4.1 ASTM C136 / AASHTO T 27: Apparatus, Specimen Masses, Stack Assembly & Sieving Sufficiency

Key Takeaways

  • ASTM C136 / AASHTO T 27 governs the determination of particle size distribution (gradation) of dry fine and coarse aggregates using nested square-mesh woven-wire cloth sieves.
  • Sieves must be assembled in order of decreasing opening sizes from top to bottom, resting over a tight-fitting catch pan with a secure cover to prevent dust loss.
  • ASTM C136 Section 8.4 defines sieving sufficiency as not more than 1% by mass of the material retained on an individual sieve passing that sieve during 1 minute of continuous hand sieving at about 150 strokes per minute; AASHTO T 27 states the same endpoint as 0.5% by mass of the total sample.
  • Fine aggregate test specimens are 300 g minimum after drying; coarse aggregate specimens run from 1 kg at 9.5 mm to 300 kg at 125 mm, and mixtures use the coarse aggregate values.
  • ASTM C136 Note 2 warns that taking more than approximately 10 minutes to achieve adequate sieving may degrade the sample, so the shaking period is established by trial rather than set by habit.
Last updated: September 2026

In concrete mix proportioning and quality assurance, particle size distribution—commonly termed gradation—is the single most influential physical characteristic of mineral aggregates. The gradation of an aggregate dictates the volumetric packing density of the skeleton, the void content between particles, the surface area requiring cement paste lubrication, the water demand for target slump, and the ultimate compressive strength and durability of hardened concrete.

ASTM C136 / C136M (and its state highway counterpart, AASHTO T 27) defines the standard test method for the sieve analysis of fine and coarse aggregates. While conceptually straightforward—separating particles through woven wire screens of progressively smaller openings—the test demands strict adherence to sample preparation, stack assembly, agitation timing, overloading limits, and mass conservation tolerances. Minor deviations in laboratory technique can produce severe grading errors, leading to improper concrete batch adjustments or wrongful rejection of aggregate shipments.


Scope and Engineering Purpose

ASTM C136 is designed to quantify the proportions of particles of different sizes within an aggregate sample by passing dry material through a vertical series of standard test sieves conforming to ASTM E11 (Standard Specification for Woven Wire Test Sieve Cloth and Test Sieves).

Primary Objectives of Sieve Analysis

  1. Specification Compliance: Verifying that manufactured sand, natural sand, or crushed stone complies with ASTM C33 grading envelopes or state Department of Transportation (DOT) project specifications.
  2. Concrete Mix Proportioning: Providing the empirical grading data necessary to calculate the Fineness Modulus (FM) of sand, determine the coarse aggregate volume fraction per ACI 211.1, and optimize combined aggregate packing curves (e.g., Coarseness Factor Chart, 0.45 Power Curve).
  3. Quality Control Monitoring: Detecting quarry segregation, screening plant wear, crusher degradation, or stockpile contamination before aggregate is charged into concrete batch plant bins.

[!IMPORTANT] Dry Sieving vs. Washing: ASTM C136 applies to dry aggregate. However, if an aggregate contains clinging dust of fracture, clay lumps, or silt that dry shaking cannot dislodge, ASTM C117 must be performed prior to ASTM C136. Dry sieving alone systematically underestimates the minus 75-µm (No. 200) fraction due to electrostatic adhesion and particle surface bonding.


Apparatus and Minimum Sample Mass Requirements

Conducting a valid sieve analysis requires calibrated laboratory equipment meeting stringent dimensional, mechanical, and thermal specifications:

1. Test Sieves and Pan

  • Sieve Construction: Square-mesh woven-wire cloth mounted on rigid cylindrical brass or stainless steel frames conforming to ASTM E11.
  • Frame Diameters: Standard commercial frames have diameters of 8 inches (203 mm) or 12 inches (305 mm). Heavy aggregate shakers may utilize large rectangular trays ($457 \times 660\text{ mm}$ or $18 \times 26\text{ in.}$). The frame height may be full-height or half-height; full-height frames are strongly preferred for coarse fractions to allow free particle bouncing without blinding the mesh.
  • Bottom Pan and Cover: A snug-fitting bottom collection pan to capture all material passing the finest sieve, and a tight-fitting top cover to prevent dust loss during violent mechanical agitation.

2. Balance Sensitivity and Readability

ASTM C136 Section 6.1 sets the balance requirements directly:

  • For Fine Aggregate: readable to 0.1 g and accurate to 0.1 g or 0.1% of the test load, whichever is greater, at any point within the range of use.
  • For Coarse Aggregate, or mixtures of fine and coarse aggregate: readable and accurate to 0.5 g or 0.1% of the test load, whichever is greater, at any point within the range of use.

3. Mechanical Sieve Shaker

A mechanical agitation device that imparts a combination of vertical and lateral motion, or orbital motion combined with a jarring/tapping action, causing particles to bounce, rotate, and present multiple orientations to the sieve openings.

4. Minimum Test Specimen Mass

Field samples obtained under ASTM D75 and reduced under ASTM C702 must satisfy the minimum dry test specimen mass requirements of ASTM C136. Minimum sample mass is strictly controlled by the Nominal Maximum Aggregate Size (NMAS) of the material:

Fine aggregate (ASTM C136 Section 7.3): the size of the test sample, after drying, shall be 300 g minimum. That is the whole rule — there is no NMAS-indexed table for fine aggregate in the current standard, and there is no 100-g row. (Editions before the 1994 revision used a two-row 100 g / 500 g split; the precision tables in Section 11 still carry a note explaining the change. Do not answer from the superseded values.)

Coarse aggregate (ASTM C136 Section 7.4):

Nominal Maximum Size, Square Openings, mm (in.)Minimum Test Sample Size, kg (lb)
9.5 (3/8)1 (2)
12.5 (1/2)2 (4)
19.0 (3/4)5 (11)
25.0 (1)10 (22)
37.5 (1-1/2)15 (33)
50 (2)20 (44)
63 (2-1/2)35 (77)
75 (3)60 (130)
90 (3-1/2)100 (220)
100 (4)150 (330)
125 (5)300 (660)

Mixtures of coarse and fine aggregate (Section 7.5): use the same specimen size as for coarse aggregate in Section 7.4.

Field sample size (Section 7.1): the field sample shall be the quantity shown in Practice D75 or four times the quantity required for the test specimen, whichever is greater.

[!WARNING] Undersized Specimen Invalidation: Testing a sample smaller than the ASTM C136 minimum threshold statistically biases the gradation results, invalidates the test report, and constitutes an immediate failure on the ACI Level 1 written and performance exams.

Sieve Stack Assembly and Mechanical Agitation

Proper assembly of the sieve nest is essential for logical particle separation and equipment protection.

+-------------------------------------------------------------+
|                 ASTM C136 SIEVE NEST ASSEMBLY               |
+-------------------------------------------------------------+
|                                                             |
|        +-----------------------------------+                |
|        |        TIGHT-FITTING COVER        |  <- Prevents   |
|        +-----------------------------------+     Dust Loss  |
|        |  3/8 in. (9.5 mm) Sieve           |  <- Coarsest   |
|        +-----------------------------------+                |
|        |  No. 4 (4.75 mm) Sieve            |                |
|        +-----------------------------------+                |
|        |  No. 8 (2.36 mm) Sieve            |   Decreasing   |
|        +-----------------------------------+    Opening     |
|        |  No. 16 (1.18 mm) Sieve           |     Sizes      |
|        +-----------------------------------+     (Top to    |
|        |  No. 30 (600 µm) Sieve            |     Bottom)    |
|        +-----------------------------------+                |
|        |  No. 50 (300 µm) Sieve            |                |
|        +-----------------------------------+                |
|        |  No. 100 (150 µm) Sieve           |                |
|        +-----------------------------------+                |
|        |  No. 200 (75 µm) Sieve            |  <- Finest     |
|        +-----------------------------------+                |
|        |       TIGHT-FITTING PAN           |  <- Catches    |
|        +-----------------------------------+     All Fines  |
|                                                             |
+-------------------------------------------------------------+

Stacking Sequence

  1. Nest the sieves in order of decreasing opening sizes from top to bottom, with the coarsest opening at the top and the finest opening (e.g., No. 200 / 75-µm) at the bottom.
  2. Place the catch pan directly beneath the bottom sieve.
  3. Pour the oven-dried sample (cooled to room temperature) onto the top sieve.
  4. Secure the top cover over the stack and clamp the nest firmly into the mechanical sieve shaker.

Mechanical Shaking Operation

  • ASTM C136 Section 8.2 requires agitation "for a sufficient period, established by trial or checked by measurement on the actual test sample," to meet the sufficiency criterion of Section 8.4. No fixed shaking time is universally valid for all shakers and materials.
  • ASTM C136 Note 2 warns in the opposite direction from what most candidates assume: excessive time — more than approximately 10 min — to achieve adequate sieving may degrade the sample. A shaker that still cannot satisfy Section 8.4 within about 10 minutes is the wrong shaker for that sample, not a reason to keep shaking.
  • Note 2 also recommends a mechanical shaker when the sample is 20 kg or greater, and permits it for smaller samples including fine aggregate.

Sieving Sufficiency: The Hand-Shaking Endpoint Criterion

How does a technician prove that mechanical shaking has separated every particle that can pass? ASTM C136 mandates a rigorous physical hand-sieving test to establish sieving sufficiency.

Hand-Shaking Protocol (Step-by-Step)

  1. After completing the initial mechanical shaking cycle, remove the sieve stack from the shaker.
  2. Disassemble the stack and test individual sieves one at a time, beginning with the coarsest sieve retaining material.
  3. Fit the individual sieve with a snug-fitting collection pan on the bottom and a cover on top.
  4. Hold the sieve in one hand in a slightly inclined position.
  5. Strike the side of the sieve frame sharply with the heel of the other hand at an approximate rate of 150 times per minute.
  6. Rotate the sieve approximately one-sixth (1/6) of a revolution (about 60 degrees) at intervals of approximately 25 strokes.
  7. Continue this standardized hand sieving for 1 minute.
  8. When checking sufficiency for sizes larger than the 4.75-mm (No. 4) sieve, limit the material on the sieve to a single layer of particles.
  9. If the size of the mounted testing sieves makes this hand motion impractical (large trays, for example), ASTM C136 directs the technician to use 203-mm (8-in.) diameter sieves to verify the sufficiency of sieving.
+-------------------------------------------------------------------------+
|                    SIEVING SUFFICIENCY HAND-SHAKE TEST                  |
+-------------------------------------------------------------------------+
|                                                                         |
|  Motion: Sharp upward strikes against the heel of the other hand,       |
|          sieve held in a slightly inclined position                     |
|  Cadence: ~150 strokes per minute                                       |
|  Rotation: 1/6 turn (~60°) every ~25 strokes                            |
|  Duration: Exactly 1 continuous minute                                  |
|                                                                         |
|  THRESHOLD CRITERION:                                                   |
|  - ASTM C136:  <= 1.0% of the residue retained ON THAT SIEVE passes     |
|  - AASHTO T 27: <= 0.5% of the TOTAL sample mass passes                  |
|  - Sizes > 4.75 mm: limit the sieve to a single layer for this check    |
|                                                                         |
|  RESULT:                                                                |
|  - If passing mass <= threshold -> Sieving is SUFFICIENT.               |
|  - If passing mass > threshold  -> Sieving INSUFFICIENT; re-establish   |
|    the shaking period by trial (Note 2: >~10 min may degrade sample)    |
+-------------------------------------------------------------------------+

Quantitative Endpoint Threshold — and the ASTM / AASHTO Split

Sieving is deemed complete and sufficient when:

  • ASTM C136 Section 8.4: Not more than 1% by mass of the material retained on that individual sieve passes that sieve during the 1 minute of hand sieving. The denominator is the residue sitting on the sieve you are testing, not the whole specimen.
  • AASHTO T 27 Section 8.4: Not more than 0.5% by mass of the total sample passes any sieve during the same 1 minute of hand sieving. Here the denominator is the whole specimen.

Both statements describe the same physical endpoint, but they are arithmetically different tests, and the ACI written exam draws questions from both documents. Read the question stem carefully: if it cites ASTM C136, the answer is 1% of the residue on the sieve; if it cites AASHTO T 27, the answer is 0.5% of the total sample.

[!CAUTION] Worked endpoint check (ASTM C136). A 500.0-g fine aggregate specimen has 112.0 g retained on the 600-µm (No. 30) sieve after mechanical shaking. One minute of hand sieving drops another 1.4 g through. The ASTM allowance is 1% of the residue: $0.01 \times 112.0 = 1.12\text{ g}$. Because $1.4\text{ g} > 1.12\text{ g}$, sieving is not sufficient. Under the AASHTO T 27 rule the allowance would instead be $0.005 \times 500.0 = 2.50\text{ g}$, and the same 1.4 g would pass. Two standards, two answers, same data — this is exactly the trap.

When sieving is insufficient, the material that came through during the hand check is added to the next finer sieve fraction, and the mechanical shaking period for that material must be re-established by trial.

Loading diagram...
ASTM C136 Sieve Analysis and Verification Protocol
Test Your Knowledge

When verifying sieving sufficiency under ASTM C136 by hand shaking an individual sieve, which condition demonstrates that the sieving operation is complete?

A
B
C
D
Test Your Knowledge

A laboratory must run a sieve analysis on a fine aggregate. After drying, how large must the test specimen be under ASTM C136, and how large must the field sample it came from be?

A
B
C
D
Test Your Knowledge

After mechanical shaking, a technician hand-sieves the 2.36-mm (No. 8) sieve of a 500.0 g fine aggregate specimen for 1 minute. The sieve held 96.0 g of residue before the hand check, and 1.2 g passes during the minute. Is sieving sufficient?

A
B
C
D