2.4 Automated Cleaning Methods
Key Takeaways
- Automated cleaning equipment includes washer-disinfectors, ultrasonic cleaners, washer-sterilizers, and cart washers
- Washer-disinfectors are the primary automated cleaning workhorses — they wash, rinse, and thermally disinfect
- Ultrasonic cleaners use high-frequency sound waves (20-40 kHz) to create cavitation that removes soil from crevices
- Ultrasonic cleaning does NOT disinfect — it is a cleaning adjunct that must be followed by further processing
- Washer-disinfectors typically reach 180-195°F (82-90°C) during the thermal disinfection phase
- Instruments must be properly loaded in automated equipment — hinges open, lumens flushed, no nesting
- Automated equipment must be validated, regularly maintained, and monitored with cleaning verification tests
- Manufacturer IFU dictates whether an instrument can be processed in automated equipment
Automated Cleaning Methods
Automated cleaning equipment provides standardized, reproducible, and validated cleaning results that are more consistent than manual cleaning alone. However, automated cleaning does not replace manual cleaning entirely — it supplements and enhances it.
Types of Automated Cleaning Equipment
1. Washer-Disinfectors (WDs)
The primary automated cleaning equipment in most CS departments.
Cycle phases:
- Pre-wash/flush — Cold water rinse to remove gross soil
- Enzymatic wash — Detergent applied at specified temperature
- Rinse — Multiple rinse cycles to remove detergent residue
- Thermal disinfection — Water heated to 180-195°F (82-90°C) for a specified time
- Drying — Heated air drying phase (some models)
Key points:
- Thermal disinfection at 180°F+ for the appropriate exposure time provides intermediate-level disinfection
- Items must still be sterilized after washer-disinfection if they are critical or semi-critical items intended for sterilization
- Proper loading is essential for effectiveness:
- Open all hinged instruments
- Position instruments with concave surfaces facing down for drainage
- Do not overload — water and detergent must reach all surfaces
- Separate dissimilar metals to prevent galvanic corrosion
- Use appropriate racks and adapters for specific instrument types
2. Ultrasonic Cleaners
Use high-frequency sound waves (20,000-40,000 Hz / 20-40 kHz) to create millions of tiny bubbles in the cleaning solution. These bubbles implode on contact with surfaces, creating a scrubbing action called cavitation.
How cavitation works:
- Sound waves create alternating high and low pressure zones in the liquid
- In low-pressure zones, microscopic bubbles form (cavitation)
- In high-pressure zones, these bubbles collapse violently against instrument surfaces
- The collapsing bubbles generate intense localized energy that dislodges soil from crevices, hinges, serrations, and other hard-to-reach areas
Key points:
- Ultrasonic cleaning is a CLEANING process only — it does NOT disinfect
- Instruments must be fully submerged in the ultrasonic solution
- Do not place hands in the ultrasonic solution while the unit is operating — cavitation can damage tissue
- Lids should be closed during operation to prevent aerosolization
- Degassing is required when filling with fresh solution — run the unit empty for the manufacturer-recommended time to remove dissolved air
- Solution should be changed at regular intervals or when visibly contaminated
- Chrome-plated instruments should not be processed in ultrasonic cleaners (plating may separate)
3. Cart Washers
- Large, drive-through or roll-in washers for case carts, transport carts, and large containers
- Operate similarly to washer-disinfectors but on a larger scale
- Ensure all surfaces are accessible to the wash spray
4. Washer-Sterilizers (Older Technology)
- Combine washing and steam sterilization in one cycle
- Being phased out in many facilities due to limitations:
- Items are sterilized while wet (reduces sterile barrier)
- Difficult to validate
- Modern washer-disinfectors + separate sterilizers provide better results
Cleaning Verification Testing
Automated cleaning equipment must be monitored regularly to verify it is functioning correctly:
| Test | What It Verifies | Frequency |
|---|---|---|
| Cleaning verification tests (protein residual tests) | Soil removal effectiveness | Per facility policy; at least periodically |
| Temperature monitoring | Thermal disinfection parameters met | Each cycle (automated recording) |
| Water quality testing | Water meets AAMI ST108 standards | Per facility policy |
| Detergent concentration | Correct amount of cleaning chemistry | Per facility policy |
| Visual inspection | Spray arms rotate, nozzles clear, gaskets intact | Daily |
Loading Best Practices
| Do | Don't |
|---|---|
| Open all hinged instruments | Load instruments closed or locked |
| Place items in appropriate racks | Stack or pile instruments loosely |
| Position concave surfaces down | Position cups/bowls upright (traps water) |
| Ensure spray arms can rotate freely | Overload or block spray arms |
| Follow manufacturer IFU for rack placement | Assume all instruments can go in any rack |
| Separate dissimilar metals | Mix stainless steel and aluminum in same wash load |
Ultrasonic cleaners remove soil through a process called:
The thermal disinfection phase in a washer-disinfector typically reaches a temperature of:
Before using an ultrasonic cleaner with fresh solution, the technician must first:
When loading instruments into a washer-disinfector, hinged instruments should be: