> 4.2.6
Laminar airflow systems and robotic enclosuresDynamic smoke test initially and at least every 6 monthsUSP <797> 4.2.3
Pressure-differential monitorsMonitor continuously; results reviewed and documented daily on compounding days; calibrated or verified per manufacturer or every 12 monthsUSP <797> 4.2.5; FAQ
Temperature and humidity devices (rooms, storage, refrigerators)Calibrated or verified per manufacturer, or every 12 months if unspecifiedUSP <797> 9.3.4, 19.1
Automated compounding devicesAccuracy assessed before first use and each day used; daily record; corrective action if out of specificationUSP <797> 9.1
BalancesChecked per SOP (commonly daily) with test weights; periodic calibration; must meet USP <41> repeatability and accuracyUSP <41>, <1251>
Impaction air samplersServiced and calibrated as the manufacturer recommendsUSP <797> 6.1
IncubatorsTemperature monitored and documented throughout incubation; located outside the sterile compounding areaUSP <797> 6.2.2, 6.3.2
Steam sterilizerBiological indicators (for example Geobacillus stearothermophilus) with each run or load; calibrated recorder for each cycle; steam from water meeting the manufacturer's specification; run and load numbers in the compounding recordUSP <797> 10.3
Dry heat sterilizerBiological indicators (Bacillus atrophaeus) with each run or load; calibrated recorder; even heat distributionUSP <797> 10.4
Depyrogenation ovenEffectiveness shown initially and at least annually with endotoxin challenge vials (at least a 3-log reduction); re-established after changes to load, time or temperatureUSP <797> 10.1
Sterilizing filtersIntegrity test (such as bubble point) after each useUSP <797> 10.2
Growth mediaSupplier COA for commercial media; growth promotion testing of each in-house batchUSP <797> 2.3

Out-of-tolerance findings

When a check fails, for example a balance reading 0.3% high on its daily test weight, the response follows a set pattern:

  1. Take the equipment out of service and label it.
  2. Assess the impact. Identify the CSPs made with it since the last acceptable check and judge whether their quality could have been affected. Quarantine or recall them under the SOPs if necessary.
  3. Repair or recalibrate, then verify before returning the equipment to service.
  4. Document the finding, the impact assessment and the corrective action, and look for trends.

This is why daily verification pays off. It limits how far back you have to look when something goes wrong.

Maintenance and servicing

  • Equipment used in classified areas stays there except for calibration, servicing or cleaning. When it comes back, it is cleaned and wiped with sterile 70% IPA or a suitable disinfectant.
  • Servicing facilities or equipment triggers microbiological air and surface sampling. That includes changing HEPA filters in the ceiling or the PEC; routine prefilter changes usually do not trigger it.
  • Preventive maintenance schedules, such as ACD tubing changes, sterilizer gaskets and HEPA filter replacement plans, come from the manufacturer's instructions and belong in the SOPs.

Records

USP <797> Section 20 requires records of equipment calibration, verification and maintenance, along with certification reports and any corrective actions for failures. Inspectors often look at these first, because gaps show quickly.

Test Your Knowledge

A pharmacy's refrigerator temperature monitor comes with no manufacturer-recommended calibration interval. Under USP <797>, how often must it be calibrated or verified for accuracy?

A

Every 12 months

B

Every 3 months

C

Only when it alarms

D

Every 5 years

Test Your Knowledge

How often does USP <797> require the effectiveness of steam sterilization of CSPs to be verified with biological indicators?

A

Once, when the autoclave is installed

B

Every 6 months, at facility certification

C

With each sterilization run or load

D

Only when a sterility test fails

Test Your Knowledge

At the morning check, a balance used to weigh nonsterile API for Category 2 CSPs reads 0.3% high on its certified test weight, well outside tolerance. It passed the previous morning. What is the best response?

A

Keep using it and correct each weight by 0.3% by hand

B

Take it out of service, assess the CSPs weighed on it since the last acceptable check, have it repaired or recalibrated and verified, and document everything

C

Keep using it, because the error is below 1%

D

Recalibrate it at the next 6-month facility certification

Sections you finish are checked off in the contents.