4.4 Immediate-Use Steam Sterilization (IUSS)
Key Takeaways
- IUSS (formerly called flash sterilization) is sterilization of an unwrapped or minimally wrapped item for immediate use
- IUSS should NOT be used for convenience, to compensate for insufficient inventory, or to save time
- IUSS is intended only for emergency situations when there is no alternative sterile instrument available
- Gravity IUSS parameters: 270°F (132°C) for 3 minutes (non-porous, non-lumened) or 10 minutes (porous/lumened)
- Prevacuum IUSS parameters: 270°F (132°C) for 3-4 minutes (non-porous/non-lumened)
- IUSS items should be used during the procedure for which they were sterilized — not stored for later use
- IUSS should be monitored with chemical and biological indicators
- Facilities should track IUSS frequency — high rates indicate a system problem (insufficient inventory, poor scheduling)
Immediate-Use Steam Sterilization (IUSS)
IUSS (Immediate-Use Steam Sterilization), formerly called "flash sterilization," is the sterilization of an unwrapped or minimally wrapped patient care item for immediate use when there is insufficient time for terminal sterilization through the normal cycle.
When IUSS Is Acceptable
IUSS is intended for emergency situations only:
- A needed instrument is accidentally dropped or contaminated during a procedure
- The only available instrument for a critical procedure becomes non-sterile
- A patient's life or safety is at immediate risk without the instrument
When IUSS Is NOT Acceptable
- Convenience — to avoid waiting for a full sterilization cycle
- Insufficient inventory — indicates a systemic problem that must be addressed by purchasing more instruments
- Routine processing — IUSS should never be a standard practice for any instrument set
- Implants — IUSS is strongly discouraged for implantable devices (BI cannot be read before use)
- Saving time — schedule and process instruments through normal cycles
AAMI, AORN, TJC, and CMS all state that IUSS should be kept to a minimum and is NOT a substitute for adequate instrument inventory.
IUSS Cycle Parameters
Gravity Displacement IUSS:
| Item Type | Temperature | Exposure Time |
|---|---|---|
| Non-porous, non-lumened (solid metal) | 270°F (132°C) | 3 minutes |
| Non-porous + porous, or lumened items | 270°F (132°C) | 10 minutes |
Prevacuum IUSS:
| Item Type | Temperature | Exposure Time |
|---|---|---|
| Non-porous, non-lumened | 270°F (132°C) | 3 minutes |
| Non-porous + porous, or lumened items | 270°F (132°C) | 4 minutes |
IUSS Monitoring Requirements
Even though IUSS is an emergency process, it still requires monitoring:
- Mechanical monitoring — verify time, temperature, pressure on the cycle printout
- Chemical indicator — place a CI with the item (at minimum, a Class 5 integrating indicator)
- Biological indicator — run a BI with each IUSS cycle or at least daily when IUSS is used
- Documentation — record the reason for IUSS, the item processed, cycle parameters, patient information, and technician ID
IUSS Tracking and Metrics
Facilities should monitor IUSS frequency as a quality indicator:
| Metric | Target |
|---|---|
| IUSS rate | As low as possible; some facilities target < 1% of all sterilization cycles |
| Reason tracking | Document every IUSS reason to identify patterns |
| Root cause analysis | If IUSS rates are high, investigate: insufficient inventory, scheduling issues, instrument damage rates |
A high IUSS rate signals a system problem — not a sterilization success. It indicates the facility lacks adequate instrumentation to support the surgical schedule through standard sterilization.
Transport of IUSS Items
Since IUSS items are unwrapped or minimally wrapped, transport must be aseptic:
- Transport in a covered, rigid container specifically designed for IUSS items
- Transport directly from the sterilizer to the sterile field
- Minimize time between sterilizer and sterile field
- The item should be used only for the patient and procedure for which it was sterilized
- Never store an IUSS-processed item — if not used immediately, it must be reprocessed through a full cycle
IUSS (formerly flash sterilization) should be used:
The gravity displacement IUSS exposure time for a non-porous, non-lumened instrument at 270°F (132°C) is: