6.3 Endoscope Reprocessing
Key Takeaways
- Flexible endoscopes are complex, expensive, semi-critical devices requiring meticulous reprocessing
- The reprocessing sequence is: point-of-use treatment, leak testing, manual cleaning, visual inspection, HLD or sterilization, rinsing, drying, storage
- AAMI ST91 is the standard for flexible and semi-rigid endoscope reprocessing
- All channels must be brushed and flushed during manual cleaning — automated cleaning alone is insufficient
- Automated endoscope reprocessors (AERs) provide more consistent HLD but do NOT replace manual pre-cleaning
- Endoscopes must be stored hanging vertically in a well-ventilated cabinet with caps removed
- Drying (including alcohol flush and forced air) is critical to prevent microbial growth in channels
- Reprocessing failures with endoscopes have been linked to patient infection outbreaks
Endoscope Reprocessing
Flexible endoscope reprocessing is one of the most complex and highest-risk processes in Central Service. The intricate design of flexible endoscopes — with multiple narrow channels, delicate components, and materials that cannot tolerate heat — makes them especially challenging to reprocess. Failures in endoscope reprocessing have been directly linked to patient infection outbreaks.
The Complete Reprocessing Sequence
Step 1: Point-of-Use Treatment (at Procedure Site)
- Immediately after removal from the patient, wipe the insertion tube
- Flush all channels with manufacturer-recommended enzymatic solution
- Suction enzymatic solution through the suction channel
- Transport to reprocessing area in a covered, leak-proof container
Step 2: Leak Testing
- Perform before every cleaning cycle — do not skip
- Pressurize the scope per manufacturer IFU and submerge in water
- Look for air bubbles indicating a breach in the outer sheath or channels
- If leak detected: Remove from service → tag as damaged → send for repair
- A scope with a leak must NEVER be processed further or used on a patient
Step 3: Manual Cleaning
This is the most critical step — automated cleaning alone is NOT sufficient:
- Brush all accessible channels with appropriately sized brushes
- Flush all channels with enzymatic detergent solution
- Clean the exterior with a soft cloth and enzymatic solution
- Clean all valves, buttons, and biopsy port caps
- Brush tips should exit the distal end of the channel — this confirms the brush traversed the entire length
- Rinse all channels and exterior thoroughly
Step 4: Visual Inspection
- Inspect the insertion tube for damage, kinks, or discoloration
- Check the distal tip for scratches or lens damage
- Verify all valves and buttons are intact
- Confirm channels are patent (not blocked)
Step 5: High-Level Disinfection or Sterilization
Manual HLD:
- Completely immerse the scope in HLD solution
- Fill ALL channels with solution (no air bubbles)
- Maintain immersion for the required contact time
- Verify MEC before each use
Automated (AER):
- Connect ALL channel adapters per manufacturer instructions
- Run the complete AER cycle
- Verify cycle completion on the AER printout
Step 6: Rinse
- Rinse thoroughly with appropriate water (critical water for final rinse)
- Flush ALL channels to remove all HLD chemical residue
- Inadequate rinsing can cause chemical injury to patients
Step 7: Drying
- Alcohol flush: Flush all channels with 70% isopropyl alcohol to facilitate drying
- Forced air: Blow compressed air through all channels to remove moisture
- Critical importance: Moisture left in channels promotes bacterial growth (especially Pseudomonas)
- Drying is NOT optional — it is a mandatory step
Step 8: Storage
- Store hanging vertically in a well-ventilated endoscope drying/storage cabinet
- Remove all valves and caps to allow air circulation through channels
- Do not coil the scope tightly — this can damage internal channels
- Storage cabinets should have HEPA-filtered air circulation
AAMI ST91: The Endoscope Reprocessing Standard
Key requirements from AAMI ST91 (Flexible and Semi-Rigid Endoscope Processing):
- Written reprocessing procedures must be available at the point of processing
- Staff must be trained and competency validated before processing endoscopes
- All steps of the IFU must be followed — no shortcuts
- Documentation of each reprocessing cycle
- Quality monitoring including periodic culture surveillance
Common Endoscope Reprocessing Errors
| Error | Consequence |
|---|---|
| Skipping manual pre-cleaning | Biofilm formation; HLD cannot penetrate soil |
| Skipping leak test | Undetected damage; fluid intrusion; contamination |
| Inadequate channel brushing | Residual bioburden in channels |
| Shortened HLD contact time | Inadequate microbial kill |
| Inadequate rinsing | Chemical injury to next patient |
| Skipping alcohol flush/drying | Bacterial growth in moist channels |
| Improper storage (coiled, in drawer) | Moisture retention; contamination |
Which step in endoscope reprocessing is considered the MOST critical?
Flexible endoscopes should be stored:
If a leak test reveals air bubbles from a flexible endoscope, the correct action is: