During an ISO 13485 audit, a Lead Auditor finds that the organization's internal audit schedule shows all MDQMS processes were audited, but the internal audit reports for three critical processes (sterilization, design controls, CAPA) contain no objective evidence — only checklist ticks with no notes. The BEST finding is:
- ANo finding — completing the schedule is sufficient
- BObservation — the format of audit reports is at the organization's discretion
- CMinor nonconformity — audit reports exist for all processes