3.1 Instrument Inspection & Functionality Testing

Key Takeaways

  • Every instrument must be inspected for cleanliness, function, and damage after cleaning and before packaging
  • Visual inspection should be performed using adequate lighting and magnification when available
  • Scissors are tested by cutting through manufacturer-specified test material (typically latex or specific fabric)
  • Hemostats and clamps are tested by closing on material — they should hold without slipping or "springing"
  • Needle holders are tested for jaw alignment and the ability to securely grip a suture needle
  • Insulated instruments must be tested with an insulation tester to detect breaches that could cause burns
  • Damaged, stained, or malfunctioning instruments must be removed from service and sent for repair or replacement
  • Instruments with pitting, corrosion, cracks, or burrs must be removed — they harbor organisms and can injure patients
Last updated: March 2026

Instrument Inspection & Functionality Testing

Inspection is the quality checkpoint between cleaning and packaging. Every instrument must be verified for cleanliness, proper function, and absence of damage before it can be packaged for sterilization. This step protects patients and ensures surgeons receive instruments that work correctly.


Visual Inspection for Cleanliness

What to Look For:

  • Remaining soil — blood, tissue, mineral deposits, stains
  • Biofilm — slimy residue in lumens, channels, or crevices
  • Detergent residue — white powder or film from incomplete rinsing
  • Water spots — mineral deposits from poor water quality

Inspection Best Practices:

  • Use bright, focused lighting (lighted magnification highly recommended)
  • Inspect all surfaces — including box locks, serrations, ratchets, lumens, and internal channels
  • Use a lumen scope or borescope to inspect internal channels of lumened instruments
  • Any instrument with visible soil must be returned to decontamination for re-cleaning
  • Protein residual testing can supplement visual inspection for quality assurance

Functionality Testing by Instrument Type

Scissors

TestMethodPass Criteria
SharpnessCut through manufacturer-specified test material (latex glove, cotton fabric) to the tipClean cut without shredding or catching
Blade alignmentHold up to light in closed positionNo light visible between blades; blades align evenly
TensionOpen and close repeatedlySmooth action without excessive looseness or tightness

Hemostats / Clamps

TestMethodPass Criteria
Jaw alignmentClose fully and examine tipsTips meet perfectly; no overlap or gap
Ratchet functionClose on each ratchet positionRatchet holds firmly; does not slip or "spring" open
Spring tensionOpen and close repeatedlyReturns to open position without sticking

Needle Holders

TestMethodPass Criteria
Jaw insertExamine carbide insert (gold handles indicate carbide)No wear, chipping, or loose inserts
Grip testPlace suture needle in jaws and close first ratchetNeedle holds securely without rotation
AlignmentClose and examine jaw tipsTips meet evenly with no offset

Forceps (Thumb/Tissue)

TestMethodPass Criteria
Tip alignmentHold up to light; close tipsTips meet evenly; teeth interlock properly
TensionSqueeze and releaseReturns to rest position with appropriate spring
Teeth conditionExamine under magnificationTeeth intact, not bent or broken

Retractors

TestMethodPass Criteria
Surface conditionVisual examinationNo bending, cracks, or sharp edges
Ratchet function (self-retaining)Engage and disengage ratchetHolds securely; releases cleanly
FinishCheck for pitting or corrosionSmooth finish without defects

Insulated Instruments (Electrosurgical)

TestMethodPass Criteria
Insulation integrityUse an insulation tester deviceNo breaches detected; no current leakage
FunctionalityTest mechanical function as aboveProper operation of all moving parts

Insulation testing is critical — tiny insulation breaches invisible to the naked eye can cause patient burns during electrosurgery. Always test with a dedicated insulation tester, not just visual inspection.


Instrument Defects — Remove From Service

DefectRiskAction
Pitting/corrosionHarbors organisms; weakens instrumentRemove; repair or replace
Cracked or broken jawsPatient injury; instrument failureRemove; discard or repair
Bent tipsCannot perform intended functionRemove; send for repair
Burrs on edgesCan tear tissue; difficult to cleanRemove; send for repair
Loose box locksInstrument will not hold properlyRemove; send for repair
Missing partsInstrument incomplete; cannot functionRemove; order replacement parts
StainingMay indicate improper cleaning or corrosionEvaluate cause; treat or replace
Dull scissorsCannot cut properly; surgeon frustrationSend for sharpening

Instrument Lubrication

  • Water-soluble, steam-penetrable lubricant (instrument milk) may be applied after cleaning
  • Lubricant keeps box locks, hinges, and moving parts functioning smoothly
  • Never use silicone-based or WD-40 type lubricants — they are NOT steam-penetrable and will prevent sterilization
  • Apply lubricant per manufacturer IFU
  • Some facilities apply lubricant through automated lubrication baths
Test Your Knowledge

When inspecting scissors, light visible between the closed blades indicates:

A
B
C
D
Test Your Knowledge

Insulated electrosurgical instruments should be tested with:

A
B
C
D
Test Your Knowledge

Which type of lubricant should be used on surgical instruments?

A
B
C
D