During an ISO 13485 audit of a design and development process, an auditor reviews design review records. The records show three design reviews were held but only one included an independent reviewer (someone not directly involved in the design activity). ISO 13485:2016 Clause 7.3.5 requires design reviews to include participants who are NOT involved in the stage under review. This finding is BEST classified as: